K211069 · Prismatik Dentalcraft, Inc. · LRK · Oct 26, 2021 · Dental
Device Facts
Record ID
K211069
Device Name
EndSnorZ Sleep Appliance
Applicant
Prismatik Dentalcraft, Inc.
Product Code
LRK · Dental
Decision Date
Oct 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EndSnorZ™ Sleep Appliance is indicated to reduce snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older. EndSnorZ™ Sleep Appliance is worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The appliance is removable by the patient.
Device Story
Mandibular advancement device; holds mandible in protrusive position during sleep to improve pharyngeal airflow; reduces snoring and mild to moderate OSA. Input: patient dental scans. Transformation: CAD software designs custom upper/lower splints based on scans; splints additively manufactured; injection-molded synthetic polymer connectors attached to splints. Output: custom-fitted oral appliance. Used at home or sleep lab; prescribed by dentist; patient self-operates. Clinician uses CAD software to design appliance based on prescription. Benefits: non-invasive reduction of snoring/OSA symptoms.
Clinical Evidence
Bench testing only. Validated design/manufacturing process, 3D print build platform positioning, and CAD software compliance (21 CFR Part 11). Mechanical testing compared tensile strength and dimensional retention in artificial saliva against predicate. Biocompatibility testing (ISO 10993 Parts 5 and 10) performed on resin and connectors.
Technological Characteristics
Mandibular advancement device; upper/lower splints additively manufactured from biocompatible light-curable methacrylate-based resin; synthetic polymer connectors (injection molded). Adjustable in 0.5 mm increments. Non-sterile. Software-driven CAD design. Biocompatibility per ISO 10993.
Indications for Use
Indicated for patients 18+ years old with snoring and mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Prismatik Dentalcraft, Inc. % Patsy Trisler Regulatory Consultant Qserve Group US, Inc. 7949 Beaumont Green East Drive Indianapolis, Indiana 46250
## Re: K211069
Trade/Device Name: EndSnorZ™ Sleep Appliance Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: September 28, 2021 Received: October 1, 2021
Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211069
Device Name
EndSnorZTM Sleep Appliance
#### Indications for Use (Describe)
EndSnorZ™ Sleep Appliance is indicated to reduce snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older. EndSnorZ™ Sleep Appliance is worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The appliance is removable by the patient.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary – K211069
| Submitter Name: | Prismatik Dentalcraft, Inc. |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 2144 Michelson Drive<br>Irvine, CA 96212 |
| Contact Person: | Herbert Crane<br>Vice President, RA/QA |
| Telephone: | 949.222.3531 |
| Date Prepared: | October 21, 2021 |
| Device Trade Name: | EndSnorZ™ Sleep Appliance |
| Common Name | Anti-Snoring Device |
| Classification Name<br>and Number | Intraoral devices for snoring and for obstructive sleep apnea<br>21 CFR 872.5570 |
| Product Code | LRK |
| Regulatory Class | II |
| Primary Predicate<br>Device: | K183270, Silent Nite sl, Prismatik Dentalcraft, Inc. |
| Reference Devices: | #1: K183598, KeyPrint KeySplint Soft, Keystone Industries<br>(Product Codes: MQC, KMY)<br>#2: K203712, The Slide, Biotex, Inc. (Product codes LQZ, LRK) |
| Indications for Use<br>Statement: | EndSnorZ™ Sleep Appliance is indicated to reduce snoring and<br>mild to moderate obstructive sleep apnea in patients 18 years of<br>age or older. EndSnorZ™ Sleep Appliance is worn while<br>sleeping to support the lower jaw in a forward position prescribed<br>by the dentist. The appliance is removable by the patient. |
| Device Description<br>and Summary of<br>Technological<br>Characteristics | The EndSnorZ Sleep Appliance is a mandibular advancement<br>device. It holds the mandible in a protrusive position determined<br>by the trained dentist. |
| | The device consists of upper and lower splints (trays), which are<br>additive manufactured using a biocompatible light curable resin,<br>and a connector, made of a biocompatible medical-grade<br>synthetic polymer with injection molding technology. The<br>connectors attach the upper and lower splints to maintain the<br>forward position of the lower jaw. The device is adjustable in 0.5<br>mm increments. |
| | The device is customized to conform to the patient's upper and<br>lower dentition, provided to Prismatik in patient scans. After<br>loading the patient's scan, the splints are designed according to<br>the scan and the clinician's prescription. Prismatik's computer<br>aided design (CAD) software is used for the design phase. The<br>connectors are applied after the splints are manufactured. |
{4}------------------------------------------------
This is a non-sterile device, to be used only by the single patient for whom it is custom-designed and is to be worn during sleep. Mechanism of Action The EndSnorZ™ Sleep Appliance is designed so the lower jaw is in an advanced (forward protruding) state during sleep. This forward position improves the flow of air through the patient's pharyngeal space during sleep, repositioning the mandible and thus reducing snoring and mild to moderate OSA. Device Testing Laboratory Testing Testing was performed by Prismatik to: . validate the design and process for manufacturing the EndSnorZ Sleep Appliance: validate the position of the devices on the build platform of . the SprintRay 3D printing machine for printing the devices with the Keysplint Soft Resin; validate that Prismatik's Wave CAD software accurately . designs the sleep appliances and is 21 CFR Part 11 compliant: and compare the following mechanical properties of EndSnorZ . Sleep Appliance to the Predicate: tensile strength of the connectors and the dimensional retention of connectors when exposed to artificial saliva. Results of all laboratory testing confirmed substantial equivalence. Biocompatibility The following ISO 10993 testing was performed by the manufacturers of the Reference devices' printer resin and the connector polymer to assess the safety and biocompatibility of the resin and synthetic polymer materials. Part 5 (Cytotoxicity Elution - MEM), Part 10 (Intracutaneous/Intradermal) Reactivity), Part 10 (Maximization for Delayed-Type Hypersensitivity) Prismatik also contracted for GLP Cytotoxicity testing (per ISO 10993, Part 5) of the injection molded connectors. The test results showed that the materials and the manufactured connectors are safe and biocompatible for the stated intended use. Animal | Human Testing No animal or human testing are required for this product type. The EndSnorZ™ Sleep Appliance as compared to the predicate Comparison to Predicate and Silent Nite sl device: Reference Devices: ■ The intended use is the same. · The mechanisms of action are the same. . The materials used to make the device components are similar to the predicate. Further, the materials and methods for fabricating the splints are the same as the materials and additive manufacturing methods used for Reference #1: The resin is the same KeyPrint KeySplint
{5}------------------------------------------------
Soft resin used for 3D printing the EndSnorZ sleep appliance splints. And, Reference #2 sleep appliance devices are also fabricated with additive manufacturing workflow and methods using a similar light-cured resin.
The following Substantial Equivalence Comparison table includes key similarities and differences of the EndSnorZ Sleep Appliance compared to the Predicate and both Reference devices.
- Based on documentation presented in the 510(k), including Substantial comparative testing between the subject and Predicate devices, Equivalence Conclusion and the presentation of similarities in technological characteristics to the Reference devices, it can be concluded that the differences in the base materials and methods of manufacture do not raise new questions of safety and effectiveness. Thus the EndSnorZ Sleep Appliance is substantially equivalent to the Predicate device.
{6}------------------------------------------------
### EndSnorZ™ Sleep Appliance Premarket Notification: Traditional 510(k)
## Substantial Equivalence Comparison Table
| Element | New Device | Primary Predicate | Reference Device<br>#1 | Reference Device<br>#2 | Differences |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name: | EndSnorZ Sleep Appliance | Silent Nite® sl | KeyPrint® KeySplint<br>Soft ™ | The Slide | N/A |
| 510(k) #: | K211069 | K183270 | K183598 | K203712 | N/A |
| Classification | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | New device, Predicate |
| Product Code: | LRK | LRK | MQC, KMY | LQZ, LRK | and Reference #2: no |
| Class: | II | II | II | II | differences |
| Submitter: | Prismatik Dentalcraft, Inc. | Prismatik Dentalcraft, Inc. | KeyStone Industries | Biotex, Inc. | N/A |
| Indications for<br>Use: | indicated to reduce<br>snoring and mild to<br>moderate obstructive<br>sleep apnea (OSA) in<br>patients 18 years of age<br>or older. EndSnorZ<br>Sleep Appliance is worn<br>while sleeping to support<br>the lower jaw in a forward<br>position prescribed by the<br>dentist. The appliance is<br>removable by the patient. | indicated to reduce<br>snoring and mild to<br>moderate obstructive<br>sleep apnea (OSA) in<br>patients 18 years of age<br>or older. Silent Nite® sl is<br>worn while sleeping to<br>support the lower jaw in a<br>forward position<br>prescribed by the dentist.<br>The appliance is<br>removable by the patient. | indicated for the<br>fabrication of<br>orthodontic and<br>dental appliances<br>such as<br>mouthguards,<br>nightguards, splints<br>and repositioners. | is for use to reduce<br>night-time snoring<br>and mild to moderate<br>obstructive sleep<br>apnea (OSA) in<br>adults. | New device, Predicate,<br>and Reference #2: no<br>differences.<br>Reference #1: may be<br>used during sleep, but<br>not for use to reduce<br>snoring or OSA. |
| Method of Use: | Single patient user,<br>removable and<br>reusable appliance | Single patient user,<br>removable and<br>reusable appliance | Single patient user,<br>removable and<br>reusable appliance | Single patient user,<br>removable and<br>reusable appliance | None |
| Rx or OTC | Rx | Rx | Rx | Rx | None |
| Environment | During sleep at home or<br>in sleep laboratory | During sleep at home or<br>in sleep laboratory | During sleep at home | During sleep at home | As noted |
| DESIGN: | | | | | |
| Rigid Trays | Upper and lower | Upper and lower | Upper and lower | Upper and lower | None |
| Adjustable | Yes | Yes | Yes | Yes | None |
| MATERIALS: | | | | | |
| Splints/Trays | Methacrylate-based light-<br>cured polymer resin used<br>to additively manufacture<br>splints. | Polyurethane or PETG<br>heat-sensitive materials<br>used to manufacture splints | Methacrylate-based<br>light-cured polymer<br>resin used to<br>additively<br>manufacture splints. | Methacrylate-based<br>light-cured polymer<br>resin used to<br>additively<br>manufacture splints | Materials & fabrication<br>method is different<br>compared to predicate<br>but identical to Ref #1<br>and similar to Ref #2. |
| Advancement<br>mechanism | Synthetic polymer nylon | Synthetic polymer nylon | N/A | Synthetic<br>polycarbonate | Compared to Predicate<br>same material type but<br>from difference sources,<br>both are manufactured<br>by injection molding<br>methods. |
| PHYSICAL<br>PROPERTIES: | | | | | |
| Connector<br>adjustment range: | 20.5mm to 26mm | 21mm to 26mm | N/A | N/A | Similar as noted |
| Maximum<br>adjustment: | 5.5mm; increments of<br>0.5 mm | 5.0 mm; increments of<br>1.0mm | N/A | N/A | Similar as noted |
| Total achievable<br>mandibular<br>advancement: | Approximately 10mm | Approximately 10mm | N/A | N/A | Similar as noted |
| STERILITY | Non-sterile | Non-sterile | Non-sterile | Non-sterile | None |
| BIOCOMPATIBLE | All materials meet<br>requirements of ISO<br>10993 | All materials meet<br>requirements of ISO 10993 | All materials meet<br>requirements of ISO<br>10993 | All materials meet<br>requirements of ISO<br>10993 | None |
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.