Violet Nitrile Powder Free Patient Examination Glove, Non Sterile
K210587 · Onetexx Sdn Bhd · LZA · Jun 28, 2021 · General Hospital
Device Facts
Record ID
K210587
Device Name
Violet Nitrile Powder Free Patient Examination Glove, Non Sterile
Applicant
Onetexx Sdn Bhd
Product Code
LZA · General Hospital
Decision Date
Jun 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device made of Nitrile latex intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable patient examination glove; manufactured from nitrile latex; ambidextrous design; non-sterile. Inner surface treated for smooth donning without powder lubricants. Used in clinical settings by healthcare professionals to provide a protective barrier between examiner and patient. Prevents contamination during medical examinations. Device is single-use.
Clinical Evidence
No clinical data. Bench testing only, including physical properties (ASTM D412), dimensions (ASTM D3767), watertightness (ASTM D5151), and residual powder (ASTM D6124). Biocompatibility testing performed per ISO 10993-10 and ISO 10993-11.
Technological Characteristics
Material: Nitrile latex. Form factor: Ambidextrous, powder-free, non-sterile. Standards: ASTM D6319 (physical properties/dimensions), ASTM D5151 (watertightness), ASTM D6124 (residual powder). Biocompatibility: ISO 10993-10 (irritation/sensitization), ISO 10993-11 (acute systemic toxicity).
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
June 28, 2021
Onetexx SDN BHD % Michael Van der Woude U.S. Agent Emergo Global Representation 2500 Bee Cave Road, Building 1 Suite 300 Austin, Texas 78746
Re: K210587
Trade/Device Name: Violet Nitrile Powder Free Patient Examination Glove, Non-Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: May 23, 2021 Received: June 1, 2021
Dear Michael Van der Woude:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ryan Ortega Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K210587
Device Name VIOLET NITRILE POWDER FREE PATIENT EXAMINATION GLOVE, NON STERILE
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "onetexx" in a sans-serif font, with the letters in dark blue. To the right of the word is a circular logo in teal and dark blue. The logo is a circle with a square cut out of the upper right quadrant.
#### 1.0 Submitter:
| Name: | Freddy Low |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Onetexx Sdn Bhd<br>No: 73-86, Jalan Logam 5, Perindustrian Kamunting 3,<br>Kamunting Raya Industrial Estate, 34600 Kamunting, Perak<br>Darul Ridzuan, Malaysia. |
| Phone No.: | +60 5 8070 666 |
| Fax No.: | +60 5 8070 666 |
Date of Summary Prepared: 27th May 2021
#### 2.0 Identification of the subject device:
| Trade Name | : Violet Nitrile Powder Free Patient Examination Glove, Non<br>Sterile |
|-----------------------|------------------------------------------------------------------------|
| Common Name | : Patient Examination Gloves |
| Classification Name | : Patient Examination Gloves |
| Device Classification | : 1 |
| Regulation Number | : 21 CFR 880.6250 |
| Product Code | : LZA. |
#### 3.0 Predicate Device:
### K143289
Non-Sterile, Powder Free Nitrile Examination Gloves - Violet Color -LZA Company: YTY INDUSTRY (MANJUNG) SDN. BHD.
#### 4.0 Description of The Device:
Violet Nitrile Powder Free Patient Examination Glove, Non-Sterile meets all requirements of ASTM standard D6319 and FDA 21 CFR 880.6250.
The powder free nitrile examination glove is manufactured from Nitrile latex. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves with ease without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand.
#### Indication for use: 5.0
A patient examination glove is a disposable device made of Nitrile latex intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the company "onetexx". The word "onetexx" is written in a bold, sans-serif font and is colored in a dark blue. To the right of the word is a circular logo that is partially filled in with dark blue and light blue.
#### 6.0 Summary of the Technological Characteristics of the Device:
The Violet Nitrile Powder Free Patient Examination Glove, Non-Sterile are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "onetexx". The word "onetexx" is written in a dark blue sans-serif font. To the right of the word is a circular logo that is partially dark blue and partially light blue. The dark blue portion of the logo is in the shape of a square.
Table 1
| CHARACTERISTICS | | STANDARDS | DEVICE PERFORMANCE | | COMPARISON<br>ANALYSIS |
|----------------------|---|------------|--------------------------------|-------------------|----------------------------------------------|
| | | | PREDICATE<br>VIOLET | CURRENT<br>VIOLET | |
| 510(k) Number | - | | K143289 | K210587 | |
| Manufacturer(s) | - | | YTY Industry (Manjung) Sdn Bhd | Onetexx Sdn Bhd | Same |
| Material | | ASTM D6319 | Nitrile | Nitrile | Same |
| Color | - | | Violet | Violet | Same |
| Product Code | - | | LZA | LZA | Same |
| Physical Properties | | ASTM D6319 | | | |
| Before Aging | | | | | |
| Tensile Strength: | | | 23.22-29.11Mpa | 32.35Mpa | Different but |
| Ultimate Elongation: | | | 520-580% | 568% | within the ASTM<br>standard |
| After Aging | | | | | |
| Tensile Strength: | | | 27.39-30.82 | 36.10Mpa | Different but |
| Ultimate Elongation: | | | 440-500% | 551% | within the ASTM<br>standard |
| Thickness: | | ASTM D6319 | | | |
| - Finger | | | 0.09-0.11mm | 0.10mm | Different but |
| - Palm | | | 0.06-0.07mm | 0.07mm | within the ASTM<br>standard |
| - Finger | | | 0.05-0.06mm | | Different but<br>within the ASTM<br>standard |
| Length: | | ASTM D6319 | 240 - 244mm | 246 - 255mm | Different but<br>within the ASTM<br>standard |
| Powder Free | | ASTM D6124 | 0.14mg/glove | 0.24 mg/glove | Different but<br>within the ASTM<br>standard |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "onetexx". The word "onetexx" is written in a dark blue sans-serif font. To the right of the word is a circular logo that is partially dark blue and partially light blue. The dark blue portion of the logo is in the shape of a square.
| | | DEVICE PERFORMANCE | | |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| CHARACTERISTICS | STANDARDS | PREDICATE | CURRENT | COMPARISON<br>ANALYSIS |
| | | VIOLET | VIOLET | |
| Biocompatibility | Primary Skin Irritation<br>- ISO 10993-10:2010<br>(E) & Consumer<br>Product Safety<br>Commission Title 16.<br>Chapter II, Part 1500 | Under the condition of the study<br>the device is non-irritant and<br>non-sensitizer | Under the conditions of this<br>study the device did not cause<br>an irritant response. The<br>Primary Irritant Response<br>Category is deemed 'Negligible' | Similar |
| | Dermal Sensitization-<br>ISO 10993-10: 2010<br>(E) & Consumer<br>Product Safety<br>Commission, Title 16,<br>Chapter II, Part<br>1500.3 (c) (4) | Under the condition of the study<br>the device is non-irritant and<br>non-sensitizer | Under the conditions of this<br>study the device did not<br>produce a skin sensitization<br>effect in the guinea pigs. | Similar |
| | Acute Systemic<br>Toxicity, ISO 10993-<br>11:2017 (E) | Not Tested | Under conditions of this study<br>the device did not induce any<br>systemic toxicity. | Different |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "onetexx". The text is in a dark blue color and is in a sans-serif font. To the right of the text is a circular icon that is partially dark blue and partially light blue.
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON<br>ANALYSIS |
|---------------------|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| | | PREDICATE<br>VIOLET | CURRENT<br>VIOLET | |
| Watertight (1000ml) | ASTM D5151:2019 | AQL 2.5<br>Result: 0 | Gloves passed AQL 1.5 | Different but<br>within the ASTM<br>standard |
| Indication for Use | - | A patient examination glove is a<br>disposable device intended for<br>medical purposes that is worn on<br>the examiner's hand or finger to<br>prevent contamination between<br>patient and examiner. | A patient examination glove is a<br>disposable device made of<br>synthetic rubber latex intended<br>for medical purposes that is<br>worn on the examiner's hand or<br>finger to prevent contamination<br>between patient and examiner. | Same |
| Size | Medical Glove<br>Guidance Manual -<br>Labeling | Medium | Extra Small<br>Small<br>Medium<br>Large<br>Extra Large | Similar |
| Single use | Medical Glove<br>Guidance Manual -<br>Labeling | Single Use | Single Use | Same |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the onetexx logo. The word "onetexx" is written in a dark blue sans-serif font. To the right of the word is a circular logo in a light blue color with a dark blue square in the upper right quadrant.
There are no significant differences between the two products and are identical in terms of intended use, materials design, physical properties, color, thickness and biocompatibility.
#### 7.0 Summary of Non-Clinical Testing
The performance test data of the non-clinical test for this powder free nitrile examination glove is summarized as per below.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "onetexx". The word "onetexx" is written in a dark blue sans-serif font. To the right of the word is a circular logo that is partially dark blue and partially light blue. The dark blue portion of the logo is in the upper right corner, and the light blue portion is in the lower left corner.
| | | | Acceptance Criteria | | | Results | | |
|------------------------|----------------------------------------------------------------|--------------------------------------------------------------|------------------------|-----------------|-----------------|-----------------|----------------|--------|
| Test<br>Method | Standard | Purpose of Testing | | Before<br>aging | After<br>aging | Before<br>aging | After<br>aging | Status |
| Physical<br>Properties | ASTM D412<br>(Standard Test<br>Method for<br>Vulcanized Rubber | To evaluate the<br>tensile (tension)<br>properties of glove. | Tensile<br>strength | Min 14.0<br>MPa | Min 14.0<br>MPa | 32.35Mpa | 36.10Mpa | Pass |
| | and Thermoplastic<br>Elastomers-Tension) | | Ultimate<br>elongation | Min<br>500% | Min<br>400% | 568% | 551% | Pass |
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | | Results | | Status |
|----------------|--------------------------------------------------------------------------------|--------------------------------------------------------------|---------------------|---------------------|-----------|--------|--------|
| Dimension | ASTM D3767<br>Standard Practice for<br>Rubber–<br>Measurement of<br>Dimensions | To measure the<br>length, width and<br>thickness of<br>glove | Length | Min 240 mm | Length | 249 mm | Pass |
| | | | Width | $95 \pm 10$ mm | Width | 98 mm | Pass |
| | | | Thickness | Finger - min 0.05mm | Thickness | 0.10mm | Pass |
| | | | | Palm - min 0.05mm | | 0.07mm | |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for "onetexx". The text "onetexx" is written in a dark blue sans-serif font. To the right of the text is a circular graphic that is partially filled in with dark blue and light blue.
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
|----------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Watertight | ASTM D5151<br>(Standard Test<br>Method for<br>Detection of Holes<br>in Medical Gloves) | To detect holes<br>that leak water<br>and thereby<br>compromise the<br>usefulness of the<br>glove. | Sample size: 500 pcs<br>Inspection level: G1<br>AQL: 1.5, Acceptance No. 10 | The batch size for this<br>sampling is 150,001 to<br>500,000. Hence,<br>according to the single<br>sampling plan GI, the<br>sample to be drawn is<br>under code M<br>equivalent to 315<br>pieces with accept 10<br>and reject 11 to be<br>accepted under AQL<br>1.5. During the test, 0<br>piece was found with<br>leaks. Hence it falls<br>within the acceptance<br>criteria. | Pass |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the onetexx logo. The word "onetexx" is written in a bold, sans-serif font. To the right of the word is a circular icon that is partially filled in with a darker color. The logo is simple and modern.
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results | Status |
|--------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------|---------------------------------------------------------------------------------|--------|
| Residual<br>Powder | ASTM D6124<br>(Standard Test<br>Method for Residual<br>Powder on Medical<br>Gloves) | To determine the<br>amount of<br>residual powder<br>and non-powder<br>solids found on<br>gloves. | Less than 2 mg per glove | Sample size : 5 pcs<br>Requirement : <2mg/glove<br>Result<br>:0.24mg/glove | Pass |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for "onetexx". The word "onetexx" is written in a bold, sans-serif font in a dark color, likely black or a very dark shade of gray. To the right of the word "onetexx" is a stylized graphic element in a lighter color, possibly teal or cyan. The graphic element appears to be a partial or stylized representation of a circle or a "C" shape, with a portion of it filled in with a darker shade, creating a sense of depth or dimension.
#### 8.0 Summary of Clinical Testing:
No clinical study is included in this submission.
#### 9.0 Conclusion
The conclusion drawn from the non-clinical tests demonstrate that the subject Violet Nitrile Powder Free Patient Examination Glove, Non-Sterile is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K143289.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.