K210845 · Jiaxing Amazing Travel-Ware Co., Ltd. · FXX · Oct 12, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210845
Device Name
Disposable Surgical Mask (Model: YYKZ-01)
Applicant
Jiaxing Amazing Travel-Ware Co., Ltd.
Product Code
FXX · General, Plastic Surgery
Decision Date
Oct 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The DISPOSABLE SURGICAL MASK is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials in infection control practices to reduce the potential exposure to blood and body fluids. It is for single-use and provided non-sterile.
Device Story
Disposable surgical mask (Model: YYKZ-01); flat-pleated design; three-layer construction (polypropylene non-woven outer/inner layers; melt-blown polypropylene filter layer); elastic ear loops (polyester/spandex); iron/polypropylene nose piece. Used in clinical or OTC settings to protect wearer and patient from microorganisms, body fluids, and particulates. Device acts as physical barrier; reduces exposure to blood/body fluids. Non-sterile; single-use.
Clinical Evidence
No clinical data. Bench testing only: BFE (99.9%), PFE (99.8%), differential pressure (2.7 mmH2O/cm²), synthetic blood penetration resistance (80 mmHg), and flammability (Class 1) per ASTM F2100-19. Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
Three-layer flat-pleated mask; non-woven polypropylene outer/inner layers; melt-blown polypropylene filter; polyester/spandex ear loops; iron/polypropylene nose piece. ASTM F2100-19 Level 1 performance. Non-sterile. No software or electronic components.
Indications for Use
Indicated for adults to protect patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials in infection control practices; reduces potential exposure to blood and body fluids. Single-use, non-sterile, OTC use.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 12, 2021
Jiaxing Amazing Travel-Ware CO., LTD Jiang Tao Head of Firm No.2196 Honggao Road, Xiuzhou Industrial Park Jiaxing, Zhejiang 314015 China
Re: K210845
Trade/Device Name: Disposable Surgical Mask (Model: YYKZ-01) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: September 6, 2021 Received: September 13, 2021
Dear Jiang Tao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210845
Device Name DISPOSABLE SURGICAL MASK(Model: YYKZ-01)
Indications for Use (Describe)
The DISPOSABLE SURGICAL MASK is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials in infection control practices to reduce the potential exposure to blood and body fluids. It is for single- use and provided non-sterile.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "JEGUN" in a stylized font. The letters are white and have a unique, interconnected design. The background is a light blue color, providing a clean and contrasting backdrop for the text. The overall design is modern and eye-catching.
# 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
#### Prepared Date: Sep. 23, 2021
# 1. Submitter's Information
The submitter of this pre-market notification is:
| Name: | JIAXING AMAZING TRAVEL-WARE CO.,LTD |
|-----------------|----------------------------------------------------------------------------------|
| Address: | No.2196 Honggao Road, Xiuzhou Industrial Park<br>Jiaxing, Zhejiang, China 314015 |
| Contact person: | Jiang Tao |
| Title: | Head of Firm |
| E-mail: | 175556919@qq.com |
| Tel: | +86-13957376195 |
# 2. Device Identification
| 510(k) Number: | K210845 |
|--------------------|--------------------------|
| Trade/Device Name: | DISPOSABLE SURGICAL MASK |
| Models: | YYKZ-01 |
| Common name: | Mask, Surgical |
| Regulation Number: | 878.4040 |
| Regulation Name: | Surgical apparel |
| Regulation Class: | Class II |
| Panel: | General Hospital |
| Product Code: | FXX |
## 3. Predicate Device
| 510(K) number: | K202061 |
|--------------------|-------------------------------------------------------|
| Device Name: | Medical Face mask |
| Manufacturer: | Jiangsu Province Jianerkang Medical Dressing Co., Ltd |
| Common name | Mask, Surgical |
| Regulation Number: | 878.4040 |
| Regulation Name: | Surgical apparel |
| Regulation Class: | Class II |
| Panel: | General Hospital |
| Product Code: | FXX |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "JEGUN" in a stylized, sans-serif font. The letters are white against a light blue background. The "G" in "JEGUN" is particularly distinctive, with a unique, geometric design that sets it apart from the other letters.
## 4. Device Description
The "DISPOSABLE SURGICAL MASK" is single use, blue color, without face shield, Flat Pleated type, utilizing elastic ear loops for wearing, and it has a Nose Piece design for fitting the facemask around the nose. The "DISPOSABLE SURGICAL MASK" is manufactured with three layers. The inner and outer layers are made of Non-woven fabric (Polypropylene), and the middle filter layer is made of a melt blown fabric (Polypropylene). The subject device is held in place over the user's mouth and nose by two ear loops welded to the facemask. The elastic ear loop is made of Polyester and Spandex. The nose piece contained in masks is in the layers of the facemask to allow the user to fit the facemask around their nose, which is made of iron strip and polypropylene. The"DISPOSABLE SURGICAL MASK" is sold non-sterile and are intended to be single-use, disposable devices. This product contains no components made with natural rubber latex.
#### 5. Indication for use
The DISPOSABLE SURGICAL MASK is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate materials in infection control practices to reduce the potential exposure to blood and body fluids. It is for single-use and provided non-sterile.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for JEGUN. The logo is white text on a blue background. The text is stylized and appears to be a custom font.
# 6. Comparison to Predicate Device
Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table
| SE<br>Comparisons | Proposed Device<br>K210845 | Predicate Device<br>K202061 | Similarities/<br>Differences |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Name | DISPOSABLE SURGICAL<br>MASK | Single-use Surgical Mask | / |
| Model | YYKZ-01 | FE2311 | / |
| Classification | Class II | Class II | Same |
| Intended use/<br>Indications for<br>Use | The DISPOSABLE<br>SURGICAL MASK is<br>intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids, and particulate<br>materials in infection control<br>practices to reduce the<br>potential exposure to blood<br>and body fluids. It is for<br>single-use and provided<br>non-sterile. | The medical face mask is<br>intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids, and particulate<br>materials in infection control<br>practices to reduce the<br>potential exposure to blood<br>and body fluids. It is for<br>single-use and provided<br>non-sterile. | Same |
| ASTM F2100<br>Level | Level 1 | Level 1 | Same |
| Mask Styles | Flat Pleated | Flat Pleated | Same |
| Design features | Ear loop | Ear loop | Same |
| Layers | 3 | 3 | Same |
| Color | Blue | Blue | Same |
| Target<br>population | Adults | Adults | Same |
| Dimension<br>(length) | $175 \pm 5$ mm | 175 mm | Same |
| Dimension<br>(width) | $95 \pm 5$ mm | 95 mm | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Use | Single use, disposable | Single use, disposable | Same |
| Anatomical site | Nose and mouth | Nose and mouth | Same |
| Technology | Self-suction filter mask | Self-suction filter mask | Same |
| Environment of<br>use | OTC | OTC | Same |
| | | | |
| Material of<br>Outer layer | Non-woven fabric<br>(Polypropylene) | polypropylene | Same |
| Material of<br>middle layer | Melt blown fabric<br>(Polypropylene) | Melt blown polypropylene | Same |
| Material of<br>inner layer | Non-woven fabric<br>(Polypropylene) | polypropylene | Same |
| Material of<br>ear loops | Spandex and nylon | Spandex and nylon | Same |
| Material of<br>Nose piece | Iron strip and Polypropylene | Polypropylene coated steel<br>wire | Note 1 |
| Performance testing Data comparison | | | |
| BFE | Average 99.9% | Average 99.8% | Meet level 1 |
| PFE | Average 99.8% | Average 96.09% | Meet level 1 |
| Differential<br>Pressure | Average 2.7 | Average 3.9 | Meet level 1 |
| Synthetic Blood<br>Penetration<br>Resistance | 80 | 80 | Meet level 1 |
| Flammability | Class 1 | Class 1 | Meet level 1 |
| Biocompatibility comparison | | | |
| In Vitro<br>Cytotoxicity | Under the conditions of the<br>test, the test article was<br>found to be non- cytotoxic | Under the conditions of the<br>test, the test article was<br>found to be non- cytotoxic | Same |
| Skin<br>Sensitization | Under the conditions of the<br>test, the test article was<br>found to be non- sensitizing | Under the conditions of the<br>test, the test article was<br>found to be non- sensitizing | Same |
| Skin Irritation<br>test | Under the conditions of the<br>test, the test article was<br>found to be non- irritating | Under the conditions of the<br>test, the test article was<br>found to be non- irritating | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "JEGUN" in a stylized, sans-serif font. The letters are white and the background is a light blue color. The letters are connected to each other, and the overall design is modern and sleek. The logo is simple and easy to read.
Note 1: The nose piece material of the subject device is iron with polypropylene covering. The predicate device uses polypropylene coated steel wire. The nose piece does not directly contact the user's skin. Biocompatibility testing was conducted on the final device, and the tests showed there was no biocompatibility risk. Performance testing was conducted according to ASTM F2100-19, and the test results show that the proposed device meets the level 1 requirements.
Otherwise, all of the specifications and materials of the subject device, DISPOSABLE SURGICAL MASK (model: YYKZ-01), are the same as the predicate device K202061 with no identified new risks.
### 7. Performance Data Clinical test:
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "JEGUN" in a stylized font. The letters are white and the background is light blue. The letters are connected to each other, and the overall design is modern and sleek.
Clinical testing is not required.
#### Non-clinical data
The proposed device DISPOSABLE SURGICAL MASK:
Performance:
1. ASTM F2100-19 Standard Specification for Performance of Materials Used in Medical Face Masks.
2. Bacterial Filtration Efficiency-Determine the bacterial filtration efficiency as directed in Test method F2101.
- 3. Differential Pressure -Determine breathing resistance or differential pressure as directed in
#### EN 14683:2019,Annex C.
- Sub-Micron Particulate Filtration-Determine particulate filtration efficiency as directed in Test 4.
#### Method F2299
- 5. Resistance to Penetration by Synthetic Blood-Determine synthetic blood penetration resistance as specified in Test Method F1862.
- 6. Flammability-Determine flammability as specified in 16 CFR Part 1610.
| standa<br>rd | Test item | Test method | Criteria | Results |
|----------------------|-------------------------------------------------|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| ASTM<br>F2100<br>-19 | BFE | ASTM F2101-19 | ≥95% | Accepted |
| | PFE | ASTM F2299-03(2017) | ≥95% | Accepted |
| | Differential<br>Pressure | EN<br>14683 :2019+AC (2019)(E),<br>Annex C | <5.0 | Accepted |
| | Synthetic<br>Blood<br>Penetration<br>Resistance | ASTM F1862M-17 | 80mmHg | Accepted |
| | Flammability | 16 CFR Part 1610<br>(As Amendment In 2008) | (A) There are no burn times;<br>or<br>(B) There is only one burn<br>time and it is equal to or<br>greater than 3.5 seconds; or<br>(C) The average burn time of<br>two or more specimens is<br>equal to or greater than 3.5<br>seconds | Class 1 |
Biocompatibility:
- 1. ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro
2. ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
| Stand ard | Test item | Test method | Criteria | Results |
|-----------|-----------|-------------|----------|---------|
| AVI | | | | |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "JEGUN" in white letters against a blue background. The font is stylized and modern, with rounded edges and a slightly futuristic look. The letters are connected to each other, creating a sense of unity and flow. The overall design is clean and simple, making it easy to read and remember.
| ISO<br>10993<br>-5:<br>2009 | In Vitro<br>Cytotoxicity | In this study, mammalian L-<br>929 cells were cultured in<br>vitro according to ISO 10993-<br>5:2009 to test the potential<br>cytotoxicity of the test article.<br>The test articles and the<br>control material were<br>separately placed in MEM<br>medium containing 10% fetal<br>bovine serum, and extracted<br>in a 37 °C incubator for 24<br>hours. After the end of the<br>extraction, the cell culture<br>medium in the 96-well plate<br>(104 cells/well) cultured for 24<br>hours was removed and<br>replaced with the<br>corresponding extract,<br>cultured in 37 °C, 5%<br>CO2, >90% humidity for 24<br>hours. After the culture, the<br>morphology and cell lysis of<br>the cells were observed<br>under the microscope, and<br>the cytotoxicity of the test<br>samples was determined by<br>MTT assay. | The 50% extract of the test<br>article should have at least<br>the same or a higher viability<br>than the 100% extract.<br>Otherwise the test should be<br>repeated.<br>The lower the Viab. % value,<br>the higher the cytotoxic<br>potential of the test article is.<br>If viability is reduced to <70%<br>of the blank, it has a<br>cytotoxic potential.<br>The Viab.% of the 100%<br>extract of the test article is<br>the final result. | Under the<br>conditions<br>of the<br>test, the<br>test article<br>was found<br>to be non-<br>cytotoxic |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| ISO<br>10993<br>-10:<br>2010 | Skin<br>Sensitization | we took guinea pigs to<br>observe the skin sensitization<br>of the test article according to<br>ISO 10993-10: 2010.<br>The test article were<br>extracted in Constant<br>Temperature Vibrator at<br>50 °C, 60 rpm for 72 h by<br>0.9 % Sodium Chloride<br>Injection and Sesame Oil.<br>Mix 50:50 (by volume) stable<br>emulsion of Freund's<br>complete adjuvant with<br>selected solvent. Intradermal<br>induction and topical<br>induction were operated in<br>the clipped intrascapular<br>region of each animal. After<br>the topical induction phase<br>was completed on day 14, all<br>test and control animals were<br>challenged with the test<br>sample. The erythema and<br>edema of the challenge site | Magnusson and Kligman<br>grades of 1 or greater in the<br>test group generally indicate<br>sensitization, provided<br>grades of less than 1 are<br>seen in control animals.<br>If grades of 1 or greater are<br>noted in control animals, then<br>the reactions of test animals<br>which exceed the most<br>severe reaction in control<br>animals are presumed to be<br>due to sensitization.<br>If the response is equivocal,<br>rechallenge is recommended<br>to confirm the results from<br>the first challenge. The<br>outcome of the test is<br>presented as the frequency<br>of positive challenge results<br>in test and control animals. | Under the<br>conditions<br>of the<br>test, the<br>test article<br>was found<br>to be non-<br>sensitizin<br>g |
| | | | | |
| | sensitization response of the<br>test article. According to the<br>Magnusson and Kligman<br>scales, the response to<br>erythema and edema at each<br>application site of the skin<br>was described and scored 24<br>hours and 48 hours after the<br>challenge phase. | | | |
| Skin<br>Irritation test | we took New Zealand white<br>Rabbits to observe the skin<br>irritation of the test article<br>according to ISO10993-<br>10:2010.<br>The test article were<br>extracted in Constant<br>Temperature Vibrator at<br>50 °C, 60 rpm for 72 h by<br>0.9 % Sodium Chloride<br>Injection and Sesame<br>Oil.Apply 0.5 ml extracts of<br>test article or control to<br>2.5 cm x 2.5 cm absorbent<br>gauze patches, and then<br>apply the patch soaked with<br>the extract of test article or<br>control directly to the skin on<br>each side of each rabbit, and<br>then wrap the application<br>sites with a bandage for a<br>minimum of 4 h.At the end of<br>the contact time, remove the<br>dressing. The describe and<br>score the skin reaction for<br>erythema and oedema for<br>each application site at each<br>time interval. Record the<br>appearance of each<br>application site at (1±0.1) h,<br>(24±2) h, (48±2) h and<br>(72±2)h following removal of<br>the patches. | Use only (24±2) h, (48±2) h<br>and (72±2) h observations for<br>calculation.<br>After the 72 h grading, all<br>erythema grades plus<br>oedema grades (24±2) h,<br>(48±2) h and (72±2) h were<br>totalled separately for each<br>test article and blank for each<br>animal. The primary irritation<br>score for an animal was<br>calculated<br>by dividing the sum of all the<br>scores by 6 (two<br>test/observation sites, three<br>time points).<br>To obtain the primary<br>irritation index for the test<br>article, add all the primary<br>irritation scores of the<br>individual<br>animals and divide by the<br>number of animals.<br>When blank or negative<br>control was used, calculate<br>the primary irritation score for<br>the controls and subtract<br>that score from the score<br>using the test material to<br>obtain the primary irritation<br>score. | Under the<br>conditions<br>of the<br>test, the<br>test article<br>was found<br>to be non-<br>irritating | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image features the word "JEGUN" in a stylized, sans-serif font. The letters are white and have a unique, interconnected design, giving the word a modern and somewhat futuristic appearance. The background is a solid light blue color, which provides a clean and contrasting backdrop for the white text, making it easily readable.
### 8. Conclusion
The conclusions drawn from the nonclinical tests demonstrates that the subject device in 510(k) submission K210845, the DISPOSABLE SURGICAL MASK is as safe, as effective, and performs as well as or better than the legally marketed device predicate cleared under K202061.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.