K210823 · Lyma Life , Ltd. · OHS · Mar 18, 2022 · General, Plastic Surgery
Device Facts
Record ID
K210823
Device Name
LYMA Laser
Applicant
Lyma Life , Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Mar 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LYMA Laser is an Over the-Counter (OTC) light based device intended for the use in treating full-face wrinkles.
Device Story
LYMA Laser is an OTC light-based device for treating full-face wrinkles. Device incorporates near-infrared (808 nm) and red (620 nm) LED light sources. Optical lens diffuses light to produce narrowband, divergent distribution across an 8 cm² treatment area. Energy output is 62.5 mW/cm². Device is intended for patient self-use. Patient applies device to skin for 3 minutes per treatment area. Mechanism involves Low Level Light Therapy (LLLT) to provide beneficial treatment for wrinkles. Device is handheld; no complex software or algorithmic processing described.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing (electrical safety per AAMI/ANSI ES60601-1, EMC per IEC 60601-1-2, laser safety per IEC 60825-1, and biocompatibility per ISO 10993-1) and a usability study. Usability study (n=26) demonstrated 88% success rate for self-selection, labeling comprehension, and proper device operation, meeting the 80% success threshold.
Technological Characteristics
Light-based device using 808 nm (near-infrared) and 620 nm (red) LEDs. Energy output 62.5 mW/cm². Optical lens for light diffusion. Treatment area 8 cm². Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60825-1. Biocompatibility evaluated per ISO 10993-1.
Indications for Use
Indicated for the treatment of full-face wrinkles in adult patients via over-the-counter use.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 18, 2022
LYMA Life Ltd. % Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Avenue, Suite B-257 San Diego, California 92110
Re: K210823
Trade/Device Name: LYMA Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: February 9, 2022 Received: February 11, 2022
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210823
Device Name LYMA Laser
Indications for Use (Describe)
The LYMA Laser is an Over the-Counter (OTC) light based device intended for the use in treating full-face wrinkles.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# L Y M A
# 510(k) Summary K210823
#### DATE PREPARED
March 17, 2022
#### MANUFACTURER AND 510(k) OWNER
LYMA Life Ltd 93 Gloucester Place Marylebone, London W1U 6JQ, United Kingdom Telephone: +44 (0)7950 396247 Official Contact: Simon Goff, CEO
## REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Matthieu Kirkland, M.S. RQM+, LLC Telephone: +1 (619) 458-9547 Email: akomiyama@RQMplus.com, mkirkland@RQMplus.com
#### DEVICE INFORMATION
| Proprietary Name/Trade Name: | LYMA Laser |
|------------------------------|----------------------------------------------------------|
| Regulation Name: | Laser surgical instrument for use in general and plastic |
| | surgery and in dermatology |
| Regulation Number: | 21 CFR 878.4810 |
| Class: | II |
| Product Code: | OHS |
| Review Panel: | General & Plastic Surgery |
| Premarket Review: | Surgical and Infection Control Devices (OHT4) |
| | General Surgery Devices (DHT4A) |
#### PREDICATE DEVICE IDENTIFICATION
The LYMA Laser is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|------------------------------------------------------|-------------------|
| K120775 | LightStim for Wrinkles / LED Intellectual Properties | ✓ |
| K120560 | Trinity Wrinkle Remover / Carol Cole Company | |
The predicate devices have not been subject to a design related recall.
{4}------------------------------------------------
## DEVICE DESCRIPTION
The LYMA Laser is an Over-the-Counter (OTC) light based device intended for the use in treating full-face wrinkles. The LYMA Laser incorporates a near-infrared LED light source within the optimum Therapeutic Optical Window at 808 nm, optimized for Low Level Light Therapy, and a red LED light source at 620 nm. An optical lens that diffuses the light to produce narrowband, divergent distribution across an 8 cm² treatment area. The energy output of the 808 nm and 620 nm wavelengths is 62.5 mW/cm² which produces a beneficial treatment for full-face wrinkles.
# INDICATIONS FOR USE
The LYMA Laser is an Over-the-Counter (OTC) light based device intended for the use in treating full-face wrinkles.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
LYMA Life believes that the LYMA Laser is substantially equivalent to the predicate devices based on the information summarized here:
- The subject device has the same intended use as the predicates (i.e., the treatment of full-face wrinkles)
- The subject device has a similar output as the predicates (i.e., 62.5 mW/cm²) ●
- The subject device utilizes a similar near-infrared treatment wavelength and a similar red light wavelength as the predicates (i.e., 808 nm and 620 nm)
- . The subject device utilizes the same treatment duration as the predicates (i.e., 3 minutes per treatment area)
- The subject device utilizes a similar irradiation, or treatment, area as the predicates (i.e., 8 cm²)
We believe we have identified appropriate predicate devices with the same intended use, same or similar technological characteristics and the same treatment parameters. Therefore, this information supports the determination of substantial equivalence in safety and effectiveness between the subject and predicate devices.
## SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the LYMA Laser. The following tests were performed to demonstrate safety based on current industry standards:
Electromagnetic Compatibility and Electrical Safety: The subject device was tested in compliance to AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2(2010)/(R)2012 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance (which complies with IEC 60601-1); and IEC 60601-1-2 Medical electrical equipment — Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
{5}------------------------------------------------
Additionally, testing to IEC 60825-1 Safety of laser products - Part 1: Equipment classification and requirements and an evaluation per ISO 10993-1 Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process were provided.
## SUMMARY OF USABILITY TESTING
To demonstrate that lay users can correctly operate the LYMA Laser, a usability study was completed. The study was split into two Parts. Part A examined the participants' abilities correctly self-identify whether they are appropriate users of the device based on the outer packaging (also known as the box label). Part B evaluated how well participants could comprehend the instructions for use and correctly operate the subject device.
This usability study evaluated the suitability of the LYMA Laser for over-the-counter (OTC) distribution. The ability of participants to correctly self-identify whether they are appropriate users of the device, comprehend labeling, and safely and effectively use the subject device was assessed in two independent parts of the study. The number of participants required to answer all questions correctly in each part was 21 out of 26, for an 80% success rate. The results of the usability testing (passing rates of 88% for the Self-Selection as well as the Labeling Comprehension and Proper Use of the Device sections) demonstrate that the LYMA Laser is suitable for OTC distribution and that lay users can correctly self-select themselves as users of the subject device.
#### CONCLUSION
Based on the testing performed and usability testing, it can be concluded that the subject device does not raise different questions of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed LYMA Laser are assessed to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.