The OsteoCentric 2.4mm Cannulated Screw is intended for fixation of fractures and non-unions of small bones and small bone arthrodesis. Examples include, but are not limited to scaphoid and other carpal fractures, metacarpal and phalangeal fusions, osteotomies, and bunionectomies. The OsteoCentric Trauma 7.0mm and 8.0mm Cannulated Screws are intended for fracture fixation of large bones and large bone fragments, femoral neck fractures, slipped capital femoral epiphyses, as an adjunct to DHS in basilar neck fractures, tibial plateau fractures, ankle arthrodesis, pediatric femoral neck fractures, intercondylar femur fractures, and subtalar arthrodesis. The OsteoCentric 7.0mm and 8.0mm Cannulated Nut is indicated for fracture fixation where fracture compression is desired, and where insertion of the nut-and-bolt system does not increase the risk of morbidity due to the need for surgical access to both sides of the compressed area. The device may also be used when compression between separate bones is desired (i.e.- the syndesmosis) as long as there is not increased morbidity associated with the insertion of a nut-and-bolt system, as discussed above.
Device Story
Cannulated screws and nuts provide mechanical fixation for bone fractures, non-unions, and arthrodesis. Devices are implanted by surgeons in clinical/OR settings to stabilize bone fragments or provide compression across fracture sites or joints. Screws are cannulated to allow placement over guide wires. Nuts are used in conjunction with bolts to achieve compression. Output is physical stabilization of bone segments, facilitating healing. Benefits include reduced fracture mobility and support for bone fusion. Devices are provided non-sterile and require sterilization before use.
Clinical Evidence
No clinical data. Bench testing only. Mechanical performance evaluated via engineering analysis of torsional strength and pullout resistance, and driving torque testing per ASTM F543. Cleaning and sterilization validations performed per AAMI TIR12:2010 and ANSI/AAMI TIR30:2011.
Technological Characteristics
Materials: Stainless steel (ASTM F138) or Titanium (ASTM F136/F1295). Design: Cannulated screws and nuts in various diameters/lengths. Principle: Mechanical bone fixation/compression. Connectivity: None. Sterilization: Provided non-sterile; validated per AAMI TIR12:2010.
Indications for Use
Indicated for fixation of fractures, non-unions, and arthrodesis of small bones (e.g., scaphoid, carpal, metacarpal, phalangeal) and large bones (e.g., femoral neck, tibial plateau, ankle, subtalar). Includes use for osteotomies, bunionectomies, slipped capital femoral epiphyses, and intercondylar femur fractures. Nut component indicated for fracture fixation requiring compression where surgical access to both sides is feasible without increased morbidity.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
April 14, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two main elements: the Department of Health and Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health and Human Services logo is a stylized representation of an eagle. The FDA acronym is in a blue square, and the full name of the agency is written in blue letters.
OsteoCentric Technologies % Nathan Wright Engineer & Regulatory Special Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918
Re: K210754
Trade/Device Name: Cannulated Fasteners and Nuts Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, HTN Dated: March 12, 2021 Received: March 15, 2021
Dear Nathan Wright:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement on last page.
510(k) Number (if known) K210754 Device Name
Cannulated Fasteners and Nuts
Indications for Use (Describe)
The OsteoCentric 2.4mm Cannulated Screw is intended for fixation of fractures and non-unions of small bones and small bone arthrodesis. Examples include, but are not limited to scaphoid and other carpal fractures, metacarpal and phalangeal fusions, osteotomies, and bunionectomies.
The OsteoCentric Trauma 7.0mm and 8.0mm Cannulated Screws are intended for fracture fixation of large bones and large bone fragments, femoral neck fractures, slipped capital femoral epiphyses, as an adjunct to DHS in basilar neck fractures, tibial plateau fractures, ankle arthrodesis, pediatric femoral neck fractures, intercondylar femur fractures, and subtalar arthrodesis.
The OsteoCentric 7.0mm and 8.0mm Cannulated Nut is indicated for fracture fixation where fracture compression is desired, and where insertion of the nut-and-bolt system does not increase the risk of morbidity due to the need for surgical access to both sides of the compressed area. The device may also be used when compression between separate bones is desired (i.e.- the syndesmosis) as long as there is not increased morbidity associated with the insertion of a nut-and-bolt system, as discussed above.
Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
| 510(K) SUMMARY |
|----------------|
| |
| Submitter's Name: | OsteoCentric Technologies |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address: | 75 West 300 North, Suite #150<br>Logan, UT 84321 |
| Submitter's Telephone: | 1-800-969-0639 |
| Contact Person: | Nathan Wright, MS<br>Empirical Testing Corp.<br>719-351-0248<br>nwright@empiricaltech.com |
| Date Summary was Prepared: | April 12, 2021 |
| Trade or Proprietary Name: | Cannulated Fasteners and Nuts |
| Common or Usual Name: | Screw, Fixation, Bone (primary) and<br>Washer, Bolt Nut |
| Device Classification Name: | Smooth or Threaded Metallic Bone Fixation Fastener<br>(primary) and<br>Single/multiple component metallic bone fixation<br>appliances and accessories |
| Classification: | Class II per 21 CFR §888.3040 (primary) and<br>§888.3030 |
| Product Code: | HWC (primary), HTN |
| Classification Panel: | Orthopedic |
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The OsteoCentric Technologies Cannulated Fasteners and Nuts within the product line of the SMV Scientific Cannulated Screws consists of cannulated screws in a variety of lengths and diameters to accommodate different anatomic sizes of patients. The screws and nuts are provided non-sterile. The screws and nuts are manufactured from Stainless Steel per ASTM F138 or from Titanium per ASTM F136 or F1295.
### INDICATIONS FOR USE
The OsteoCentric 2.4mm Cannulated Screw is intended for fixation of fractures and non-unions of small bones and small bone arthrodesis. Examples include, but are not limited to scaphoid and other carpal fractures, metacarpal and phalangeal fusions, osteotomies, and bunionectomies.
The OsteoCentric Trauma 7.0mm and 8.0mm Cannulated Screws are intended for fracture fixation of large bones and large bone fragments, femoral neck fractures, slipped capital femoral epiphyses, as an adjunct to DHS in basilar neck fractures, tibial plateau fractures, ankle arthrodesis, pediatric femoral neck fractures, intercondylar femur fractures, and subtalar arthrodesis.
The OsteoCentric 7.0mm and 8.0mm Cannulated Nut is indicated for fracture fixation where fracture compression is desired, and where insertion of the nut-and-bolt system does not increase
{4}------------------------------------------------
the risk of morbidity due to the need for surgical access to both sides of the compressed area. The device may also be used when compression between separate bones is desired (i.e. - the syndesmosis) as long as there is not increased morbidity associated with the insertion of a nutand-bolt system, as discussed above.
## TECHNICAL CHARACTERISTICS
The indications for use for the OsteoCentric Cannulated Fasteners and Nuts are same as that of the SMV Scientific Cannulated Bone Screws. The OsteoCentric Cannulated Fasteners and Nuts are made from stainless steel per ASTM F138 and from titanium per ASTM F136 or F1295. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:
- . Indications for Use
- Materials of manufacture .
- Principles of operation
- Sizes
#### Table 5-1 Predicate Devices
| 510k Number | Trade or Proprietary or Model Name | Manufacturer | Predicate Type |
|-------------|-------------------------------------------------------------------------|--------------------------------|----------------|
| K170021 | SMV Scientific Cannulated Bone Screws | SMV Scientific | Primary |
| K202680 | ExtremiFix Mini & Small<br>Cannulated Screw System | OsteoMed LLC | Additional |
| K191344 | Arthrex Mini Comprehensive<br>Fixation System – 2.0mm & 2.4mm<br>Module | Arthrex Inc. | Additional |
| K000080 | Asnis III Cannulated Screw System | Howmedica<br>Osteonics Corp. | Additional |
| K081813 | CompresSURE Fracture Repair<br>System | U.S. Implant<br>Solutions, LLC | Additional |
| K150981 | SMV Scientific Bone Screws | SMV Scientific | Additional |
# PERFORMANCE TESTING SUMMARY
In support of this Special 510(k) Device Modification Premarket Notification, OsteoCentric Technologies has conducted mechanical testing with engineering analyses and geometric comparisons to demonstrate that the modifications to the SMV Scientific Cannulated Screws provide adequate and substantially equivalent mechanical performance (torsional strength, driving torque, and pullout strength) for their intended use. Non-clinical analysis and testing included:
- Engineering analysis of torsional strength .
- Engineering analysis of pullout resistance .
{5}------------------------------------------------
- Driving torque testing per ASTM F543 .
- Instrument Cleaning Validation per AAMI TIR12:2010 and ANSI/AAMI TIR30:2011 .
- Sterilization Validation per AAMI TIR12:2010 .
# CONCLUSION
The subject OsteoCentric Cannulated Fasteners and Nuts are very similar to previously cleared Cannulated Screws (K170021). The indications for use for the subject are the same to that of the previously cleared predicates. The subject screws have similar intended uses, indications, technological characteristics, and principles of operation as the predicate devices. The modifications of additional sizes and inclusion of nuts raise no new types of safety or effectiveness questions. The overall technology characteristics and mechanical performance evaluation lead to the conclusion that the OsteoCentric Cannulated Fasteners and Nuts are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.