K210804 · Ivoclar Vivadent, AG · KLE · Jun 7, 2022 · Dental
Device Facts
Record ID
K210804
Device Name
Adhese Universal DC, Cention Primer
Applicant
Ivoclar Vivadent, AG
Product Code
KLE · Dental
Decision Date
Jun 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
Missing tooth structure Defective restorations
Device Story
Adhese Universal DC is a dual-curing, single-component dental adhesive for enamel and dentin; compatible with self-etch, selective enamel etch, and etch & rinse techniques. Used in clinical dental settings by professionals for adhesive cementation of indirect restorations and direct restorative procedures. Device supplied in Free Stand Single Dose units; applicator coated with co-initiators for self-curing reaction. Clinician dips applicator into single dose unit to mix initiator and adhesive; applies to tooth surface. Dual-curing mechanism allows for polymerization in areas with limited light access. Benefits include improved versatility for indirect restorations compared to light-only curing adhesives.
Clinical Evidence
Bench testing only. Evaluated shear bond strength per ISO 29022 (≥20 MPa for self-curing mode). Biocompatibility evaluated per ISO 10993-1:2018, ISO 7405:2018, and ISO 14971:2012. No clinical data presented; equivalence supported by toxicological evaluation and historical clinical use of similar methacrylate-based dental materials.
Technological Characteristics
Dual-curing, single-component resin adhesive. Materials: HEMA, MDP, Bis-GMA, D3MA, ethanol, methacrylate-modified polyacrylic acid, silicon dioxide, potassium hydroxide, camphorquinone. Standards: EN 1641:2009, ISO 29022. Form factor: Single Dose units or bottle. No sterilization required (consumable).
Indications for Use
Indicated for patients requiring treatment for missing tooth structure or defective restorations.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2022
Ivoclar Vivadent, AG % Lori Aleshin Director of Quality & Regulatory Affairs Ivoclar Vivadent, Inc 175 Pineview Drive Amherst. New York 14228
Re: K210804
Trade/Device Name: Adhese Universal DC, Cention Primer Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE, LBH Dated: March 5, 2022 Received: March 10, 2021
Dear Lori Aleshin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K210804
Device Name Adhese® Universal DC
Indications for Use (Describe)
- Missing tooth structure
- Defective restorations
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Ivoclar Vivadent, a dental company. The logo features the company name in a stylized font, with the words "passion vision innovation" underneath. The logo also includes a series of green and blue dots above and to the right of the company name. The logo is simple and modern, and it conveys a sense of innovation and professionalism.
### Adhese Universal DC- K210804
| Contact: | Robert Streber, Director of Quality & Regulatory Affairs<br>Ivoclar Vivadent, Inc.<br>175 Pineview Drive<br>Amherst, New York 14228<br>716-264-2045<br>robert.streber@ivoclar.com |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company: | Ivoclar Vivadent, AG<br>Bendererstrasse 2, Schaan, FL-9494, Liechtenstein<br>+423-235-3535 |
| Date Prepared: | May 5, 2022 |
| Proprietary Name: | Adhese® Universal DC (K210804) |
| Primary Classification Name: | Agent, Tooth Bonding, Resin (872.3200)<br>(Classification Code KLE) |
| Secondary Classification Name: | Varnish, Cavity (872.3260)<br>(Classification Code LBH) |
| Predicate Device: | Adhese Universal (K133318) by Ivoclar Vivadent, AG |
Device Description: Adhese Universal DC is a dual-curing, single-component dental adhesive for enamel and dentin that is compatible with all etching techniques (self-etch, selective enamel etch, etch & rinse techniques). The areas of application include adhesive cementation of indirect restorations and direct restorative procedures. Adhese Universal DC is available in Free Stand Single Dose units. The applicator is coated with co-initiators required for the self-curing reaction. The initiator and adhesive are mixed together by dipping the applicator into the single dose unit containing the liquid Adhese Universal DC.
### Indications for Use:
- Missing tooth structure -
- -Defective restorations
Comparison to Predicate: The primary predicate device to which Adhese Universal DC has been compared is Ivoclar Vivadent, AG Adhese Universal (K133318).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Ivoclar Vivadent, a dental company. The logo features the company name in a stylized font, with the words "passion vision innovation" underneath. Above the company name are a series of green and blue dots and squares. To the right of the logo is the number 2.
| Predicate Device: | | Proposed Device: | Deviation | | |
|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|----|--|
| Device | Ivoclar Vivadent AG: Adhese<br>Universal (K133318) | Ivoclar Vivadent AG: Adhese<br>Universal DC (K210804) | Yes | No | |
| Indications for Use | - Direct-placed light-curing composite<br>and compomer restorations.<br>- Direct-placed core build-ups with light-<br>, self- and dual-curing composites.<br>- Repairs of fractured composite and<br>compomer restorations.<br>– Adhesive cementation of indirect<br>restorations with light- and dual-curing<br>luting composites.<br>- Sealing of prepared tooth surfaces<br>before temporary / permanent<br>cementation of indirect restorations.<br>- Desensitization of hypersensitive<br>cervical areas. | - Missing tooth structure<br>- Defective restorations | ☑ | | |
| Precaution<br>Measures/<br>Contraindications/<br>Processing<br>restrictions/ Side<br>effects | Contraindication<br>- Do not use Adhese Universal if the<br>patient is known to be allergic to any of<br>the materials' ingredients or if the<br>stipulated working technique<br>cannot be employed.<br>- Applications in which sufficient<br>illumination cannot be ensured (e.g.<br>cementation of root canal posts).<br>- Direct pulp cappings. | Contraindications:<br>Do not use Adhese Universal DC if the<br>patient is known to be allergic to any of<br>the materials' ingredients or if the<br>stipulated working technique cannot be<br>employed.<br>Limitations for use:<br>- Direct pulp capping<br>– Do not use Adhese Universal DC as a<br>primer for ceramic restorative materials.<br>A suitable ceramic primer<br>must be used (e.g. Monobond Etch &<br>Prime). | | ☑ | |
| Summary of<br>Indications,<br>Precaution<br>Measures/<br>Contraindications/<br>Processing<br>restrictions/<br>Side effects | In the Instructions for Use for the predicate Adhese Universal the application of the medical device<br>was assigned as "indications." According to the official definition of indications as clinical<br>conditions, the list of indications was updated in the IfU for Adhese Universal DC. The applications<br>are still part of the IfU under section "Description": Adhese Universal DC covers the same<br>application as Adhese Universal.<br>The contraindications are basically the same. | | | | |
| Working Principle | Adhese Universal is a light-curing<br>single-component dental adhesive for<br>enamel and dentin and is compatible<br>with all etching techniques (self-etch,<br>selective-enamel-etch and etch & rinse<br>techniques). | Adhese Universal DC is a dual-curing,<br>single-component dental adhesive for<br>enamel and dentin that is compatible<br>with all etching techniques (self-etch,<br>selective enamel-etch, etch & rinse<br>techniques).<br>The areas of application include<br>adhesive cementation of indirect<br>restorations and direct restorative<br>procedures.<br>Adhese Universal DC is available in<br>Free Stand Single Dose units. The<br>applicator is coated with co-initiators<br>required for the self-curing reaction. The<br>initiator and adhesive are mixed<br>together by dipping<br>the applicator into the single dose unit<br>containing the liquid Adhese Universal<br>DC. | ☑ | | |
| Summary of<br>Working Principle | The main difference between Adhese Universal and Adhese Universal DC is that the new device is<br>dual curing in comparison to the predicate which can be cured with light only. | | | | |
| Delivery<br>forms/dosage | Bottle of liquid 5g<br>Vivapen 0.5 ml and 2 ml | Single Dose 0.1 g<br>Bottle 6 g with a separate brush | ☑ | ☐ | |
| Summary of<br>Delivery<br>forms/dosage | Single doses and bottle available; difference in the amount only; | | | | |
| Storage Conditions | 24 months at 2-28 °C / 36-82 °F | 24 months at 2-28 °C / 36-82 °F | ☐ | ☑ | |
| Summary of<br>Storage Conditions | No difference. | | | | |
| Principles of<br>Operation | Step-by-step:<br>1. Isolation<br>2. Pulp protection/cavity liner<br>3. Conditioning with phosphoric acid gel (optional)<br>4. Handling VivaPen and bottle<br>5. Application of the adhesive<br>6. Light-curing the adhesive<br>7. Application of the restorative or luting composite | Step-by-step application:<br>1. Isolation<br>2. Pulp protection/cavity liner<br>3. Conditioning with phosphoric acid gel<br>4. Activation of the Free Stand Single Dose<br>5. Application of the Adhesive<br>6. Curing the adhesive<br>7. Application of the restorative or luting composite | ☑ | ☐ | |
| Summary<br>Principles of<br>operation | Basically, the step-by-step application is the same for both devices. | | | | |
| Chemical<br>Composition | Methacrylates, ethanol, water, highly<br>dispersed silicon dioxide, initiators and<br>stabilizers | HEMA, MDP, Bis-GMA, D3MA, ethanol,<br>methacrylate-modified polyacrylic<br>acid, silicon dioxide, potassium<br>hydroxide and camphorquinone. | ☑ | ☐ | |
| Summary of<br>Chemical<br>Composition | The chemical composition slightly differs between Adhese Universal and Adhese Universal DC, as<br>the new device is dual curing. The results of the Biocompatibility Assessment for Adhese Universal<br>DC is substantially equivalent to the results of the Biocompatibility Assessment for the predicate<br>device Adhese Universal. | | | | |
| Finished Device<br>Specification | EN 1641:2009- Dentistry – Medical<br>devices for dentistry – Materials<br><br>ISO 29022- Dentistry – Adhesive –<br>Notched-edge sheer bond strength test | EN 1641:2009- Dentistry – Medical<br>devices for dentistry – Materials<br><br>ISO 29022- Dentistry – Adhesive –<br>Notched-edge sheer bond strength test | ☐ | ☑ | |
| Summary of<br>Finished Device<br>Specification | No difference. | | | | |
| Sterilization | Not applicable. No sterilization<br>recommended. | Not applicable. No sterilization<br>recommended. | ☐ | ☑ | |
| Single use | Consumable material | Consumable material | ☐ | ☑ | |
| Performance-<br>Specification<br>Adhesion | Shear bond strength (Dentin): ≥25 MPa | Shear bond strength (Dentin) ≥20 MPa<br>Self-curing | ☑ | ☐ | |
| Summary of<br>Performance<br>Specification | The specification for Adhese Universal DC describes lower values (20 MPa) for shear bond<br>strength (dentin) compared to Adhese Universal (25 MPa) due to the self-curing mode of Adhese<br>Universal DC which generally shows lower values. | | | | |
| Biocompatibility | These values are comparable to the specification of Adhese Universal.<br>evaluated for Biocompatibility according<br>to ISO 10993-1:2018, ISO 7405:2018<br>and ISO 14971:2012 | evaluated for Biocompatibility according<br>to ISO 10993-1:2018, ISO 7405:2018<br>and ISO 14971:2012 | ☐ | ☑ | |
| Discussion of<br>Biocompatibility | The results of the Biocompatibility Assessment for Adhese Universal DC is substantially<br>equivalent to the results of the Biocompatibility Assessment for the predicate device | | | | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Ivoclar Vivadent logo. The logo is blue and green, and it features the company name in a stylized font. There are green and blue dots above the company name. There is a small number 3 on the right side of the logo.
on
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Ivoclar Vivadent. The logo is blue and green, with the words "ivoclar vivadent" in a stylized font. Below the company name is the slogan "passion vision innovation" in a smaller font. To the right of the company name is the number 4.
### Substantial Equivalence to the predicate:
Adhese Universal and Adhese Universal DC are both dental adhesives. The main difference between the two devices is that Adhese Universal is light-curing compared to Adhese Universal DC which is dual-curing. Therefore, Adhese Universal DC is substantially equivalent to the predicate device, Adhese Universal.
### Differences:
The main difference between Adhese Universal and Adhese Universal DC is that the new device is dual-curing compared to the light-curing Adhese Universal. As the new device is dual-curing the chemical composition has been slightly adapted for Adhese Universal DC.
### Non-clinical performance testing:
Bench testing was performed to test the physical property (shear bond strength) included in the Finished Device Specification for the subject device according to EN ISO 1641:2009 -Dentistry- Medical devices for dentistry - Materials and ISO 29022- Dentistry - Adhesive -Notched-edge sheer bond strength test.
### Biocompatibility:
The subject device was also evaluated for Biocompatibility according to ISO 10993-1:2018, ISO 7405:2018 and ISO 14971:2012. Uncured Adhese Universal as well as extracts of polymerized Adhese Universal DC possess a cytotoxic potential in the XTT assay. The clinical experience with a previous generation of the product (Adhese Universal) indicates no adverse effect in clinical use. The product, particularly in the uncured state may cause sensitization against methacrylates. This is typical for all resin-based dental materials. On the basis of the clinical experience with a previous generation of the product it can be concluded that the products do not cause oral mucosal or gingival irritation. There is no material mediated pyrogenicity. According to the information available, the product have no acute nor subacute system toxicity. Subchronic or chronic toxicity is not expected. According to the data available Adhese Universal DC is not genotoxic nor carcinogenic. According to the current knowledge, it can be concluded that Adhese Universal DC does not cause adverse reactions upon implantation in teeth. Adhese Universal DC are expected not to cause pulp or dentine damage. There is no indication that toxic degradation products might be released during body contact.
On the basis of the toxicological evaluation of the product and the longstanding worldwide clinical use of similar materials it can be concluded that the benefits provided by the final product will exceed any potential risks produced by device materials providing that the instructions for use have been followed. The results of the Biocompatibility Assessment for Adhese Universal DC is substantially equivalent to the results of the Biocompatibility Assessment for the predicate device Adhese Universal.
### Conclusion:
Adhese Universal and Adhese Universal DC are both dental adhesives. The main difference between the two devices is that Adhese Universal is light-curing compared to Adhese Universal DC which is dual-curing. The indications / contraindications are basically the same. The device performance is substantially equivalent. The biocompatibility and storage stability of the new formulation was fully assessed and is substantially equivalent to Adhese Universal.
Therefore, Adhese Universal DC is substantially equivalent to the predicate device, Adhese Universal.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.