The Rocket IPC Platinum Cured Catheter is intermittent, long-term drainage of symptomatic, recurrent, effusion or ascites, including malignant effusion or ascites and other recurrent effusions or ascites that do not respond to medical management of underlying disease. This device should only be used by or under the supervision of trained personnel and in conjunction with current clinical practice.
Device Story
Fenestrated silicone drainage catheter; includes polyester cuff for patient attachment; features one-way silicone valve to prevent fluid/air reflux; includes radiopaque barium sulfate stripe. Used for intermittent, long-term drainage of pleural effusions or abdominal ascites. Operates by connecting to external drainage line and vacuum bottle; vacuum provides negative pressure to remove fluid from pleural/peritoneal space. Used by trained personnel in clinical settings. Output is fluid drainage; enables symptom relief for patients with recurrent effusions without repeated hospital visits for drainage procedures.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via ISO 10993 biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation), sterilization validation (EtO residuals), and mechanical testing (EN 1617/1618 standards for deformation, leakage, flow rate, and impact resistance).
Indicated for intermittent, long-term drainage of symptomatic, recurrent pleural effusions or abdominal ascites, including malignant cases, in patients unresponsive to medical management of underlying disease.
Regulatory Classification
Identification
A patient care suction apparatus is a device used with an intrathoracic catheter to withdraw fluid from the chest during the recovery period following surgery.
Predicate Devices
Rocket Indwelling Pleural Catheter (IPC) System (K123033)
Pleurx Peritoneal Catheter Kit and Pleurx Drainage Kits (K051711)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 9, 2021
Rocket Medical Plc Tracy Charlton Regulatory Affairs Manager 2-4 Sedling Road Washington, Tyne and Wear NE38 9BZ UNITED KINGDOM
Re: K210509
> Trade/Device Name: Rocket Platinum Cured Catheter Regulation Number: 21 CFR 870.5050 Regulation Name: Patient Care Suction Apparatus Regulatory Class: II Product Code: DWM, PNG Dated: August 6, 2021 Received: August 16, 2021
Dear Tracy Charlton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Glenn B Bell, Ph.D. Director THT3A1: Renal, Gastrointestinal, Obesity and Transplantation Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210509
Device Name Rocket Platinum Cured Catheter
#### Indications for Use (Describe)
The Rocket IPC Platinum Cured Catheter is intermittent, long-term drainage of symptomatic, recurrent, effusion or ascites, including malignant effusion or ascites and other recurrent effusions or ascites that do not respond to medical management of underlying disease. This device should only be used by or under the supervision of trained personnel and in conjunction with current clinical practice.
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K210509 Page 1 of 7
Image /page/3/Picture/1 description: The image contains the logo for Rocketmedical. The logo consists of a blue icon on the left and the word "Rocketmedical" in blue text on the right. The icon appears to be a stylized letter "R" with medical symbols incorporated into its design.
Traditional 510(k) for
Rocket IPC Platinum Cured Catheter Section 05: 510(k) Summary
# SECTION 5: 510(K) SUMMARY
## 1. SUBMITTER
| Name:<br>Address: | Rocket Medical Plc.<br>Sedling Road<br>Washington, Tyne & Wear<br>NE38 9BZ<br>United Kingdom<br>Phone: 00 44 191 419 4488<br>Fax: 00 44 191 416 5693 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Tracy Charlton<br>Regulatory Affairs Manager |
| Summary<br>Preparation Date: | February 15th, 2021 |
## 2. DEVICE
| Trade Name: | Rocket Platinum Cured Catheter |
|-------------------------|------------------------------------------------------------------------------|
| Common Name: | Rocket Platinum Cured Catheter |
| Classification<br>Name: | Patient Care Suction Apparatus<br>Peritoneal dialysis system and accessories |
| Classification: | Class II, 21 CFR 870.5050<br>Class II, 21 CFR 876.5630 |
| Product Code: | DWM<br>PNG |
{4}------------------------------------------------
K210509 Page 2 of 7
Image /page/4/Picture/1 description: The image shows the logo for Rocket Medical. The logo consists of a blue circle with a white "R" inside, and the word "Rocketmedical" in blue text to the right of the circle. The font is sans-serif and the text is all lowercase.
Traditional 510(k) for
Rocket IPC Platinum Cured Catheter Section 05: 510(k) Summary
## 3. PREDICATE DEVICE
- 3.1 The Rocket Platinum Cured Catheter is claimed to be substantially equivalent to the following
legally marketed predicate devices:
- 3.1.1 Rocket Indwelling Pleural Catheter (IPC) System (K123033) & Rocket Indwelling Peritoneal Catheter (IPC) Insertion Kit (K152105).
Verification and validation of the Rocket IPC Platinum Cured Catheter was performed through extensive bench testing, sterilization, packaging and shelf life testing. Results of the testing demonstrated that the Rocket IPC Catheter design met all specifications and is adequate for its intended use. Additionally, the test results demonstrated substantial equivalence of the Rocket IPC Catheter to its predicate device:
In conclusion, the Rocket IPC Platinum Cured Catheter is substantially equivalent intended use, technological characteristics, safety, and performance in characteristics to the following legally marketed predicate device:
### K123033 - Rocket Indwelling Pleural Catheter (IPC) System,
## K152105 – Rocket Indwelling Peritoneal Catheter (IPC) Insertion Kit
### 4. DEVICE DESCRIPTION
- 4.1 The pleural space consists of two thin membranes, one lining the lung and the other lining the chest wall. Draining away the fluid can help relieve breathlessness for a short period, but the fluid often builds up again. While it is possible to have repeated drainage with a chest drain insertion, this could require repeated hospital visits. The IPC is a way of allowing fluid to be drained repeatedly without the painful drainage procedures and without having to come to hospital. The Rocket IPC Platinum Cured Catheter is a fenestrated silicone drainage catheter indicated for intermittent, long-term drainage of symptomatic, recurrent, effusion or ascites, including malignant effusion or ascites and other recurrent effusions or ascites that do not respond to medical management of underlying disease. There is a polyester cuff for attachment to the patient and a silicone one-way valve to prevent collected air and fluid from migrating back into the body. The Rocket platinum Cured Catheter is used alongside the Rocket IPC drainage line and Rocket IPC bottles (these we included in the Rocket Indwelling Pleural Catheter (IPC) System (K123033) & Rocket Indwelling Peritoneal Catheter (IPC) Insertion Kit (K152105).
- 4.2 The intended patient population is patients requiring external drainage of pleural effusions or abdominal ascites (patients suffering from fluid build-up with within the pleural/peritoneal space).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Rocket Medical. The logo consists of a blue circle with a white "R" inside, with a medical cross incorporated into the "R". To the right of the logo is the text "Rocketmedical" in blue sans-serif font.
Rocket IPC Platinum Cured Catheter Section 05: 510(k) Summary
#### INDICATIONS FOR USE 5.
- 5.1 The Rocket IPC Platinum Cured Catheter is indicated for intermittent, long-term drainage of symptomatic, recurrent, effusion or ascites, including malignant effusion or ascites and other recurrent effusions or ascites that do not respond to medical management of underlying disease. This device should only be used by or under the supervision of trained personnel and in conjunction with current clinical practice.
#### 6. COMPARISON TECHNOLOGICAL CHARACTERISTICS WITH OF THE PREDICATE DEVICE
- 6.1 The Rocket Platinum Cured Catheter has the same intended use and the same technological characteristics as the identified predicate device.
- 6.1.1 The system employs the same technologies as the identified predicate including:
- 6.1.1.1 Fenestrated silicone catheter with a one-way silicone valve mechanism to prevent the reflux of fluid or air, a polyester cuff, and a radiopaque barium sulfate stripe.
- 6.1.1.2 The catheter system uses the vacuum from a drainage bottle as a neqative pressure to remove fluid from the pleural / peritoneal space quickly and efficiently.
- 6.1.2 The system has the same technical characteristics including:
- Materials: Biocompatible Silicone tubing, polyester cuff, and 6.1.2.1 silicone adhesive is implanted in the pleural / peritoneal space.
- 6.1.2.2 Sterility Assurance Level: 1x10-6.
### 7. PERFORMANCE DATA
- 7.1 Performance bench testing of the Rocket IPC was conducted in accordance with all applicable FDA Guidance documents and ISO standards, including:
- 7.1.1 ISO 10993-1:2009 - Biological evaluation of medical devices Part 1: Evaluation and Testing.
- ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene 7.1.2 oxide Sterilization Residuals.
- 7.1.3 ISO 11135-1:2007 - Sterilization of health care products -- Ethylene oxide --Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
- 7.1.4 ASTM F1980- 07(2011): Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices batch testing.
- 7.1.5 EN 1617:1997 Sterile Drainage Catheters and Accessory Devices for Single Use.
- 7.1.6 EN 1618:1997 Catheters Other than Intravascular Catheters - Test Methods for Common Properties.
- 7.1.7 ISO 15223:2000 Medical Devices - Symbols to be used with medical device labels, labeling, and information to be supplied.
- 7.1.8 EU MDD Applicable sections of the European Medical Device Directive
{6}------------------------------------------------
Traditional 510(k) for
Image /page/6/Picture/1 description: The image shows the logo for Rocket Medical. The logo consists of a blue circle with a white "R" inside, with a plus sign in the upper left quadrant of the "R". To the right of the logo is the word "Rocketmedical" in blue, with the "Rocket" portion in a bolder font than the "medical" portion.
(MDD) 93/42/EEC as amended by 2007/47/EC.
- 7.1.9 EU MDR Applicable Sections of the European Medical Device Requlation (MDR) 2017/745/EU amended 2020/561.
- 7.1.10 2016 FDA Biocompatibility Guidance Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process", Guidance for Industry and Food and Drug Administration Staff, June 16, 2016.
- 7.1.11 ISO 10993-1 (2018) *Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process.
- 7.1.12 ISO 10993-2 (2006) Biological Evaluation of Medical Devices Part 2: Animal welfare requirements.
- 7.1.13 ISO 10993-3 (2014) Biological Evaluation of Medical Devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity.
- 7.1.14 ISO 10993-5 (2009) Biological Evaluation of Medical Devices Part 5: Tests for in vitrocytotoxicity.
- 7.1.15 ISO 10993-6 (2016) Biological Evaluation of Medical Devices Part 6: Tests for local effects after implantation.
- 7.1.16 ISO 10993-10 (2010) Biological Evaluation of Medical Devices - Part10: Tests for irritation and skin sensitization.
- 7.1.17 ISO 10993-11 (2017) Biological Evaluation of Medical Devices- Part 11: Tests for systemic toxicity.
- 7.1.18 ISO 10993-12 (2012) Biological Evaluation of Medical Devices - Part 12: Sample preparation and reference materials.
- 7.1.19 ISO 10993-17 (2002) Biological Evaluation of Medical Devices - Part 17: Establishment of allowable limits for leachable substances.
- 7.1.20 ISO 10993-18 (2020) Biological Evaluation of Medical Devices Part 18: Chemical characterization of materials.
- 7.1.21 ASTM Standard D4169 Standard Practice for Performance Testing of Shipping Containers and Systems
- 7.2 A list of Performance Testing conducted on the Rocket IPC Catheter includes:
- 7.2.1 Sterilization Validation
- 7.2.2 Biocompatibility Validation
- 7.2.3 Packaging Validation
- 7.2.4 All bench testing, unless otherwise specified, was conducted on the finished devices, which were sterilized by the final validated sterilization process.
{7}------------------------------------------------
Traditional 510(k) for
Image /page/7/Picture/1 description: The image shows the logo for Rocketmedical. The logo consists of a blue circle with a white cross inside, followed by the word "Rocketmedical" in blue. The font is sans-serif and the text is all lowercase except for the "R" in Rocket.
Rocket IPC Platinum Cured Catheter Section 05: 510(k) Summary
- 7.3 A Summary of Standards Based compliance testing conducted on the Rocket IPC Catheter is shown in Table 9.1 and 9.2 below:
| Test | Test Method or Standard Reference | Sample Size | Final Report | Accept/Reject Criteria | Results |
|-----------------------------------------------------------------------|-------------------------------------------|-------------|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Resistance to deformation | EN 1617:1997 4.2 (including Annex A) | 3 | NPD QA 1050 | The drainage system or any component intended to operate under negative pressure shall not show deformation sufficient to impair the function of the device at the maximum negative pressure stated by the manufacturer | PASS |
| Force at break - Drainage catheters and all other parts of the system | EN 1617:1997 4.3.2; EN 1618: 1997 Annex B | 3 | NPD QA 1054 | The minimum force at break for catheter and all other parts of system shall be 20 N. (Nominal outside diameter >4 mm) | PASS |
| Freedom from leakage - During aspiration or vacuum | EN 1617:1997 4.5; EN 1618:1997 Annex D | 3 | NPD QA 1051 | Neither the drainage system nor any components shall leak at the maximum negative pressure stated by the manufacturer (73.2-73.6 cmHg) | PASS |
| Impact resistance - Collection device | EN 1617:1997 4.6 (including Annex B) | 3 | NPD QA 1052 | The collection device shall not leak | PASS |
| Impact resistance - Suction Source | EN1617:1997 4.6 (including Annex B) | 3 | NPD QA 1052 | The suction source shall not show any loss of vacuum greater than 2 % | PASS |
| Flow Rate | EN 1618:1997 Annex D | 3 | NPD QA 1053 | Calculate the arithmetic average of three determinations and express it as water flow rate through the catheter in millilitres per minute | PASS |
## Table 9.1: Summary of Verification Testing
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Rocket Medical. The logo consists of a blue circle with a white "R" inside, and the word "Rocketmedical" in blue text to the right of the circle. The "R" in the circle has a cross-like shape in the middle.
Rocket IPC Platinum Cured Catheter Section 05: 510(k) Summary
| Ref No. | Test Performed | Results |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| A1 584464 | MEM Elution GLP Report | PASS |
| A2 584476 | ISO Guinea Pig Maximization Sensitization Test | Did not elicit a<br>sensitization<br>response |
| A3 584477 | ISO Intracutaneous Reactivity Test | PASS |
| A4 584454 | ISO Acute Systemic injection Test | PASS |
| A5 584455 | Rabbit Pyrogen Test (Material Mediated) ISO | Requirements Met |
| A6 584456 | Subacute 14-day Intra Peritoneal Toxicity Study in<br>Mice | PASS |
| A7 584465 | The Salmonella Typhimurium Reverse Mutation<br>Assay | Non-Mutagenic |
| A8 584466 | Chromosome Aberration Assay | PASS |
| A9 584462 | In Vivo Mouse Micronucleus Assay | Non-Mutagenic |
| A10 584463 | In Vitro Mouse Lymphoma Assay | Non-Mutagenic |
| A11 584453 | Intramuscular Implant Test Thirteen Week Duration | Non-Irritant |
| A12 584452 | Intramuscular Implant Test Six Month Duration | No-Irritant |
| 19T 35058-11,<br>12, 13 | Chemical Characterization | Refer to Test |
| 267037 | RA Final Evaluation of Local Tissue Effects of<br>ZASEM222 Catheter Assembly following<br>intramuscular Implantation in the Rabbit.<br>(ZASEM222 is now renamed as ZASEM297). | Refer to Test |
| UK000326 -<br>Rev 1 | Biological Risk Assessment, ZASEM222 Platinum<br>Cured Silicone Catheter including ZDLTR043 Steel<br>Dilator). | Refer to Test |
| 54400-16-MT Transport Study Report J15474-TX-1 | | |
| EtO Residual Test Reports IPC | | Original Catheter in<br>substantial<br>equivalence 510k |
# Table 9.2: Summary of Further Testing Performed
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Rocket Medical. The logo consists of a blue circle with a white stylized "R" inside, followed by the text "Rocketmedical" in blue. The font is sans-serif and the text is all lowercase.
K210509 Page 7 of 7 Traditional 510(k) for
Rocket IPC Platinum Cured Catheter Section 05: 510(k) Summary
#### CONCLUSIONS 8
K152033 - The technology characteristics of the Rocket Medical IPC system are slightly different in the following ways. Although the premise of inserting a silicone catheter to drain pleural fluid and the technique of insertion are identical to the Pleurx version, the one-way valve design itself is slightly different, however the performance and its intended use are identical. The way the catheters are connected differs in that the Pleurx catheter has a push fit system, whereas the Rocket Medical system has a push fit and bayonet cap to ensure that dislocation does not occur. Verification and validation of the Rocket IPC System was performed through extensive bench testing, sterilization, packaging and shelf life testing. Results of the testing demonstrated that the Rocket IPC design met all specifications and is adequate for its intended use. Additionally, the test results demonstrated substantial equivalence of the Rocket IPC to its predicate device: In conclusion, the Rocket IPC is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the following legally marketed predicate device: PleurX Pleural Catheter System (K112831), manufactured by CareFusion
K152105 - Verification and validation of the Rocket IPC was performed through extensive bench testing, sterilization, packaging and shelf life testing. Results of the testing demonstrated that the Rocket IPC design met all specifications and is adequate for its intended use. In conclusion, the Rocket IPC is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the following legally marketed predicate device: 10.1.1 Pleurx Peritoneal Catheter Kit and Pleurx Drainage Kits (K051711), manufactured by CareFusion.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.