K210629 · Interrad Medical, Inc. · OKC · Apr 12, 2021 · General Hospital
Device Facts
Record ID
K210629
Device Name
SecurAcath
Applicant
Interrad Medical, Inc.
Product Code
OKC · General Hospital
Decision Date
Apr 12, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
The SecurAcath device is intended to secure catheters by means of a subcutaneous anchor.
Device Story
SecurAcath is a sterile, single-use subcutaneous anchoring system for percutaneous indwelling catheters. It secures catheters via a blunt nitinol anchor deployed into the subcutaneous space at the access site and a base/cover assembly that clamps the catheter shaft. Used in clinical settings to reduce catheter migration and accidental pull-out; does not impede fluid flow. Operated by healthcare professionals. Device is MR conditional and compatible with standard adhesive dressings and skin cleaning solutions. Current version incorporates a laser-marking additive in the polypropylene base for improved marking durability compared to previous pad-printed versions. Benefits include enhanced catheter stability for both intravenous and drainage applications.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional verification, joint tensile strength, base/cover interaction, hinge performance, and marking durability. Packaging validation performed per ISTA 2A and ASTM D4169. Biocompatibility testing conducted per ISO 10993.
Indicated for catheter securement to the access site by means of subcutaneous anchors in: a) Short or long-term securement of percutaneous indwelling catheters for intravenous use; b) Short and long-term securement of percutaneous indwelling catheter for abscess/general drainage.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
Interrad Medical Inc % Denise Lenz Regulatory Consultant Libra Medical, Inc 8401 73rd Ave North, Suite 63 Brooklyn Park, Minnesota 55428
Re: K210629
Trade/Device Name: SecurAcath Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: Class II Product Code: OKC, KMK Dated: February 19, 2021 Received: March 2, 2021
Dear Denise Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# James M. Simpson Jr -S7
for Paval Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210629
Device Name SecurAcath
Indications for Use (Describe)
The SecurAcath Device is indicated for catheter securement to the access site by means of subcutaneous anchors in:
a) Short or long-term securement of percutaneous indwelling catheters for intravenous use
b) Short and long-term securement of percutaneous indwelling catheter for abscess/general drainage
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY 8
#### 8.1 ADMINISTRATIVE INFORMATION
#### February 19, 2021 Date of Summary Preparation:
#### Contact Information 8.1.1
Submitter/Manufacturer Interrad, Inc 181 Cheshire Lane, Suite 100 Plymouth, MN 55441 Tel: 763-225-6699 Fax: 763-225-6695
## Primary Submission Contact
# Primary Contact
Denise Lenz
Regulatory Consultant, Libra Medical, Inc. 8401 73rd Ave North, Suite 63 Brooklyn Park, MN 55428 Cell: 612-965-3445 Email: dlenz@libramed.com
# Secondary Submission Contact
# Secondary Contact
Sew-Wah Tay Regulatory Consultant, Libra Medical, Inc. 8401 73rd Ave North, Suite 63 Brooklyn Park, MN 55428 Cell: 612-801-6782 Email: swtay@libramed.com
#### 8.1.2 Device Information
| Trade Name | SecurAcath |
|---------------------------|-----------------------------------------------------------------------------------|
| Common Name | Subcutaneous securement device or Subcutaneous<br>Engineered Stabilization Device |
| Classification Name | Implanted Subcutaneous Securement Catheter |
| Classification Regulation | 21 CFR 880.5970/880.5210/878.4200 |
| Class | Class II |
| Panel | General Hospital |
| Product Code | OKC / KMK |
#### 8.2 PREDICATE DEVICE
The modified device is substantially equivalent to the Interrad Medical SecurAcath (K180994).
{4}------------------------------------------------
#### 8.3 DEVICE DESCRIPTION
The SecurAcath device is a standalone subcutaneous anchoring securement system used to secure the catheter to the access site. The securement is achieved by means of a blunt nitinol anchor deployed into the subcutaneous space at the catheter access site, and the clamping of the catheter shaft between the Base Assembly and Cover of the device.
#### 8.4 INTENDED USE
The SecurAcath device is intended to secure catheters by means of a subcutaneous anchor.
#### INDICATIONS FOR USE 8.5
The SecurAcath Device is indicated for catheter securement to the access site by means of subcutaneous anchors in:
- a) Short or long-term securement of percutaneous indwelling catheters for intravenous use
- b) Short and long-term securement of percutaneous indwelling catheter for abscess/general drainage
#### TECHNOLOGICAL CHARACTERISTICS 8.6
The SecurAcath is a single use, sterile device for securing indwelling catheters. The device is a stand-alone accessory to percutaneous indwelling catheters. The securement to the catheter access site is achieved by means of a blunt nitinol Anchor deployed into the subcutaneous space at the catheter access site. The securement of the catheved by the clamping of the catheter shaft between the Base Assembly and Cover of the device. This reduces catheter migration and accidental pull-out while not significantly affecting fluid flow. The SecurAcath has the same technological characteristics as the predicate device beside the predicate device is pad printed and does not contain the laser marking additive in the polypropylene portion to optimize the laser marking of the device (K180994).
| Table 8-1: Comparison of Device Characteristics to Marketed Predicate Device | | |
|------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Items | Predicate SecurAcath (K180994) | SecurAcath (Current Device) |
| Product Name | Interrad Medical SecurAcath | Same-No change |
| Intended Use | The SecurAcath device is intended to secure catheters by means of a subcutaneous anchor. | Same - No change |
| Indication for Use | The SecurAcath Device is indicated for catheter securement to the access site by means of<br>subcutaneous anchors in:<br>a) Short or long-term securement of percutaneous indwelling catheters for intravenous use<br>b) Short and long-term securement of percutaneous indwelling catheter for abscess/general drainage | Same - No change |
| Regulation<br>Number and<br>Product Code | 21 CFR 880.5970<br>21 CFR 880.5210<br>21 CFR 878.4200<br>OKC/KMK | Same - No change |
| Classification | Class II | Same - No change |
| Regulatory<br>Classification<br>Name | Device, Vascular and Non-Vascular Catheter<br>Securement | Same - No change |
| Technological<br>Characteristics | Single use, sterile device for securing<br>indwelling catheters. The device is a stand-<br>alone accessory to percutaneous indwelling<br>catheters and consists of a subcutaneous<br>nitinol anchor that is deployed in the<br>subcutaneous space at the catheter access site<br>to reduce catheter migration and pull-out. | Same - No change |
| Technology-<br>Materials | PP Mevopur/Compound 100% SG702<br>Poppy V2 (first shot material) TPE Mevopur/Counpound 100%<br>G2706SLRBL V2 (second shot material) Nitinol (anchor) Marabu TPR 980 (Black ink) | PP Mevopur-Orange/Compound 100%<br>SG702 (first shot material) with laser<br>marking additive. Material passed ISO<br>10993, design verification and shelf life<br>testing and does not impact safety and<br>effectiveness. TPE Mevopur/Compound 100%<br>G2706SLRBL V2 (second shot material)-<br>Same, no change. Nitinol with/without premium surface<br>finish (anchor). Material passed ISO<br>10993 and design verification testing and<br>the change does not impact safety and<br>effectiveness. |
| Dressing<br>Compatibility | Compatible with standard adhesive dressing<br>including Biopatch and Tegaderm. | Same - No change |
| Cleaning solution<br>compatibility | Compatible with standard skin cleaning<br>solution including Chlorhexidine, alcohol,<br>and iodine solution. | Same - No change |
| Principles of<br>Operation | Catheter shaft held between the Base and Cap<br>of the SecurAcath with nitinol anchors<br>extending into the subcutaneous space from<br>the base. | Same - No change |
| Use duration | For the life of the catheter secured unless<br>anchors get accidently pulled out. | Same - No change |
| MRI<br>Compatibility | MR Conditional | Same - No change |
| Device<br>Configuration | Single use, sterile device | Same - No change |
| Packaging<br>Configuration | 1) Device on a backercard and an Extension<br>Tube Tag inside a Tyvek pouch. Ten<br>devices in Tyvek Pouch and ten Patient<br>Guide Cards inside the 10-up shelf box. | 1) Device in a Tyvek pouch. 10 devices in<br>Tyvek Pouch inside the 10-up shelf box.<br>Patient Guide Cards available upon<br>request. New packaging configuration |
| | 2) Tyvek pouch contains anti-block additive<br>(CAS number 7631-86-9) in the nylon. | passed ISTA 2A, ASTM D4169, sterile<br>barrier performance testing and ISO<br>10993-7 testing and the new packaging<br>configuration does not impact safety and<br>effectiveness. |
| | 2) Tyvek pouch contains anti-block<br>additive (CAS number 1344-01-0) in the<br>nylon. Tyvek pouch supplier data for<br>sterilization validation and shelf life<br>were evaluated and pouch sealing<br>process validation testing passed. The<br>anti-block additive in the nylon layer<br>does not impact safety and effectiveness. | |
| Labeling<br>Configuration | 1) Extension Tube Tag included in Tyvek<br>pouch.<br>2) Label content on device applied as an<br>ink.<br>3) Labeling on Pouch and Shelf Box. | 1) Extension Tube Tag available upon<br>request. New packaging configuration<br>passed ISTA 2A, ASTM D4169, sterile<br>barrier performance testing and ISO<br>10993-7 testing. The new packaging<br>configuration does not impact safety and<br>effectiveness.<br>2) Label content on device applied as a<br>laser marking. Laser marking passed<br>testing per IEC 60601-1 for durability of<br>markings and the laser marking does not<br>impact safety and effectiveness.<br>3) Labeling on Pouch and Shelf Box-Same,<br>no change. |
| Shelf Life | Shelf life of 2 years. | Shelf life of 4 years. Shelf life product was<br>2X sterilized, passed design verification<br>testing including pouch seal peel and bubble<br>leak tests. The extension of the shelf life does<br>not impact safety and effectiveness. |
| Size | 3F, 4F, 5F, 6F, 7F, 8F, 10F, 12F | Same - No change |
| Biocompatibility | Meets ISO 10993-1 | Same - No change |
| Sterilization<br>method and<br>sterility | EtO and SAL of 10-6 | Same - No change |
#### 8.7 DEVICE COMPARISON
Table 8-1; Comparison of Dovice Charactoristics to Marketed Prodicate Device
{5}------------------------------------------------
{6}------------------------------------------------
#### PERFORMANCE DATA 8.8
Performance tests include dimensional verification, functional tests and marking durability. The company performed testing to demonstrate that the device meets product specifications and is not negatively impacted by the material change or marking process change. The device uses a modified Polypropylene material formulation to optimize the device markings using a laser instead of pad printing the device markings. All other materials are the same as its predicate device. The device continues to meet the same specifications as its predicate devices. Test results demonstrate that the device functions as intended. The following tests were performed:
Dimensional
{7}------------------------------------------------
Joint Tensile Strength Base & Cover Interaction Hinge performance Device Marking Durability
#### SUBSTANTIAL EQUIVALENCE 8.9
The SecurAcath device covered by this submission is substantially equivalent to the predicate Interrad Medical SecurAcath device K180994.
The Polypropylene material formulation change and change from pad printing to laser marking to the SecurAcath device results in the same indication for use, same principles of operation, same technological characteristics, incorporates the same design, and are packaged and sterilized using the same materials and processes as the previously cleared predicate device. The changes do not raise new questions of safety or efficacy.
# 8.10 CONCLUSION
Based on the test data, the same intended use, and same indications for use, the modified SecurAcath device is found to be substantially equivalent to its predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.