Straumann® RidgeFit Implants Ø2.4 mm are for oral endosteal implantation in the upper and/or lower jaw of fully edentulous patients. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved for all implants. Straumann® RidgeFit Implants are intended for the stabilization of removable dentures. - For mandibular restorations, at least 4 Straumann® RidgeFit Implants Ø2.4 mm should be placed. - For maxillary restorations, at least 6 Straumann® RidgeFit Implants ø2.4 mm should be placed.
Device Story
Tapered endosseous dental implants; manufactured from Roxolid (TiZr) with SLA surface; integrated Optiloc attachment portion with Titanium Nitride (TiN) coating. Used for stabilization of removable dentures in fully edentulous patients. Placed in upper/lower jaw by dental clinicians. Provides retention for dentures via mechanical attachment. Supports immediate or conventional loading based on primary stability. Benefits patients by stabilizing removable prosthetics.
Clinical Evidence
No new clinical testing performed. Substantial equivalence supported by literature review (Lemos et al., 2017) regarding the use of mini implants for overdenture retention.
Technological Characteristics
Material: Roxolid (TiZr). Surface: SLA. Coating: TiN. Dimensions: 2.4 mm diameter; 10-14 mm lengths. Sterilization: Gamma irradiation (ISO 11137-1:2006, ISO 11137-2:2013). Mechanical: Anchor ball attachment. MR Conditional.
Indications for Use
Indicated for oral endosteal implantation in the upper and/or lower jaw of fully edentulous patients for the stabilization of removable dentures. Requires at least 4 implants for mandibular restorations and at least 6 implants for maxillary restorations.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 08, 2021
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
### Re: K211052
Trade/Device Name: Straumann RidgeFit Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: April 8, 2021 Received: April 9, 2021
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K211052
Device Name
Straumann® RidgeFit Implants
Indications for Use (Describe)
Straumann® RidgeFit Implants 02.4 mm are for oral endosteal implantation in the upper and/or lower jaw of fully edentulous patients. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved for all implants. Straumann® RidgeFit Implants are intended for the stabilization of removable dentures.
- For mandibular restorations, at least 4 Straumann® RidgeFit Implants ø2.4 mm should be placed.
- For maxillary restorations, at least 6 Straumann® RidgeFit Implants Ø2.4 mm should be placed. ●
Type of Use (Select one or both, as applicable) 図Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
## Straumann® RidgeFit Implants
510(k) Summary
# 510(k) Summary
### Submitter's Contact Information
| Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 9005052 |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs and Quality<br>Phone Number: +1 978 747-2509<br>Fax Number: +1 978 747-0023 |
| Prepared By &<br>Alternate Contact: | Chanrasmey White, MS<br>Regulatory Affairs Specialist<br>Straumann, USA, LLC |
| Date of Submission: | July 8, 2021 |
### Name of the Device
| Trade Names: | Straumann® RidgeFit Implants |
|------------------------|---------------------------------------------|
| Common Name: | Endosseous dental implant (21 CFR 872.3640) |
| Classification Name: | Endosseous dental implant (21 CFR 872.3640) |
| Regulation Number: | §872.3640 |
| Device Classification: | II |
| Product Code(s): | DZE |
| Classification Panel: | Dental |
### Predicate Device(s)
Primary Predicate:
- K191895 Straumann® Mini Implants .
{4}------------------------------------------------
### Straumann® RidgeFit Implants
510(k) Summary
### Device Description
The Straumann® RidgeFit Implants are tapered implants manufactured from Roxolid (TiZr) with a finished SLA surface. The implant neck is machined with an Optiloc® attachment portion resulting in a one-piece implant system acting as a retention feature for dentures, this feature is coated in a Titanium Nitride (TiN) coating. The Straumann® RidgeFit Implants have an external diameter of 2.4 mm and are available in implant lengths 10 mm, 12 mm and 14 mm with a gingival height of 2.8 mm and implant lengths 10 mm and 12 mm for gingival heights 3.8 mm and 4.8 mm. The Straumann® RidgeFit Implants are identical to the Straumann® Mini Implants cleared under K191895 apart from proposed labeling changes. Note for US Markets, Straumann has identified Straumann® Mini Implants as Straumann® RidgeFit Implants to distinguish it from other markets, this change has been documented through Memo to File. The purpose of this traditional 510(k) is to propose labeling changes to the cleared indications and the addition of contraindications to the existing devices cleared under K191895.
#### Intended Use
The Straumann® RidgeFit Implants are intended for the stabilization of removable dentures. The removable dentures are connected to the RidgeFit Implants through the incorporated Optiloc® attachment element.
#### Indications for Use
Straumann® RidgeFit Implants Ø2.4 mm are for oral endosteal implantation in the upper and/or lower jaw of fully edentulous patients. The implants can be placed with immediate function when primary stability is achieved for all implants or with conventional loading if primary stability is not achieved for all implants. Straumann® RidgeFit Implants are intended for the stabilization of removable dentures.
- For mandibular restorations, at least 4 Straumann® RidgeFit Implants Ø2.4 mm should . be placed.
- For maxillary restorations, at least 6 Straumann® RidgeFit Implants ø2.4 mm should be . placed.
{5}------------------------------------------------
# Straumann® RidgeFit Implants
510(k) Summary
### Technological Characteristics
The Straumann® RidgeFit Implants are identical to the Straumann® Mini Implants cleared under K191895 apart from proposed labeling changes under this submission. The purpose of this traditional 510(k) is to propose changes to the cleared indications and to add additional contraindications to the devices cleared under K191895. The technological characteristics of the subject devices are compared to the primary predicate devices in Table 1.
| FEATURE | SUBJECT DEVICES | PRIMARY PREDICATE DEVICES |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K211052 | K191895 |
| Indications for<br>Use | Straumann® RidgeFit Implants Ø2.4 mm are<br>for oral endosteal implantation in the upper<br>and/or lower jaw of fully edentulous patients.<br>The implants can be placed with immediate<br>function when primary stability is achieved for<br>all implants or with conventional loading if<br>primary stability is not achieved for all<br>implants. Straumann® RidgeFit Implants are<br>intended for the stabilization of removable<br>dentures.<br>· For mandibular restorations, at least 4<br>Straumann® RidgeFit Implants Ø2.4 mm<br>should be placed.<br>· For maxillary restorations, at least 6<br>Straumann® RidgeFit Implants Ø2.4 mm<br>should be placed. | Straumann® Mini Implants Ø2.4 mm are<br>suitable for oral endosteal implantation in the<br>upper and lower jaw of fully or partially<br>edentulous patients. The implants can be<br>placed with immediate function when good<br>primary stability is achieved.<br>Straumann® Mini Implants are intended for<br>the stabilization of removable dentures. |
| Implant<br>Diameter | 2.4 mm | 2.4 mm |
| Implant Length | 10, 12, 14 mm | 10, 12, 14 mm |
| Gingival heights | 2.8 mm | 2.8, 3.8 and 4.8 mm |
| Material | Roxolid (TiZr) | Roxolid (TiZr) |
| Coating | TiN coated | TiN coated |
| Surface<br>Treatment | SLA | SLA |
| Abutment-to-<br>restoration<br>connection | Anchor Ball | Anchor Ball |
| Recommended<br>restoration | Stabilization of removable dentures | Stabilization of removable dentures |
| Sterilization<br>Method | Gamma Irradiation | Gamma Irradiation |
Table 1 - Comparison of the Subject and the Primary Predicate Devices
{6}------------------------------------------------
## Straumann® RidgeFit Implants
510(k) Summary
### Performance Testing
The following performance data support the substantial equivalence determination.
### Sterilization Validation and Shelf Life
The Straumann® RidgeFit Implants are single patient devices, are provided sterile via gamma irradiation at a dose of 25 kilogray (2.5 Mrad) Minimum and will be sterilized after final packaging. A sterility assurance level (SAL) of 10° has been validated in accordance with ISO 11137-1:2006, Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. The validation method used was the VDmax 25method in accordance with ISO 11137-2:2013, Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose. The sterilization validation and shelf life data have been submitted under K191895. The shelf life of the subject RidgeFit Implants is 5 years. Sterilization method and shelf life are identical to the primary predicate devices cleared under K191895.
#### Biocompatibility Testing
There are no changes to the materials, surface treatments, or manufacturing methods from the currently marketed predicate devices. Therefore, no new issues of biocompatibility are raised for the subject devices and no additional biocompatibility testing was required.
#### Electromagnetic Compatibility
There are no changes to the materials, surface treatments, or manufacturing methods from the currently marketed predicate devices. Therefore, no new issues of electromagnetic compatibility are raised for the subject devices and they can be considered MR Conditional.
#### Performance Testing - Bench
The RidgeFit (Mini) Implants with 2.8 mm of gingival height have been cleared under K191895. The gingival heights of 3.8 and 4.8 mm for the lengths 10 and 12mm have been introduced via Memo to File. A fatigue test on the new 4.8 mm gingival height has demonstrated a performance equivalent to the primary predicate device of the K191895.
Relevant bench testing, including insertion torque, wear, and dynamic fatigue testing, was performed and provided in K191895 in accordance with the FDA guidance document Guidance
{7}------------------------------------------------
## Straumann® RidgeFit Implants
510(k) Summary
for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments, Document issued on: May 12, 2004.
This submission only intends to propose changes in the labeling of the devices. The subject devices are identical to the previously cleared devices (primary predicate and Memo to File), therefore, no additional performance testing is required.
### Performance Testing - Clinical
The change in the Indications for Use to specifically indicate the RidgeFit (Mini) Implants for a minimum of 4 implants in the mandible and a minimum of 6 implants in the maxilla is supported by a clinical summary article by Lemos et al.1
### Conclusion
The documentation submitted in this premarket notification demonstrates the Straumann® RidgeFit Implants are substantially equivalent to the primary predicate.
<sup>1</sup> Lemos CA, Verri FR, Batista VE, Júnior JF, Mello CC, Pellizzer EP. Complete overdentures retained by mini implants: A systematic review. J Dent. 2017 Feb; 57:4-13. Doi: 10.1016/i,ident.2016.11.009. Epub 2016 Nov 22.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.