K210968 · Led Technologies, Inc. · OLP · Dec 21, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210968
Device Name
reVive Light Therapy Essentials
Applicant
Led Technologies, Inc.
Product Code
OLP · General, Plastic Surgery
Decision Date
Dec 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The reVive Light Therapy® Essentials is an Over the Counter (OTC) LED device intended for use in treating mild to moderate inflammatory acne.
Device Story
The reVive Light Therapy® Essentials is an OTC LED device for treating mild to moderate inflammatory acne. It utilizes 630nm red and 415nm blue LEDs to deliver light energy directly to the skin. The device is battery-powered or USB-powered and requires no user settings or adjustments. It is intended for patient self-use by applying the device directly to the skin for a 3-minute treatment duration. The device provides consistent light administration to the treatment area. By delivering specific wavelengths of light, the device aims to treat acne, offering a non-invasive therapeutic benefit to the user.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing, including functional performance, software validation, and a usability study with 16 participants. Usability results showed 100% success in device preparation and usage. Biocompatibility testing was performed per ISO 10993 and USP standards.
Technological Characteristics
Materials: ABS plastic and PC. Energy: LED-based light therapy (630nm red, 415nm blue). Power: Battery or USB. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1, ANSI/AAMI/IEC 60601-1-2, ISO 10993-5, ISO 10993-10, ISO 10993-11, USP 151. Connectivity: None. Software: Embedded firmware.
Indications for Use
Indicated for the treatment of mild to moderate inflammatory acne in the general population using an over-the-counter (OTC) device.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
reVive Light Therapy® LED Cleansing System (K180447)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 21, 2021
LED Technologies, Inc. Jelena Barbaric Compliance Manager 12821 Starkey Rd., Suite 4900 Largo, Florida 33773
Re: K210968
Trade/Device Name: reVive Light Therapy Essentials Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: November 24, 2021 Received: November 29, 2021
Dear Jelena Barbaric:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210968
Device Name reVive Light Therapy® Essentials
Indications for Use (Describe)
The reVive Light Therapy® Essentials is an Over the Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) Summary – K210968
This summary of 510 (k) information is being submitted in accordance with the requirements of 21 CFR § 807.92.
Submission Date: December 16th, 2021
| 1. Submitter Information: | LED Technologies, Inc. - Jelena Barbaric |
|---------------------------|------------------------------------------|
| | 12821 Starkey Rd., Suite 4900 |
| | Largo, FL 33773 |
| | Tel: 727-285-8300 Ext 203 |
| | Email: jbarbaric@ledtechnologies.com |
For Specification Developer: LED Technologies, Inc. Attn: Lloyd Nelson 12821 Starkey Rd., Suite 4900 Largo, FL 33773 Tel.: 727-285-8300. Ext 201 Email: Inelson@ledtechnologies.com
## 2. General Information
- 2.1 Classification Name: Over the Counter Powered Light Based Laser for Acne
- 2.2 Common/usual name: reVive Light Therapy® Essentials
- 2.3 Proprietary Names: reVive Light Therapy® Essentials
- Classification: Class II 2.4
- Classification Number: 878.4810 2.5
- 2.6 Product Code: OLP
- 2.7 Review Panel: General & Plastic Surgery
#### 3. Device Description
The reVive Light Therapy® Essentials is an over-the counter light emitting diode (LED) device, that emits energy for use in dermatology for the treatment of mild to moderate inflammatory acne. The device uses two types of LEDs: 630nm red and 415nm blue.
{4}------------------------------------------------
The reVive Light Therapy® Essentials components include the device containing the LED module.
There are no user settings or adjustments required.
The unit is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components.
## Indications/Intended Use:
The reVive Light Therapy® Essentials is an Over the Counter (OTC) LED device intended for use in treating mild to moderate inflammatory acne.
## 4. Predicate Devices:
Primary Predicate - K180447 – reVive Light Therapy® LED Cleansing System (LED Technologies, Inc.)
K180320 – dpl® IIa Panel (LED Technologies, Inc.)
# Comparison Chart
| Device | reVive Light<br>Therapy® Essentials<br>K210968 | dpl® Ila Panel<br>LED Technologies,<br>Inc.<br>K180320 | Primary Predicate<br>reVive Light Therapy® LED<br>Cleansing System<br>LED Technologies, Inc.<br>K180447 |
|-------------------------|-----------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Wavelengths | 630nm, 415nm | 605nm, 630nm,<br>660nm, 880nm<br>415nm | 630nm, 415nm |
| Irradiance<br>source | LED | LED | LED |
| Treatment Area<br>(cm²) | 7 | 415 | 18.86 |
| Treatment Time | 3 minutes per<br>treatment | 3 minutes per<br>treatment | 3 minutes per treatment |
| Type/Class | OTC | OTC | OTC |
| IFU | For treatment of<br>mild to moderate<br>inflammatory acne | For treatment of<br>wrinkles & mild to<br>moderate<br>inflammatory acne | For treatment of mild to<br>moderate inflammatory<br>acne |
{5}------------------------------------------------
## Summary of the technological characteristics of the device compared to predicate device:
- 1. Has the same intended use as the predicate devices (i.e., treatment of mild to moderate inflammatory acne).
- 2. Has similar technical characteristics.
The device has same control mechanism, same operating principle, and energy type. It uses the same method of operation as described in K180447 and K180320.
The subject device is smaller in size. The changes do not have significant effect on the use of the device. The risk assessment did not identify any new or significantly modified risks. Moreover, there were no unexpected issues from validation and verification activities.
Therefore, the minor changes in the device appearance do not affect the substantial equivalence of the subject device with the predicates.
- 3. Has the same method of operation.
- 4. Utilizes the same treatment duration as the predicate devices.
The reVive Light Therapy® Essentials and the above referenced predicate devices are Over the Counter Devices used to treat mild to moderate inflammatory acne as defined in 21 CFR § 878.4810. These devices utilize red and blue diodes with 630 nm and 415 nm to provide narrow bands of light energy to treat acne. The performance achieved by these devices is same with similar power output.
The reVive Light Therapy® Essentials is powered by Batteries) or via Universal USB power source.
The devices are intended to be placed directly on the skin. They are manufactured out of similar materials. Based upon comparison to the predicate devices, the reVive Light Therapy® Essentials has the same intended uses, with similar technological characteristics as predicate devices. The system performs as intended and does not raise any new safety or effectiveness issues.
## 5. Performance Testing and Standards:
Testing of the reVive Light Therapy® Essentials, included functional performance testing, software validation, testing, and user safety testing.
Safety and functionality testing demonstrate that the reVive Light Therapy® Essentials conforms to various international consensus standards.
ANSI/AAMI ES 60601-1:2005® + 2012 and A1:2012 IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012: Medical Electrical Equipment part 1: General Requirements for Basic Safety and Essential Performance.
ANSI/AAMI/IEC 60601-1-2 (2014): Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance: Collateral Standard: Electromagnetic Compatibility.
{6}------------------------------------------------
#### Biocompatibility
ISO 10993-5:2009 – Cytotoxicity Test ISO 10993-10:2010 - Intracutaneous reactivity test ISO 10993-10:2010 – Skin Sensitization Test ISO 10993-11:2017 – Acute Systemic Toxicity USP General Chapter 151 – Material Medicated Pyrogenicity
Additionally, the subject device tissue contacting components are the ABS plastic and the PC. The materials data from previous submissions is identical in materials selection, manufacturing process, chemical composition of the materials, as well as nature of patient contact, proving the ABS and PC of the reVive Light Therapy® Essentials medical device in its final finished form is identical to the ABS and PC material of the predicate devices.
The reVive Light Therapy® Essentials software was tested and validated in accordance with FDA's "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices".
A Usability Study was conducted with 16 participants.
The results of the study found that:
100% of the participants were able to demonstrate the correct preparation of device for use.
100% of the participants were able to demonstrate correct device usage.
The conclusions drawn from nonclinical tests demonstrate that the device is safe, as effective, and performs as well as the legally marketed devices.
## 6. Conclusion
After analysis of safety, indications, intended uses, dose rates, performance, features, design materials, power output, technological properties, treatment regiment regimens and methods of operation, the manufacturer asserts that no significant differences exist between the subject device and predicate, and no different questions of safety and effectiveness arise. Therefore, the subject device is substantial equivalence to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.