E-Dent 1000 is a light-curable resin indicated for the fabrication of: - individual and fixed permanent full single crowns, permanent partial crowns in front and posterior area, - individual and fixed single veneers, - artificial teeth for dental prostheses, which are used for removable permanent full dentures, - individual and removable monolithic full and partial dentures in dental laboratories. The material is an alternative dental material. E-Dent 1000 is intended exclusively for professional dental work. Fabrications with E-Dent 1000 requires a computer aided and manufacturing (CAD/CAM) system that includes the following components: digital dental files based on a digital impression, or in case of artificial teeth for dental prostheses the digital dental files based on manufacturer's data, a digital light processing (DLP) printer, and curing light equipment.
Device Story
E-Dent 1000 is a light-curable resin for additive manufacturing of dental prosthetics. Input consists of digital dental files (from impressions or manufacturer data) processed via CAD/CAM software. The system utilizes a Digital Light Processing (DLP) printer and curing light equipment to transform liquid resin into solid, customized dental devices (crowns, veneers, dentures). Used in dental laboratories by professional technicians. The resin is packaged with an RFID chip (TAG) to ensure compatibility with validated EnvisionTEC 3D printers (e.g., EnvisionOne cDLM, Micro, Vida, P4K, D4K series). Output is a physical dental restoration used by dentists to restore natural teeth or replace missing teeth, potentially improving patient oral function and aesthetics.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated against DIN EN ISO 10477:2018, DIN EN ISO 4049:2019-09, DIN EN ISO 7491, and DIN EN ISO 20795-1:2009-02. Biocompatibility testing conducted per ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity).
Technological Characteristics
Light-curable resin composed of acrylates, methacrylates, methacrylated oligomers/monomers, photo initiators, colorants, and absorbers. Form factor: 1 kg PE bottles with RFID TAG. Manufacturing: DLP 3D printing. Standards: DIN EN ISO 10477, 4049, 7491, 20795-1. Biocompatibility: ISO 10993 compliant.
Indications for Use
Indicated for professional dental laboratory fabrication of permanent full/partial crowns, single veneers, artificial teeth for removable permanent full dentures, and monolithic full/partial dentures. Intended for adult/pediatric patients requiring dental restoration.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
IvoBase CAD for Zenotec, IvoBase CAD Bond and Modelling Liquid (K151142)
Reference Devices
e-DENT Temporary Resin and Extra-Oral Curing System (K102776)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
May 25, 2021
EnvisionTEC GmbH Patsy Trisler Regulatory Consultant Oserve Group US, Inc. 7949 Beaumont Green East Drive Indianapolis, Indiana 46250
Re: K210977
Trade/Device Name: E-Dent 1000 Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF, EBI, ELM Dated: March 29, 2021 Received: April 1, 2021
Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210977
Device Name
E-Dent 1000
Indications for Use (Describe)
E-Dent 1000 is a light-curable resin indicated for the fabrication of:
- · individual and fixed permanent full single crowns, permanent partial crowns in front and posterior area,
- · individual and fixed single veneers,
- · artificial teeth for dental prostheses, which are used for removable permanent full dentures,
- · individual and removable monolithic full and partial dentures
in dental laboratories. The material is an alternative dental material. E-Dent 1000 is intended exclusively for professional dental work. Fabrications with E-Dent 1000 requires a computer aided and manufacturing (CAD/CAM) system that includes the following components: digital dental files based on a digital impression, or in case of artificial teeth for dental prostheses the digital dental files based on manufacturer's data, a digital light processing (DLP) printer, and curing light equipment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| I. SUBMITTER | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Submitter Name: | EnvisionTEC GmbH | | |
| Submitter Address: | Bruesseler Straße 51<br>D-45968 Gladbeck<br>Germany | | |
| Contact Person: | Ruediger van Bernum, Head of Application | | |
| Email: | ruediger.v.bernum@envisiontec.de | | |
| Telephone: | +49 2043 9875 45 | | |
| Date Prepared: | March 29, 2021 | | |
| II. DEVICE | | | |
| Trade Name: | E-Dent 1000 | | |
| Common Name | Tooth shade resin material,<br>Denture relining, repairing, rebasing resin,<br>Preformed plastic denture tooth | | |
| Classification<br>Numbers Names<br>Product Codes<br>Device Class | 21 CFR 872.3690<br>Material, Tooth<br>Shade, Resin<br>EBF<br>II | 21 CFR 872.3760<br>Resin, Denture,<br>Relining, Repairing,<br>Rebasing<br>EBI<br>II | 21 CFR 872.3590<br>Denture, Plastic,<br>Teeth<br>ELM<br>II, 510(k)-Exempt |
| III. PREDICATE DEVICE | | | |
| Predicate Devices | | | |
| Primary Predicate | K201668, VarseoSmile Crown Plus, BEGO Bremer Goldschlagerei<br>Wilh. Herbst GmbH & Co. KG (EBF product code) | | |
| Secondary<br>Predicate | K151142, IvoBase CAD for Zenotec, IvoBase CAD Bond and<br>Modelling Liquid, Ivoclar Vivadent, Inc. (EBI product code) | | |
| Reference<br>Device(s) | K102776: e-DENT Temporary Resin and Extra-Oral Curing System,<br>DeltaMed GmbH | | |
| IV. INDICATIONS FOR USE STATEMENT | | | |
| E-Dent 1000 is a light-curable resin indicated for the fabrication of:<br>• individual and fixed permanent full single crowns, permanent partial crowns in front<br>and posterior area. | | | |
- individual and fixed single veneers,
- . artificial teeth for dental prostheses, which are used for removable permanent full dentures,
- individual and removable monolithic full and partial dentures ●
in dental laboratories. The material is an alternative to traditional restorative dental material. E-Dent 1000 is intended exclusively for professional dental work. Fabrication of dental applications with E-Dent 1000 requires a computer aided and manufacturing (CAD/CAM) system that includes the following components: digital dental files based on a digital impression or in case of artificial teeth for dental prostheses the digital dental files
{4}------------------------------------------------
| based on manufacturer's data, a digital light processing (DLP) printer, and curing light<br>equipment. | |
|--------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| V. DEVICE DESCRIPTION | |
| Device<br>Identification | The E-Dent 1000 system combines a scanner with design software, the light-curable resin, a 3D printer and a curing unit. These components are used together during the manufacture of the customized E-Dent 1000 restorative dental products. |
| Technological<br>Characteristics | The light-curable resin is a proprietary composition of acrylates, methacrylates, methacrylated oligomers and monomers, photo initiators, colorants/dyes and absorbers. It is used by dental laboratories to make the customized temporary crowns and bridges for patients who need restoration of their natural teeth. |
| | E-Dent 1000 is available in six different colors. |
| | The resin is packaged in lightproof 1 kg PE bottles along with a programmed RFID chip (referred to as TAG), which is required for use with the validated 3D printers. The TAG contains information identifying the resin material, name and amount. |
| | E-Dent 1000 resin is an alternative material to heat-curable and auto-polymerizable resins. |
| | EnvisionTECs Perfactory® 3D-Printer DLP models designed and validated for use with the E-Dent 1000 light cured resin are:<br>• EnvisionOne cDLM, with LED<br>• Micro series, with LED<br>• Vida Series, with LED<br>• P4K Series, with LED<br>• D4K Series, with LED |
| VII PERFORMANCE AND SAFETY TESTING | |
| Animal Testing: | This product category does not require animal testing. |
| Clinical Testing: | This product category does not require human clinical testing. |
| Laboratory<br>Testing: | Testing was conducted to evaluate the performance of a manufactured crowns, artificial teeth, veneers and dentures, according to requirements of: |
| | • DIN EN ISO 10477:2018: Dentistry—Polymer-based crown and veneering materials (type 2, class II)<br>• DIN EN ISO 4049:2019-09: Dentistry—Polymer-based restorative materials (type 1, class II, group 2),<br>• DIN EN ISO 7491: Dental materials—Determination of colour stability.<br>• DIN EN ISO 20795-1:2009-02 Dentistry-Resins Part 1: Prosthetic resins. |
| | Including biocompatibility requirements, the following specification requirements of the 3D-printed material samples were tested and |
| | Surface quality Dimensional stability Color and color stability Translucency Flexural strength and flexural modulus Freedom from porosity Water Sorption Water Solubility |
| Shelf Life Testing: | Validated accelerated shelf life testing of the E-Dent 1000 resin at<br>time of 510(k) submission is 4 months. The resin is on real-time<br>validation testing for an ultimate shelf life of 24 months, stored in<br>the original packaging at temperatures at 30° C. Properties being<br>tested include material viscosity, material photoreactivity and color<br>change. The E-Dent 1000 resin also was tested for good transport<br>stability. |
| Biocompatibility<br>Testing: | A biocompatibility risk assessment was developed and presented<br>in the 510(k). As a result, the following ISO 10993 tests were<br>conducted according to Good Laboratory Practices.<br><br>Testing showed the E-Dent 1000 printed and tested samples are<br>biocompatible and non-toxic and meet the requirements for a<br>device in contact with mucosal membrane for >30 days. |
| | Cytotoxicity Study Using ISO Elution Method (Part 5) Guinea Pig Maximization Sensitization Test (Part 10) Tests for Irritation and Skin Sensitization – Intracutaneous<br>Injection in Rabbits (Part 10) Acute Systemic Toxicity Study in Mice (Part 11) |
| Additive<br>Manufacturing | Testing, according to FDA's guidance <i>Technical Considerations for<br/>Additive Manufactured Medical Devices</i> , was performed and<br>results were provided in the 510(k). These tests included<br>evaluation of all relevant properties of the printed resin using the<br>permitted machines. Further, tests based on considerations of the<br>orientation during manufacturing were performed. |
{5}------------------------------------------------
### VIII COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES
The intended use, technological characteristics and critical specifications, are similar to both Predicate devices. The additive method of manufacturing E-Dent 1000 is similar to Primary Predicate.
The intended use of fabrication of dental prosthetics is the same as both Predicates. The manufacture of the permanent crowns and veneers is the same as the Primary Predicate, K201668. The manufacture of the removable dentures is the same as the Secondary Predicate, K151142. The manufacture of artificial teeth is Class II, but 510(k)-exempt (product code ELM), therefore no Predicate is necessary.
While the E-Dent 1000 resin is different from Primary Predicate resin, each is a photo-
{6}------------------------------------------------
curable resin used in additive manufacturing of dental restorative devices and are of the same material category.
The Reference device, K102776, is presented because of similar technological characteristics to E-Dent 1000; it also is a photo-curable resin and the 3D printer and associated software are the same as used for E-Dent 1000.
The additive manufacturing processes for Primary Predicate and Reference devices use a photo-curable resin, scanner, 3D printer, curing unit and associated validated software.
The noted differences, in comparison to the Predicate devices, raise no new questions of safety and effectiveness.
## VIX CONCLUSION
Based on the comparisons provided and the data submitted in this 510(k), it can be concluded E-Dent 1000 substantially equivalent to the Predicate devices. EnvisionTEC's analysis of E-Dent 1000 shows it has the same intended use, and technological parameters that meet the requirements of ISO 20795-1:2009-02.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.