K210782 · Outset Medical, Inc. · FJK · Nov 26, 2021 · Gastroenterology, Urology
Device Facts
Record ID
K210782
Device Name
Tablo Cartridge
Applicant
Outset Medical, Inc.
Product Code
FJK · Gastroenterology, Urology
Decision Date
Nov 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tablo Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extra-corporeal access during hemodialysis. The Tablo Cartridge is compatible only with the Tablo Hemodialysis System.
Device Story
Tablo Cartridge is single-use, disposable arterial/venous blood tubing set; functions as extracorporeal blood circuit for Tablo Hemodialysis System. Cartridge inserts onto console front panel for each treatment. Device facilitates blood flow between patient and hemodialysis machine. E-beam sterilized. Used in clinical or home settings where Tablo Hemodialysis System is operated. Healthcare providers or patients use the cartridge to establish blood access; output is the physical blood path for dialysis. Benefits include safe, disposable blood access for hemodialysis therapy.
Clinical Evidence
No clinical studies were performed. Substantial equivalence supported by bench performance testing, sterilization process validation (ISO 11137-1/2, ISO 11737-1/2, USP <71>, ANSI/AAMI ST72), shelf-life verification (ISO 11607-1/2), and biocompatibility testing (ISO 10993-1).
Technological Characteristics
Single-use, disposable blood tubing set; E-beam sterilized. External communicating device with prolonged contact (>24 hours to 30 days) with blood path. Complies with ISO 11137, ISO 11737, ISO 11607, ISO 10993-1, and USP <71> standards. Interfaces with Tablo Hemodialysis System console.
Indications for Use
Indicated for patients requiring hemodialysis; provides extracorporeal blood access via arterial and venous bloodline set. Compatible only with Tablo Hemodialysis System.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a larger font size than the name. The name is written as "U.S. FOOD & DRUG ADMINISTRATION".
November 26, 2021
Outset Medical, Inc. Claire Bao Sr. Regulatory Affairs Specialist 3052 Orchard Drive San Jose. California 95134
Re: K210782
Trade/Device Name: Tablo® Cartridge Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: FJK Dated: October 27, 2021 Received: Ocobter 28, 2021
Dear Claire Bao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Glenn B. Bell, Ph.D. Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210782
Device Name Tablo Cartridge
Indications for Use (Describe)
The Tablo Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extra-corporeal access during hemodialysis. The Tablo Cartridge is compatible only with the Tablo Hemodialysis System.
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(21 CFR 807.92)
### I. SUBMITTER
| Name: | Outset Medical, Inc.<br>3052 Orchard Drive<br>San Jose, CA 95134 |
|------------------|------------------------------------------------------------------|
| Phone: | (669) 231-8200 |
| Primary Contact: | Jennifer Mascioli-Tudor |
| Prepared by: | Claire Bao |
| Date Prepared: | March 24, 2021 |
## II. DEVICE
| Trade/Device Name: | Tablo Cartridge |
|-----------------------|-------------------------------------|
| Common or Usual Name: | Blood tubing set |
| Regulation Name: | Hemodialysis system and accessories |
| Regulation Number: | 21 CFR § 876.5820 |
| Product Code: | FJK |
| Regulatory Class: | II |
## III. PREDICATE DEVICE
Tablo Cartridge (K190793) The predicate device has not been subjected to a design related recall. No reference devices were used in this submission.
## IV. DEVICE DESCRIPTION
The Tablo Cartridge is a single use blood tubing set attached to an organizer tray. The E-beam sterilized and disposable cartridge is inserted onto the front panel of the console for each dialysis treatment (Figure 1).
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a medical device with a cartridge installed on the console front panel. The device has a screen on top and a door that opens to reveal the cartridge and other components. The cartridge is connected to the device with tubes and wires. The device is on wheels, making it easy to move around.
Figure 1. Tablo Cartridge Inserted into Tablo Hemodialysis Delivery System
### V. INDICATIONS FOR USE
The Tablo Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extra-corporeal access during hemodialysis. The Tablo Cartridge is compatible only with the Tablo Hemodialysis System.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The modified Tablo Cartridge has the same fundamental technology, principle of operation, and principal functionality as the predicate device. The Tablo Cartridge is a blood tubing set which functions as part of the extracorporeal blood system of the Tablo Hemodialysis machine.
The Tablo Cartridge with E-beam sterilization (Modified Device) and the Predicate Device are equivalent in technological characteristics:
- . Intended use - To provide extra-corporeal access during hemodialysis.
- Operating principle – Inserted onto Tablo Console as part of the extracorporeal blood system.
- . Compatibility - Interface with the Tablo Console (Hemodialysis System).
The following differences exist between the subject and predicate device:
- Sterilization method changed from Ethylene Oxide (EO) gas to E-beam radiation.
- . Minor design changes for improvement.
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### E-Beam Sterilization Process Validation Testing
The E-beam sterilization process validation testing for the Tablo Cartridge was conducted in accordance with FDA guidance document, "Hemodialysis Blood Tubing Sets – Premarket Notification [510(k)] Submissions," dated April 23, 2008 and "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," dated January 21, 2016.
{5}------------------------------------------------
The Tablo Cartridge complies with following standards:
- ISO 11137-1:2006 Sterilization of health care products Radiation Part 1: Requirements for ● development, validation and routine control of a sterilization process for medical devices
- . ISO 11137-2:2013 Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose
- ISO 11737-1:2018 Sterilization of health care products – Microbiological methods – Part 1: Determination of a population of microorganisms on products
- . ISO 11737-2:2019 Sterilization of health care products – Microbiological methods – Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- . USP <71> Sterility Tests standard
- . ANSI/AAMI ST72:2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing
The E-beam sterilization validation included the following tests to support the sterilization validation:
- Bioburden Recovery Test
- Bioburden Determination
- Verification Dose Determination
- Dose Map Study
- Method Suitability Determination
- Test of Sterility
### Shelf-Life Testing (Accelerated Aging and Transit)
The shelf-life verification for the Tablo Cartridge was conducted as recommended by FDA's guidance document, "Hemodialysis Blood Tubing Sets – Premarket Notification [510(k)] Submissions," dated April 23, 2008.
The shelf-life verification testing included the following tests to ensure that the Tablo Cartridge complies with the ISO 11607-1:2019 and ISO 11607-2:2019 standards:
- . Accelerated aging testing
- Climatic conditioning testing
- Gross leak detection (Bubble) testing
- . Seal strength (Peel) testing
- Package performance testing
- Product functional testing
### Biocompatibility Testing
The biocompatibility evaluation for the Tablo Cartridge was conducted in accordance with the FDA guidance document, "Hemodialysis Blood Tubing Sets – Premarket Notification [510(k)] Submissions," dated April 23, 2008.
The battery of testing included the following tests and continues to be supportive for the modified device, per ISO 10993-1:2018, as an external communicating device with prolonged direct and indirect contact (>24 hours to 30 days) with the blood path:
{6}------------------------------------------------
- Cytotoxicity
- Sensitization
- Irritation
- Acute systemic toxicity
- Hemocompatibility
- Genotoxicity
- Pyrogenicity
### Bench Performance Testing
The bench performance testing for the Tablo Cartridge was conducted in accordance with the FDA guidance document, "Hemodialysis Blood Tubing Sets – Premarket Notification [510(k)] Submissions," dated April 23, 2008.
Nonclinical bench performance tests were conducted to demonstrate that the Tablo Cartridge meets the system requirements and performs as intended.
### Animal Study
No animal studies were performed in support of the modifications.
### Clinical Studies
No clinical studies were performed in support of the modifications.
### VIII. CONCLUSION
The performance testing demonstrates that the Tablo Cartridge meets all performance specifications and complies with applicable standards and FDA Guidance Documents. The Tablo Cartridge (Modified Device) is substantially equivalent to the Predicate Device, and the minor difference between the Modified and the Predicate Device do not raise any new or different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.