K211086 · Ortho Development Corporation · HSB · Dec 21, 2021 · Orthopedic
Device Facts
Record ID
K211086
Device Name
The Progen™ Trochanteric Nail System
Applicant
Ortho Development Corporation
Product Code
HSB · Orthopedic
Decision Date
Dec 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Progen™ short trochanteric nail is indicated for fixation of various types of stable and unstable neck, intertrochanteric, and peritrochanteric fractures. The Progen™ long trochanteric nail is intended for fixation of stable femoral fractures occurring from the base of the femoral neck extending distally to a point approximal to the intercondylar notch including fractures of the basilar neck, intertrochanteric fractures, subtrochanteric fracture, femoral shaft fractures, pathological fractures, impending pathological fractures, tumor resection, nonunion, and revisions.
Device Story
Progen™ Trochanteric Nail System is a set of single-use intramedullary nails and instrumentation for femoral fracture fixation. Implants are available in various lengths/diameters to accommodate patient anatomy. System components include nail, lag screw, secondary screw, lag cap, locking screws, set screws, and nail end cap. Used in clinical settings by surgeons for stabilization of stable/unstable fractures. Implants are secured via screws transecting proximal/distal nail holes. Device provides mechanical stabilization of bone segments to facilitate healing.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including full construct fatigue and static cantilever loading, bone screw axial pullout, driving torque, torsional properties, static four-point bend, bending fatigue, and static torsional stiffness tests.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V ELI) with Type II anodized surface treatment on nails and lag screws. Mechanical testing per ASTM F384:2017, ASTM F543:2017, and ASTM F1264:2016(e1). Sterilization: Gamma radiation (ISO 11137-1/2). Packaging: ISO 11607-1/2. Biocompatibility: ISO 10993-1.
Indications for Use
Indicated for patients requiring fixation of stable and unstable femoral fractures, including neck, intertrochanteric, peritrochanteric, subtrochanteric, and femoral shaft fractures; also indicated for pathological fractures, impending pathological fractures, tumor resection, nonunion, malunion, and revisions.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
Ortho Development® Corporation Drew Weaver Director of Regulatory Affairs 12187 South Business Park Drive Draper, Utah 84020
Re: K211086
Trade/Device Name: The Progen™ Trochanteric Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, JDS Dated: November 23, 2021 Received: November 24, 2021
Dear Drew Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
December 21, 2021
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Jiping Chen, Ph.D., M.P.H. Acting Division Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211086
Device Name The Progen™ Trochanteric Nail System
#### Indications for Use (Describe)
The Progen™ short trochanteric nail is indicated for fixation of various types of stable and unstable neck, intertrochanteric, and peritrochanteric fractures.
The Progen™ long trochanteric nail is intended for fixation of stable femoral fractures occurring from the base of the femoral neck extending distally to a point approximal to the intercondylar notch including fractures of the basilar neck, intertrochanteric fractures, subtrochanteric fracture, femoral shaft fractures, pathological fractures, impending pathological fractures, tumor resection, nonunion, and revisions.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the text "K211086 Page 1 of 4" at the top. Below the text is a logo for Ortho Development. The logo features a stylized purple circle with curved lines emanating from the left side, followed by the word "ORTHO" in bold, purple letters. Underneath "ORTHO" is the word "DEVELOPMENT" in a smaller, lighter font.
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
| Name of the Sponsor: | Ortho Development® Corporation |
|-----------------------------|-------------------------------------------------------|
| | 12187 South Business Park Drive |
| | Draper, Utah 84020 |
| 510(k) Primary Contact: | Name: Drew Weaver |
| | Position: Director of Regulatory and Clinical Affairs |
| | Telephone: (801) 619-3419 |
| | Email: dweaver@orthodevelopment.com |
| 510(k) Secondary Contact: | Name: Stan Despres |
| | Position: V.P. Product Development |
| | Telephone: (801) 376-6231 |
| | Email: stan@orthodevelopment.com |
| Date Prepared: | December 14, 2021 |
| Submission Type: | Traditional 510(k) |
| Proprietary Name: | The Progen™ Trochanteric Nail System |
| Common Name: | Intramedullary fixation rod and accessories |
| Classification: | 21 CFR 888.3020 |
| Device Class: | Class II |
| Device Product Code: | HSB, JDS |
| Primary Predicate Device: | Gamma3® Nail System (K043431) |
| Secondary Predicate Device: | TriGen InterTAN (K040212) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the text "K211086 Page 2 of 4" at the top. Below the text is a logo for "ORTHO DEVELOPMENT". The logo consists of a purple circle with three curved lines emanating from the left side, followed by the word "ORTHO" in large purple letters and the word "DEVELOPMENT" in smaller letters below.
## 5.1 Device Description:
The Progen™ Trochanteric Nail System consists of temporary fixation intramedullary nails and their accompanying instrumentation designed for fracture fixation and stabilization of the femur. The implants are available in various lengths and diameters to accommodate a range of patient anatomy. Each of the intramedullary nails is secured by a sequence of screws that transect through holes in the proximal and distal sections of each nail.
The Progen™ Trochanteric Nail System consists of single-use intramedullary nails for stable and unstable neck, intertrochanteric, pertrochanteric, and subtrochanteric fractures and combinations of these fractures. The system consists of Nail, Lag Screw, Secondary Screw, Lag Cap, Locking Screws, Set Screws, and Nail End Cap. The nails and accompanying components are manufactured from titanium alloy (Ti-6Al-4V ELI). Additionally, the Nails and Lag Screws have a Type II anodized surface treatments.
### 5.2 Indication for Use:
The Progen™ short trochanteric nail is indicated for fixation of various types of stable and unstable neck, intertrochanteric, and peritrochanteric fractures.
The Progen™ long trochanteric nail is intended for fixation of stable and unstable femoral fractures occurring from the base of the femoral neck extending distally to a point approximately 10cm proximal to the intercondylar notch including fractures of the basilar neck, intertrochanteric fractures, peritrochanteric fractures, subtrochanteric fracture, femoral shaft fractures, pathological fractures, impending pathological fractures, tumor resection, nonunion, malunion, and revisions.
### 5.3 Comparison of Technological Characteristic:
The Progen™ Trochanteric Nail System is technologically similar to the already cleared predicate device Gamma3® Nail System (K043431) and TriGen InterTAN (K040212) in terms of indication for use/ intended use, technological characteristics, basic design, device material, mechanical performance and principle of operation.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the text "K211086 Page 3 of 4" at the top. Below the text is a logo for a company called "ORTHO DEVELOPMENT". The logo features a stylized purple circle with curved lines emanating from it, followed by the company name in a bold, sans-serif font.
## 5.4 Performance Data:
## Sterilization
The Progen™ Trochanteric Nail System is gamma radiation sterilized and was validated to a sterility assurance level of 106 in accordance with the following standards:
- ISO 11137-1:2006, Am1:2013, Sterilization of health care products Requirements . for validation and routine control - Radiation sterilization; and
- . ISO 11137-2:2013, Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose.
Validation results indicate that the Progen™ Trochanteric Nail System complies with the standards.
### Shelf Life
The packaging of The Progen™ Trochanteric Nail System was validated in accordance with the following standards:
- . ISO 11607-1:2006 Packaging for terminally sterilized medical devices - Part 1: requirements for materials, sterile barrier systems and packaging systems; and
- . ISO 11607-2:2006 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes.
Validation results indicate that the packaging for the Progen™ Trochanteric Nail System complies with the standards.
### Biocompatibility
The Progen™ Trochanteric Nail System contact materials were verified in accordance with the following standards:
- . ISO 10993-1:2009, Biological evaluation of medical device – Part 1: Evaluation and testing within a risk management process.
Validation results indicate that the biological evaluation complies with the standard.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the text "K211086 Page 4 of 4" at the top, followed by a logo for "ORTHO DEVELOPMENT". The logo features a stylized purple circle with curved lines emanating from it, followed by the word "ORTHO" in bold purple letters. Below "ORTHO" is the word "DEVELOPMENT" in smaller, lighter letters with a trademark symbol.
# Mechanical Testing
The following mechanical testing were performed on The Progen™ Trochanteric Nail System
- . Full construct fatigue cantilever loading test per ASTM F384:2017
- . Full construct static cantilever loading test per ASTM F384:2017
- . Bone screw axial pullout strength test per ASTM F543:2017
- Bone screw driving torque per ASTM F543:2017
- Bone screw torsional properties test per ASTM F543:2017
- Static Four-Point bend test per ASTM F1264:2016(e1)
- Bending Fatigue test per ASTM F1264:2016(e1)
- . Static Torsional Stiffness Test ASTM F1264:2016(e1)
## Clinical Testing
None provided for basis of substantial equivalence.
## 5.5 Substantial Equivalence Conclusion:
Verification and Validation activities were conducted to establish the performance of The Progen™ Trochanteric Nail System. The results of these activities demonstrate that The Progen™ Trochanteric Nail System is as safe, as effective, and performs as well as legally marketed predicates.
Bases on similarities in indication for use/intended use, technological characteristic, basic design, device material, and principle of operation, The Progen™ Trochanteric Nail System is considered substantially equivalent to the previously cleared predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.