← Product Code [GEX](/productcode/GEX) · K210951

# Leonardo Mini Blue (K210951)

_Ceramoptec GmbH · GEX · Sep 1, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210951

## Device Facts

- **Applicant:** Ceramoptec GmbH
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Sep 1, 2021
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Leonardo Mini Blue 445nm Laser is intended for incision, excision, vaporization, ablation, hemostasis and coagulation of soft tissue.

## Device Story

Leonardo Mini Blue 445nm Laser is a diode laser system for soft tissue surgery. Device converts electrical energy into 445 nm coherent laser radiation via semiconductor diodes; 635 nm aiming beam provided. System includes touch screen control panel for setting laser power and treatment parameters. Energy delivered to tissue via fiber-optic delivery system. Features automatic optical fiber recognition via coding. Operated by physicians in clinical settings. Provides continuous and pulsed wave modes. Output affects tissue through thermal effects (incision, ablation, coagulation). Benefits include precise soft tissue management. Powered by mains or Li-ion rechargeable batteries.

## Clinical Evidence

No clinical data. Substantial equivalence established through bench testing and compliance with international standards (IEC 60601 series, IEC 60825-1, IEC 62304).

## Technological Characteristics

Diode laser system; 445 nm wavelength; 10W max power; 635 nm aiming beam. Features touch screen interface, internal fan cooling, and Li-ion battery/mains power. Automatic fiber recognition system. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and IEC 62304. Software validated for moderate level of concern.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Wolf 445nm ([K192272](/device/K192272.md))

## Reference Devices

- MegaBeam ENT Probe ([K113858](/device/K113858.md))
- MegaBeam 400 Micron Forward fire ([K113709](/device/K113709.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.

September 1, 2021

CeramOptec GmbH % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114

Re: K210951

Trade/Device Name: Leonardo Mini Blue Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 28, 2021 Received: March 30, 2021

Dear Daniel Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K210951

Device Name Leonardo Mini Blue 445nm Laser

Indications for Use (Describe)

The Leonardo Mini Blue 445nm Laser is intended for incision, vaporization, ablation, hemostasis and coagulation of soft tissue.

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary: 510(k) Number K210951 CeramOptec GMBH Siemensstrabe 44 Bonn Nordrhein-Westfalen, Bonn, Germany D-53121 Registration Number: 3003340844 Phone Number: +49 (228) 97 967 - 0 Date Prepared: August 31, 2021 Contact: Dr. Roland Dreschau, Director

- 1) Identification of the Device: Trade/Device Name: Leonardo® Mini Blue 445nm Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX
- 2) Equivalent legally marketed device: K192272, A.R.C. Laser GmbH Trade/Device Name: Wolf 445nm Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX
- 3) Reference Devices (Sterile Single Patient Use Fiberoptic Probes) Biolitec Medical Devices, Inc.: MegaBeam ENT Probe: K113858., Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX and Biolitec Medical Devices, Inc.: MegaBeam 400 Micron Forward fire: K113709. Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX
- 4) Indications for Use: The Leonardo Mini Blue 445nm Laser is intended for incision, excision, vaporization, ablation, hemostasis and coagulation of soft tissue.
- 5) Description of the Device: The laser family LEONARDO® Mini Blue is a laser system with functions and ergonomics specially developed for medical applications. A control panel is used to set treatment

{4}------------------------------------------------

parameters, such as laser power. User-friendly menu navigation ensures a reliable operation while allowing physicians to concentrate on the essential aspects of treatment. The fiber-coupled semiconductor laser diodes convert electrical energy to coherent laser radiation with the wavelength of 445 nm (aiming beam 635 nm). A beam transporting system delivers this energy to affected surfaces and organs. Depending on the model, the LEONARDO® Mini Blue laser has the following maximum laser output power:

LEONARDO® Mini Blue: 10W@445nm

LEONARDO® Mini Blue is available as a single-wavelength device with 445 nm. All LEONARDO® Mini Blue laser can be operated in continuous mode. Additional special treatment modes for specific treatment procedures or in combination with corresponding application fibers are available, depending on the configuration of the unit. For safety reasons, the LEONARDO® Mini Blue laser is equipped with a system for automatic recognition of the optical fibers used. Application fibers from CeramOptec have coding for communicating with the laser device.

| Characteristic         | A.R.C. Laser GmbH<br>Device Wolf 445nm K192272                                                                               | CeramOptec GMBH<br>Leonardo® Mini Blue 445nm Laser                                                                                                      |
|------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Wolf 445 nm is intended for incision,<br>excision, vaporization, ablation, hemostasis<br>and coagulation of soft tissue. | The Leonardo Mini Blue 445nm Laser is<br>intended for incision, excision, vaporization,<br>ablation, hemostasis and coagulation of soft<br>tissue. SAME |
| Photo                  | Image: A.R.C. Laser GmbH Device Wolf 445nm K192272                                                                           | Image: CeramOptec GMBH Leonardo® Mini Blue 445nm Laser                                                                                                  |
| Display                | Touch Screen                                                                                                                 | Touch Screen SAME                                                                                                                                       |
| Mode                   | Diode Laser                                                                                                                  | Diode Laser SAME                                                                                                                                        |
| Laser power            | 0.5 W up to 10 W                                                                                                             | Up to 10 W SAME                                                                                                                                         |
| Laser type             | Diode Laser                                                                                                                  | Diode Laser SAME                                                                                                                                        |
| Wavelength             | 445 nm                                                                                                                       | 445 nm SAME                                                                                                                                             |

#### Substantial Equivalence Chart

{5}------------------------------------------------

| Characteristic            | A.R.C. Laser GmbH<br>Device Wolf 445nm K192272                           | CeramOptec GMBH<br>Leonardo® Mini Blue 445nm Laser                                                                                                  |
|---------------------------|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Aiming beam               | 532 nm                                                                   | 635 nm Both predicate and proposed<br>device aiming beams are within the<br>readily visible spectrum. No adverse<br>safety or effectiveness impact. |
| Aiming beam<br>power      | ≤ 5 mW                                                                   | Max. 4 Mw. Comparable. No adverse safety<br>or effectiveness impact.                                                                                |
| Laser Class               | 4                                                                        | 4                                                                                                                                                   |
| Operation<br>mode         | Pulsed, Continuous Wave (CW)                                             | Pulsed, Continuous Wave (CW) SAME                                                                                                                   |
| Pulse length/<br>duration | ≤ 4 Watt: 1 ms to 45 sec and CW<br>> 4 Watt: 1 ms to 60 ms               | 10 Watt: 10 ms to 1 sec and CW.<br>Useful range for the indications. No adverse<br>safety or effectiveness impact.                                  |
| Pulse<br>frequency        | ≤ 4 Watt: 0.01 Hz to 500 Hz and SP<br>> 4 Watt: 0.02 Hz to 6.6 Hz and SP | 10 Watt: 0.5Hz - 50Hz<br>Useful range for the indications. No adverse<br>safety or effectiveness impact.                                            |
| Cooling                   | Air                                                                      | Internal Fan. No adverse safety or<br>effectiveness impact.                                                                                         |
| Main power<br>Supply      | 100-240 V~, 47-63 Hz, 1.06-0.45 A                                        | Mains operated OR Li-ion rechargeable<br>battery. No adverse safety or effectiveness<br>impact.                                                     |
| Dimensions of<br>system   | H 10.06 cm (3.9 inches), W 20.3 cm (7.9<br>inches), 23.9 cm (9.4 inches) | H 6 cm, W 9 cm, L 21,5 cm. SMALLER. No<br>adverse safety or effectiveness impact.                                                                   |
| Weight                    | 2.8 kg (6.17 pounds)                                                     | 0.960kg (without batteries)<br>2x 0.045kg (battery weight)<br>LIGHTER WEIGHT No adverse safety or<br>effectiveness impact.                          |

- 6) The technological characteristics are substantially the same, so there are no issues impacting safety and effectiveness. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new device is as safe and effective as the predicate device.
### 7) Summary of non-clinical testing:

Firmware was validated according to the FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005. Software validation documents were provided for a moderate level of concern. This device complies with the applicable portions of the standards listed below:

21 CFR Part 1040 PERFORMANCE STANDARDS FOR LIGHT-EMITTING PRODUCTS

EN 1041:2008 Information Supplied By The Manufacturer Of Medical Devices

ISO 14971: 2007 Medical devices — Application of risk management to medical devices FDA # 5-40

{6}------------------------------------------------

IEC ISO 15223-1:2016 (Corrected Version 2017-03) Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements FDA # 5-117

IEC 60601-1:2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 60601-1-6:2010+A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability

IEC 60601-2-22: 2007+A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment FDA # 12-268

IEC 62304: 2006 Medical device software - Software life cycle processes

IEC 62366-1:2007+A1:2014 Medical devices - Part 1: Application of usability engineering to medical devices FDA # 5-114

IEC 60825-1: 2007 Safety of laser products - Part 1: Equipment classification, and requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)] FDA # 12-273

The IEC standards testing was conducted by an accredited testing laboratory "EZU".

- 8) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence.
- 9) Conclusion: After analyzing bench and clinical tests, it is the conclusion of CeramOptec GMBH that the new CeramOptec GMBH Leonardo® Mini Blue 445nm Laser is as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K210951](https://fda-staging.innolitics.com/device/K210951)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
