The PREFERX Delivery System is intended to be used for the delivery of hydrated allograft, or synthetic bone graft material to an orthopedic surgical site.
Device Story
Manual orthopedic surgical instrument; delivers hydrated allograft, autograft, or synthetic bone graft material. System components: feeder (body, stand, funnel), reduction sleeve, biologic cartridge, plunger, and ratcheting gun. User fills biologic cartridge with graft material, attaches to gun/plunger assembly; trigger mechanism advances plunger to express graft from cartridge tip directly into surgical site. Single-use, disposable, pre-sterilized. Used in operating room by surgeons. Enables precise graft placement; cartridge can be removed and refilled during procedure if additional volume is required.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility, burst/functionality testing, and sterilization validation.
Indicated for delivery of hydrated allograft, autograft, or synthetic bone graft material to orthopedic surgical sites.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Graftgun Universal Graft Delivery System (K170675)
Reference Devices
Graftgun Universal Graft Delivery System (K180937)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
September 7, 2021
MD3 LLC Mary Ann Greenawalt Head of Quality 3650 Coral Ridge Drive. Suite 107 Coral Springs, Florida 33065
Re: K210997
Trade/Device Name: PREFERX Delivery System Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: July 29, 2021 Received: July 30, 2021
Dear Mary Ann Greenawalt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K210997
Device Name
PREFERX Delivery System
Indications for Use (Describe)
The PREFERX Delivery System is intended to be used for the delivery of hydrated allograft, or synthetic bone graft material to an orthopedic surgical site.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary: PREFERX™ Delivery System
In accordance with 21 CFR 807.92 of the Federal Code of Regulations
| Date Prepared | July 29, 2021 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By | MD3 LLC<br>3650 Coral Ridge Drive Ste 107<br>Coral Springs, FL 33065<br>1.561.703.2311 |
| Primary Contact | Mary Ann Greenawalt<br>3650 Coral Ridge Drive Ste 107<br>Coral Springs, FL 33065<br>1.503.318.7722<br>maryann@md3inc.com |
| Trade Name | PREFERX™ Delivery System |
| Common Name | Graft Delivery Device |
| Classification Name | Syringe, Piston |
| Class | II |
| Product Code | FMF |
| CFR Section | 21 CFR section 880.5860 |
| Device Panel | General and Plastic Surgery |
| Primary Predicate Device | K170675 Graftgun Universal Graft Delivery System, SurGenTec, LLC |
| Reference Device | K180937 Graftgun Universal Graft Delivery System, Surgentec, LLC |
| Device Description | The PREFERX Delivery System is a single use pre-sterilized and disposable system consisting of<br>a Feeder including a CC Body, Stand, and Funnel; a Reduction Sleeve; a BC-Sleeve and BC-Sleeve<br>Cap, used to load the graft material into the Biologic Cartridge. The Biologic Cartridge contains and<br>delivers the graft material to the surgical site. A Plunger expresses the graft material from the<br>Biologic Cartridge through an actuating trigger on the Gun that advances the Plunger down the<br>length of the Biologic Cartridge via a ratcheting mechanism. The system is designed such that the<br>Biologic Cartridge can be filled with the desired graft material, attached to the Gun and Plunger for use, then removed and refilled during the same procedure.<br><br>The Biologic Cartridge does not have a Luer lock mechanism; the device does not require a needle<br>or similar attachment; the Biologic Cartridge contents being dispensed directly from the tip of the<br>Biologic Cartridge into the graft site.<br><br>The device is packaged in a thermoformed tray with a Tyvek lid. Each tray is then packaged individually in an outer paperboard carton.<br><br>The PREFERX Delivery System is a single use pre-sterilized, disposable system intended to be used for the delivery of hydrated allograft, autograft, or synthetic bone graft material to an orthopedic surgical site. |
{4}------------------------------------------------
| Materials | Clear Polycarbonate - ASTM F997-18<br>ABS GA850 (Acrylonitrile Butadiene Styrene) - USP Class VI<br>Stainless steel - ASTM F899-12b | | | |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Apply bone graft to an orthopedic surgical site in an operating room environment. | | | |
| Substantial<br>Equivalence Claimed<br>to Predicate Devices | The PREFERX Delivery System is substantially equivalent to the predicate devices in terms of<br>intended use, design, and materials used. | | | |
| Indications for Use | The PREFERX Delivery System is intended to be used for the delivery of hydrated allograft,<br>or synthetic bone graft material to an orthopedic surgical site. | | | |
| | GraftGun<br>Universal Graft<br>Delivery<br>(predicate) | PREFERX<br>Delivery System<br>(proposed) | Same/Similar/Different<br>as predicate | |
| | Sponsor/Submitter | SurGenTec | MD3 LLC | N/A |
| | FDA Regulation | 21 CFR 880.5860<br>Piston Syringe | 21 CFR 880.5860<br>Piston Syringe | Same |
| | FDA Product Code | FMF | FMF | Same |
| | FDA Classification | Class II | Class II | Same |
| | K Number | K170675 | TBD | N/A |
| Summary of the<br>technological<br>characteristics<br>compared to<br>predicate | Indications for Use | The GraftGun<br>Delivery System<br>is intended to be<br>used for the<br>delivery of<br>hydrated<br>allograft,<br>autograft, or<br>synthetic bone<br>graft material to<br>an orthopedic<br>surgical<br>site. | The PREFERX<br>Delivery System is<br>intended to be used<br>for the delivery of<br>hydrated allograft,<br>autograft, or<br>synthetic bone graft<br>material to an<br>orthopedic surgical<br>site. | Same |
| | Single Use | Yes | Yes | Same |
| | Sterilization Method | Gamma<br>irradiation to<br>SAL of 10-6 | Gamma irradiation<br>to SAL of 10-6 | Same |
| | Patient Contact<br>Material (transient use<br>in surgical suite) | Medical Grade:<br>• Polycarbonate<br>• ABS<br>• Polypropylene<br>• Stainless Steel<br>- 316L, 316F,<br>304H, 304HC<br>• Radiopaque ring | Medical Grade:<br>• Polycarbonate<br>• ABS | The proposed device is<br>comprised of fewer patient<br>contacting materials than the<br>predicate device. The<br>absence of patient contacting<br>material: polypropylene and a<br>stainless-steel radiopaque<br>ring is not a significant<br>change in materials or design<br>and does not raise questions<br>of safety and effectiveness.<br>All component materials have<br>a long history of use in<br>medical devices. Both<br>products use ABS &<br>polycarbonate. |
| | Volume | Up to 7.5cc in<br>graft tube | Up to 5.0cc in graft<br>tube | If more graft is needed the<br>PREFERX cartridge may be<br>reloaded during the same<br>procedure. |
| | Operating Principle | Graft material<br>expressed from<br>graft<br>tube via a plunger,<br>operated by a<br>ratchet-actuated<br>handle. Material<br>expressed from<br>graft tube to graft<br>site. | Graft material<br>expressed from<br>graft tube via a<br>plunger, operated by<br>a ratchet-actuated<br>handle. Material<br>expressed from graft<br>tube to graft site. | Same |
| | Performance data, including worst-case scenario, was provided in support of the substantial<br>equivalence determinations. To further support substantial equivalence, we used the FMEA method<br>to identify risks and characterize the severity and probability. The risks were reduced and are ranked<br>as Acceptable. MD3 design control procedures, design verification and validation testing of the device<br>were performed based on the results of the risk analysis. The following summarizes the identified<br>risks and the applicable testing that was performed.<br>Summary of Identified Risk and Verification Testing | | | |
| | Potential Effect of Failure | Potential Cause | | Action Taken |
| Non-clinical Test<br>Summary | Inflammation, redness and<br>swelling, sometimes<br>accompanied by heat and pain | Incompatible materials | | Biocompatibility Testing<br>Certification included in DHR |
| | Graft is not applied to surgical<br>site; graft must be applied<br>another way - prolonged<br>surgery | Gun is not properly<br>assembled<br>Biologic Cartridge is plugged | | Perform visual/functionality<br>test at incoming inspection<br>Perform burst and<br>functionality test |
| | Fusion of bone may not occur,<br>or heterotopic bone may form | Dimensions of cartridge are<br>not correct resulting in<br>incorrect volume of graft<br>Volume markings are not<br>applied accurately<br>Teeth on plunger too small | | Perform visual/functionality<br>test at incoming inspection<br>Records of compliance for<br>dimensional certifications |
| | Patient becomes infected | No or inadequate sterilization<br>method or cycle | Certificate of Sterilization<br>included in DHR<br>Sterilization validation<br>Ink Test | |
| | Chills to fever to swelling or<br>sepsis | Endotoxins remain after<br>sterilization | Bacterial endotoxin test | |
| | Delivery system is plugged<br>from contents - prolonged<br>surgery | Wrong material placed in<br>cartridge | The device has a fail-safe<br>mechanism to avert breakage<br>of cartridge. | |
| Clinical Test Summary | No clinical studies were performed | | | |
| Conclusion: | MD3 LLC considers the PREFERX Delivery System to be as safe and effective and perform at least<br>as well as the legally marketed predicate device when used as intended. The systems, predicates<br>and proposed, are designed so that the Biologic Cartridge can be filled with the desired graft material,<br>attached to the Gun and the Plunger, and delivered to the orthopedic surgical site. Substantial<br>equivalence to the predicate device is based upon the similarities in design, principles of operation,<br>technology, materials and indications for use. | | | |
{5}------------------------------------------------
{6}------------------------------------------------
-end-
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.