← Product Code [FXX](/productcode/FXX) · K210813

# Medical Surgical Mask (K210813)

_Sichuan Pharmaceutical., Inc. · FXX · Oct 8, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210813

## Device Facts

- **Applicant:** Sichuan Pharmaceutical., Inc.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Oct 8, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

## Device Story

Single-use, flat-pleated surgical mask; three-layer construction (spun-bond polypropylene outer/inner layers, melt-blown polypropylene filter middle layer). Features nylon/spandex ear loops and malleable polyethylene wire nose piece for fit. Used in clinical settings by healthcare personnel and patients to provide a physical barrier against blood, body fluids, and particulates. Device is non-sterile and disposable. Performance relies on material filtration properties to reduce infection transmission risk.

## Clinical Evidence

No clinical data. Bench testing only: Synthetic blood penetration (ASTM F1862), particulate filtration efficiency (ASTM F2299 >98%), bacterial filtration efficiency (ASTM F2101 >98%), differential pressure (EN 14683 ≤ 6.0 mmH2O/cm²), and flammability (16 CFR 1610 Class 1). Biocompatibility testing (ISO 10993-5, 10) confirmed non-cytotoxic, non-irritating, and non-sensitizing properties.

## Technological Characteristics

Three-layer flat-pleated mask. Materials: Spun-bond polypropylene (outer/inner), melt-blown polypropylene (filter), nylon/spandex (ear loops), polyethylene wire (nose piece). ASTM F2100 Level 3 performance. Biocompatibility per ISO 10993-5 and ISO 10993-10. Non-sterile, single-use.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- DemeMASK Surgical Mask ([K201479](/device/K201479.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

October 8, 2021

Sichuan Pharmaceutical. Inc. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China

Re: K210813

Trade/Device Name: Medical Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: August 25, 2021 Received: September 9, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, Ph.D. Assistant Director DHT4B: Division of Infection Control And Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210813

Device Name Medical Surgical Mask

#### Indications for Use (Describe)

The Medical Surgical Mask is intended to be worn to protect and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable):

| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CER 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# (K210813)

This summary is submitted in accordance with 21 CFR 807.92.

# 1.0 Submitter's Information

| Name:    | Sichuan Pharmaceutical,Inc.                                                                              |
|----------|----------------------------------------------------------------------------------------------------------|
| Address: | Sichuan Pharmaceutical.Inc No. 18 Baiye Road,<br>Hi-Tech West Zone, Chengdu, Sichuan Province,<br>China. |
| Contact: | Yunrui Zhang                                                                                             |
| Tel:     | 86-13678040510                                                                                           |

### Designated Submission Correspondent

| Contact: | Mr. Boyle Wang                                                                                                                |
|----------|-------------------------------------------------------------------------------------------------------------------------------|
| Address: | Shanghai Truthful Information Technology Co., Ltd.<br>Room 1801, No. 161 East Lu Jiazui Rd., Pudong<br>Shanghai, 200120 China |
| Tel:     | +86-21-50313932                                                                                                               |
| Email:   | info@truthful.com.cn                                                                                                          |

Date of Preparation: Oct.8,2021

# 2.0 Device Information

| Trade name:          | Medical Surgical Mask |
|----------------------|-----------------------|
| Common name:         | Surgical Face Mask    |
| Classification name: | Surgical Face Mask    |
| Production code:     | FXX                   |
| Regulation number:   | 21CFR 878.4040        |
| Classification:      | Class II              |
| Review Panel:        | Surgical Apparel      |

# 3.0 Predicate Device Information

| Manufacturer:  | DemeTECH Corporation.  |
|----------------|------------------------|
| Device:        | DemeMASK Surgical Mask |
| 510(k) number: | K201479                |

#### 4.0 Device Description

{4}------------------------------------------------

The Medical Surgical Mask is a single use, three-layer, flat-pleated style surgical mask with ear loops and nose piece. The mask is manufactured with three layers, the inner and outer layers are made of spun-bond polypropylene, and the middle layer is made of melt blown polypropylene filter. The surgical masks is held over the user's mouth and nose via earloops made of nylon with spandex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of malleable polyethylene wire. The medical surgical mask includes blue colorant, Shining Blue R603. The medical surgical mask is sold non-sterile and is intended to be single use, and disposable.

## 5.0 Indication for Use Statement

The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

| Item                                             | Subject Device<br>K210813                                                                                                                                                                                                                                                                                                                                | Predicate Device<br>K201479                                                                                                                                                                                                                                                                     | Comparison                                                                                    |            |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|------------|
| Product Name                                     | Medical Surgical Mask                                                                                                                                                                                                                                                                                                                                    | DemeMASK Surgical Mask                                                                                                                                                                                                                                                                          | --                                                                                            |            |
| Product Code                                     | FXX                                                                                                                                                                                                                                                                                                                                                      | FXX                                                                                                                                                                                                                                                                                             | Same                                                                                          |            |
| Regulation No.                                   | 21 CFR 878.4040                                                                                                                                                                                                                                                                                                                                          | 21 CFR 878.4040                                                                                                                                                                                                                                                                                 | Same                                                                                          |            |
| Class                                            | II                                                                                                                                                                                                                                                                                                                                                       | II                                                                                                                                                                                                                                                                                              | Same                                                                                          |            |
| Intended Use&<br>Indications for<br>use          | The Medical Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), | The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to | Same                                                                                          |            |
|                                                  |                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                 |                                                                                               |            |
|                                                  |                                                                                                                                                                                                                                                                                                                                                          | provided non-sterile.                                                                                                                                                                                                                                                                           | blood and body fluids.<br>This is a single use,<br>disposable device<br>provided non-sterile. |            |
| Design features                                  |                                                                                                                                                                                                                                                                                                                                                          | Ear Loops, 3 layers                                                                                                                                                                                                                                                                             | Ear Loops, 3 layers                                                                           | Same       |
| Mask Styles                                      |                                                                                                                                                                                                                                                                                                                                                          | Flat pleated                                                                                                                                                                                                                                                                                    | Flat pleated                                                                                  | Same       |
| Material                                         | Outer<br>facing<br>layer                                                                                                                                                                                                                                                                                                                                 | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                      | Spun-bond<br>polypropylene                                                                    | Same       |
|                                                  | Middle<br>layer                                                                                                                                                                                                                                                                                                                                          | Melt blown<br>polypropylene filter                                                                                                                                                                                                                                                              | Melt blown<br>polypropylene filter                                                            | Same       |
|                                                  | Inner<br>Facing<br>layer                                                                                                                                                                                                                                                                                                                                 | Spun-bond<br>polypropylene                                                                                                                                                                                                                                                                      | Spun-bond<br>polypropylene                                                                    | Same       |
|                                                  | Nose<br>piece                                                                                                                                                                                                                                                                                                                                            | Malleable polyethylene<br>wire                                                                                                                                                                                                                                                                  | Galvanized wire with<br>coated polyethylene                                                   | *Different |
|                                                  | Ear<br>loops                                                                                                                                                                                                                                                                                                                                             | Nylon and spandex                                                                                                                                                                                                                                                                               | Nylon and spandex                                                                             | Same       |
| Color                                            |                                                                                                                                                                                                                                                                                                                                                          | Blue                                                                                                                                                                                                                                                                                            | Blue                                                                                          | Same       |
| Dimension<br>(Length)                            |                                                                                                                                                                                                                                                                                                                                                          | 17.5 cm +/- 0.2 cm                                                                                                                                                                                                                                                                              | 17.5 cm +/- 1cm                                                                               | *Different |
| Dimension<br>(Width)                             |                                                                                                                                                                                                                                                                                                                                                          | 9.5 cm +/- 0.2 cm                                                                                                                                                                                                                                                                               | 9.5 cm +/- 1cm                                                                                | ent        |
| OTC use                                          |                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                             | Yes                                                                                           | Same       |
| Sterility                                        |                                                                                                                                                                                                                                                                                                                                                          | Non-Sterile                                                                                                                                                                                                                                                                                     | Non-Sterile                                                                                   | Same       |
| Single Use                                       |                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                             | Yes                                                                                           | Same       |
| Sterile                                          |                                                                                                                                                                                                                                                                                                                                                          | No                                                                                                                                                                                                                                                                                              | No                                                                                            | Same       |
| ASTM F2100<br>Level                              |                                                                                                                                                                                                                                                                                                                                                          | Level 3                                                                                                                                                                                                                                                                                         | Level 3                                                                                       | Same       |
| Biocompatibility<br>ISO 10993-5/<br>ISO 10993-10 |                                                                                                                                                                                                                                                                                                                                                          | Non-cytotoxic,<br>non-sensitizing, and<br>non-irrtating                                                                                                                                                                                                                                         | Non-cytotoxic,<br>non-sensitizing, and<br>non-irrtating                                       | Same       |

## 6.0 Comparison to the Predicate Device

{5}------------------------------------------------

* Performance testing including biocompatibility evaluation has been performed on the final finished device which includes all construction materials. The difference in the materials does not raise additional questions for safety and effectiveness.

# 7.0 Non-Clinical Test Conclusion

The proposed device was tested and conformed to the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004.

{6}------------------------------------------------

| Test                                                                         | Subject Device                                                  | Predicate Device                                             | Results |
|------------------------------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------|---------|
| Synthetic<br>Blood<br>Penetration<br>Resistance<br>ASTM F1862                | No penetration<br>at 160 mmHg                                   | No penetration at<br>160 mmHg                                | Pass    |
| Particulate<br>Filtration<br>Efficiency<br>ASTM F2299                        | Particle filtration<br>efficiency at 0.1<br>micron<br>(%) >98 % | Particle filtration<br>efficiency at 0.1<br>micron (%) >98 % | Pass    |
| Bacterial<br>Filtration<br>Efficiency<br>ASTM F2101                          | >98 %                                                           | >98 %                                                        | Pass    |
| Differential<br>Pressure<br>(Delta P)<br>EN<br>14683:2019+AC:2019<br>Annex C | Differential<br>pressure ≤ 6.0<br>mmH2O/cm²                     | Differential<br>pressure ≤ 6.0<br>mmH2O/cm²                  | Pass    |
| Flammability<br>16 CFR 1610                                                  | Class 1                                                         | Class 1                                                      | Pass    |

## Table 2 - Performance Testing

## Table 3 - Biocompatibility Testing

| Item                          | Subject Device                                                   | Result |
|-------------------------------|------------------------------------------------------------------|--------|
| Cytotoxicity<br>ISO 10993-5   | Under the conditions of the study, the device is non-cytotoxic.  | Pass   |
| Irritation<br>ISO 10993-10    | Under the conditions of the study, the device is non-irritating. | Pass   |
| Sensitization<br>ISO 10993-10 | Under the conditions of the study, the device is non-sensitizing | Pass   |

# 8.0 Clinical Test Conclusion

No clinical study is included in this submission.

## 9.0 Conclusion

The conclusion drawn from the non-clinical tests demonstrates that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device in K201479.

---

**Source:** [https://fda-staging.innolitics.com/device/K210813](https://fda-staging.innolitics.com/device/K210813)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).