K210983 · Becton, Dickinson and Company · CAZ · May 19, 2022 · Anesthesiology
Device Facts
Record ID
K210983
Device Name
BD Epilor Syringe
Applicant
Becton, Dickinson and Company
Product Code
CAZ · Anesthesiology
Decision Date
May 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
BD Epilor™ syringes are intended for use with either air or liquid in conjunction with an epidural needle for identifying the epidural space. These types of syringes facilitate the "loss of resistance" technique for identifying the epidural space by reducing subjectivity when locating this space and the potential for complications when administering epidural anesthesia to patients. Not for spinal applications. These devices are intended for adult and pediatric patients.
Device Story
Manual syringe used by anesthesiologists to identify epidural space; operates via 'loss of resistance' technique; clinician attaches syringe filled with air or saline to epidural needle; advancement of needle tip into epidural space results in sudden decrease in resistance to plunger movement; provides tactile feedback to clinician; used in clinical settings for epidural anesthesia procedures; benefits patient by facilitating accurate needle placement and reducing potential complications.
Clinical Evidence
No clinical data. Bench testing only. Verification included ISO 594-1/2 mechanical testing (separation force, torque, leakage, stress cracking), biocompatibility (cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, hemolysis), and particulate matter testing per USP <788>.
Technological Characteristics
Manual 7mL syringe; materials: polypropylene barrel/plunger, silicone stopper/lubricant; Luer-Lok or Luer Slip tip per ISO 594-1:1986 and ISO 594-2:1998; ethylene oxide sterilization (SAL 10^-6).
Indications for Use
Indicated for adult and pediatric patients requiring epidural anesthesia; used with air or liquid and an epidural needle to identify the epidural space via the loss of resistance technique. Contraindicated for spinal applications.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
May 19, 2022
Becton Dickinson and Company Huwien Yang Senior Regulatory Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K210983
Trade/Device Name: BD Epilor Syringe Regulation Number: 21 CFR 868.5140 Regulation Name: Anesthesia conduction kit Regulatory Class: Class II Product Code: CAZ Dated: April 15, 2022 Received: April 18, 2022
Dear Huwien Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210983
Device Name BD Epilor™ Syringe
#### Indications for Use (Describe)
BD Epilor™ syringes are intended for use with either air or liquid in conjunction with an epidural needle for identifying the epidural space. These types of syringes facilitate the "loss of resistance" technique for identifying the epidural space by reducing subjectivity when locating this space and the potential for complications when administering epidural anesthesia to patients. Not for spinal applications. These devices are intended for adult and pediatric patients.
| Type of Use (Select one or both, as applicable) | | | |
|-------------------------------------------------|---------------------------------------------|--|--|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K210983
Image /page/3/Picture/2 description: The image shows the logo for BD, a global medical technology company. The logo consists of an orange circular symbol with radiating lines, resembling a sun or starburst, followed by the letters "BD" in blue. The letters are bold and capitalized, with a sans-serif font, and are positioned to the right of the circular symbol.
Submitter
Information
Premarket Notification - Traditional Section 7 – 510K Summary
# 510(k) Summary (21 CFR §807.92) BD Epilor™ Syringe
| Submitter Name: | Becton, Dickinson and Company |
|--------------------|--------------------------------------|
| Submitter Address: | 1 Becton Drive |
| | Franklin Lakes, NJ 07417 |
| Contact Person: | Huiwen Yang |
| | Senior Regulatory Affairs Specialist |
| Email Address: | Huiwen.yang@bd.com |
| Phone Number: | Phone: (201) 847-4408 |
| | Fax: (201) 847-5397 |
| | Email Address: | Huiwen.yang@bd.com |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| | Phone Number: | Phone: (201) 847-4408 |
| | Fax Number: | Fax: (201) 847-5307 |
| | Date of<br>Preparation: | April 15, 2022 |
| Subject Device | Trade Name: | BD Epilor™ Syringe |
| | Common Name: | Loss of Resistance Syringe |
| | Regulation<br>Number: | 21 CFR §868.5140 |
| | Regulation Name: | Anesthesia Conduction Kit |
| | Regulatory Class: | Class II device |
| | Product Code: | CAZ (Anesthesia Conduction Kit) |
| | Classification Panel: | Anesthesiology |
| Predicate<br>Device | Trade Name: | B-D Loss of Resistance Syringe |
| | 510(k) Reference: | K925902 |
| | Common Name: | Loss of Resistance Syringe |
| | Regulation<br>Number: | 21 CFR §868.5140 |
| | Regulation Name: | Anesthesia Conduction Kit |
| | Regulatory Class: | Class II device |
| | Product Code: | CAZ (Anesthesia Conduction Kit) |
| | Classification Panel: | Anesthesiology |
| Reason For<br>Change | The purpose of the submission is to re-baseline the Predicate<br>device B-D Loss of Resistance Syringe. | |
| Device<br>Description | The subject BD Epilor™ (Loss of Resistance Syringe) is used in<br>various types of epidural anesthesia procedures. The purpose of<br>this syringe is to help the anesthesiologist locate the epidural<br>space prior to administering either single shot or continuous<br>epidural anesthesia. The syringe assembly consists of a<br>lubricated polypropylene barrel imprinted with a graduated<br>scale, a polypropylene plunger rod, and a silicone rubber<br>stopper which is a double-ribbed to prevent leakage. BD | |
{4}------------------------------------------------
Premarket Notification - Traditional Section 7 – 510K Summary
| | Epilor ™ is used in conjunction with an epidural needle for the<br>purpose of identifying the epidural space. BD Epilor™ syringes<br>are available in 7ml sizes and supplied in plastic configurations<br>and as sterile or Bulk Non-Sterile. |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Indications for Use</b> | BD Epilor™ syringes are intended for use with either air or<br>liquid in conjunction with an epidural needle for identifying the<br>epidural space. These types of syringes facilitate the “loss of<br>resistance” technique for identifying the epidural space by<br>reducing subjectivity when locating this space and the potential<br>for complications when administering epidural anesthesia to<br>patients. These devices are intended for adult and pediatric<br>patients. |
## Technological Characteristics
The subject devices are equivalent to the predicate devices in intended use, materials and performance characteristics:
| Element of<br>Comparison | Subject Device | Predicate Device | Comparison |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| 510K # | K210983 | K925902 | Not applicable |
| Indications for<br>Use/Intended<br>Use | BD Epilor™ syringes<br>are intended for use<br>with either air or liquid<br>in conjunction with an<br>epidural needle for<br>identifying the epidural<br>space. These types of<br>syringes facilitate the<br>"loss of resistance"<br>technique for<br>identifying the epidural<br>space by reducing<br>subjectivity when<br>locating this space and<br>the potential for<br>complications when<br>administering epidural<br>anesthesia to patients.<br>Not for spinal<br>application, | B-D Loss of Resistance<br>Syringe is used in<br>combination with an<br>Epidural Needle in the<br>first phase of the<br>Epidural Anesthesia<br>Procedure. The "loss of<br>resistance" technique in<br>epidural anesthesia is<br>used for identifying the<br>epidural space prior to<br>administration of<br>medication or<br>placement of an<br>Epidural Catheter. By<br>attaching a LOR<br>Syringe filled with air or<br>saline to the epidural<br>needle the clinician can<br>identify arrival of the<br>needle tip in the<br>epidural space by a<br>dramatic Loss of<br>Resistance to syringe<br>plunger movement. | Equivalent,<br>the<br>indications for<br>use/intended<br>use has been<br>modified to<br>provide better<br>clarity. |
{5}------------------------------------------------
Premarket Notification - Traditional
| | Barrel | Polypropylene | Polypropylene | same |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Syringe materials | Barrel<br>Lubricant | Silicone | Silicone | same |
| | Plunger<br>Rod | Polypropylene+Colorant<br>(Blue) | Polypropylene+Colorant<br>(Blue) | same |
| | Stopper | Self- lubricated silicone | Self-lubricated silicone | same |
| | Stopper<br>Lubricant | Silicone | Silicone | same |
| | Barrel ink | Black Ink | Black Ink | same |
| | Syringe Type | 3 Pieces (barrel,<br>stopper and plunger) | 3 Pieces (barrel,<br>stopper and plunger) | same |
| Tip type | | Luer-Lok™ or Luer Slip<br>per ISO 594-1: 1986<br>and ISO 594-2:1998<br>7mL | Luer-Lok™ or Luer Slip<br>per ISO 594-1: 1986<br>and ISO 594-2:1998<br>3mL, 5mL, 10mL ,<br>20mL | Equivalent;<br>Only 7mL BD<br>Epilor™<br>syringes are<br>offered |
| Dose<br>Setting/Volumes | | | | |
| | Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | same |
| | SAL | 10-6 | 10-6 | same |
| Shelf Life | | 5 Years | 5 Years | same |
| | | | | |
| | Functional Tests | | | |
| | Ink Permanency | Per BD internal<br>requirements | Per BD internal<br>requirements | same |
| Luer-Lok<br>Separation<br>Force | | Meets ISO 594-2<br>requirement; | Meets ISO 594-2<br>requirement; | same |
| Luer Slip<br>Separation force | | Meets ISO 594-1<br>requirement | Meets ISO 594-1<br>requirement | same |
| | Unscrewing<br>Torque(Luer-<br>Lok) | Meets ISO 594-2<br>requirement | Meets ISO 594-2<br>requirement | same |
| | Resistance to<br>Overriding<br>(Luer-Lok) | Meets ISO 594-2<br>requirement | Meets ISO 594-2<br>requirement | same |
| Luer Leakage -<br>Positive<br>Pressure Decay<br>(Luer-Lok and | | Meets ISO 594-1 and<br>594-2 requirement | Meets ISO 594-1 and<br>594-2 requirement | same |
| Luer-Slip ) | | | | |
| Sub-<br>Atmospheric<br>Pressure Air<br>Leakage (Luer-<br>Lok and Luer-<br>Slip ) | Meets ISO 594-1 and<br>594-2 requirement | Meets ISO 594-1 and<br>594-2 requirement | same | |
| Stress Cracking | Meets ISO 594-1 and<br>594-2 requirement | Meets ISO 594-1 and<br>594-2 requirement | same | |
| Fit Test | Meet BD internal<br>requirements | Meet BD internal<br>requirements | same | |
| Stopper<br>Leakage | Meet BD internal<br>requirements | Meet BD internal<br>requirements | Same | |
| Biocompatibility tests | | | | |
| Cytotoxicity | Per ISO 10993-5, ISO<br>10993-12, & USP<br><87>:<br>Non-cytotoxic | Non-cytotoxic | Equivalent | |
| Sensitization | Per ISO 10993-10:<br>Non-sensitizer | Non-sensitizer | Equivalent | |
| Intracutaneous<br>Reactivity | Per AAMI ISO 10993-10<br>& USP<88>:<br>Non-irritant | Non-irritant | Equivalent | |
| Acute Systemic<br>Toxicity | Per ISO 10993-11 &<br>USP<88> :<br>Non-toxic | Non-toxic | Equivalent | |
| Material-<br>mediated<br>Pyrogenicity | Per ISO 10993-<br>11:2017 & USP<151>:<br>Non-pyrogenic | Non-pyrogenic | Equivalent | |
| Extractables/<br>Leachables | Per ISO 10993-18:<br>Acceptable | Heavy metal tests;<br>Pass | Equivalent | |
| Hemolysis | Per ISO 10993-4, ASTM<br>Guideline F619-14,<br>ASTM Guideline F756-<br>17:Non-hemolytic | Non-hemolytic | Equivalent | |
| Additional Tests:<br>Particulate<br>Matter | Per USP 788:<br>Particulate number is<br>under the limit. | Below the limit | Equivalent | |
| Performance<br>Tests | BD has performed the following non-clinical/design verification<br>testing based on the risk analysis conducted and the results of<br>these tests demonstrate that the BD EpilorTM performed in an | | | |
{6}------------------------------------------------
# Premarket Notification - Traditional
{7}------------------------------------------------
Premarket Notification - Traditional Section 7 – 510K Summary
### equivalent manner to the predicate device.
Per ISO 594-1 and 594-2:
- LL Separation Force ●
- Unscrewing Torque ●
- Overriding Torque ●
- Luer Leakage – Positive Pressure Decay
- Sub-Atmospheric Pressure Air Leakage ●
- Stress Cracking ●
Per BD internal requirements:
- Ink Permanency ●
- Fit Test
- Stopper Leakage ●
A biocompatibility evaluation was conducted on the subject device per ISO 10993-1:2018, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within a Risk Management Process. Based on the evaluation, the following biological tests were conducted:
- Cytotoxicity ●
- Sensitization ●
- Intracutaneous Reactivity ●
- Acute Systemic Toxicity
- Material- mediated pyrogenicity ●
- Hemolysis
- Chemical Extractables Analysis ●
Additionally, the following tests were performed:
- Particulate Matter per USP <788> .
The device is sterilized using ethylene oxide process and was validated per ISO 11135.
Per the design control requirements specified in 21 CFR 820.30, the subject device met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate device.
## Clinical Testing
Clinical testing was not required for the subject device this submission.
{8}------------------------------------------------
Premarket Notification - Traditional
| Summary of<br>Substantial<br>Equivalence | Based on the intended use, technological characteristics and<br>performance testing, the subject device meets the<br>requirements that is considered sufficient for its intended use.<br>Therefore, BD EpilorTM syringe is substantially equivalent to the<br>predicate device in principles of operation, technology, design,<br>materials and performance. The indications for use/intended<br>use has been modified to provide better clarity. |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.