# implaFit® hip stems (K210678)

_Implantcast GmbH · LZO · Aug 17, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210678

## Device Facts

- **Applicant:** Implantcast GmbH
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Aug 17, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The implaFit® hip stems are indicated for use in cases of: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - Rheumatoid arthritis: - Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable by other techniques; and - Revision of previously failed total hip arthroplasty. The implaFit® hip stems, when used in conjunction with the ic-Bipolar Heads, are intended for the following indications: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis - Rheumatoid arthritis - Correction of functional deformity - Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques. The implaFit® hip stems cementless are intented, press-fit fixation. The implaFit® hip stems cemented are intended for cemented fixation.

## Device Story

The implaFit® hip stems are femoral components for total or hemi-hip arthroplasty. The system includes two configurations: cementless and cemented. The cementless version is a monoblock, collarless, tapered straight stem made of titanium alloy with a partial commercially pure titanium (cpTi) coating, designed for press-fit fixation. The cemented version is a monoblock, collared, tapered straight stem made of cobalt chromium (CoCrMo) alloy, designed for use with bone cement. The device is used by orthopedic surgeons in a clinical/OR setting to replace the femoral head and neck. The surgeon selects the appropriate stem based on patient anatomy and pathology. The device provides structural support for the hip joint, aiming to restore joint function, reduce pain, and improve mobility in patients with degenerative or traumatic hip conditions.

## Clinical Evidence

Bench testing only. Evaluations included fatigue testing (ISO 7206-4/6), impingement testing (ASTM F2582-14), taper disassembly (ASTM F1820), range of motion (ISO 21535), coating characterization (ASTM F1147, F1044, F1978, F1854, F1160), and wear/corrosion evaluation. All testing met predetermined acceptance criteria.

## Technological Characteristics

Cementless stems: Titanium alloy with cpTi coating; press-fit fixation. Cemented stems: CoCrMo alloy; cemented fixation. Both are monoblock, tapered straight stems with 12/14 tapers. Standards: ISO 7206-4/6 (fatigue), ASTM F2582-14 (impingement), ASTM F1820 (taper), ISO 21535 (ROM), ASTM F1147/F1044/F1978/F1854/F1160 (coatings).

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- EcoFit® Hip System ([K163577](/device/K163577.md))

## Reference Devices

- EcoFit® Hip System ([K180263](/device/K180263.md))
- EcoFit® Hip System ([K191569](/device/K191569.md))
- DePuy Orthopaedics, Inc. Summit DuoFix Hip ([K193398](/device/K193398.md))
- Summit Cemented Hip Prosthesis ([K013352](/device/K013352.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

August 17, 2022

implantcast, GmbH % Dave McGurl Vice President, Regulatory Affairs - Orthopedics MCRA, LLC. 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001

Re: K210678

Trade/Device Name: implaFit® hip stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, KWY, JDI Dated: July 20, 2022 Received: July 21, 2022

Dear Dave McGurl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

#### K210678

Device Name implaFit® hip stems

Indications for Use (Describe)

The implaFit® hip stems are indicated for use in cases of:

- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis:

- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable by other techniques; and

- Revision of previously failed total hip arthroplasty.

The implaFit® hip stems, when used in conjunction with the ic-Bipolar Heads, are intended for the following indications:

- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis

- Rheumatoid arthritis
- Correction of functional deformity

- Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable using other techniques.

The implaFit® hip stems cementless are intented, press-fit fixation. The implaFit® hip stems cemented are intended for cemented fixation.

| Type of Use (Select one or both, as applicable)                                                                        |                                                                                                                       |
|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

| Manufacturer:                    | implantcast GmbH<br>Lueneburger Schanze 26<br>21614 Buxtehude<br>Germany                                                                                                                                                                                                                                                                                                                    |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                         | Ms. Juliane Höppner<br>Phone: +49 4161 744-135<br>Fax: +49 4161 744-200<br>j.hoeppner@implantcast.de                                                                                                                                                                                                                                                                                        |
| Prepared By:                     | Mr. Dave McGurl<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Phone: 202.552.5797<br>dmcgurl@mcra.com                                                                                                                                                                                                                                                             |
| Date Prepared:                   | August 15, 2022                                                                                                                                                                                                                                                                                                                                                                             |
| Device Trade Name:               | implaFit® hip stems                                                                                                                                                                                                                                                                                                                                                                         |
| Device Class and<br>Common Name: | Class II, Total Hip Replacement                                                                                                                                                                                                                                                                                                                                                             |
| Classification:                  | 21 CFR 888.3353: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis<br>21 CFR 888.3350: Hip joint metal/polymer semi-constrained cemented prosthesis.<br>21 CFR 888.3390: Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis.                                                                                           |
| Product Codes:                   | LZO, KWY, JDI                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for Use:             | The indications for use for the proposed implaFit® hip stems are as follows:<br><br>The implaFit® hip stems are indicated for use in cases of:<br>- Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis;                                                                                                                                            |
|                                  | -             Treatment of non-union, femoral neck fracture,<br>and trochanteric fractures of the proximal femur<br>with head involvement that are unmanageable by<br>other techniques; and<br>-             Revision of previously failed total hip<br>arthroplasty.                                                                                                                       |
|                                  | The implaFit® hip stems, when used in conjunction with<br>the ic-Bipolar Heads, are intended for the following<br>indications:                                                                                                                                                                                                                                                              |
|                                  | -             Non-inflammatory degenerative joint disease<br>including osteoarthritis and avascular necrosis<br>-             Rheumatoid arthritis<br>-             Correction of functional deformity<br>-             Treatment of non-union, femoral neck fracture and<br>trochanteric fractures of the proximal femur with<br>head involvement, unmanageable using other<br>techniques. |
|                                  | The implaFit® hip stems cementless are intended for<br>uncemented, press-fit fixation. The implaFit® hip stems<br>cemented are intended for cemented fixation.                                                                                                                                                                                                                              |
| Device Description:              | The purpose of this Traditional 510(k) is introduce into<br>interstate commerce in the United States the implaFit® hip<br>stems. The implaFit® hip stems are femoral stems for total or<br>hemi hip arthroplasty. The implaFit® hip stems include<br>cementless and cemented configurations (i.e., the implaFit®<br>hip stems cementless and the implaFit® hip stems cemented).             |
|                                  | The implaFit® hip stems cementless are monoblock,<br>collarless tapered straight stems made from titanium alloy<br>with a partial coating of commercially pure titanium (cpTi).<br>They are fixed by press-fitting.                                                                                                                                                                         |
|                                  | The implaFit® hip stems cemented are monoblock, collared,<br>tapered straight stems made from cobalt chromium<br>(CoCrMo). They are fixed using cement.                                                                                                                                                                                                                                     |
| Predicate Devices:               | Primary Predicate:<br>EcoFit® Hip System (K163577)<br>Reference Devices:<br>-             EcoFit® Hip System (K180263, K191569)<br>-             DePuy Orthopaedics, Inc. Summit DuoFix Hip<br>(K193398)<br>-             Summit Cemented Hip Prosthesis (K013352)                                                                                                                          |
| Substantial<br>Equivalence:      | The implaFit® hip stems cementless are similar to<br>components previously cleared with the EcoFit® Hip System                                                                                                                                                                                                                                                                              |

- Rheumatoid arthritis; -

{4}------------------------------------------------

{5}------------------------------------------------

(K191569). Both are cementless femoral hip stems that are monoblock, collarless, tapered (12/14), and made from titanium alloy. In addition, both have a commercially pure titanium (cpTi) coating on the proximal end side of the stem, starting at the upper shoulder.

There are some differences in stem material, geometry, and dimensions implaFit® hip stems as compared to the EcoFit® Hip System predicate. These differences include the material, cross-sectional geometry, head/neck angle, stem length (lower shoulder to tip), neck length, and offset length.

The implaFit® hip stems are substantially equivalent to the hip stems of the legally marketed predicate device systems, Hip System, the implantcast the DePuy Orthopaedics, Inc. Summit Cemented Hip Prosthesis, and the DePuy Orthopaedics, Inc. Summit DuoFix Hip with respect to intended use and design.

- Performance Testing: All necessary testing has been performed for the "worst-case" components of the implaFit® hip stems to assure substantial equivalence to its predicates and to demonstrate the subject devices perform as intended. All testing was performed on test units representative of or worst-case compared to the finished device. The following evaluations were conducted to characterize the devices:
	- . Fatigue Testing (ISO 7206-4 and ISO 7206-6)
	- Impingement Testing (ASTM F2582-14)
	- . Taper Disassembly Testing (ASTM F1820)
	- Range of Motion (ROM) Evaluation (ISO 21535)
	- . Coating Testing (ASTM F1147, F1044, F1978, F1854, F1160)
	- Wear / Corrosion Evaluation

Conclusions: The implaFit® hip stems subject to this submission possess the same intended use and technological characteristics as the predicate device system components. All performance testing conducted for the implaFit® hip systems met the predetermined acceptance criteria or or were otherwise considered acceptable. As such, the implaFit® hip stems components are substantially equivalent to the predicate devices for the intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K210678](https://fda-staging.innolitics.com/device/K210678)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).