← Product Code [MCW](/productcode/MCW) · K210586

# Diamonback 360 Peripheral Orbital Atherectomy System, Diamondback 360 Stealth Orbital Atherectomy system (K210586)

_Cardiovascular Systems, Inc. · MCW · Sep 14, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210586

## Device Facts

- **Applicant:** Cardiovascular Systems, Inc.
- **Product Code:** [MCW](/productcode/MCW.md)
- **Decision Date:** Sep 14, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4875
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Diamondback 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectorny system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty. The Stealth 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectorny system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

## Device Story

Percutaneous orbital atherectomy system designed to remove or reduce occlusive material and restore luminal patency in peripheral arteries and AV dialysis fistulae. System comprises a reusable saline pump, single-use orbital atherectomy device (OAD), lubricant, and guide wire. Mechanism utilizes an orbiting, diamond-coated, eccentrically mounted crown; centrifugal force presses the rotating crown against calcified plaque for differential and bi-directional sanding. Diameter of orbit increases with each pass. Operated by clinicians in a clinical setting. Output is the physical removal of stenotic material, enabling improved blood flow. Benefits include restoration of vessel patency in patients with peripheral arterial disease or graft occlusion.

## Clinical Evidence

Bench testing only. Testing included start-up torque, dynamic stall, OAD life test, sound level verification, and tight stenosis testing to confirm device specifications.

## Technological Characteristics

System utilizes an orbiting, diamond-coated, eccentrically mounted crown. Components include a reusable saline pump and single-use OAD, lubricant, and 0.014" guide wire. Modifications include ABEC 3 tolerance bearings and LY551 grease. Sterilization method is unchanged from predicate.

## Regulatory Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

## Predicate Devices

- Diamondback 360 [Stealth] Peripheral Orbital Atherectomy System ([K190634](/device/K190634.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 14, 2021

Cardiovascular Systems Inc. Kris Miller Sr. Regulatory Specialist, Regulatory Affairs 1225 Old Highway 8 NW New Brighton, Minnesota 55112

## Re: K210586

Trade/Device Name: Diamondback 360® Peripheral Orbital Atherectomy System, Diamondback 360® Stealth Orbital Atherectomy system Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: August 13, 2021 Received: August 16, 2021

Dear Kris Miller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210586

#### Device Name

Diamondback 360® Peripheral Orbital Atherectomy System Stealth 360® Peripheral Orbital Atherectomy System

#### Indications for Use (Describe)

The Diamondback 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectorny system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

The Stealth 360® Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The system is a percutaneous orbital atherectorny system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CSI. The logo consists of the letters "CSI" in a sans-serif font, with the "C" and "S" in a darker gray color and the "I" in a lighter gray. Below the letters is a curved green line, resembling a smile or an arc. The overall design is simple and modern.

#### K210586 510(K) SUMMARY

| Submitter:                                                                                                                                                                                                                                                                                                                                 | Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>Saint Paul, MN 55112                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                                                                                                                                                                                                                                                                                                            | Kris Miller<br>Sr. Regulatory Specialist<br>Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>St. Paul, MN 55112<br>Ph: 612-999-3749<br>kmiller@csi360.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Date Prepared:                                                                                                                                                                                                                                                                                                                             | February 25, 2021                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Trade Name:                                                                                                                                                                                                                                                                                                                                | Diamondback 360® Peripheral Orbital Atherectomy System,<br>Stealth 360® Peripheral Orbital Atherectomy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Common Name:                                                                                                                                                                                                                                                                                                                               | Intraluminal Artery Stripper                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulation Number:                                                                                                                                                                                                                                                                                                                         | 870.4875                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification:                                                                                                                                                                                                                                                                                                                            | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code:                                                                                                                                                                                                                                                                                                                              | MCW                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Predicate Device(s):                                                                                                                                                                                                                                                                                                                       | K190634 - Diamondback 360 [Stealth] Peripheral Orbital<br>Atherectomy System (Cardiovascular Systems, Inc.)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Description:                                                                                                                                                                                                                                                                                                                        | The Diamondback 360 and Stealth 360 Peripheral Orbital<br>Atherectomy Systems (OAS) are designed to remove or reduce<br>occlusive material and restore luminal patency by using an orbiting,<br>diamond-coated, eccentrically mounted crown.<br><br>The OAS consists of the following main components:<br>1. Reusable Saline Pump (provided non-sterile)<br>2. Single-use Orbital Atherectomy Device (OAD) (provided sterile)<br>3. Single-use Atherectomy lubricant (provided sterile)<br>4. Single-use Atherectomy guide wire (provided sterile)<br><br>Mechanism of Action<br>The Diamondback and Stealth OAS mechanism of action is defined by:<br>• Centrifugal force                                                                                                                                                                                                                                                                       |
| Orbital rotation Differential sanding Bi-directional sanding              The rapidly rotating eccentric crown creates a centrifugal force that             presses the diamond-coated crown against the calcified plaque.             With each pass of the crown, plaque is reduced and the diameter of             the orbit increases. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Indications for Use:                                                                                                                                                                                                                                                                                                                       | The Diamondback 360® [Stealth 360®] Peripheral Orbital             Atherectomy System is a percutaneous orbital atherectomy system             indicated for use as therapy in patients with occlusive             atherosclerotic disease in peripheral arteries and who are acceptable             candidates for percutaneous transluminal atherectomy.<br>The OAS supports removal of stenotic material from artificial             arteriovenous dialysis fistulae (AV shunt). The system is a             percutaneous orbital atherectomy system indicated as a therapy in             patients with occluded hemodialysis grafts who are acceptable             candidates for percutaneous transluminal angioplasty.                                                                                                                                                                                                                    |
| Comparison to<br>Predicate Device:                                                                                                                                                                                                                                                                                                         | The modified Diamondback and Stealth OAS devices is identical to             the predicate device as follows:  Same regulation number, product code and classification Same intended use and indications for use Same principles of operation Same 0.014" guidewire compatibility Same sterilization method and SAL Same number of uses per device (single use)              The modified Diamondback and Stealth OAS device is different             from the predicate device as follows:  Alternate supplier for the current motor Alternate supplier for the current driveshaft bearings. The                     new bearings have a lower specification from an ABEC                     tolerance from ABEC 7P to ABEC3 that aligns with the                     short duration of use of our device. A change in the grease                     used in the bearing from a LY75 to LY551 is included in                     this change. |
| Functional and Safety<br>Testing:                                                                                                                                                                                                                                                                                                          | The following testing was performed to confirm pre-determined             device specifications were met:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                                                                            | • Start-Up Torque test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                                                                                                                                                                                                                                                                                                            | • Dynamic Stall test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                                                                                                                                                                                                                                                                                            | • OAD Life test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                                                                                                                                                                                                                                                                                            | o Sound Level Verification                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                                                                                                                                                                                                                                                                            | • Tight Stenosis test (as needed, based on results of stall test)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Conclusion:                                                                                                                                                                                                                                                                                                                                | Testing demonstrates substantial equivalence. No new questions of safety or effectiveness were identified compared to the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for cSi, which is written in gray letters. Below the logo is a green curved line. The text "K210586" is written in black font below the green line.

Cardiovascular Systems, Inc.

Special 510(k) Pre-Market Notification for Diamondback 360® [Stealth 360®] Peripheral Orbital Atherectorny System

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letters "cSi." in gray, with a green curved line underneath. Below the logo is the text "K210586" in black. The logo appears to be for a company or organization, and the text below may be a product or identification number.

---

**Source:** [https://fda-staging.innolitics.com/device/K210586](https://fda-staging.innolitics.com/device/K210586)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).