← Product Code [ODC](/productcode/ODC) · K210625

# JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve (K210625)

_Fujifilm Medwork GmbH · ODC · Sep 10, 2021 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210625

## Device Facts

- **Applicant:** Fujifilm Medwork GmbH
- **Product Code:** [ODC](/productcode/ODC.md)
- **Decision Date:** Sep 10, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The JAZZ Disposable Valve Kit is comprised of the JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve. The JAZZ Suction Valve is intended to control the suction (aspiration of air and fluids) of a FUIFILM gastrointestinal endoscope during endoscopy procedures. The JAZZ Air/Water Valve is intended to control air of a FUJIFILM gastrointestinal endoscope for insufflations and water delivery (for distal lens cleaning) during endoscopy procedures. The JAZZ Forceps Valve is intended to facilitate passage of an endotherapy device, to prevent a leak or backflow of air and/or fluids, and to enable the suction function of a FUJIFILM gastrointestinal endoscope.

## Device Story

JAZZ Disposable Valve Kit consists of three single-use, sterile valves (Suction, Air/Water, Forceps) designed for FUJIFILM G7 gastrointestinal endoscopes. Valves replace reusable components, eliminating manual cleaning/reprocessing. Suction valve controls aspiration of air/fluids; Air/Water valve manages insufflation and lens cleaning; Forceps valve seals biopsy port to prevent leakage while allowing instrument passage. Used in clinical endoscopy settings by physicians/endoscopy staff. Output is mechanical control of endoscope functions; benefits include reduced cross-contamination risk and simplified workflow via single-use design.

## Clinical Evidence

No clinical data. Bench testing only. Evaluated per ISO 10993 (biocompatibility) and ISO 11135/11607 (sterility). Functional testing included suction pressure, insufflation/flushing pressure, tightness, functional life, and endoscope compatibility. Validated for backflow prevention per FDA guidance on cross-contamination.

## Technological Characteristics

Single-use, sterile (EO) valves. Materials evaluated per ISO 10993. Mechanical operation via spring-loaded mechanisms. Designed for specific compatibility with FUJIFILM G7 series GI endoscopes. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- DEFENDO Disposable Suction Valve Model 100305 ([K102581](/device/K102581.md))
- DEFENDO Disposable Air/Water Valve Model 100304 ([K102409](/device/K102409.md))
- DEFENDO Disposable Biopsy Valve Model 100301 ([K090851](/device/K090851.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 10, 2021

FUJIFILM medwork GmbH % Jeffrey Wan Manager, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 81 Hartwell Avenue, Suite 300 Lexington, MA 02421

K210625 Re:

> Trade/Device Name: JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: ODC, FDF Dated: August 6, 2021 Received: August 9, 2021

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210625

Device Name

JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve

Indications for Use (Describe)

The JAZZ Disposable Valve Kit is comprised of the JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve.

The JAZZ Suction Valve is intended to control the suction (aspiration of air and fluids) of a FUIFILM gastrointestinal endoscope during endoscopy procedures.

The JAZZ Air/Water Valve is intended to control air of a FUJIFILM gastrointestinal endoscope for insufflations and water delivery (for distal lens cleaning) during endoscopy procedures.

The JAZZ Forceps Valve is intended to facilitate passage of an endotherapy device, to prevent a leak or backflow of air and/or fluids, and to enable the suction function of a FUJIFILM gastrointestinal endoscope.

**Type of Use (Select one or both, as applicable)**

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY FUJIFILM medwork GmbH JAZZ Disposable Valve Kit

Date: Auqust 6, 2021

## Submitter's Information:

FUJIFILM medwork GmbH Medworkring 1 91315 Höchstadt, Germany

### Contact Person:

Jeffrey Wan Manager, Regulatory Affairs Telephone: (201) 675-8947 E-Mail: jeffrey.wan@fujifilm.com

### ldentification of the Proposed Device:

Device Name: JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve Common Name: Endoscope Channel Accessory Device Class: Class II Classification Number: 21 C.F.R. § 876.1500 Classification Name: Endoscope and accessories Device Panel: Gastroentereology/Urology Product Codes: ODC, FDF

### Predicate Devices:

- DEFENDO Disposable Suction Valve Model 100305 (K102581) ●
- DEFENDO Disposable Air/Water Valve Model 100304 (K102409)
- DEFENDO Disposable Biopsy Valve Model 100301 (K090851) .

#### Intended Use / Indications for Use

The JAZZ Disposable Valve Kit is comprised of the JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve.

The JAZZ Suction Valve is intended to control the suction function of air and fluids) of a FUJIFILM gastrointestinal endoscope during endoscopy procedures.

The JAZZ Air/Water Valve is intended to control air of a FUJIFILM gastrointestinal endoscope for insufflations and water delivery (for distal lens cleaning) during endoscopy procedures.

The JAZZ Forceps Valve is intended to facilitate passage of an endotherapy device, to prevent a leak or backflow of air and/or fluids, and to enable the suction function of a FUJIFILM gastrointestinal endoscope.

{4}------------------------------------------------

#### Device Description

The JAZZ Disposable Valve Kit is comprised of the JAZZ Suction Valve, Air/Water Valve, and Forceps Valve. These devices are intended for single-use and are supplied sterile, eliminating the need for manual cleaning and reprocessing. These valves are designed to the corresponding port of a FUJIFILM G7 gastrointestinal endoscope. The air/water valve can be activated to control air and water flow, while the suction valve can be activated to aspirate excess fluids or other debris. The forceps valve forms a tight seal with the biopsy port to prevent leakage of biomaterial, provide easy passage of endotherapy devices, and support thesuction function.

### Comparison of Technological Characteristics

A comparison of technological characteristics between the JAZZ Disposable Valve Kit and the DEFENDO disposable valves is provided below:

|                        | JAZZ Suction Valve                                                                                                                                                                           | DEFENDO Disposable<br>Suction Valve 100305                                                                                                                   | Substantial<br>Equivalence |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| 510(k) number          | To be assigned                                                                                                                                                                               | K102581                                                                                                                                                      |                            |
| Product Code           | ODC                                                                                                                                                                                          | ODC, FDF                                                                                                                                                     | Similar                    |
| Regulation No.         | 21CFR § 876.1500                                                                                                                                                                             | 21CFR § 876.1500                                                                                                                                             | Identical                  |
| Classification         | 2                                                                                                                                                                                            | 2                                                                                                                                                            | Identical                  |
| Manufacturer           | FUJIFILM medwork<br>GmbH                                                                                                                                                                     | Byrne Medical                                                                                                                                                |                            |
| Supplied Sterile       | Yes                                                                                                                                                                                          | Yes                                                                                                                                                          | Identical                  |
| Sterile method         | EO                                                                                                                                                                                           | EO                                                                                                                                                           | Identical                  |
| Single use             | Yes                                                                                                                                                                                          | Yes                                                                                                                                                          | Identical                  |
| Compatibility          | FUJIFILM G7 Series GI<br>Endoscopes                                                                                                                                                          | Olympus and Pentax 90<br>Series GI Endoscopes                                                                                                                | Similar                    |
| Indications for<br>Use | The JAZZ Suction Valve is<br>intended to control the<br>suction function<br>(aspiration of air and<br>fluids) of a FUJIFILM<br>gastrointestinal<br>endoscope during<br>endoscopy procedures. | The DEFENDO Disposable<br>Suction Valve is intended to<br>be used to control the suction<br>function of an endoscope<br>during a GI endoscopic<br>procedure. | Identical                  |
| Packaging              | Suction, Air/Water, and<br>Forceps Valves are<br>housed in a single tray<br>and sealed with a Tyvek<br>sheet                                                                                 | Suction and air/water valves<br>are housed in a single tray<br>and packaged in a sealed<br>Tyvek pouch.                                                      | Similar                    |

{5}------------------------------------------------

|                        | JAZZ Air/Water Valve                                                                                                                                                                                        | DEFENDO Disposable<br>Air/Water Valve 100304                                                                                                             | Substantial<br>Equivalence |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| 510(k) number          | To be assigned                                                                                                                                                                                              | K102409                                                                                                                                                  |                            |
| Product Code           | ODC                                                                                                                                                                                                         | ODC, FDF                                                                                                                                                 | Similar                    |
| Regulation No.         | 21CFR § 876.1500                                                                                                                                                                                            | 21CFR § 876.1500                                                                                                                                         | Identical                  |
| Classification         | 2                                                                                                                                                                                                           | 2                                                                                                                                                        | Identical                  |
| Manufacturer           | FUJIFILM medwork GmbH                                                                                                                                                                                       | Byrne Medical                                                                                                                                            |                            |
| Supplied Sterile       | Yes                                                                                                                                                                                                         | Yes                                                                                                                                                      | Identical                  |
| Sterile method         | EO                                                                                                                                                                                                          | EO                                                                                                                                                       | Identical                  |
| Single use             | Yes                                                                                                                                                                                                         | Yes                                                                                                                                                      | Identical                  |
| Compatibility          | FUJIFILM G7 Series GI<br>Endoscopes                                                                                                                                                                         | Olympus and Pentax 90<br>Series GI Endoscopes                                                                                                            | Similar                    |
| Indications for<br>Use | The JAZZ Air/Water Valve<br>is intended to control air of<br>a FUJIFILM gastrointestinal<br>endoscope for insufflations<br>and water delivery (for<br>distal lens cleaning) during<br>endoscopy procedures. | The Single Use Air/Water<br>Valve is intended to be used<br>to control the air/water<br>function on an endoscope<br>during a GI endoscopic<br>procedure. | Identical                  |
| Packaging              | Suction, Air/Water, and<br>Forceps Valves are housed<br>in a single tray and sealed<br>with a Tyvek sheet.                                                                                                  | Suction and air/water valves<br>are housed in a single tray<br>and packaged in a sealed<br>Tyvek pouch.                                                  | Similar                    |

|                  | JAZZ Forceps Valve                  | DEFENDO Disposable<br>Biopsy Valve Model 100301 | Substantial<br>Equivalence |
|------------------|-------------------------------------|-------------------------------------------------|----------------------------|
| 510(k) number    | To be assigned                      | K090851                                         |                            |
| Product Code     | ODC                                 | ODC                                             | Identical                  |
| Regulation No.   | 21CFR § 876.1500                    | 21CFR § 876.1500                                | Identical                  |
| Classification   | 2                                   | 2                                               | Identical                  |
| Manufacturer     | FUJIFILM medwork<br>GmbH            | Byrne Medical                                   |                            |
| Supplied Sterile | Yes                                 | Yes                                             | Identical                  |
| Sterile method   | EO                                  | EO                                              | Identical                  |
| Single use       | Yes                                 | Yes                                             | Identical                  |
| Compatibility    | FUJIFILM G7 Series GI<br>Endoscopes | Olympus and Fujifilm GI<br>Endoscopes           | Similar                    |

{6}------------------------------------------------

| Indications for<br>Use | *The JAZZ Forceps<br>Valve is intended to<br>facilitate passage of an<br>endotherapy device, to<br>prevent a leak or<br>backflow of air and/or<br>fluids, and to enable the<br>suction function of a<br>FUJIFILM<br>gastrointestinal<br>endoscope. | The Single Use Biopsy Valve is<br>intended for covering the<br>endoscope biopsy port during<br>an endoscopy procedure. The<br>Single Use Biopsy Valve<br>provides access for endoscopic<br>device passage and exchange,<br>helps maintain sufflation, and<br>minimizes leakage of<br>biomaterial from the biopsy<br>port throughout the endoscopic<br>procedure. | Identical |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Packaging              | Suction, Air/Water, and<br>Forceps Valves are<br>housed in a single tray<br>and sealed with a Tyvek<br>sheet.                                                                                                                                      | Suction and air/water valves<br>are housed in a single tray and<br>packaged in a sealed Tyvek<br>pouch.                                                                                                                                                                                                                                                          | Similar   |

# Performance Data

Biocompatibility of the subject device was evaluated using the following consensus standards: ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-7:2008, ISO 10993-10:2010. Biocompatibility testing was performed in accordance with FDA's guidance, "Use of International StandardISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," published June 16, 2016.

Sterility of the subject devices was evaluated using the following consensus standards: ISO 11135:2014 and ISO 11607-1:2019.

The JAZZ Air/Water Valve was validated for its efficiency in preventing backflow to the proximal irrigation system as defined by the FDA Guidance, "Mitigating the Risk of Cross-Contamination from Valves and Accessories Used for Irrigation Through Flexible Gastrointestinal Endoscopes" published November 29, 2016.

Spring force testing was performed to evaluate the springs used in the subject devices.

The subject devices met performance specifications in the following additional non-clinical tests:

### Suction Valve

- Suction pressure
- -Tightness
- Functional life -
- Compatibility with endoscope -

### Air/Water Valve

- Insufflation pressure ।

{7}------------------------------------------------

- Flushing pressure -
- Tightness -
- Functional life -
- -Compatibility with endoscope

### Forceps Valve

- Tightness -
- -Compatibility with endoscope
- Compatibility with endoscopic instruments -
- -Ease of use

### Conclusions

The subject device JAZZ Disposable Valve Kit shares the same intended use and substantially similar indications to the predicate devices. Bench testing demonstrates that the subject device is as safe and effective as the predicate devices. Thus, JAZZ Disposable Valve Kit is substantially equivalent to the listed predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K210625](https://fda-staging.innolitics.com/device/K210625)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
