ECOMA Level 2 Disposable Surgical Mask with Ear Loops

K210698 · Xianning Eco Medical Articles Co., Ltd. · FXX · Aug 19, 2021 · General, Plastic Surgery

Device Facts

Record IDK210698
Device NameECOMA Level 2 Disposable Surgical Mask with Ear Loops
ApplicantXianning Eco Medical Articles Co., Ltd.
Product CodeFXX · General, Plastic Surgery
Decision DateAug 19, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Story

ECOMA Level 2 Disposable Surgical Mask is a 3-layer, flat-pleated, single-use, non-sterile device. Construction includes an outer polypropylene spunbond layer, a middle polypropylene melt-blown filter layer, and an inner polypropylene spunbond layer, sonically welded. Features include spandex/polyester ear loops and a malleable iron wire nosepiece with plastic cover for fit. Used in clinical settings by healthcare personnel and patients to provide a barrier against blood, body fluids, and particulates. Device is disposable and intended for direct skin contact for less than 24 hours.

Clinical Evidence

Bench testing only. Performance testing included Bacterial Filtration Efficiency (BFE) ≥ 98%, Particulate Filtration Efficiency (PFE) ≥ 98%, differential pressure < 6.0 mm H2O/cm2, and fluid resistance at 120 mmHg. Biocompatibility testing (cytotoxicity, irritation, sensitization) conducted per ISO 10993-5 and ISO 10993-10 confirmed the device is non-cytotoxic, non-irritating, and non-sensitizing. Flammability testing per 16 CFR 1610 confirmed Class 1 rating.

Technological Characteristics

3-layer flat-pleated mask. Materials: Polypropylene spunbond (outer/inner), polypropylene melt-blown (filter). Components: Spandex/polyester ear loops, malleable iron wire nosepiece with plastic cover. Standards: ASTM F2100-19, ASTM F1862/F1862M-17, ASTM F2101-19, ASTM F2299/F2299M-17, 16 CFR Part 1610, ISO 10993-5, ISO 10993-10. Non-sterile, single-use, disposable.

Indications for Use

Indicated for use by healthcare personnel and patients to reduce exposure to microorganisms, body fluids, and particulate material in infection control settings.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 19, 2021 Xianning Eco Medical Articles Co., Ltd. Wei Zhou General Manager No.16,Shutai Street High-Tech Industrial Zone Xianning, Hubei 437000 China Re: K210698 Trade/Device Name: ECOMA Disposable Surgical Mask (Ear loops) Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: July 14, 2021 Received: July 14, 2021 Dear Wei Zhou: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210698 Device Name ECOMA Level 2 Disposable Surgical Mask with Ear Loops Indications for Use (Describe) The surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. **Type of Use (Select *one or both*, as applicable)**Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "ECOMA" in large, bold, blue letters. The letters are evenly spaced and appear to be a sans-serif font. The background is a light color, providing contrast for the blue text. Tel: +86-715-8215222 Web:www.ecoma.com.cn Add:No.16, Shutai Street, High-Tech Industrial Zone, Xianning, China Fax:+86-715-8217333 E-mail:service@ecoma.com.cn ## 510(k) SUMMARY 510(k) Premarket Notification for ECOMA Level 2 Disposable Surgical Mask with Ear Loops In accordance with the requirements set forth in Title 21 CFR §807.92 - 1. Submitter: XIANNING ECO MEDICAL ARTICLES CO., LTD. No.16, Shutai Street High-Tech Industrial Zone Xianning, China 437000 Tel: +86-715-8215222 8018 # Fax: +86-715-8217333 FDA Registration Number: 3008240985 - 2. XIANNING ECO MEDICAL ARTICLES CO., LTD. Manufacturing Location: No.16, Shutai Street High-Tech Industrial Zone Xianning, China 437000 Registration Number: 3008240985 - 3. Packaging Location: Same as Manufacturing Location - 4. Regulatory Affairs Contact: Wei Zhou General Manager Tel: +86-715-8215222 8018 Fax: +86-715-8217333 Email: zhouwei@ecoma.com.cn ട. Date Summary Prepared: Feb-08-2021 - 6. Name of Device: ECOMA Level 2 Disposable Surgical Mask with Ear Loops {4}------------------------------------------------ Tel: +86-715-8215222 Web:www.ecoma.com.cn Add: No.16, Shutai Street, High-Tech Industrial Zone, Xianning, China Fax:+86-715-8217333 E-mail:service@ecoma.com.cn - 7. Trade Name: ECOMA Level 2 Disposable Surgical Mask with Ear Loops - 8. Common/Classification Name: Surgical Mask - 9. Regulation Number: 21 CFR §878.4040 - 10. Device Class: Class II - Surgical Apparel 11. Regulation Name: - 12. Product Code: FXX 13. Predicate Device: BH Medical Products Co., Ltd. Level 2 Surgical Mask with ear loops (K133070). Device Description: The ECOMA surgical mask is composed of 3-layers and is flat-pleated. The mask 14. materials consist of an outer layer (polypropylene spunbond, blue), filter layer (polypropylene melt blown, white), and inner layer (polypropylene spunbond, white). The three layers of the mask body are collated and sonically welded around the edges. The surgical mask contains ear loops attached by welding to secure the mask over the user's mouth and face and includes a malleable nosepiece to provide a firm fit over the nose. The surgical face mask is a single use, disposable device, provided non-sterile. Over the counter use: Yes Duration and type of contact: Direct contact, less than 24 hours, skin contact Single use disposable device: Yes 15. Packaging: 50 masks/box 16. Indications for Use: The surgical masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile. ## 17. Technological Characteristics Comparison: - : For the Surgical Face Masks with Ear loops: - Surgical Face Mask with Ear Loops {5}------------------------------------------------ Tel: +86-715-8215222 Fax:+86-715-8217333 Web:www.ecoma.com.cn E-mail:service@ecoma.com.cn Add:No.16, Shutai Street, High-Tech Industrial Zone, Xianning, China - K133070 - Submitter of 510(K)/holder: BH Medical Products Co., Ltd ● ## Technological Characteristics Comparison Table (Ear loop): 18. | Element of<br>Comparison | Predicted Device<br>K133070 | Subject Device | Comparison | |------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------| | Manufacturer | BH Medical Products Co., Ltd | Xianning ECO Medical<br>Articles Co, Ltd. | N/A | | Proprietary or<br>Model Name | Surgical Face Mask, Ear<br>Loops, Model<br>101B, 101G, 136B, 136G,<br>137B, 137G | Surgical Face Mask, Ear<br>Loops,<br>Model LV2001W-B | N/A | | Intended Use | The surgical face masks are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids and<br>particulate material. These<br>facemasks are intended for use<br>in infection control practices to<br>reduce the potential exposure to<br>blood and body fluids. This is a<br>single use, disposable device(s),<br>provided nonsterile. | The surgical face masks are<br>intended to be worn to protect<br>both the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and<br>particulate material. These<br>facemasks are intended for use<br>in infection control practices to<br>reduce the potential exposure<br>to<br>blood and body fluids. This is a<br>single use, disposable device(s),<br>provided nonsterile. | Same | | ASTM Level | 1, 2, 3 | 2 | Similar | | Design feature | Ear Loops and Tie on | Ear Loops | Similar | | Mask styles | 3 flat pleated | 4 Flat pleated | Similar | | Product<br>Performance<br>Specifications | Meet ASTM F1862-07<br>Meet ASTM F2101-07<br>Meet ASTM F2299-03<br>Meet MIL-M-36954C<br>Meet 16 CFR Part 1610 | ASTM F1862/F1862-M-17<br>Meet ASTM F2101-19<br>and EN 14683:2019<br>ASTM F2299/F2299M-17<br>Meet ASTM F2100-19<br>Meet 16 CFR Part 1610 | Similar<br>Similar<br>Similar<br>Similar<br>Similar | | 16 CFR Part 1610 | Class 1 | Class 1 | similar | | Single Use | Yes | Yes | same | | Disposable | Yes | Yes | same | | Non-sterile | Yes | Yes | Same | | Material<br>Composition | Three-layer mask constructed<br>of:<br>Outer layer: spunbond<br>polypropylene<br>Filter layer: melt blown<br>polypropylene<br>Inner layer: spunbond<br>polypropylene<br>Binding: Spunbond<br>Polypropylene<br>Ear loops: Polyester<br>Nose clip: Aluminum Wire | Three-layer mask constructed<br>of:<br>Outer layer: Polypropylene<br>Spunbond<br>Filter layer: Polypropylene<br>Melt blown<br>Inner layer: Polypropylene<br>Spunbond<br>Ear loops: Spandex elastic,<br>polyester<br>Nose Piece: Malleable Iron wire<br>with plastic<br>cover/Polypropylene | Similar | | ASTM Level | 1, 2, 3 | 2 | Similar | | Design feature | Ear Loops and Tie on | Ear Loops | Similar | | Mask styles | 3 flat pleated | 4 Flat pleated | Similar | | Color | Blue, Green | Blue | Same | | Dimensions(L*W) | 6.8" +/-0.25"x 4.2" +/-0.25" | 17.5 x 9.5 cm (+/- 0.5cm) | Same | | Use | Single Use; Disposable | Single Use; Disposable | Same | | Biocompatibility | The surgical mask was tested in<br>accordance with ISO10993 and<br>passed acceptance<br>criteria. | The surgical mask was tested<br>with<br>following standards:<br>• AAMI /ANSI/ ISO<br>10993-5:2009,<br>Under the testing conditions,<br>the subject surgical mask did<br>not show potential cytotoxicity<br>• AAMI /ANSI/ ISO<br>10993-10:2010,<br>Under the testing conditions,<br>the subject surgical mask did<br>not cause significant irritation<br>or sensitization<br>reaction to the test animals. | Same | {6}------------------------------------------------ # 11 Tel: +86-715-8215222 Web:www.ecoma.com.cn Fax:+86-715-8217333 E-mail:service@ecoma.com.cn Add:No.16, Shutai Street, High-Tech Industrial Zone, Xianning, China {7}------------------------------------------------ Tel: +86-715-8215222 Fax:+86-715-8217333 Web:www.ecoma.com.cn E-mail:service@ecoma.com.cn Add:No.16, Shutai Street, High-Tech Industrial Zone, Xianning, China ## Summary of Non-Clinical Testing (ASTM F2100 Performance Testing and Biocompatibility Testing Results) 19. | Test Methodology | Purpose | Acceptance Criteria | Test Results | |---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | ASTM F2101-<br>Bacterial Filtration<br>Efficiency (BFE) | To measure the bacterial<br>filtration efficiency (BFE) of<br>medical face mask materials,<br>employing a ratio of the<br>upstream bacterial challenge<br>to downstream residual<br>concentration to determine<br>filtration efficiency of medical<br>face mask materials. | ≥ 98% Average BFE<br>for all samples tested | 96 /96 samples passed ≥ 98% Average BFE (3 lots, 32<br>samples per lot with a lot size<br>of 200,000 each) | | ASTM F2299-<br>Particulate<br>Filtration<br>Efficiency | To determine the Initial<br>Efficiency of Materials<br>Used in Medical Face<br>Masks to Penetration by<br>Particulates Using Latex.<br>Spheres | ≥ 98% Average PFE<br>for all samples tested | 96 /96 samples passed ≥ 98% Average PFE (3 lots, 32<br>samples per lot with a lot size<br>of 200,000 each) | | Differential Pressure | To determine the pressure<br>required to breathe through<br>the final manufactured face<br>mask | Samples must be <<br>$6.0 mm H_2O/cm^2$ | 96 /96 samples passed<br>differential pressure < $6.0 mm H_2O/cm^2$ (3 lots, 32<br>samples per lot with a lot<br>size of 200,000 each) | | ASTM F1862-Fluid<br>Resistance | To evaluate the<br>resistance of medical<br>face masks to<br>penetration by the<br>impact of a small volume<br>(~2mL) of a high-velocity<br>stream of synthetic<br>blood. | At least 29 out of 32<br>specimens per lot<br>show passing results<br>at 120 mmHg | 95 /96 samples passed fluid<br>resistance at 120 mmHg (3<br>lots, 32 samples per lot with<br>a lot size of 200,000 each) | | 16 CFR<br>1610-Flammability | The purpose of this standard is<br>to reduce danger of injury and<br>loss<br>of life by providing, on a<br>national basis, standard<br>methods of testing and rating<br>the flammability of textiles and<br>textile products for clothing<br>use, thereby prohibiting the<br>use of any dangerously | All samples must be<br>Class I | 96 /96 samples passed Class I<br>flammability test (96 /96<br>samples passed Class I<br>flammability test ) | | | | | | | Sensitization | Under the research conditions,<br>determine whether the non-<br>polar and polar extracts of the<br>target device are irritating. | The sample is non-<br>sensitizing. | The device was tested in<br>accordance with ISO 10993<br>and passed acceptance<br>criteria | | Irritation | Under the research conditions,<br>determine whether the non-<br>polar and polar extracts of the<br>target device are sensitive. | The sample is non-<br>irritating. | | | Cytotoxicity | Under the research conditions,<br>determine whether the target<br>device extract is cytotoxic. | The sample is non-<br>cytotoxic. | | {8}------------------------------------------------ # Tel: +86-715-8215222 Fax:+86-715-8217333 Web:www.ecoma.com.cn E-mail:service@ecoma.com.cn Add:No.16, Shutai Street, High-Tech Industrial Zone, Xianning, China ## 20. Conclusion The conclusion drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device, BH Medical Products Surgical Face Mask cleared under K133070.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...