K211068 · Entac Medical, Inc. · DQD · Aug 4, 2021 · Cardiovascular
Device Facts
Record ID
K211068
Device Name
PrevisEA Device
Applicant
Entac Medical, Inc.
Product Code
DQD · Cardiovascular
Decision Date
Aug 4, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Indications for Use
PrevisEA is a compact, non-invasive device placed on the abdomen to capture the digestive sounds of patients. PrevisEA displays information which assists physicians in assessing the digestive health of patients. This device is for prescription use only and should be used under the direction of a licensed healthcare practitioner. The PrevisEA device has not been tested for and it is not intended for pediatric use.
Device Story
PrevisEA is a single-use, disposable, self-contained electronic stethoscope for abdominal auscultation. It features an adhesive backing for direct attachment to the patient's abdomen. The device captures digestive sounds via an embedded microphone and processes them using a proprietary algorithm to detect the MH4 acoustic biomarker. It quantifies MH4 detections over a 4-minute interval and compares the result against a threshold to assess digestive health, specifically for gastrointestinal impairment (GII) following intestinal surgery. The device includes an integrated CPU, display screen, and buttons; it operates without external cables. Physicians use the displayed information to assist in clinical assessment of digestive health. The device is intended for use under the direction of a licensed healthcare practitioner.
Clinical Evidence
Clinical trial (NCT03505476) conducted with 75 subjects undergoing intestinal resection surgery. The device was applied post-surgery to evaluate its capability to listen to and capture the MH4 acoustic biomarker. Results demonstrated the device successfully captures the biomarker in a clinical setting, validating its intended use.
Technological Characteristics
Single-use, disposable, self-contained electronic stethoscope. Materials: biocompatible per ISO 10993-1:2018. Sensing: standard electrical microphone. Power: Lithium-Ion battery. Connectivity: standalone (no PC interface). Standards: IEC 60601-1:2005 (safety), EN 60601-1-2 (EMC).
Indications for Use
Indicated for adult patients to assist physicians in assessing digestive health by capturing and quantifying digestive sounds (MH4 biomarker) on the abdomen. Not for pediatric use. Prescription use only.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
August 4, 2021
Entac Medical Inc. % Adam Harris Director, Regulatory Affairs Target Health LLC 261 Madison Avenue, 24th Floor New York, NY 10016
Re: K211068
Trade/Device Name: PrevisEA Device Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DOD Dated: May 5, 2021 Received: May 7, 2021
Dear Adam Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211068
Device Name PrevisEA Device
Indications for Use (Describe)
PrevisEA is a compact, non-invasive device placed on the abdomen to capture the digestive sounds of patients. PrevisEA displays information which assists physicians in assessing the digestive health of patients. This device is for prescription use only and should be used under the direction of a licensed healthcare practitioner. The PrevisEA device has not been tested for and it is not intended for pediatric use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
PrevisEA Device K211068 Premarket Notification
# 5. 510(k) Summary
## 5.1
Entac Medical Inc. 680 Oakleaf Office Lane, Suite 201 Memphis, TN 38117
#### 5.2 Contact Person
Adam Harris, MM, RAC Director, Regulatory Affairs Target Health LLC 261 Madison Avenue, 24th Floor New York, NY 10016 Telephone: (646) 218-2009 aharris@targethealth.com
Date of Summary: July 8, 2021
| Device Proprietary Name | PrevisEA™ Device | | |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|-------------|
| Common/Usual Name | Electronic Stethoscope | | |
| Classification Names /<br>Numbers and Code | 21 CFR<br>870.1875 | Classification Name<br>Electronic Stethoscope | Code<br>DQD |
| Regulatory Class | 2 | | |
| Prescription Status | Prescription Device | | |
| Classification Panel | Gastroenterology | | |
| Predicate Device | K150782 | GI Logic AbStats® Gateway | |
| Description of Device | PrevisEA is a non-invasive, self-contained, medical device with a<br>proprietary algorithm used to assess the digestive health of patients. The<br>device includes a stethoscope diaphragm, sound chamber, embedded<br>microphone and CPU, as well as a display screen and buttons. The<br>PrevisEA device is a convenient, single-use, disposable unit, which<br>attaches to the abdomen of the patient via an adhesive wafer on the back<br>of the device. The PrevisEA is a single-use, disposable unit, which avoids<br>the need for cleaning and disinfection.<br><br>The PrevisEA requires no cable connections for operation of the device.<br>Integration of the above components allow auscultation and digital<br>capture of digestive sounds, such as MH4.<br><br>Messages are transmitted to the easy-to-read display screen on the front<br>of the device. The PrevisEA detects the acoustic biomarker MH4 and<br>quantifies the number of MH4 detections over a 4-minute interval, which it<br>compares to a predefined threshold to determine the presence of abnormal | | |
{4}------------------------------------------------
| | digestive health in patients following intestinal surgery, such as gastrointestinal<br>impairment (GII). |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use/Intended<br>Use | PrevisEA is a compact, non-invasive device placed on the abdomen to<br>capture the digestive sounds of patients. PrevisEA displays information<br>which assists physicians in assessing the digestive health of patients. This<br>device is for prescription use only and should be used under the direction<br>of a licensed healthcare practitioner. The PrevisEA device has not been<br>tested for and it is not intended for pediatric use. |
{5}------------------------------------------------
| | Subject Device | Predicate Device | | |
|----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|--|
| | Entac PrevisEA Device | GI Logic Abstats Gateway | Comparison | |
| 510(k) Number | K211068 | K150782 | Same | |
| Product Code | DQD | DQD | Same | |
| Regulation | 21 CFR 870.1875 Stethoscope | 21 CFR 870.1875 Stethoscope | Same | |
| Intended Use | PrevisEA is a compact, non-<br>invasive device placed on the<br>abdomen to capture the digestive<br>sounds of patients. PrevisEA<br>displays information which assists<br>physicians in assessing the<br>digestive health of patients.<br>This device should be used under<br>the direction of a licensed<br>healthcare practitioner. The<br>PrevisEA device has not been<br>tested for and it is not intended for<br>pediatric use. | The AbStats Gateway is a compact<br>device with integrated sensor<br>interfaces and embedded<br>computing system (stethoscope)<br>that in conjunction with external<br>sensors constitutes the AbStats<br>system. The AbStats system<br>external sensors are placed on the<br>abdomen and identify the vibratory<br>signals associated with digestive<br>processes. This device should be<br>used under the direction of a<br>licensed healthcare practitioner<br>when it is required to determine<br>this patient digestive state. The<br>device has not been tested for and<br>it is not intended for pediatric use. | Substantially equivalent | |
| Number of Sensors | 1 - self contained | 2 - external | Different number and location but<br>substantially equivalent<br>functionality | |
| Sensor Size (mm) Active Area | 3 inches | 30 mm x 20 mm<br>400 mm2 | Different size but substantially<br>equivalent functionality | |
| | Subject Device | Reference Device | Comparison | |
| | Entac PrevisEA Device | GI Logic Abstats Gateway | | |
| Algorithm | The PrevisEA utilizes<br>algorithm technology to<br>specifically evaluate<br>gastrointestinal auditory<br>biomarker MH4 | The Abstats Gateway utilizes algorithm<br>technology to recognize and classify<br>gastrointestinal signals | Different but Subject and<br>Reference devices have similar<br>functionality | |
| Sensor Technology | Standard Electrical<br>Microphone | Standard Electrical Microphone | Substantially equivalent | |
| System Technology | Sensor data signal processing<br>and computing | Sensor data signal processing<br>and computing | Substantially equivalent | |
| Interface to PC | None required (no PC<br>Interface) | None required (no PC Interface) | Substantially equivalent | |
| Power Supply | Lithium-Ion Battery | 5V USB Power Source | Different but substantially<br>equivalent functionality | |
| Sensor Cable Length (m) | N/A – none required | 2.0 m | Different but substantially<br>equivalent functionality | |
| IEC60601 1-1:2005 3rd Edition<br>Medical Electrical Equipment Part<br>1: General Requirement or Safety | Conforms | Conforms | Substantially equivalent | |
| EN60601-1-2, 2007/03, EMC<br>Requirements for Safety, 2.<br>Collateral Standard –<br>Electromagnetic Compatibility<br>Requirements and Tests | Conforms | Conforms | Substantially Equivalent | |
| Immunity Requirements for Medical<br>Electrical Equipment Part 1:<br>General | Conforms | Conforms | Substantially Equivalent | |
| | Subject Device | Reference Device | Comparison | |
| | Entac PrevisEA Device | GI Logic Abstats Gateway | | |
| Biocompatibility | Testing demonstrated all<br>patient contacting materials<br>for the PrevisEA device to be<br>safe and biocompatible | No biocompatibility testing performed as the<br>only patient contact consists of commercially<br>available dermal adhesive dressings | Substantially Equivalent | |
| Software | The PrevisEA software<br>processes audio signals for<br>display and presentation to<br>the user. | The AbStats software processes<br>audio signals for display<br>and presentation to the user. | Substantially Equivalent | |
| Auscultation Performance Tests:<br>Bench Testing | Passed/Conforms | Passed/Conforms | Substantially Equivalent | |
| Clinical performance validation | Listens for the MH4 acoustic<br>biomarker | Listens for acoustic event rates to determine<br>an intestinal motility | Different mechanism/parameter<br>of assessment but does not raise<br>new issues of safety and efficacy. | |
| Abdomen Placement | The PrevisEA is placed in<br>any quadrant of the patient's<br>abdomen.<br>Image: PrevisEA device on a human torso | AbStats external sensors placed on the<br>abdomen,<br>Image: AbStats external sensors placed on a human abdomen | Substantially Equivalent | |
# Table 1: Substantial Equivalence Comparison Table
{6}------------------------------------------------
# Table 1: Substantial Equivalence Comparison Table (Continued)
{7}------------------------------------------------
# Table 1: Substantial Equivalence Comparison Table (Continued)
{8}------------------------------------------------
| Characteristics | PrevisEA | Abstats Gateway |
|-------------------------------|----------|-----------------|
| Electronic Stethoscope | Yes | Yes |
| Non-invasive system | Yes | Yes |
| Attach to abdomen | Yes | Yes |
| Auscultation device | Yes | Yes |
| Assess digestion specifically | Yes | Yes |
| Algorithm technology | Yes | Yes |
| Self-contained sensors | Yes | No |
| Batter powered | Yes | No |
### Table 2: Summary of Technological Characteristics Compared to the Predicate Device
### 5.3 Biocompatibility Testing
Biocompatibility tests were conducted with the PrevisEA Device according to ISO 10993-1:2018 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. These studies demonstrated the PrevisEA Device is biocompatible.
#### 5.4 Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical product safety, electromagnetic compatibility, and battery safety testing were performed for the PrevisEA Device. All testing met acceptance criteria and demonstrated the PrevisEA Device to be safe and electromagnetically compatible.
### ર્સ્ડ Clinical Testing - PrevisEA Optimization Trial
Clinical data was submitted in support of this 510k Premarket Notification. Clinical testing demonstrated the PrevisEA Device accomplishes its intended use. Trial Registration: ClinicalTrials.gov NCT03505476
75 subjects undergoing intestinal resection surgery had PrevisEA applied at the completion of surgery.
This study demonstrated the capability of the PrevisEA Device in listening/capturing the acoustic biomarker, MH4 while in clinical use (attached to patient abdomens).
#### 5.6 Conclusion
Since the comparison of bench testing to clinical outcomes is still not well understood for this type of device, clinical testing was required to support substantial equivalence.
{9}------------------------------------------------
The clinical data submitted validates the intended use of the PrevisEA Device. The nonclinical data support the safety and effectiveness of the device, and the bench demonstrates the device meets its own specifications. The hardware and software verification and validation demonstrate that the PrevisEA device performs as intended in the specified use conditions. Though there are minor differences in technological characteristics, the data provided in this submission and the comparisons demonstrate these differences do not raise new questions of safety and efficacy. It is the conclusion of the sponsor that the subject device is substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.