K210692 · Philips Medical Systems Dmc GmbH · KPR · Apr 2, 2021 · Radiology
Device Facts
Record ID
K210692
Device Name
DigitalDiagnost
Applicant
Philips Medical Systems Dmc GmbH
Product Code
KPR · Radiology
Decision Date
Apr 2, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DigitalDiagnost is intended to acquire, process, store, display and export digital radiographic images. The DigitalDiagnost is suitable for all routine radiography examinations, including specialist areas like intensive care, trauma or pediatric work, excluding fluoroscopy, angiography and mammography
Device Story
DigitalDiagnost is a stationary digital radiography system for routine clinical imaging. System components include a height-adjustable patient support table, ceiling-suspended X-ray tube, and vertical stands. Input consists of X-ray radiation captured by a fixed flat-panel detector (Pixium 4343RCE) or portable wireless detectors (SkyPlates). The system uses a high-voltage generator and collimator to produce images, which are processed, stored, and viewed via the Eleva workstation. Operated by clinicians in clinical settings (e.g., trauma, ICU), the device provides digital images to assist in diagnostic decision-making. Modifications from the predicate include mechanical tube head controls, updated service logging features, and software version 42. The device facilitates patient care by enabling rapid, high-quality radiographic assessment.
Clinical Evidence
No clinical data. Bench testing only, including system verification, software verification, risk control, image quality, and usability engineering, performed per FDA-recognized standards (e.g., IEC 60601-1, IEC 60601-2-54, IEC 62304).
Technological Characteristics
Stationary X-ray system; Class II. Components: height-adjustable table, ceiling suspension, X-ray generator, Pixium 4343RCE fixed detector, SkyPlates wireless detectors. Connectivity: DICOM-compliant, networked for service logs/PerformanceBridge. Software: Eleva v42, Windows 10. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-54, IEC 62304.
Indications for Use
Indicated for patients requiring routine radiography examinations, including intensive care, trauma, or pediatric imaging. Contraindicated for fluoroscopy, angiography, and mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Medical Systems DMC GmbH % Supriya Dalvi Regulatory Affairs Specialist Roentgenstrasse 24-26 Hamburg, Hamburg 22335 GERMANY
### Re: K210692
Trade/Device Name: DigitalDiagnost Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR, MQB, LLZ Dated: March 3, 2021 Received: March 8, 2021
Dear Supriya Dalvi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
April 2, 2021
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210692
Device Name DigitalDiagnost
Indications for Use (Describe)
The DigitalDiagnost is intended to acquire, process, store, display and export digital radiographic images. The DigitalDiagnost is suitable for all routine radiography examinations, including specialist areas like intensive care, trauma or pediatric work, excluding fluoroscopy, angiography and mammography
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
| Preparation<br>Date: | March 3rd 2021 | | |
|----------------------|-----------------------------------------------|----------------------------------------------------------------|--|
| 510(k)<br>Owner: | Philips Medical Systems DMC GmbH | | |
| | Roentgenstrasse 24 | | |
| | 22335 Hamburg | | |
| | GERMANY | | |
| | Establishment registration number: 3003768251 | | |
| Contact: | Dr. Supriya A. Dalvi | | |
| | Regulatory Operations Specialist | | |
| | Phone: +91 9825604544 / +91 8733918445 | | |
| | Fax: +49 40 5078-2425 | | |
| | E-mail: supriya.dalvi@philips.com | | |
| Proposed<br>Device | Device Name | DigitalDiagnost | |
| | Legal Manufacturer | Philips Medical Systems DMC GmbH | |
| | Classification Name: | Stationary x-ray system | |
| | Classification Regulation: | 21 CFR 892.1680 | |
| | Classification Panel: | 90 - Radiology | |
| | Device Class: | Class II | |
| | Primary Product Code: | KPR | |
| | Secondary Product Codes: | MQB, LLZ | |
| Predicate<br>Device | Device Name | DigitalDiagnost C90 (K202564, cleared September<br>30th, 2020) | |
| | Legal Manufacturer | Philips Medical Systems DMC GmbH | |
| | Classification Name: | Stationary X-Ray System | |
| | Classification Regulation: | 21 CFR Part 892.1680 | |
| | Classification Panel: | 90 - Radiology | |
| | Device Class: | Class II | |
| | Primary Product Code: | KPR | |
| | Secondary Product Codes: | MQB, LLZ | |
| Reference<br>Device | Device Name | CombiDiagnost R90 (K203087, cleared December 3rd,<br>2020) | |
| | Legal Manufacturer | Philips Medical Systems DMC GmbH | |
| | Classification Name: | Image-intensified fluoroscopic x-ray System | |
| | Classification Regulation: | 21 CFR Part 892.1650 | |
This 510(k) summary of safety and effectiveness is prepared in accordance with 21 CFR §807.92.
{4}------------------------------------------------
| Classification Panel: | 90 - Radiology | | |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class: | Class II | | |
| Primary Product Code: | JAA | | |
| Secondary Product Codes: | KPR, MQB | | |
| Device Description: | The proposed DigitalDiagnost is a high-end digital radiography system consisting of a height adjustable patient support table and a ceiling suspension consisting of a tube including a control handle used to acquire images with a flat panel fixed RAD detector. Additionally, different vertical stands for the radiography examinations are available. The ceiling suspension can be moved in longitudinal and lateral directions and additionally the tube can be tilted and rotated as well. The system is configured with a Philips x-ray generator and a flat panel fixed RAD detector, Pixium 4343RCE. Together with the tube these components form the radiography Image Chain. The additional option of the portable wireless detector i.e. the SkyPlates family can be used for free exposures as well as in the patient support table or in the vertical stand.<br>The proposed DigitalDiagnost is a modification of the predicate device, DigitalDiagnost C90 (K202564). The modifications include change of colour of the table, stand and ceiling suspension; changes in ceiling suspension, service features and a software update. The changes related to the software updates and service features have been recently cleared in the reference device, CombiDiagnost R90 (K203087). | | |
| Indications for Use: | The DigitalDiagnost is intended to acquire, process, store, display and export digital radiographic images. The DigitalDiagnost is suitable for all routine radiography examinations, including specialist areas like intensive care, trauma or pediatric work, excluding fluoroscopy, angiography and mammography. | | |
| Fundamental Scientific Technology: | The proposed DigitalDiagnost employs the same basic construction, fundamental scientific technology and workflow as the predicate device, DigitalDiagnost C90 (K202564) with regards to the functionality of all its components. It has the same high voltage generator, X-ray tube, Collimator, detectors, workstation (ELEVA) for images post-processing, storage and viewing.<br><br>The solid state imaging including the detector for proposed device are same as predicate device (K202564) having the same physical, functional and operational characteristics. Also, other image chain components like X-ray tube and generator, which are used for exposure characteristics and clinical performance evaluation remains same. Hence all the features and characteristics potentially influencing image quality are same as predicate device, in accordance to FDA guidance document 'Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, dated September 1, 2016'.<br><br>The changes with respect to the predicate device are mainly to accommodate users with a focused need for core functionality, some of the optional features of the predicate device have been replaced or modified to result in the proposed DigitalDiagnost. These changes do not affect the clinical or functional outcome of the device. The modifications have been evaluated per the ISO 14971 risk assessment. Design control and risk management activities show that all risks are sufficiently mitigated and that the overall residual risks are acceptable. Refer Table 1 below for comparison of the technological characteristics of the proposed device and predicate device, DigitalDiagnost C90 | | |
| | Predicate device,<br>DigitalDiagnost C90<br>(K202564) | Proposed device,<br>DigitalDiagnost | |
| Legal<br>Manufacturer | Philips Medical Systems DMC GmbH | Identical | |
| Classification | Class II per 21 CFR 892.1680,<br>Product codes KPR, MQB LLZ | Identical | |
| Regulation<br>Name | Stationary X-ray system | Identical | |
| Class | II | Identical | |
| Review Panel | Radiology | Identical | |
| Indications<br>for Use | The DigitalDiagnost C90 is intended to acquire, process, store, display and export digital radiographic images. The DigitalDiagnost C90 is suitable for all routine radiography examinations, including specialist areas like intensive care, trauma or pediatric work, excluding fluoroscopy, angiography and mammography. | Identical except updated trade name | |
| Principle of<br>Operation | DigitalDiagnost C90 is intended to acquire, process, store, display and export digital radiographic images. It is suitable for all routine radiography examinations, including specialist areas like intensive care, trauma or pediatric work, excluding fluoroscopy, angiography and mammography. The system consists of several components that can be combined to create a variety of different Xray configurations. | Identical | |
| Components | Height adjustable table (TH2) | Identical except color of table<br>(Change #1) | |
| | Single side suspended table (TH-S) | Identical except color of table<br>(Change #1) |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.