Nitrile Examination Gloves

K210706 · Niujian Technology Co., Ltd. · LZA · Oct 19, 2021 · General Hospital

Device Facts

Record IDK210706
Device NameNitrile Examination Gloves
ApplicantNiujian Technology Co., Ltd.
Product CodeLZA · General Hospital
Decision DateOct 19, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Nitrile Examination Gloves are non-sterile disposable devices intended for medical purposes that are worn on the examiner's hands or finger to prevent contamination between patient and examiner.

Device Story

Nitrile examination gloves; non-sterile; disposable; single-use barrier protection. Worn by clinicians on hands or fingers during medical examinations. Prevents contamination between patient and examiner. Manufactured from nitrile compound; blue color; powder-free. Complies with ASTM D6319-19 specifications for physical dimensions, tensile strength, and elongation. Provides physical barrier against contaminants.

Clinical Evidence

No clinical data. Bench testing only. Compliance with ASTM D6319-19 (dimensions, physical properties), ASTM D5151-19 (watertightness/holes), ASTM D6124-06 (residual powder), and ISO 10993 series (biocompatibility: irritation, sensitization, acute systemic toxicity, and cytotoxicity) demonstrated performance equivalent to predicate.

Technological Characteristics

Material: Nitrile compound. Form factor: Ambidextrous, blue, powder-free. Standards: ASTM D6319-19 (dimensions, tensile properties), ASTM D5151-19 (holes), ASTM D6124-06 (powder content), ISO 10993-5/10/11 (biocompatibility). Energy source: None (mechanical barrier). Connectivity: None. Sterilization: Non-sterile.

Indications for Use

Indicated for use as a non-sterile, disposable medical glove worn on the examiner's hands or fingers to prevent cross-contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. October 19, 2021 Niujian Technology Co., Ltd. Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong, Shanghai, 200120 CHN Re: K210706 Trade/Device Name: Nitrile Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 6, 2021 Received: September 14, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210706 Device Name Nitrile Examination Gloves #### Indications for Use (Describe) The Nitrile Examination Gloves are non-sterile disposable devices intended for medical purposes that are worn on the examiner's hands or finger to prevent contamination between patient and examiner. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| | <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | escription Use (Part 21 CFR 801 Subpart D) |X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary (K210706) This summary of 510(k) is being submitted in accordance with 21 CFR 807.92. Date Prepared: October 5, 2021 # 1.0 Submission Sponsor | Manufacturer Name | Niujian Technology Co., Ltd. | |-------------------|-------------------------------------------------------------------------------------------------------| | Address | Room 401 Building 1, Anfangcheng No. 919, East<br>Qunxian Road, Yuecheng District, Shaoxing, Zhejiang | | Tel | 86-18210518920 | | Email | 385206015@qq.com | | Contact Person | Xiuqiang Zhou | ### Designated Submission Correspondent | Company Name | Shanghai Truthful Information Technology Co., Ltd | |-----------------|-------------------------------------------------------------------------| | Address | Room 1801, No. 161 East Lujiazui Rd., Pudong,<br>Shanghai 200120, China | | Tel | +86-21-50313932 | | Email | Info@truthful.com.cn | | Contact Person: | Mr. Boyle Wang | #### 2.0 Device Identification | Classification Name | Polymer Patient Examination | |-----------------------|-----------------------------| | Glove Trade Name | Nitrile Examination Gloves | | Device Classification | Class I | | Regulation Number | 21 CFR 880.6250 | | Panel | General Hospital | | Product Code | LZA | | Previous Submissions | None | #### 3.0 Classification Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I {4}------------------------------------------------ Panel: General Hospital #### 4.0 Predicate Device Information Ever Growth (Vietnam) Co., Ltd. Sponsor: Device: Disposable Powder Free Nitrile Examination Glove, Pink Color Disposable Powder Free Nitrile Examination Glove, Black Color 510(k) number: K190942 ### 5.0 Indication for Use The Nitrile Examination Gloves are non-sterile disposable devices intended for medical purposes that are worn on the examiner's hands or finger to prevent contamination between patient and examiner. #### 6.0_Device Description Nitrile Examination Gloves are patient examination gloves made from nitrile compound, non-sterile (as per 21 CFR 880.6250, Class I). The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirements specifications for barrier protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves. The subject device is powder free nitrile examination gloves. The subject device is in blue color. #### 7.0 Technological Characteristic Comparison Table | Item | Subject Device<br>(K210706) | Predicate Device<br>(K190942) | |----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | LZA | LZA | | Regulation No. | 21CFR880.6250 | 21CFR880.6250 | | Class | I | I | | Intended Use | The Nitrile Examination<br>Gloves are non-sterile<br>disposable device intended<br>for medical purposes that are<br>worn on the examiner's<br>hands or finger to prevent<br>contamination between<br>patient and examiner. | The Nitrile Powder Free<br>patient examination glove is a<br>non-sterile disposable device<br>intended for medical purposes<br>that is worn on the examiner's<br>hands or finger to prevent<br>contamination between patient<br>and examiner. | #### Table1-General Comparison {5}------------------------------------------------ | Material | | Nitrile | Nitrile | | | |-----------------------------|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|------------| | Powdered or<br>Powered free | | Powdered free | Powdered free | | | | Design Feature | | Ambidextrous | Ambidextrous | | | | Colorant | | Blue | Blue | | | | Labeling Information | | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile | Single-use indication, powder free, device color, device name, glove size and quantity, Non-Sterile | | | | Dimensions(mm) | | Complies with ASTM D6319-19:<br>Length: $\ge$ 230;<br>Width:<br>S: 80 $\pm$ 10;<br>M: 95 $\pm$ 10;<br>L: 110 $\pm$ 10;<br>XL: 120 $\pm$ 10 | Complies with ASTM D6319-19:<br>Length: $\ge$ 230;<br>Width:<br>XS:70 $\pm$ 10;<br>S: 80 $\pm$ 10;<br>M: 95 $\pm$ 10;<br>L: 110 $\pm$ 10;<br>XL: 120 $\pm$ 10 | | | | Thickness(mm) | | Complies with ASTM D6319-19:<br>Finger: $\ge$ 0.05;<br>Palm: $\ge$ 0.05 | Complies with ASTM D6319-19:<br>Finger: $\ge$ 0.05;<br>Palm: $\ge$ 0.05 | | | | Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14MPa, min | Tensile<br>Strength | 14MPa, min | | | | Ultimate<br>Elongation | 500% min | Ultimate<br>Elongation | 500% min | | | After<br>Aging | Tensile<br>Strength | 14MPa, min | Tensile<br>Strength | 14MPa, min | | | | Ultimate<br>Elongation | 400%min | Ultimate<br>Elongation | 400%min | | Freedom from<br>Holes | | In accordance with ASTM<br>D6319-19 and ASTM D5151-<br>19, G-1, AQL 2.5 | In accordance with ASTM<br>D6319-19 and ASTM D5151-<br>19, G-1, AQL 2.5 | | | | Powder Content | | Complies with ASTM D6319-<br>19:< 2mg per glove | Complies with ASTM D6319-<br>19: < 2mg per glove | | | | Biocompatibility | | ISO 10993-10;<br>Under the conditions of the<br>study, not an irritant or a<br>sensitizer<br>ISO 10993-11;<br>Under the<br>condition of acute | ISO 10993-10;<br>Under the conditions of the<br>study, not an irritant or a<br>sensitizer<br>/ | | | {6}------------------------------------------------ | | systemic toxicity test,<br>the test article did not<br>show acute systemic<br>toxicity in vivo. | | |--|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------| | | ISO 10993-5<br>Under conditions of the<br>study, device extract is<br>cytotoxic | ISO 10993-5<br>Under conditions of the study,<br>device extract is cytotoxic | #### 8.0 Summary of Non-clinical Testing Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization. ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application. | Test Methodology | Purpose | Acceptance Criteria | Results | |------------------|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------| | ASTM D6319 | Physical<br>Dimensions<br>Test | Length(mm): $\geq$ 230;<br>Width(mm):<br>S: 80 $\pm$ 10;<br>M: 95 $\pm$ 10;<br>L: 110 $\pm$ 10;<br>XL: 120 $\pm$ 10; | Length: $\geq$ 230;<br>Width:<br>S: 80-84<br>M: 95-99<br>L: 108-111<br>XL: 117-121<br>Pass | | | | Thickness (mm):<br>Finger: $\geq$ 0.05<br>Palm: $\geq$ 0.05 | Finger: 0.08-0.10<br>Palm: 0.06-0.08<br>Pass | | ASTM D5151 | Watertightness<br>Test<br>for<br>Detection<br>of | Meet the requirements of<br>ASTM D5151 AQL 2.5 | 0/125 leaks<br>Pass | #### Table 5 Summary of non-clinical performance testing {7}------------------------------------------------ | | Holes | | | | | |-------------------------|------------------------------------|------------------------------------------------|------------------------|--------------|--------------------------------------------------------------------------------------------------------------------------------------------| | ASTM D6124 | Powder<br>Content | Meet the requirements of<br>ASTM D6124 < 2.0mg | | | 0.02 mg<br>Pass | | ASTM D6319<br>ASTM D412 | Physical<br>properties | Before<br>Aging | Tensile<br>Strength | $\geq14$ MPa | 26-35<br>Pass | | | | | Ultimate<br>Elongation | $\geq500$ % | 509-545<br>Pass | | | | After<br>Aging | Tensile<br>Strength | $\geq14$ MPa | 19-36<br>Pass | | | | | Ultimate<br>Elongation | $\geq400$ % | 447-510<br>Pass | | ISO 10993-5 | Cytotoxicity | cytotoxic | | | Under conditions of the<br>study, show slight<br>potential toxicity to L-<br>929 cells. | | ISO 10993-11 | Acute<br>systemic<br>toxicity test | Non- toxicity | | | Under the<br>condition of acute<br>systemic toxicity test,<br>the test article did not<br>show acute systemic<br>toxicity in vivo.<br>Pass | | ISO 10993-10 | Irritation | Non-irritating | | | Under the conditions of<br>the study, not an irritant/<br>Pass | | ISO 10993-10 | Sensitization | Non-sensitizing | | | Under conditions of the<br>study, not a sensitizer./<br>Pass | # 9.0 Summary of Clinical Testing Clinical testing is not needed for this device. #### 10.0 Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device in K190942.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...