Philips Electrocardiograph, PageWriter TC cardiograph (TC20, TC30, TC50, and TC70) is intended to acquire multi-channel ECG signals from adult and pediatric patients from body surface ECG electrodes and to record, display, analyze and store these ECG signals for review by the user. It is to be used in healthcare facilities by trained healthcare professionals. Analysis of the ECG signals is accomplished with algorithms that provide measurements, data presentations, graphical presentations and interpretations for review by the user. The interpreted ECG with measurements and interpretive statements is offered to the clinician on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the patient, the results of the physical examination, the ECG tracings, and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
Device Story
PageWriter TC series (TC20, TC30, TC50, TC70) acquires multi-channel ECG signals via body surface electrodes; processes signals using DXL 12/16-Lead ECG algorithm to provide measurements, interpretations, and graphical presentations; outputs include standard 12-lead ECG reports and configurable vectorcardiogram (VCG) reports. VCG functionality derives Frank lead signals (X, Y, Z) from a representative beat to render vector loops for frontal, horizontal, and sagittal planes. Used in healthcare facilities by trained professionals; supports wired/wireless connectivity to IntelliSpace ECG Management System or third-party systems for order download and DICOM report transmission. Clinicians use computer-generated interpretations as advisory information, requiring physician over-read and validation. Benefits include enhanced diagnostic visualization through VCG loops alongside standard ECG data.
Clinical Evidence
No clinical data required. Substantial equivalence supported by bench testing, including software functional testing to verify VCG report generation, loop direction, and magnitude accuracy, and compliance with electrical safety and EMC standards (IEC 60601-2-25, IEC 60601-1-2, IEC 62366-1).
Technological Characteristics
Multi-channel ECG acquisition system; battery-operated (1-2 batteries depending on model); wired/wireless connectivity (DICOM, IntelliSpace); software-based DXL 12/16-Lead ECG algorithm; VCG report generation via Cardiograph Report Manager Module; compliant with IEC 62304 (software lifecycle) and IEC 60601 series (safety/EMC).
Indications for Use
Indicated for adult and pediatric patients requiring multi-channel ECG signal acquisition, recording, display, analysis, and storage in healthcare facilities by trained professionals.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 30, 2021
Philips Medical Systems Hong Zhu Sr. Regulatory Affairs Manager Building A2, #718 Lingshi Road, Jingan District Shanghai, Shanghai 200072 China
Re: K210560
Trade/Device Name: Page Writer TC20 Cardiograph, Page Writer TC30 Cardiograph, Page Writer TC50 Cardiograph, Page Writer TC70 Cardiograph Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: February 10, 2021 Received: March 22, 2021
Dear Hong Zhu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K210560
#### Device Name
Philips Electrocardiograph Page Writer TC20, TC30, TC50, TC70
#### Indications for Use (Describe)
Philips Electrocardiograph, Page Writer TC cardiograph (TC20, TC30, and TC70) is intended to acquire multichannel ECG signals from adult and pediatic patients from body surface ECG electrodes and to record, display, analyze and store these ECG signals for review by the used in healthcare facilities by trained healthcare professionals. Analysis of the ECG signals is accomplished with algorithms that provide measurements, data presentations, graphical presentations and interpretations for review by the user.
The interpreted ECG with measurements and interpretive statements is offered to the clinician on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the results of the physical examination, the ECG tracings, and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is horizontally oriented and fills most of the frame.
# 5. 510(k) Summary
| 510(k) Summary | | | |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|-----------------|
| 5.1 Submitter | | | |
| Date Prepared | 30 June 2021 | | |
| Submitter/Owner | Philips Healthcare-VSS-Shanghai CDC<br>A division of Philips Medical Systems<br>Building A2, #718 Lingshi Road, Jingan District,<br>Shanghai, China 200072 | | |
| Key Contact | Hong Zhu<br>Sr. Regulatory Affairs Manager<br>Telephone: +86 18501616553<br>Email: hong.zhu_2@philips.com | | |
| 510(k) Submission Type | This is a Special 510(k). | | |
| 5.2 Device | | | |
| Trade Name | PageWriter TC20, TC30, TC50, TC70 | | |
| Common Name | Cardiograph | | |
| Catalog Number | TC20 (860332), TC30 (860306), TC50 (860310), TC70 (860315) | | |
| Classification Name | Panel & Name: Electrocardiograph<br>Subpart & Division: 21 CFR § 870.2340<br>Regulatory Class: Class II<br>Product Code: DPS | | |
| 5.3 Predicate Device | | | |
| | 510(k) No. | Company Name<br>Device Name | Product<br>Code |
| Predicate Device | K191738 | Philips Medical Systems<br>PageWriter TC20, TC30, TC50, TC70 | DPS |
| Reference Device | K052049,<br>K073376,<br>K132068 | Philips Medical Systems<br>Philips DXL 12/16-Lead ECG Algorithm | DPS |
| PageWriter TC20, TC30, TC50, TC70 with the addition of vectorcardiograph (VCG) function is | | | |
substantially equivalent to the legally marketed predicate devices, Philips PageWriter TC20, TC30, TC50, TC70 (K191738).
Image /page/3/Picture/6 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, the word "PHILIPS" is written in white, bold letters at the top. Below the word, there are three horizontal wavy lines, also in white. Two white stars are positioned below the wavy lines, adding a decorative element to the logo.
{4}------------------------------------------------
# 5.4 Device Description
#### PageWriter TC20, TC30, TC50, TC70 with VCG function - description of the device per 21 CFR 807.92(a) (4)
Philips Electrocardiograph, PageWriter TC cardiograph is a product family intended to acquire, record, display, analyze and store multi-channel ECG signals from adult and pediatric patients through a body surface ECG electrode. It includes PageWriter TC20. PageWriter TC30. PageWriter TC50 and PageWriter TC70. Each device under this product family is integrated with the algorithms to provide measurements, data presentations, graphical presentations and interpretations for review by the clinician on an advisory basis. The interpreted ECG with measurements and interpretive statements are to be used in conjunction with clinician's knowledge of the patient, the results of the physical examination, the ECG tracings and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
PageWriter TC cardiograph can be operated on the battery when the AC power cord or AC power connector is not in an operable condition, depending on the model of the cardiograph (TC20 - single battery only; TC30/TC50 - either one or two batteries: TC70 - two batteries only). When operating a PageWriter TC cardiograph with one battery or two, only the approved batteries with the same part number can be used.
PageWriter TC cardiograph has various configuration provided for the preference of the user, and can be updated with the compatible options/modules from the corresponding Upgrade Kit for use with all approved accessories and spare parts, including patient data cables, Patient Information Modules (PIM), lead sets, electrodes, print papers and trolleys. None of the accessories and spare parts are provided sterile.
Once configured. PageWriter TC cardiograph can provide integrated connectivity (wired or wireless) with the compatible Philips IntelliSpace ECG Management System, or other third party ECG system, for patient order download and ECG transmission. It also supports integrated connectivity with an ADT Order Update system to manage the patient demographic data, and transmit ECG reports in compliance with DICOM protocol once configured with DICOM option.
#### MODIFICATIONS
The modification to the subject devices was to provide a configurable report, vectorcardiogram (VCG) report in addition to the cleared 12-Lead Auto ECG reports. The VCG data (derived Frank lead signals for a single representative beat) is from an existing DXL 12/16-Lead ECG algorithm to provide three-dimensional signals X, Y, and Z (Frank leads) on vector angle (in degree) and vector magnitude (in mV). The cardiograph turns those signals into vector loops for the frontal plane (X,Y), the horizontal plane (X,Z) and the sagittal plane (Y,Z). The ability of the cardiograph to render and present the VCG report was done by collating VCC data and printing to the final ECG report through Cardiograph Report Manager Module under Business Logic Layer.
## 5.5 Indications for Use
Intended Use as required per 21 CFR 807.92(a)(5)
Image /page/4/Picture/13 description: The image shows the Philips logo, which is a blue shield shape with the word "PHILIPS" in white at the top. Below the word, there are three horizontal wavy lines in white, and two white stars are placed on either side of the lines. The logo is simple and recognizable, representing the brand's identity.
{5}------------------------------------------------
Philips Electrocardiograph, PageWriter TC cardiograph (TC20, TC30, TC50, and TC70) is intended to acquire multi-channel ECG signals from adult and pediatric patients from body surface ECG electrodes and to record, display, analyze and store these ECG signals for review by the user. It is to be used in healthcare facilities by trained healthcare professionals. Analysis of the ECG signals is accomplished with algorithms that provide measurements, data presentations, graphical presentations and interpretations for review by the user.
The interpreted ECG with measurements and interpretive statements is offered to the clinician on an advisory basis only. It is to be used in conjunction with the clinician's knowledge of the patient, the results of the physical examination, the ECG tracings, and other clinical findings. A qualified physician is asked to over read and validate (or change) the computer generated ECG interpretation.
## 5.6 Comparison of Intended Uses for Subject Device and Predicate
The intended use for the subject devices and the predicates is identical.
The Indication for Use statement for PageWriter TC20, TC30, and TC70 with the proposed software version is identical to the predicates. There is no change to the indication for use statement in this submission.
## 5.7 Comparison of Technological Characteristics with the Predicate Device
#### Similarities
Philips Electrocardiograph, PageWriter TC20, TC30, TC70 with the proposed modifications VCG report has similar technological and performance characteristics as the predicate devices cleared under K191738.
It is intended to acquire, record, display, analyze, store and print the multi-channel ECG signals from adult and pediatric patients through a body surface ECG electrode. It is integrated with algorithm that provide measurements, data presentations, graphical presentations and interpretations for review by the clinician on an advisory basis. The basic technological and performance characteristics are kept same for the subject devices with the proposed modifications and the predicates.
- ECG Acquisition ●
- ECG Quality Monitor ●
- ECG Memory and Transfer ●
- ECG Interpretation
- Print Preview Capability ●
- Clinical Workflow Networked Orders ●
#### Differences
Image /page/5/Picture/19 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, there are two wavy lines and two four-pointed stars. The word "PHILIPS" is written in white at the top of the shield.
Philips GSC
{6}------------------------------------------------
The cardiographs with the proposed modification provide a configurable report presentation, VCG report, in addition to the existing 12-Lead Auto ECG reports. Display formats presented by the existing ECG reports are not changed as the result of the proposed modification.
The VCG data (derived Frank lead signals for a single representative beat) is a part of DXL algorithm outputs that was cleared in DXL 12/16-Lead ECG Algorithms (K052049, K073376, K132068) and presented in three-dimensional signals X, Y, and Z (Frank leads) on vector angle (in degree) and vector magnitude (in mV). The cardiograph turns those signals into vector loops for the frontal plane (X, Y), the horizontal plane (X,Z) and the sagittal plane (Y,Z). The ability of the cardiographs to render and present the VCG report was done by collating VCG data from DXL algorithm and printing to final ECG report through Cardiograph Report Manager Module under Business Logic Layer. Final presentation of VCG is an additional page to auto ECG reports.
## 5.8 Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### Electrical safety, Electromagnetic Compatibility (EMC) and Usability
The change was evaluated for impact on electrical safety, electromagnetic compatibility (EMC) and usability according to the consensus standards, and determined that the proposed modification, the addition of vectorcardiograph report did not require additional testing. PageWriter TC20, TC50, TC70 with VCG function are in compliance with the consensus standards, including IEC 60601-2-25, IEC 60601-1-2 and IEC 62366-1.
#### Software and Cybersecurity
The change to the software was reviewed through risk management process and determined that software functional testing was required. The testing was needed to verify the ability of the cardiographs to present VCG report, and to ensure the presented vectorcardiograph (VCG) loop is in the correct direction (rotation) and has the accurate magnitude. PageWriter TC20, TC30, TC50, TC70 with VCG function was developed in compliance with IEC 62304:2006/A1:2015. The proposed change did not impact the effectiveness of the security measures in place to the subject devices.
No other functions of the subject devices were impacted by the proposed changes. No other performance data is required as a result of this modification.
## 5.9 Clinical Studies
This premarket submission did not require clinical study to demonstrate the substantial equivalence.
Image /page/6/Picture/14 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, the word "PHILIPS" is written in white at the top. Below the word, there are three wavy lines and two stars, also in white. The logo is simple and recognizable, representing the brand's identity.
Philips GSC
{7}------------------------------------------------
# a
# 5.10 Conclusions
The verification and validation demonstrate that the PageWriter TC20, TC30, TC50, TC70 with VCG function is substantially equivalent to the predicate PageWriter TC20, TC30, TC50 and TC70 cleared through K191738. The fundamental technology and intended use of the subject devices have not changed. The proposed change, the addition of vectorcardiograph (VCG) report is presented in a VCG loop with a correct direction and an accurate magnitude, which does not raise different questions of safety and effectiveness to the subject devices when compared to the subject devices can operate as intended.
Image /page/7/Picture/5 description: The image shows the Philips logo, which is a blue shield shape. Inside the shield, the word "PHILIPS" is written in white at the top. Below the word, there are two wavy lines and two stars, also in white. The overall design is simple and recognizable, representing the brand's identity.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.