K211114 · Hdx Will Corp. · LLZ · Dec 15, 2021 · Radiology
Device Facts
Record ID
K211114
Device Name
Will3D
Applicant
Hdx Will Corp.
Product Code
LLZ · Radiology
Decision Date
Dec 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Will3D is a software application used for the display and 3D visualization of medical image files from scanning devices, such as Dental CT scanner. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, store, print, assist in diagnosis, and distribute images utilizing standard PC hardware. Additionally, Will3D is a preoperative software application and evaluation and evaluation of dental implants, orthodontic planning and surgical treatments.
Device Story
Will3D is a standalone software application for medical image management and processing. It ingests DICOM files from scanning devices like Dental CT scanners. The software provides tools for 2D/3D visualization, multi-planar reconstruction (MPR), panoramic/cross-sectional reconfiguration, nerve tracing, virtual implant placement, cephalometric analysis, surgical simulation, and airway endoscopy. It operates on standard PC hardware (Windows 7+/macOS Yosemite+). Clinicians use the software to view, manipulate, and analyze images to assist in diagnosis and preoperative planning. Outputs include rendered 3D models, surgical simulations, and patient reports. The device facilitates clinical decision-making by allowing providers to visualize anatomical structures, simulate implant positioning, and compare pre- and post-surgical volumes via superimposition. It is intended for use in clinical settings by qualified healthcare professionals.
Clinical Evidence
Bench testing only. The device underwent software verification and validation according to FDA guidance for moderate level of concern software. No clinical data was required or presented.
Technological Characteristics
Standalone software application; compatible with Windows 7+ and macOS Yosemite+. Supports DICOM standard for image communication. Features include MPR, volume/surface rendering, 3D cephalometric analysis, and virtual implant planning. Software development follows established verification and validation processes.
Indications for Use
Indicated for radiologists, clinicians, and referring physicians to display, visualize, and process medical images (e.g., Dental CT) for diagnostic assistance, preoperative dental implant evaluation, orthodontic planning, and surgical treatment simulation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
December 15, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The words "U.S. FOOD & DRUG" are on the first line, and "ADMINISTRATION" is on the second line.
Hdx Will Corp. % Kim In-Young Staff #105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu Cheongju-si, Chungcheongbuk-do 28161 REPUBLIC OF KOREA
Re: K21114
Trade/Device Name: Will3D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: November 2, 2021 Received: November 12, 2021
Dear Kim In-Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211114
Device Name Will3D
### Indications for Use (Describe)
Will3D is a software application used for the display and 3D visualization of medical image files from scanning devices, such as Dental CT scanner. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, store, print, assist in diagnosis, and distribute images utilizing standard PC hardware. Additionally, Will3D is a preoperative software application and evaluation and evaluation of dental implants, orthodontic planning and surgical treatments.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|-----------------------------------------------|
| X Prescription I Ise (Part 21 CER 801 Subnart D) | Over-The-Counter I lea (21 CER 801 Subnart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green square with a white "W" inside, followed by the text "HDX Will" in blue. The logo is simple and modern, and the colors are bright and eye-catching.
# 510(k) Summary
[As required by 21 CFR 807.92]
K211114
# 1. Date Prepared [21 CFR 807.92(a)(1)]
November 2th 2021
## 2. Submitter's Information [21 CFR 807.92(a)(1)]
| - Name of Manufacturer: | HDX WILL Corp. |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| - Address: | #105, 106, 201, 202, 203, 204, 38, Osongsaengmyeong<br>4-ro, Osong-eup, Heungdeok-gu, Cheongju-si,<br>Chungcheongbuk-do, 28161, Republic of Korea |
| - Contact Person: | In Young Kim / Staff |
| - Telephone No.: | +82-43-710-7318 |
| - Fax No.: | +82-43-710-7312 |
| - Email Address: | kiy@iwillmed.com |
| - Registration No.: | 3013511605 |
# 3. Trade Name, Regulation Name, Classification [21 CFR 807.92(a)(2)]
| Device Name | Will3D |
|---------------------|------------------------------------------------|
| Regulation Number | 21 CFR 892.2050 |
| Common/Usual Name | Medical Image Management and Processing System |
| Regulatory Class | Class II |
| Product Code | LLZ |
| Classification Name | System, Image Processing, Radiological |
| Panel | Radiology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green and blue square with the letter "W" in white inside the square. To the right of the square is the text "HDX Will" in blue. The logo is simple and modern.
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[
The identified predicate devices within this submission are shown as follow;
### Predicate device
K170180 HDX WILL CORP. Will3D
There are no significant differences between the Will3D and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, and technical characteristics
## 5. Description of the Device [21 CFR 807.92(a)(4)]
Will3D is one of the components of a Picture Archiving and Communications System(PACS). Will3D is the software application that provides image viewing and manipulation in the diagnostic imaging setting.
| No. | Feature /<br>Functionality | Description |
|-----|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Authentication | Software user authentication process function |
| 2 | License | Software license check function |
| 3 | Main | Software configuration and execution function<br>View |
| 4 | File | Register/search the data desired by the user in the<br>database and load the selected data |
| 5 | OTS<br>DICOM | Scan, organize image files and convert from software to<br>DICOM image files, process offline media, and send<br>images over a network connection |
| 6 | MPR | Display the image visualized in 2D/3D and show the<br>image from a selected direction<br>A function that visualizes the input DICOM image in<br>2D/3D and outputs it on the screen, allowing the user to<br>view the image from the desired direction. |
| 7 | Panorama | Reconfigure the panoramic and cross-sectional images<br>and draw the nerve in the panoramic image |
| 8 | Implant | Display the Implant Model registered in a database and<br>virtually place implants in the ideal position |
| 9 | TMJ | Display the 2D/3D images of the TMJ field on the screen<br>and provide an interface to easily observe the TMJ area |
| 10 | 3D Ceph | Provide the cephalometric analysis and surgical<br>simulation functions |
| 11 | Face Simulation | A surface model can be created from the input volume |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a blue square with a green upper-right corner and a white "W" inside the square. To the right of the square is the text "HDX Will" in a stylized blue font. The text is slightly slanted to the right.
| | | and 2D photo can be mapped, and the 3D photo created<br>in this way or the 3D photo created by the equipment can<br>be mapped to the input volume.<br>In addition, the function of comparing the image result of<br>surgical simulation performed by 3D Ceph with the<br>existing image. | |
|--------|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 12 | Super imposition | Visually compare and observe changes before and after<br>surgery by outputting two different volumes on one<br>screen | |
| 13 | Endoscopy | Provide a virtual endoscopic function for airway<br>observation | |
| 14 | Report | Create the patient report | |
| | Common | | |
| 15 | View Tool | 'View' motion control function | |
| 16 | Measure Tool | Measure function | |
| 17 | Common Tool | Commonly used functions such as project saving,<br>printing, screen capture, and software information | |
| 18 | OTF Tool | Adjust the color, transparency, and color range of Volume<br>Rendering video | |
| Engine | | | |
| 19 | MPR Engine | Generate MPR image | |
| 20 | VR Engine | Create Volume Rendering and Surface Rendering<br>images (*VR means Volume Rendering) | |
| 21 | OTS 3D<br>Rendering | Create a 2D/3D screen with texture mapping, special<br>effects, and other visual functions | |
# 6. Indications for Use [21 CFR 807.92(a)(5)]
Will3D is a software application used for the display and 3D visualization of medical image files from scanning devices, such as Dental CT scanner. It is intended for use by radiologists, clinicians, referring physicians and other qualified individuals to retrieve, process, render, review, store, print, assist in diagnosis, and distribute images utilizing standard PC hardware. Additionally, Will3D is a preoperative software application used for the simulation and evaluation of dental implants, orthodontic planning and surgical treatments.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for HDX Will. The logo consists of a green and blue square with the letter "W" in white inside the square. To the right of the square is the text "HDX Will" in blue.
# 7. Determination of Substantial Equivalence
Summary of technological characteristics of the device compared to the predicate device. [21 CFR 807.92(a)(6)]
# a) Technological Characteristics
| | Subject Device | Predicate Device | SE Note |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Model Name | Will3D | Will3D (K170180) | - |
| Manufacturer | HDX WILL CORP. | HDX WILL CORP. | - |
| K Number | K211114 | K170180 | - |
| Classification<br>Name | System, Image Processing,<br>Radiological | System, Image Processing,<br>Radiological | Same |
| Regulatory<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Product Code | LLZ | LLZ | Same |
| Classification | II | II | Same |
| Indication for<br>Use | Will3D is a software application<br>used for the display and 3D<br>visualization of medical image<br>files from scanning devices,<br>such as Dental CT scanner. It<br>is intended for use by<br>radiologists, clinicians,<br>referring physicians and other<br>qualified individuals to retrieve,<br>process, render, review, store,<br>print, assist in diagnosis, and<br>distribute images utilizing<br>standard PC hardware.<br>Additionally, Will3D is a<br>preoperative software<br>application used for the<br>simulation and evaluation of<br>dental implants, orthodontic<br>planning and surgical<br>treatments. | Will3D is a software application<br>used for the display and 3D<br>visualization of medical image<br>files from scanning devices,<br>such as Dental CT scanner. It<br>is intended for use by<br>radiologists, clinicians,<br>referring physicians and other<br>qualified individuals to retrieve,<br>process, render, review, store,<br>print, assist in diagnosis, and<br>distribute images utilizing<br>standard PC hardware.<br>Additionally, Will3D is a<br>preoperative software<br>application used for the<br>simulation and evaluation of<br>dental implants, orthodontic<br>planning and surgical<br>treatments. | Same |
| Type of Use | Prescription Use | Prescription Use | Same |
| Component | Standalone software | Standalone software | Same |
| Modality<br>Support | CT image data | CT image data. | Same |
| Operating<br>System | Window 7 or higher and Mac<br>os Yosemite or higher | Window 7 or higher and Mac<br>os Yosemite or higher | Same |
| | Subject Device | Predicate Device | SE Note |
| Model Name | Will3D | Will3D (K170180) | - |
| Image<br>Communication<br>Standard | DICOM | DICOM | Same |
| Feature/<br>Functionality | - File<br>- MPR<br>- Panorama<br>- Implant<br>- TMJ<br>- 3D Ceph<br>- Face Simulation<br>- Super imposition<br>- Endoscopy<br>- Report | - File<br>- MPR<br>- Panorama<br>- Implant<br>- TMJ<br>- Orthodontic<br>- 3D Ceph<br>- Face Simulation<br>- Super imposition<br>- Endoscopy<br>- Report | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for HDXWILL. The logo consists of a green square with a white "W" inside, followed by the text "HDXWILL" in a bold, blue font. The logo appears to be for a company or organization.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for HDXWill. The logo consists of a green square with a white "W" inside, followed by the text "HDXWill" in blue. The font is sans-serif and the overall design is modern and professional.
## b) Differences between Subject and Predicate Devices
The Will3D is substantially equivalent to the predicate device identified above with respect to intended use and technological characteristics. From the information provided in table above, it is understood that the subject device does not introduce any new technology and/or indications of use. Therefore, the Will3D is considered substantially equivalent to the predicate device.
# 8. Non-Clinical Test Summary
The Will3D contains MODERATE level of concern software. Software was designed and developed according to a software development process and was verified and validated.
Software information is provided in accordance with FDA guidance: "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005."
Cybersecurity information is provided in accordance with FDA guidance: "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, issued on October 2, 2014"
# 9. Conclusion [21 CFR 807.92(b)(3)]
The Will3D has same indications for use and technical characteristic to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness. In addition, performance testing conducted demonstrate that the subject device is as safe as effective as the predicate. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.