A retrospective validation study was conducted using 186 clinical echocardiography studies to assess the accuracy of the automated ejection fraction software compared to the biplane method of disks.
Retrospective, non-interventional validation study; Retrospective validation study
Adult patients; dataset constructed to provide a balanced range of gender, ejection fraction values, and body mass index levels.; Sample Size: 186 acquired studies; Number of Sites: Not specified (mentions multiple ultrasound system manufacturers)
Biplane method of disks ejection fraction
Root Mean Square Deviation (RMSD) of EF%
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Left ventricular ejection fraction
Machine learning-based algorithm with simple pooling layers
Root Mean Square Deviation below a set threshold
7.21 RMSD EF% [95% Cl]
—
—
Retrospective, non-interventional validation study using over 186 acquired studies.
—
Indications for Use
The Caption Interpretation Automated Ejection Fraction software is used to process previously acquired transthoracic cardiac ultrasound images, to store images, and to manipulate and make measurements on images using an ultrasound device, personal computer, or a compatible DICOM-compliant PACS system in order to provide automated estimation of left ventricular ejection fraction. This measurement can be used to assist the clinician in a cardiac evaluation. The Caption Interpretation Automated Ejection Fraction Software is indicated for use in adult patients.
Device Story
Software processes previously acquired transthoracic cardiac ultrasound images (DICOM clips) to automate left ventricular ejection fraction (EF) calculation. Operates on ultrasound devices, PCs, or DICOM-compliant PACS. Machine learning algorithms select appropriate apical four-chamber, apical two-chamber, or parasternal long-axis views; perform EF estimation; and output a percentage value with a confidence metric. Clinicians review the result and associated video clip to confirm or reject the estimate. Assists in cardiac evaluation by providing automated measurements, potentially improving workflow efficiency and consistency across diverse ultrasound platforms.
Clinical Evidence
Retrospective, non-interventional validation study using 186 acquired studies. Primary endpoint: Root Mean Square Deviation (RMSD) of EF compared to biplane method of disks ground truth. Subject device achieved 7.21 RMSD EF% (95% CI), demonstrating improved performance over predicate (7.94 RMSD EF%). Outlier rates (EF error >15%: 1.09% vs 1.61%; EF error >20%: 0.55% vs 0%) confirm comparable safety and effectiveness profile.
Technological Characteristics
Software-based medical image processing system. Utilizes machine learning algorithms for automated EF calculation. Key technical features: simple pooling layers, linear support vector machines for image quality, and optimized image preprocessing. Operates on DICOM-compliant systems. No hardware components; software-only device.
Indications for Use
Indicated for use in adult patients to provide automated estimation of left ventricular ejection fraction from previously acquired transthoracic cardiac ultrasound images to assist clinicians in cardiac evaluation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Caption Health % Savannah Hari Quality Systems Manager Manatt, Phelps & Phillips, LLP 1 Embarcadero Center, 30th Floor San Francisco, California 94111
January 19, 2022
# Re: K210747
Trade/Device Name: Caption Interpretation Automated Ejection Fraction Software Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: January 18, 2022 Received: January 18, 2022
## Dear Savannah Hari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K210747
Device Name
Caption Interpretation Automated Ejection Fraction Software
#### Indications for Use (Describe)
The Caption Interpretation Automated Ejection Fraction software is used to process previously acquired transhoracic cardiac ultrasound images, to store images, and to make measurements on images using an ultrasound device, personal computer, or a compatible DICOM-compliant PACS system in order to provide automation of left ventricular ejection fraction. This measurement can be used to assist the clinician in a cardiac evaluation.
The Caption Interpretation Automated Ejection Fraction Software is indicated for use in adult patients.
Type of Use (Select one or both, as applicable)
🇿 Prescription Use (Part 21 CFR 801 Subpart D)
[ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An aqency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Caption Health, Inc. Caption Interpretation Automated Ejection Fraction K210747
### Submitter
Caption Health, Inc. 2000 Sierra Point Parkway, 8th Floor Brisbane, CA 94005
Contact Person: Savannah Hari, Quality Systems Manager
Phone: (415) 671-4711 Email: savannah@captionhealth.com Date Prepared: January 14, 2022
## Device
Proprietary Name: Caption Interpretation Automated Ejection Fraction Software
Common Name: Caption Interpretation Automated Ejection Fraction Software
Classification Name and Number: Medical Image Management and Processing System, 21 CFR 892.2050
Regulatory Class: II
Product Code: QIH, Automated Radiological Image Processing Software
## Predicate Device
Caption Health, Inc. Caption Interpretation Automated Ejection Fraction Software (K200621)
## Device Description:
The Caption Interpretation Automated Ejection Fraction Software ("AutoEF") applies machine learning algorithms to process echocardiography images in order to calculate left ventricular ejection fraction. The cleared Caption Interpretation AutoEF performs left ventricular ejection fraction measurements using apical four chamber or apical two chamber cardiac ultrasound views, or the parasternal long-axis cardiac ultrasound view in combination with an apical four chamber view. The software selects the image clips to be used, performs the AutoEF calculation, and forwards the results to the desired destination for clinician viewing. The output of the Ejection Fraction estimate stated as a percentage, along with an indication of confidence regarding that estimate.
{4}------------------------------------------------
## Intended Use / Indications for Use:
No differences exist between the subject device and the predicate device with respect to intended use or indications for use. The intended use / indications for use are provided below:
The Caption Interpretation Automated Ejection Fraction software is used to process previously acquired transthoracic cardiac ultrasound images, to store images, and to manipulate and make measurements on images using an ultrasound device, personal computer, or a compatible DICOM-compliant PACS system in order to provide automated estimation of left ventricular ejection fraction. This measurement can be used to assist the clinician in a cardiac evaluation.
The Caption Interpretation Automated Ejection Fraction Software is indicated for use in adult patients.
## Summary of Technological Characteristics:
The Caption Interpretation Automated Ejection Fraction Software is an updated version of the predicate device and features very similar technological characteristics. Specific changes between the subject device and the predicate include:
- An additional 30% of training data from three ultrasound devices and two clinical sites for retraining of algorithms
- Simplification of network architecture via simple pooling layers and quantization
- Simplification of image quality algorithms via linear support vector machines
- Optimization of image quality thresholds to maintain performance of the subject device ●
- Runtime speed optimization of image preprocessing and removal of vestigial code
These particular changes are intended to:
- 1) Improve the function of the product across the diversity of ultrasound devices in use and patient types in the population.
- 2) Reduce the complexity of computations, enabling the algorithm to run faster and allow it potentially to operate on lower computing power hardware platforms.
- 3) Maintain optimal clinical performance and acceptability due to additional network training.
The technological principle underlying both the current version of Caption Interpretation Automated Ejection Fraction Software and the predicate device remains the same, in that both the subject device and predicate enable the calculation of ejection (EF) on previously acquired cardiac scans using machine learning-based algorithms and biplane apical echocardiographic images. Further details can be found in the comparison table below.
{5}------------------------------------------------
### Performance Data:
The Caption Interpretation Automated Ejection Fraction Software was developed and tested in accordance with Caption Health's Design Control processes and has been subjected to extensive safety and performance testing. Verification and validation testing was conducted to demonstrate the substantial equivalence of the subject device to the predicate. The primary success criterion was that the subject device would produce an ejection fraction number with a Root Mean Square Deviation below a set threshold as compared to the reference ground truth EF. The same test protocols, acceptance criteria and endpoints were used between the predicate and the subject device to ensure that performance could be appropriately compared.
Non-clinical verification and validation test results established that the device meets its design requirements and intended use. Extensive algorithm development and software verification testing assessed the performance of the software's image video clip selection function, performance characteristics of the algorithm, including AutoEF accuracy and overall functional performance.
Images and cases used for verification and validation testing were carefully separated from training datasets.
A formal retrospective, non-interventional validation study was conducted using over 186 acquired studies where the biplane method of disks ejection fraction was reported. This patient dataset was constructed to provide a balanced range of gender, ejection fraction values, and body mass index levels. Testing included a wide array of ultrasound system manufacturers to verify that the subject device performs acceptably across multiple scanner platforms. The ejection measurements from the subject device were compared to the biplane method ejection fraction, and a root mean square deviation was calculated and used as the primary endpoint for the subject device was met (results: 7.21 RMSD EF% [95% Cl]), and demonstrated slightly improved performance compared to the predicate device (results: 7.94 RMSD EF % [95% Cl]). Performance improvements between the subject device and the predicate device did not lead to significant differences in the outlier rate, with comparable performance between the subject device [1.09% for EF error >15%, 0.55% for EF error >20%] and the predicate [1.61% for EF error >15%, 0% for EF error >20%]. Based on the clinical performance as documented in this retrospective validation study, the device has a safety and effectiveness profile that is substantially equivalent to the predicate device.
## Conclusions:
Performance testing demonstrated that the Caption Interpretation Automated Ejection Fraction Software performs as expected and in a manner that is substantially equivalent to the predicate device. The Caption Interpretation Automated Ejection Fraction software has the same intended use, indications for use, and principles of operation as its predicate device. Thus, the Caption Interpretation Automated Ejection Fraction software is substantially equivalent to its predicate.
{6}------------------------------------------------
| Comparison of Features between Proposed Subject Device and Predicate Device | | | |
|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Caption Health, Inc., Caption<br>Interpretation Automated<br>Ejection Fraction ("AutoEF<br>2.5") (K210747) - Proposed<br>Device | Caption Health, Inc., Caption<br>Interpretation Automated<br>Ejection Fraction Software<br>Application ("AutoEF 2.0")<br>(K200621) - Predicate Device | |
| Product Code | QIH | QIH | |
| Intended Use | The Caption Health, Inc.<br>Caption Interpretation<br>Automated Ejection Fraction<br>software is used to process<br>previously acquired<br>transthoracic cardiac<br>ultrasound images, to store<br>images, and to manipulate<br>and make measurements on<br>images using an ultrasound<br>device, personal computer, or<br>a compatible DICOM-<br>compliant PACS system in<br>order to<br>provide automated estimation of<br>left ventricular ejection fraction.<br>This measurement can be used<br>to assist the clinician in a cardiac<br>evaluation.<br>The Caption Interpretation<br>Automated Ejection Fraction<br>Software is indicated for use in<br>adult patients. | The Caption Health, Inc. Caption<br>Interpretation Automated Ejection<br>Fraction software is used to process<br>previously acquired transthoracic<br>cardiac ultrasound images, to store<br>images, and to manipulate and make<br>measurements on images using an<br>ultrasound device, personal<br>computer, or a compatible DICOM-<br>compliant PACS system in order to<br>provide automated estimation of left<br>ventricular ejection fraction. This<br>measurement can be used to assist<br>the clinician in a cardiac evaluation.<br>The Caption Interpretation<br>Automated Ejection Fraction<br>Software is indicated for use in<br>adult patients. | |
| General Principles of Operation | | | |
| Machine Learning- Based<br>Algorithm | Yes | Yes | |
| Operates on DICOM clips | Yes | Yes | |
| Automation level | Fully automated, including clip<br>selection | Fully automated, including clip<br>selection | |
| User Alert to Use Multiple<br>Views | Yes | Yes | |
| Automated Ejection<br>Fraction Calculation | Yes | Yes | |
| Ejection Fraction reported | Whole number estimate | Whole number estimate | |
| Quantitative feedback<br>to enable clinician to<br>assess EF calculation | • Confidence Metric<br>• Qualitative Bin Likelihood | • Confidence Metric<br>• Qualitative Bin Likelihood | |
| EF Result shown with<br>video clip | Yes | Yes | |
| User confirmation/<br>rejection of result | Yes | Yes | |
| Technological Characteristics | | | |
| Network Architecture | Simple Pooling | Advanced Pooling | |
| Regressor algorithm | Linear Support Vector Machine | Radial-basis-function (RBF) Support<br>Vector Machine | |
| Image Processing | Grayscaling done first and integer<br>resizing. | Floating point resizing and grayscaling<br>done at the end | |
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.