The Medicalgorithmics' Q Patch is intended to be used by patients who have a demonstrated need for extended cardiac monitoring and patients with symptoms that may be due to cardiac arrhythmias such as, dizziness, lightheadedness, shortness of breath, palpitations, dyspnea (shortness of breath), anxiety, syncope of unknown etiology in which arrhythmias are suspected or need to be excluded. It is indicated for use on adult patients. The sensor records single ECG channel for up to 15 days and can be used on patients with implanted pacemakers but is not intended to record pacemaker activity.
Device Story
Q Patch is a single-channel ambulatory ECG recorder; snaps onto off-the-shelf Ag/AgCl electrodes; worn on sternum for up to 15 days. Powered by non-rechargeable CR2032 battery. Operated by patients (home/daily life) or professionals (healthcare facilities). Q Patch Mobile App initiates/monitors recording status; Q Patch Downloader (PC) transfers stored ECG data via USB to remote storage. Device does not perform automated analysis; data intended for review by trained professionals using third-party software (e.g., PocketECG IV). Defibrillation safe but not defibrillation proof. Benefits include extended, continuous cardiac monitoring for arrhythmia detection in non-hospitalized patients.
Clinical Evidence
Bench testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and software validation. Biocompatibility testing (ISO 10993-1) confirmed safety for prolonged skin contact. Usability testing involved 60 participants. A clinical study with 30 participants compared Q Patch ECG signals against a reference device, demonstrating no clinically significant performance differences and confirming suitability for human interpretation.
Technological Characteristics
Single-channel ECG recorder; Ag/AgCl electrodes; CR2032 3V lithium battery. Dimensions/form factor: sternum-worn patch. Connectivity: USB for data download, mobile app for status/setup. Sterilization: Not specified. Software: Moderate level of concern for device, minor for apps. Standards: ANSI AAMI ES60601-1, IEC 60601-1-11, IEC 62304, IEC 60601-2-47, IEC 60601-1-2.
Indications for Use
Indicated for adult patients requiring extended cardiac monitoring due to symptoms of potential arrhythmias (dizziness, lightheadedness, dyspnea, palpitations, anxiety, syncope). Contraindicated for patients requiring inpatient monitoring for life-threatening arrhythmias (e.g., ventricular fibrillation, sustained ventricular tachycardia) or asystole.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Unified Arrhythmia Diagnostic System PocketECG IV (K193104)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.
June 28, 2022
Medicalgorithmics S.A. % Agnieszka Romowicz Product Compliance Manager Medicalgorithmics US Holding Corporation Corporation Service Company 251 Little Falls Drive Wilmington, Delaware 19808
Re: K210758
Trade/Device Name: Q Patch Regulation Number: 21 CFR 870.2800 Regulation Name: Electrocardiograph, Ambulatory (Without Analysis) Regulatory Class: Class II Product Code: MWJ
Dear Agnieszka Romowicz:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 2, 2022. Specifically, FDA is updating the SE Letter correspondent contact information as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Shih Kozen, OHT2: Office of Cardiovascular Devices, (301) 796-5813, Jennifer.Shih(@fda.hhs.gov.
Sincerely,
Jennifer W.Shih-S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 2, 2022
Medicalgorithmics S.A. % Przemyslaw Tadla Operations & Product Compliance Director Medicalgorithmics US Holding Corporation Corporation Service Company 251 Little Falls Drive Wilmington, Delaware 19808
Re: K210758
Trade/Device Name: O Patch Regulation Number: 21 CFR 870.2800 Regulation Name: Electrocardiograph, Ambulatory (Without Analysis) Regulatory Class: Class II Product Code: MWJ Dated: April 29, 2022 Received: May 2, 2022
Dear Przemyslaw Tadla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210758
Device Name Q Patch
### Indications for Use (Describe)
The Medicalgorithmics' Q Patch is intended to be used by patients who have a demonstrated need for extended cardiac monitoring and patients with symptoms that may be due to cardiac arrhythmias such as, dizziness, lightheadedness, shortness of breath, palpitations, dyspnea (shortness of breath), anxiety, syncope of unknown etiology in which arrhythmias are suspected or need to be excluded. It is indicated for use on adult patients. The sensor records single ECG channel for up to 15 days and can be used on patients with implanted pacemakers but is not intended to record pacemaker activity.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:100%;"> <b> <span style="font-size:100%;">☑</span> </b> </span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Medical Algorithms. The logo has the word "MEDICAL" in gray and the word "Algorithms" in red. Below the words is the phrase "Innovative Solutions in Medicine" in a smaller font.
June 2, 2022
# 510(k) Summary
#### l. Submitter's name and address:
| Medicalgorithmics S.A. | Aleje Jerozolimskie 81, |
|------------------------|----------------------------------|
| | 02-001 Warsaw, Poland |
| Contact Person: | Agnieszka Romowicz |
| Phone: | (+1) 302 2615184 |
| Email: | a.romowicz@medicalgorithmics.com |
| Date Prepared: | 2022-06-02 |
#### II. Device
| Trade/Device name: | Q Patch |
|--------------------|---------------------------------------------------------|
| Type: | P5HP-AA-ADS |
| Regulation number: | 21 CFR 870.2800 |
| Regulation name: | Electrocardiograph, Ambulatory (Without Analysis) |
| Regulatory Class: | Class II |
| Product codes: | MWJ |
| Classification: | Electrocardiograph, Ambulatory (Without Analysis) - MWJ |
#### III. Predicate Device
The selected predicate device is:
myPatch®sl, K163535
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Medical Algorithms. The logo has the word "MEDICAL" in black and "ALGORITHMIC" in red. Below the word, it says "INNOVATIVE SOLUTIONS IN MEDICINE".
#### Device description IV.
The Q Patch is a single channel ECG recorder. The device is intended to be placed on the sternum (in the middle of the chest). The Q Patch ECG recorder snaps onto the two off-the-shelf electrodes (Solid gel, Ag/AgCl) and records patient's ECG for up to 15 days powered from single disposable, non-rechargeable battery. The Q Patch Mobile Application is used to initiate recording session, for checking Q Patch status and to finish/stop the session if needed. When the recording is finished, the ECG data stored in Q Patch's memory can be downloaded through a USB port using the Q Patch Downloader (PC application).
The product and it's accessories are intended to be used by:
- Patients: adults which may require long-term, continuous heart activity ● monitoring to possible cardiac arrhythmias. Patients use the Q Patch recorder and Q Patch Mobile Application operating in patient mode
- o Professional users:
- o Healthcare professionals: physician or other qualified healthcare professionals (Nurses/RNs and Certified Medical Assistants (CMA/AAMA))
- o Technicians: electronic device maintenance staff
Professionals use the Q Patch recorder, Q Patch Mobile Application operating in professional mode and Q Patch Downloader.
Use environment includes Healthcare facilities, Home, Office, Car, Outside under proper clothing.
The Q Patch recorder is accompanied with following accessories:
- Q Patch Mobile Application (Mobile application) - The Q Patch Mobile Application is intended to be used for interacting with the Q Patch recorder for transmission patient data to set up ECG recording session and displaying the Q Patch's status. The application is obligatory for healthcare professionals and optional for patients.
- Q Patch Downloader (PC application) The intended use of the Q Patch ● Downloader application is to download the ECG data stored in the Q Patch memory after recording session is completed and then to upload files containing ECG data to the predefined remote storage.
The device is powered from a lithium coin non-rechargeable battery type CR2032, rated 3V, 230mAh. Battery capacity is sufficient to provide power throughout entire ECG monitoring period (15-days). The battery should be replaced by technicians of monitoring center. The Q Patch is designed to be defibrillation safe device, meaning that the Q Patch does not impact the defibrillation procedure performance and does not introduce additional risk. However, the Q Patch is not defibrillation proof device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Medical Algorithms. The logo has the word "MEDICAL" in gray and the word "Algorithms" in red. Below the words is the phrase "Innovative Solutions in Medicine" in a smaller font. The logo is simple and modern.
thus it may not function or recovery after patient defibrillation. The Q Patch defibrillation safe performance has been proved via laboratory testing.
Although ECG analysis and visualization software are not included in this product. Q Patch, Q Patch Mobile Application and Q Patch Downloader are designed to be integrated with third-party data analysis and visualization software.
The ECG recording can be transferred in an available file format and can be reviewed using software being a part of Unified Arrhythmia Diagnostic System PocketECG IV (K193104). The ECG data can be reviewed by healthcare professionals or ECG technicians trained in the identification and treatment of arrhythmia events. However, the subject device is not intended for automated analysis and then arrhythmia classification algorithm being a part of Unified Arrhythmia Diagnostic System PocketECG IV (K193104) system should not be used to produce beat annotations to be considered during patient diagnosis.
Potential patient cardiac abnormalities, must be confirmed by a qualified ECG technician or by a physician with other relevant clinical information. The Q Patch does not provide interpretive statements.
The performance and integration tests were carried out to prove performance of Q Patch operation with Unified Arrhythmia Diagnostic System PocketECG IV (K193104).
The Q Patch is not intended for patients who should be hospitalized, patients with potentially life-threatening arrhythmias who require inpatient monitoring such as primary ventricular fibrillation or sustained ventricular tachycardia and for individuals where asystole represents the terminal rhythm of a cardiac arrest.
#### Indications for use V.
The Medicalgorithmics' Q Patch is intended to be used by patients who have a demonstrated need for extended cardiac monitoring and patients with symptoms that may be due to cardiac arrhythmias such as, dizziness, lightheadedness, shortness of breath, palpitations, dyspnea (shortness of breath), anxiety, syncope of unknown etiology in which arrhythmias are suspected or need to be excluded. It is indicated for use on adult patients. The sensor records single ECG channel for up to 15 days and can be used on patients with implanted pacemakers but is not intended to record pacemaker activity.
The Indications for Use statement for the Q Patch device is not identical to the predicate device; however, the differences do not alter the intended use nor do they affect the safety and effectiveness of the device relative to the predicate.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the text "VI. Technological comparison to predicate device". The text is in black font and is left-aligned. The text is likely a heading or title for a document or section.
The intended use and technological features of the proposed Q Patch do not substantially differ from the legally marketed predicate device. Both devices are comprised of essentially identical configuration where the main component is a recorder. The operational characteristics are identical for the proposed and predicate device as they are both a small Holter recorders operated by professionals to acquire ECG data from a patient at a patient setting or wherever the patient may go. Both sensors are placed on the patient's sternum and record ECG signal continuously. The patient's ECG signal is managed similarly for both devices as it is recorded by the ECG recorder and at the end of recording can be downloaded through a USB port.
There are some differences between these two devices:
- The patient population indicated for the predicate device includes adults and ● all pediatric subgroups, whereas the pediatric use is restricted for the Q Patch. The subject device can be used in adults in the same way as the predicate device.
- The predicate device records ECG in 1 or 2 channel mode for up to 14 days or in 3 channel mode for up to 9 days. The subject device records ECG in one of these modes - single channel mode; for up to 15 days. Longer monitoring time is a minor difference and does not pose any additional risks and has no impact on the substantial equivalence of the product.
- The predicate sensor uses a rechargeable Li-Ion battery versus off the shelf, disposable, non-rechargeable lithium coin battery used in the subject device.
- When the recording is finished the ECG data stored in Q Patch's memory can ● be downloaded through a USB port using the Q Patch Downloader (PC application). While in the predicate device the patient's ECG is recorded to the device and then transferred via the USB data transfer cable to Holter analysis system (which is not a part of predicate's 510(k))
- . The subject device is also equipped with an additional application the Q Patch Mobile Application for interacting with the Q Patch recorder for transmission patient data to set up ECG recording session and displaying the Q Patch's status.
Differences between the predicate device and Q Patch are minor and have no impact on the substantial equivalence of the product. Descriptive characteristics and performance testing are used to demonstrate substantial equivalence. The 510(k) Premarket Notification demonstrates that the proposed Q Patch is substantially equivalent to the predicate device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "MEDICALgorithmics". The word "MEDICAL" is in gray, with a dark gray square to the left of it. The word "Algorithmics" is in red. Below the company name is the text "INNOVATIVE SOLUTIONS IN MEDICINE" in gray.
#### Performance data VII.
The following performance data were provided in support of the substantial equivalence determination.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Q Patch device. The Q Patch was designed and tested for compliance with the applicable clauses of the following standards:
- ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And ● A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod);
- ANSI AAMI HA60601-1-11:2015, Medical Electrical Equipment -- Part 1-11: ● General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment (IEC 60601-1-11:2015 Mod);
- IEC 62304:2006/A1:2015, Medical Device Software Software life cycle ● processes [Including Amendment 1 (2016)];
- ANSI AAMI IEC 60601-2-47:2012 /(R)2016, Medical Electrical Equipment -- Part ● 2-47: Particular Requirements For The Basic Safety And Essential Performance Of Ambulatory Electrocardiographic Systems:
- ANSI AAMI IEC 60601-1-2:2014, Medical Electrical Equipment Part 1-2: General . Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (Edition 4);
# Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Mav 11, 2005). The software for Q Patch device was considered a Moderate Level of Concern as a design flaw or failure could directly result in minor injury to the patient or operator. Both, Q Patch Mobile Application and The Q Patch Downloader (MDDS), are classified as Minor
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for MedicAlgoritmics. The logo has the word "MEDIC" in gray and the word "ALgoritmics" in red. Below the words is the phrase "INNOVATIVE SOLUTIONS IN MEDICINE" in a smaller font.
Level of Concern, which means that potential failures or latent design flaws are unlikely to cause any injury to the patient or operator.
### Usability Engineering
The following recommendations and requirements related to the Human Factors and Usability Engineering to Medical Devices were applied:
- . The FDA Guidance for Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016)
- IEC 60601-1-6 Edition 3.1 2013-10, Medical Electrical Equipment Part 1-6: ● General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability;
- ANSI AAMI IEC 62366-1:2015, Medical devices Part 1: Application of usability ● engineering to medical devices;
There was a total of 60 participants involved in the summative usability testing of the Q Patch recorder and accessories. The usability testing covers all aspects of the product including the Q Patch recorder, Q Patch Mobile Application, and Q Patch Downloader.
### Biocompatibility
The biocompatibility evaluation for the Q Patch device was conducted in accordance with the FDA Guidance for Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (September 4, 2020), and International Standard ISO 10993-1 Fifth edition 2018-08 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Q Patch device is intended to direct skin contact up to 15 days, according to ISO 10993-1 is category B- prolonged time contact therefore the following requirements were found applicable:
- Cytotoxicity
- Sensitization ●
- Irritation
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Medical Algorithms. The logo has the word "MEDICAL" in gray and the word "ALGORITHMS" in red. Below the words is the phrase "INNOVATIVE SOLUTIONS IN MEDICINE" in gray.
### Performance tests
Tests which confirmed that the subject device, when used in conjunction with compatible electrodes, adequately adheres to the patient's skin throughout the wearing period were carried out. Study involving 18 participants, proved that the standard off-the-shelf Ag/AgCl electrodes when replaced regularly can be used with the Q Patch device to record ECG for up to 15 days.
Clinical study where ECG signals were recorded by the subject device and reference device simultaneously have been performed. The results of this analysis confirmed that there were no clinically significant differences in performance of the Q Patch against reference device. It has been demonstrated that non-standard ECG lead recorded by the Q Patch device is appropriate for ECG evaluation performed by the trained human interpreter. There were 30 participants involved in this study.
The subject device acquires ECG data and does not perform automatic arrhythmia analysis.
### Summary
Based on the performance study results the Q Patch system was found to have a safety and effectiveness profile that is similar to the predicate device and reference devices.
#### VIII. Conclusion
The Q Patch, type P5HP-AA-ADS is substantially equivalent to the predicate device as supported by the descriptive information and the performance testing. The verification and validation of hardware and software demonstrate that the Q Patch devices perform as intended in the specified use conditions. The performance data demonstrate that the Q Patch device performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.