CPM Medical Consultants Tibial Revision Knee System
K210695 · Cpm Medical Consultants, LLC · JWH · Aug 16, 2021 · Orthopedic
Device Facts
Record ID
K210695
Device Name
CPM Medical Consultants Tibial Revision Knee System
Applicant
Cpm Medical Consultants, LLC
Product Code
JWH · Orthopedic
Decision Date
Aug 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CPM Medical Consultants Tibial Revision Knee System is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. The device is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock and/or require constrained stabilization for tibiofemoral joint due to soft tissue in-balance. The tibial augments are to be attached to their respective components with a fixation screw or screws. The CPM Medical Consultants Tibial Revision Knee System may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. The CPM Medical Consultants Tibial Revision Knee System is designed for cemented use only. The CPM Medical Consultants Tibial Revision Knee System is to be used in conjunction with the Progressive Orthopaedic Company Total Knee System cleared via K142649 & K150783. See below for a compatibility chart between systems.
Device Story
System provides tibial revision components for knee arthroplasty; includes cobalt-chromium-molybdenum (Co-Cr-Mo) metallic baseplates, tibial augments, stems, and offset adapters. Used in conjunction with Progressive Orthopaedic Company Total Knee System (K142649/K150783). Designed for cemented implantation by orthopedic surgeons in clinical settings. Components provide constrained stabilization for tibiofemoral joint in patients with inadequate bone stock or soft tissue imbalance. Augments attach via fixation screws. System restores femoral anatomy, stress distribution, and rotation. Benefits include pain relief and improved knee function for patients with severe joint degeneration or failed prior surgeries.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included fatigue performance of tibial tray, interlock mechanism strength, shear fatigue strength of tibial insert post, tibiofemoral contact area/stress, tibiofemoral constraint, and taper lock testing.
Indicated for skeletally mature patients with severe knee pain/disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, collagen disorders, avascular necrosis, pseudogout, or posttraumatic joint loss. Includes patients requiring augmentation/stem extensions for inadequate bone stock or constrained stabilization for soft tissue imbalance. Indicated for salvage of failed surgical attempts. Designed for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Progressive Orthopaedic Company Total Knee System (K142649/K150783)
United Orthopedic Corporation U2 Total Knee System (K082424/K122183)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 16, 2021
CPM Medical Consultants, LLC. Andy Rynearson Director of Engineering 1565 N. Central Expressway Richardson, Texas 75080
Re: K210695
Trade/Device Name: CPM Medical Consultants Tibial Revision Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: May 14, 2021 Received: May 18, 2021
Dear Andy Rynearson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210695
#### Device Name
CPM Medical Device Consultants Revision Tibial Knee System
#### Indications for Use (Describe)
The CPM Medical Consultants Tibial Revision Knee System is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthrius, osteoarthritis, primary and secondary traumatic arthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. The device is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock and/or require constrained stabilization for tibiofemoral joint due to soft tissue in-balance. The tibial augments are to be attached to their respective components with a fixation screw or screws.
The CPM Medical Consultants Tibial Revision Knee System may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. The CPM Medical Consultants Tibial Revision Knee System is designed for cemented use only.
The CPM Medical Consultants Tibial Revision Knee System is to be used in conjunction with the Progressive Orthopaedic Company Total Knee System cleared via K142649 & K150783. See below for a compatibility chart between systems.
Image /page/2/Figure/10 description: The image is a compatibility chart for CPM Medical Consultants Tibial Revision Components and Progressive Orthopaedic Company Components. The chart shows the compatibility between PS+ Tibial Insert, Tibial Stem, Tibial Augments, and Revision Tibial Tray with CR Femoral, PS Femoral, CR Tibial Insert, PS Tibial Insert, Tibial Tray, and Patella. Green indicates compatibility, while red indicates non-compatibility. For example, PS Femoral is compatible with PS+ Tibial Insert, and Tibial Tray is compatible with PS+ Tibial Insert and Revision Tibial Tray.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following 510(k) Summary is provided in accordance with 21 CFR 807.92.
## 510(k) Owner and Registration
| Owner's Name: | CPM Medical Consultants, LLC. |
|------------------------------------|----------------------------------------------------------------|
| Address: | 1565 N. Central Expressway, Suite 200, Richardson,<br>TX 75080 |
| Phone Number: | (321) 316-2601 |
| Fax Number: | N/A |
| Date Summary Prepared: | April 30, 2021 |
| Establishment Registration Number: | N/A |
## 510(k) Contact
| Contact: | Andy Rynearson |
|-----------------|----------------------------------------------------------------|
| Address: | 1565 N. Central Expressway, Suite 200, Richardson,<br>TX 75080 |
| Phone Number: | (321) 316-2601 |
| Fax Number: | N/A |
| Contact Person: | Andy Rynearson |
## Device Name and Classification
| Device Trade Name: | CPM Medical Consultants Revision Tibial Knee System |
|------------------------------------|-----------------------------------------------------|
| Device Common Name: | Revision Tibial Knee Replacement |
| Regulation Number and Description: | 21 CFR 888.3560 |
| Device Class: | Class II |
| Product Codes: | JWH |
| Advisory Panel: | 87 (Orthopedic) |
## Legally Marketed Predicate
CPM Medical Consultants is utilizing the Modal Manufacturing Total Knee System, previously cleared as the Progressive Orthopaedic Company Total Knee System (K142649/K150783), and the United Orthopedic Corporation U2 Total Knee System as the predicate devices (K082424/K122183). The CPM Medical Consultants Tibial Revision Knee System features component designs, materials, indications, and manufacturing methods that are similiar to the Modal Manufacturing Total Knee System and the United Orthopedic Corporation U2 Total Knee System. The CPM Medical Consultants Revision Tibial Knee System is be used in conjunction with the previously cleared Progressive Orthopaedics/Modal to Manufacturing Total Knee System (K142649/K150783).
{4}------------------------------------------------
### Device Description
The CPM Medical Consultants Tibial Revision Knee System is an extended design of The Modal Manufacturing Total Knee System (K142649/K150783). It is a cobalt-chromium-molybdenum (Co-Cr-Mo) metallic baseplate with a variety of components including tibial augments, stems, and offset adapters that provide more choice for surgeon treatment of their patients. The CPM Medical Consultants posterior stabilized plus inserts is a line extension to the previously cleared Modal Manufacturing posterior stabilized tibial insert (K142649). The CPM Medical Consultants Tibial Revision Knee System is to be used in conjunction with the previously cleared Modal Manufacturing Total Knee System (K142649/K150783) which is a fixed bearing implant available in posterior-stabilized (PS) and cruciate-retaining (CR) configurations. It is a patellofemorotibial, polymer/metal/polymer, semi-constrained, cemented knee prosthesis that consists of a femoral component, tibial insert, tibial tray and patellar component. The PS version has tibial inserts with posts and femoral components with cams that interact with the posts to help stabilize the knee after the posterior cruciate ligament is removed. The CR version of the Progressive Orthopaedic Total Knee System has tibial inserts and femoral components without posts or cams, allowing the posterior cruciate ligament to be "retained" and provide stability to the knee joint. The femoral component articulates with the tibial insert component. The underside of the tibial insert component is flat and "snaps" into the tibial baseplate component. The design and sizing of the femoral components correspond to the natural femoral anatomy, enhancing stress distribution and restoring original femoral dimensions and normal rotation, and flexion. Each femoral component has the same intercondylar distance and radius of curvature. Each tibial insert component is complimentarily shaped to conform to the femoral components. This allows any size femoral component to be matched with any size tibial component. The dome shape of each UHMWPE patellar component provides excellent contact with the femoral component and evenly distributes stresses. The dome shape of each patellar component also simplifies implantation by eliminating the need for rotation.
## Intended Use
The CPM Medical Consultants Tibial Revision Knee System is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion deformities. The device is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock and/or require constrained stabilization for tibiofemoral joint due to soft tissue in-balance. The tibial augments are to be attached to their respective components with a fixation screw or screws.
The CPM Medical Consultants Tibial Revision Knee System may also be indicated in the salvage of previously failed surgical attempts if the knee can be satisfactorily balanced and stabilized at the time of surgery. The CPM Medical Consultants Tibial Revision Knee System is designed for cemented use only.
The CPM Medical Consultants Tibial Revision Knee System is to be used in conjunction with the Progressive Orthopaedic Company Total Knee System cleared via K142649 & K150783. See below for a compatibility chart between systems.
{5}------------------------------------------------
| | | CPM Medical Consultants Tibial Revision<br>Components | | | |
|-----------------------------------------------------|---------------------|-------------------------------------------------------|-------------|-----------------|----------------------|
| | | PS+ Tibial Insert | Tibial Stem | Tibial Augments | Revision Tibial Tray |
| Progressive<br>Orthopaedic<br>Company<br>Components | CR<br>Femoral | | | | |
| | PS<br>Femoral | | | | |
| | CR Tibial<br>Insert | | | | |
| | PS Tibial<br>Insert | | | | |
| | Tibial<br>Tray | | | | |
| | Patella | | | | |
### Compatibility Chart
## Summary of Technological Characteristics
The CPM Medical Consultants Tibial Revision Knee System is similiar to the predicate Modal Manufacturing Total Knee System. Both devices are manufactured from identical materials, possess the same sizes, and feature the same packaging and sterilization processes. Extensive preclinical testing was performed on the predicate devices per K142649 and K150783 and found substantially equivalent. The performance tests are listed below and used herein to establish substantial equivalence (Section 19 Performance Testing - Bench)
Given that the subject device is identical to the subject system is substantially equivalent to the predicate systems (K142649/K150783).
## Performance Testing
Extensive preclinical performance testing was conducted on the CPM Medical Consultants Tibial Revision Knee System and substantial equivalence to the predicate device was determined. The subject device are similar to the predicate device, and therefore the predicate device testing demonstrates substantial equivalence for the subject device. The results confirm that all components of the CPM Medical Consultants Tibial Revision Knee System exhibit the appropriate mechanical characteristics for tibial revision knee joint replacement and are substantially equivalent to the predicate devices.
- Fatigue performance of the tibial tray
- Interlock mechanism strength of the tibial tray and insert utilizing Modal Manufacturing design and testing.
- Shear fatigue strength of the tibial insert post utilizing Modal Manufacturing design and testing.
{6}------------------------------------------------
- Tibiofemoral contact area and stress utilizing Modal Manufacturing design and testing.
- Tibiofemoral constraint utilizing Modal Manufacturing design and testing.
- Taper Lock Testing
## Conclusions
The subject device has the same design features, materials, and indications for use as the predicate devices. The testing performed for the predicate device indicates that the CPM Medical Consultants Tibial Revision Knee System is safe for clinical use.
The CPM Medical Consultants Tibial Revision Knee System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.