Alpha Dent Implants Dental Implants System is intended for surgical placement in the maxillary and/or the mandibular arch, to support crowns, bridges, or over dentulous patients. It is intended to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. The System is intended to be used in either single tooth or multiple teeth applications. The prostheses can be screw or cement retained to the abutment. The Alpha Dent Implants Dental Implants System is indicated also, for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
System consists of endosseous root-form dental implants (Active Konus, Classic Konus, Active Bio) and prosthetic components (abutments, healing caps, ball attachments, locators). Implants are surgically placed into maxillary/mandibular bone by dental professionals to support prosthetic restorations (crowns, bridges, overdentures). Implants feature internal hexagon/conical connections for abutment interface. Prosthetic components are screw or cement-retained. System restores chewing function and aesthetics. Immediate loading is permitted given sufficient primary stability. Device provides structural support for artificial teeth; clinical decision-making relies on surgeon assessment of bone quality and occlusal loading requirements.
Clinical Evidence
No clinical studies were performed. Evidence is based on bench testing, including static and dynamic compression performance testing per ISO 14801, biocompatibility evaluation per ISO 10993-1, and sterilization validation per ISO 11137 and ISO 17665.
Technological Characteristics
Implants manufactured from Ti 6Al 4V ELI titanium alloy (ASTM F 136). Features tapered/conical/spiral designs with self-tapping threads. Internal hexagon and conical implant/abutment connection. Sterilization via gamma irradiation (implants) or moist heat (abutments). Dimensions vary by diameter (3.3-6.0 mm) and length (8-16 mm).
Indications for Use
Indicated for edentulous or partially edentulous patients requiring surgical placement of dental implants in the maxillary or mandibular arch to support crowns, bridges, or overdentures and restore chewing function. Suitable for single or multiple tooth applications and immediate loading when primary stability and appropriate occlusal loading are achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
IS-III active System, Neobiotech Co., Ltd (K181138)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Alpha Dent Implants GmbH Simha Sibony Regulatory Affairs Consultant Hanauer Street 8 Pforzheim, 75181 GERMANY
November 1, 2021
Re: K210499
Trade/Device Name: Alpha Dent Implants Dental Implants System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 20, 2021 Received: September 27, 2021
Dear Simha Sibony:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210449
Device Name Alpha Dent Implants Dental Implants System
### Indications for Use (Describe)
Alpha Dent Implants Dental Implants System is intended for surgical placement in the maxillary and/or the mandbular arch, to support crowns, bridges, or over dentulous patients. It is intended to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. The System is intended to be used in either single tooth or multiple teeth applications. The prostheses can be screw or cement retained to the abutment.
The Alpha Dent Implants Dental Implants System is indicated also, for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
C)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font, and to the right of the word "IMPLANTS" is a small German flag.
# 510(k) Summary
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Alpha Dent Implants Dental Implants System
# 1. GENERAL INFORMATION
| Date Prepared: | October 18th, 2021 |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Alpha Dent Implants Dental Implants System |
| Classification Name: | Implant, Endosseous, Root-Form |
| Class: | II |
| Product Code: | DZE |
| Subsequent product code | NHA |
| CFR section: | 21 CFR872.3640 |
| Device panel: | Dental |
| Legally Marketed Primary<br>Predicate Device: | K181138 - IS-III active System, Neobiotech Co., Ltd |
| Legally Marketed<br>Reference Devices: | K180968 – Alpha Dent Implants Ltd ; K181381-A.B. Dental Devices Ltd |
| Submitter: | Dr Boris Simanovski -CEO<br>Alpha Dent Implants GmbH<br>Hanauer Str.8, 75181 Pforzheim, Germany<br>E: dr.simanovski@gmail.com<br>Tel: +4917678531156 |
| Contact: | Simha Sibony- Regulatory Affairs Consultant<br>GMRE Ltd<br>21 Hazamir St.<br>Nahariya 2226024<br>Israel<br>Email: simhasibony@gmail.com<br>Tel: +972-52-654-6625 |
# 2. DEVICE DESCRIPTION
The Alpha Dent Implant Active Conus (IAK), Implant Classic Conus (ICK), Implant Active Bio (IAB) Dental Implants System consists of one or two stage Endosseous form dental implants, with same platform of implant/prosthetics abutments connection- internal hexagon for anti-rotation and internal cone, and It is intended to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small image of the German flag.
#### IMPLANTS 2.1
#### IMPLANT ACTIVE KONUS 2.1.1
Alpha Dent Active Konus are implants with tapered external conical shape that mimics the shape of the natural tooth root. These implants are used at two stage protocol for all types of bones. Alpha Dent Active Konus are spiral, conic, with deep and sharp double threads for self- retention.
Active Konus implant has a platform with an inner cone and a hexagon for positioning. This type of platform allows creating a conical connection with suprastructures.
Narrow Platform 2,1 mm
| IAK- Implant Active Konus | Predicate K181138 |
|-------------------------------------------|-------------------|
| Length 8, 10, 11.5, 13 mm Diameter 3.3 mm | |
| Length 8, 10, 11.5,13 mm Diameter 3.75 mm | |
| Length 8, 10, 11.5,13 mm Diameter 4.2 mm | |
| Length 8, 10, 11.5,13 mm Diameter 5.0 mm | |
| Length 8, 10, 11.5,13 mm Diameter 6.0 mm | |
| Made of titanium alloy Ti 6Al 4V ELI | |
### 2.1.2 IMPLANT CLASSIC KONUS
Classic Konus implant has a platform with an inner cone and a hexagon for the positioning of dental structures. This type of platform allows creating a conical connection with suprastructures. The 3.3 mm diameter implants are installed in front and lateral parts (canine teeth, incisors, premolar teeth).
Narrow Platform 2.1 mm
| ICK- Implant Classic Konus | Predicate K181138 |
|--------------------------------------------|-------------------|
| Length 10, 11.5, 13 mm Diameter 3.3 mm | |
| Length 8, 10, 11.5, 13 mm Diameter 3.75 mm | |
| Length 8, 10, 11.5, 13 mm Diameter 4.2, mm | |
| Length 8, 10, 11.5,13 mm Diameter 5.0 mm | |
| Length 8, 10, 11.5,13 mm Diameter 6.0 mm | |
| Made of titanium alloy Ti 6Al 4V ELI | |
## 2.1.3 IMPLANT ACTIVE BIO
Active Bio is a conical-shaped implant with a helical thread. It is used under one- or two-steps procedure in all bone types.
Narrow Platform 2,1 mm
| IAB - Implant Active Bio | Predicate K181138 | |
|--------------------------|-------------------|--|
| | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small graphic that appears to be the German flag.
Image /page/5/Picture/1 description: The image shows a close-up of a dental implant. The implant is made of metal and has a screw-like shape. The threads of the screw are visible, and there are also some holes in the implant. The implant is designed to be inserted into the jawbone to support a false tooth.
Length 8, 10, 11.5,13 mm Diameter 3.3 mm Length 8, 10, 11.5, 13 mm Diameter 3.75 mm Length 8, 10, 11.5, 13 mm Diameter 4.2, mm Length 8, 10, 11.5, 13 mm Diameter 5.0 mm Length 8, 10, 11.5, 13 mm Diameter 6.0 mm Made of titanium alloy Ti 6Al 4V ELI
#### 2.2 PROSTHETIC COMPONENTS
The Alpha Dent Implants Dental Implants System includes prosthetics components that consist of healing caps, Cemented restorations: straight and angular abutments (regular/narrow/ wide/shoulder/esthetic abutment); Screw retained restorations: Multi unit, Titanium Esthetic abutments; Removable restorations: Ball attachments, Locators.
For all the prosthetic components of the subject device described below, the Implant – abutment connection is same platform with an inner cone and a hexagon for positioning:
NARROW CONICAL PLATFORM 2.1 mm (3.0)
This implant/abutment connection is compatible to all subject device implants: ICK, IAK, IAB
## Straight Abutments- cemented retained reconstruction
A variety of titanium straight abutments are available for use in different cases. They are used in the fabrication of cement-retained restorations, single crowns or bridges.
Standard abutments with an extended body are used for thick gums, or when the implant is very deep.
- . Narrow abutments are used in minimal prosthetic space
- . Wide abutments are used in wide prosthetic spaces, mainly for posterior teeth.
● Anatomic abutments have a shape that is contoured to the gingiva to allow for individualization of a prosthetic unit.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha, followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small image of the German flag.
| Image: Abutment 1 | STRAIGHT<br>CONICAL | Straight titanium<br>antirotational conical<br>abutment. Abutment<br>includes a screw.<br>Available in sizes: 7<br>mm, 9 mm, 11 and 12<br>mm. | |
|------------------------------------------------------------------------|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image: Abutment 2 | STRAIGHT<br>CONICAL | Wide antirotational<br>titanium conical<br>abutment with hexagon.<br>Mainly used in the area<br>of posterior teeth. | |
| Image: Abutment 3 | STRAIGHT<br>CONICAL | Anatomical titanium<br>shoulder abutment<br>straight conical. | |
| Image: Abutment 4 | STRAIGHT<br>CONICAL | Anatomical titanium<br>shoulder wide abutment<br>straight conical.<br>Available in three<br>shoulder sizes: 1, 2, 3<br>mm. | |
| TAK3 – Titanium<br>Abutment Konus | | | |
| TAWK 3/9 -<br>Titanium Abutmet<br>Wide Konus | | | |
| TAK3001/002/003;<br>Anatomical<br>Titanium Shoulder<br>Abutment | | | |
| TAWK3001/002/003<br>– Anatomical<br>Titanium Shoulder<br>Wide Abutment | | | |
| | TAAK3/015 | 15°,<br>CONICAL | Angular antirotational<br>titanium conical<br>abutment 15° without a<br>shoulder.<br><br>It is used in cases, when<br>the implant is installed<br>at an angle. |
| Image: abutment | | | |
Angular Abutments - cemented retained reconstruction
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the company name, the word "IMPLANTS" is written in a smaller font. To the right of the word "IMPLANTS" is a small graphic of the German flag.
Angular abutments are used when a change to the axis of the implant is required. Normally used for constructing cement-retained single crowns or bridges. The abutments are available with angles of 15° and 25°.
- Angular, narrow abutments are used in minimal prosthetic space.
- . Angular anatomic abutments have a shape that is contoured to the gingiva. This enables an customization of the prosthetic unit. • Narrow top angular abutment allows minimal technical preparation
Titanium Angulated Abutments (shoulder)
| Image: dental abutment | | | Angular anatomical<br>titanium conical<br>abutment 15°and 15°<br>with shoulder 1-3 mm. |
|------------------------|-----------------|-----------------|-------------------------------------------------------------------------------------------------------------------|
| | TAAK3/01501/02/ | 15°,<br>CONICAL | It is used for aesthetic<br>tasks solving in<br>vestibular area, and<br>replicating the shape of<br>gingival line |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha, followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font, and to the right of the word "IMPLANTS" is a small German flag.
| Image: Dental abutment | TAAK3/025<br>TAAK3/02502 | 25°, CONICAL | Angular anatomical titanium conical abutment 25° and 25° with shoulder 2 mm.<br>It is used for aesthetic tasks solving in vestibular area, and replicating the shape of gingival line. |
|------------------------|--------------------------|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------|--------------------------|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## Ball attachment - removable prosthesis
The ball attachment prosthetic is intended to secure a removable prosthesis. The attachment is used in conjunction with a stainless-steel cap and an intermediate silicone insert. Not used for single unit reconstruction
| Image: Dental implant | BAK- Ball Attachment Konus<br>Length; 2, 3,4,5,6mm<br>Platform :<br>NARROW CONICAL PLATFORM 2.1 mm (3.0)Made of titanium alloy Ti 6Al 4V ELI | A structure with conical ball<br>attachment is used for<br>fixation of removable<br>denture. A cap made of<br>stainless<br>steel can be included, as well<br>as an intermediate silicone<br>cap of three different<br>degrees of hardness: soft,<br>standard,<br>and strong. |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
## Locators with a conical connection is used for removable denture fixation.
| Image: dental implant | LCK3- Locators<br>Length; 2, 3,4,5,6 mm<br>Platform :<br>• NARROW CONICAL PLATFORM 2.1 mm (3.0)<br>Made of titanium alloy Ti 6Al 4V ELI | Locator with a conical<br>connection is used for<br>removable denture fixation.<br>A cap made of stainless steel<br>can be included, as well as an<br>intermediate silicone cap. |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### 2.3 AESTHETIC ANTIROTATIONAL ABUTMENT (STRAIGHT MULTI-UNIT)
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small graphic that appears to be the German flag.
Aesthetic antirotational abutment (straight multi-unit) with a sleeve is used in order to restore a bridge with a screw retention.
Screw in with a maximum force of 20 newton
Image /page/9/Picture/3 description: The image shows an AEAK3 - Aesthetic Antirotational Abutment. The abutment has lengths of 1, 2, and 3 mm. The platform is a NARROW CONICAL PLATFORM with a size of 2.1 mm (3.0). The image shows a side view of the abutment.
#### AESTHETIC ABUTMENT (STRAIGHT MULTI-UNIT) 2.4
Aesthetic abutment (straight multi-unit) with a sleeve is used in order to restore a bridge with a screw retention.
Screw in with a maximum force of 20 newton
All aesthetic abutments are intended only for Multi Unit loaded restoration.
No single unit implant use with aesthetic abutments.
Image /page/9/Picture/9 description: The image shows a 3D rendering of a dental implant. The implant has a screw-like structure at the bottom, a cylindrical body in the middle, and a hexagonal head at the top. The implant is silver in color and appears to be made of metal.
- AEAK+ Aesthetic Abutment Length: 1, 2, 3 mm Platform : ● NARROW CONICAL PLATFORM 2.1 mm (3.0)
#### 2.5 ONE-PIECE ANGULAR MULTI-UNIT 17° /30° CONICAL
Multi-unit system is used for a prosthesis with screw retention.
One-piece angular multi-unit allows rehabilitation at non-parallel implants by correcting the angle.
| 17°/30°Conical | |
|-----------------------|-------------------------------------------------------------------------------------------|
| Image: Dental implant | MUBK3.0 – One Piece Angular Multi-Unit 17°/30° Conical<br>Length:1, 2, 3 mm<br>Platform : |
| | • NARROW CONICAL PLATFORM 2.1 mm (3.0) |
- 2.6 HEALING CAP KONUS 3
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small image of the German flag.
The titanium conical healing cap is installed on the implant and is designed to form the gingival margin and gingival papilla. Available in the following heights: 2, 3, 4, 5, 7mm. The size is determined by the height of the gums.
Image /page/10/Picture/2 description: The image shows a 3D rendering of a dental implant. The implant has a cylindrical head with a slightly rounded top and several horizontal grooves. Below the head, the implant tapers down to a narrower, threaded screw-like section, which is designed to be inserted into the jawbone.
Healing cap konus 3 (HCK3) Length: 2, 3, 4, 5, 7 mm
#### HEALING CAP WIDE KONUS 3 2.7
Image /page/10/Picture/5 description: The image shows a close-up of a metal screw with a large, flat head. The screw has a cylindrical body with threads that run along its length. The head of the screw is wider than the body and has a smooth, rounded surface. The screw appears to be made of a shiny metal, possibly stainless steel or aluminum. The image is well-lit, and the screw is in focus.
Healing cap wide konus 3 (HCWK3) Length: 2, 3, 4, 5 mm
## 3. MATERIALS:
The implants are manufactured from Titanium alloy (Ti 6Al 4V ELI) complying with standard ASTM F 136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for surgical implant applications.
# 4. INDICATION FOR USE
Alpha Dent Implants Dental Implants System is intended for surgical placement in the maxillary and/or the mandibular arch, to support crowns, bridges, or over dentulous patients. It is intended to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. The System is intended to be used in either single tooth or multiple teeth applications. The prostheses can be screw or cement retained to the abutment.
The Alpha Dent Implants Dental Implants System is indicated also for immediate loading when primary stability is achieved and with appropriate occlusal loading.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small graphic that appears to be the German flag.
# 5. SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES Alpha Dent Implants Dental
Implants System subject to this submission is substantially equivalent to K181138 - IS-III active System, Neobiotech Co., Ltd in terms of intended use, implant/abutment connection design, materials used, safety and performance testing.
The surface treatment, packaging and sterilization are substantially equivalent to K180968 – Alpha Dent Dental Implants Ltd.
Evidence of equivalence has been demonstrated through:
* The Alpha Dent Implants Dental Implants System subject device's intended use and indications for use were previously cleared by FDA for the reference device K180968 – Alpha Dent Dental Implants Ltd.
- * The technical characteristics of the Alpha Dent Implants System subject devices' are similar to those of reference device K180968 – Alpha Dent Dental Implants Ltd.
- * Safety and performance testing of the Alpha Dent Implants System subject device's are similar to those of the predicate device K181138 - IS-III active System, Neobiotech Co.
Therefore, the Alpha Dent Implants Dental Implants System subject device is substantially equivalent to the predicate devices in terms of intended use, materials used, and technological characteristics.
| Device Name: | Implant IS-III active<br>Neobiotech<br>Primary Predicate | Implant<br>ALPHA DENT<br>Reference Predicate<br>Implant Active (IA)<br>ALPHA DENT<br>Reference Predicate | Implant Classic (IC)<br>ALPHA DENT<br>Reference Predicate | Implant Active Bio (IAB)<br>Implant Active Conus (IAK)<br>Implant Classic Conus (ICK)<br>ALPHA DENT |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| 510K Number | K181138 | K180968 | K180968 | Current Submission |
| Indications for use | [1] below this table | [2]below this table | [2]below this table | [2]below this table |
| Patient Population | Edentulous<br>or<br>partially | Edentulous<br>or<br>partially | Edentulous<br>or<br>partially | Edentulous<br>or<br>partially |
| | edentulous individuals | edentulous individuals | edentulous individuals | edentulous individuals |
| Material | TI CP4 of ASTM F67. | Ti Grade 5 (Ti6AL4V- | Ti Grade 5 (Ti6AL4V-ELI) | Ti Grade 5 (Ti6AL4V-ELI) |
| | | ELI) ASTM F136-13 | ASTM F136-13 | ASTM F136-13 |
| Outer Diameters<br>(mm) | 3.5; 4.0; 4.5; 5.0; 6.0; 7.0" | 3.3; 3.75; 4.2; 5.0 | 3.3; 3.75; 4.2; 5.0 | 3.3; 3.75; 4.2; 5.0; 6.0 |
| Length (mm) | 7.3;8.5; 10.0; 11.5; 13.0;<br>15.0" | 8.0; 10.0; 11.5; 13.0; 16.0 | 8.0; 10.0; 11.5; 13.0; 16.0 | 8.0; 10.0; 11.5; 13.0; 16.0 |
| Thread Design | Self-tapping | Self-tapping | Self-tapping | Self-tapping |
| Recommended<br>type of Bone | Bone - All Types | Soft Bone<br>(Type 3-4) | Hard Bone<br>(Type 1-2) | Bone - All Types |
| Implant/abutment<br>connection | Internal Hexagon for anti-<br>rotation and Internal Cone | Internal Hex Connection<br>2.4mm | Internal Hex Connection<br>2.4mm | Internal Hexagon for anti-<br>rotation and Internal Cone |
| Placement in Bone | Bone Level | Bone Level | Bone Level | Bone Level |
| Surface<br>treatments | SLA | Anodized layer | Anodized layer | Anodized layer |
| Clinical procedure | This product is a root type<br>fixture which is inserted in<br>the alveolar bone. It<br>replaces the functions of<br>the missing teeth as a<br>dental implant fixture. | Immediate loading or for<br>loading after a<br>conventional healing<br>period | Immediate loading or for<br>loading after a<br>conventional healing<br>period | Immediate loading or for<br>loading after a<br>conventional healing<br>period |
| Sterility | Gamma irradiation | Gamma irradiation | Gamma irradiation | Gamma irradiation |
### Table 1: Implants Comparison Table
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha, followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small image of the German flag.
[1] The IS-III Active Fixture is intended to be surgically placed in the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the wing function. It is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
[2] Alpha Dent Implants Dental Intended for surgical placement in the maxillary and/or the mandibular arch, to support crowns, bridges, or over dentures, in edentulous patients. It is intended to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. The System is intended to be used in either single teeth applications. The prostheses can be screw or cement. The Alpha Dent Implants Dental Implants System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading
### Table 2: Implants' Prosthetics Comparison Table
| Device Name: | Healing caps for implants with internal hexagon connection<br>Manufacturer: AlphaDent | Healing caps for implants with conical connection.<br>Manufacturer: AlphaDent |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K Number | K180968 | Current submission |
| Connection | Internal hexagon connection | Conical hexagon |
| Material | Grade 5 (Ti6AL4V-ELI) | Grade 5 (Ti6AL4V-ELI) |
| Size | HC:<br>Diameter: 4.5 mm<br>Length: 2,3,4,5 and 7 mm<br>Platform: 3.75 mm<br>HCW:<br>Diameter: 5.5 mm<br>Length: 2,3,4,5<br>Platform: 3.75 mm | HCK 3:<br>Diameter: 4.2 mm<br>Length: 2,3,4,5 and 7 mm<br>Platform: 2.80 mm<br>HCWK 3:<br>Diameter: 5.5 mm<br>Length: 2,3,4,5<br>Platform: 2.80 mm |
| Principle of Operation | This product is healing cap to formation appropriate gingival shape during the soft tissue healing period combined with implant. This product should be removed when the superstructure is set up. | This product is healing cap to formation appropriate gingival shape during the soft tissue healing period combined with implant. This product should be removed when the superstructure is set up. |
| Device Name: | Abutments for implants with internal hexagon connection<br>Manufacturer: AlphaDent | Abutments for implants with conical connection.<br>Manufacturer: AlphaDent |
| 510K Number | K180968 | Current submission |
| Connection | Internal Hexagon connection | Conical hexagon |
| Sterility: | Non-Sterile | Non-Sterile |
| Material | Grade 5 (Ti6AL4V-ELI) | Grade 5 (Ti6AL4V-ELI) |
| Size/geometry | TA - Straight anti rotation abutment<br><br>TA-Standard:<br>Diameter: 4.5 mm;<br>Length: 7, 9, 12 mm<br>Platform: 3.75 mm<br><br>TA0XX - Titanium Abutment Shoulder<br><br>Diameter: 4.5 mm<br>Length: 1, 2, 3 mm<br>Platform: 3.75 mm<br><br>TAW-Straight anti rotation abutment<br>TAW-Wide:<br>Diameter: 5.5mm;<br>Length: 9, 12 mm<br>Platform: 3.75 mm<br><br>TA0XXW- Titanium Abutment Shoulder Wide<br>Diameter: 5.5 mm | TAK 3XX - Titanium Abutment Konus.<br><br>Diameter: 3.35 mm<br>Length: 7, 9, 12 mm<br>Platform: 2.80 mm<br><br>TAK 3/0XX - Titanium Abutment Konus Shoulder<br><br>Diameter: 4.2 mm<br>Length: 1, 2, 3 mm<br>Platform: 2.80 mm<br><br>TAWK 3/9 - Titanium Abutment Wide Konus<br><br>Diameter: 5.0 mm<br>Length: 9 mm<br>Platform: 2.80 mm<br><br>TAWK 3/0XX - Titanium Abutment Shoulder Wide Konus<br>Diameter: 5.5 mm<br>Shoulder: 1, 2, 3 mm |
| | TAA0XXYY - Titanium Angulated<br>Abutment-shoulder<br>Diameter: 4.5 mm<br>Shoulder: 1, 2 , 3 mm<br>Platform: 3.75 mm<br>Angle: 15 deg and 25 deg.<br>TAA - Angular Abutments<br>Diameter: 4.80 mm<br>Platform: 3.75 mm<br>Angle: 15 deg and 25 deg. | TAAK3/0XXYY - Angulated Titanium Abutment Konus<br>Diameter: 4.2 mm<br>Shoulder: 1, 2 , 3 mm<br>Platform: 2.80 mm<br>Angle: 15 deg and 25 deg.<br>TAAK3/0XX - Angulated Titanium Abutment Konus<br>Diameter: 3.35 mm<br>Platform: 2.80 mm<br>Angle: 15 deg and 25 deg. |
| Principle of Operation | It is indicated for screw retained single<br>tooth or cement retained single tooth<br>and bridge restorations. | It is indicated for screw retained single tooth or cement<br>retained single tooth and bridge restorations. |
| Device Name: | Ball Attachments for implants with<br>internal hexagon connection<br>Manufacturer: AlphaDent | Ball Attachments for implants with conical connection.<br>Manufacturer: AlphaDent |
| 510K Number | K180968 | Current submission |
| Connection | Internal hexagon connection | Conical hexagon |
| Material | Grade 5 (Ti6AL4V-ELI) | Grade 5 (Ti6AL4V-ELI) |
| Size | BA- Ball attachment<br>Diameter: 4.0 mm<br>Heights: 1, 2, 3, 4, 5, 6 mm<br>Platform: 3.75 mm | BAK3 - Ball Attachment Konus<br>Diameter 4.2 mm<br>Heights: 2, 3, 4, 5, 6 mm<br>Platform: 2.80 mm |
| Principle of Operation | Used for implant retained mucosa-<br>supported restorations, such as<br>overdentures where the patient is fully | Used for implant retained mucosa-supported restorations,<br>such as overdentures where the patient is fully edentulous<br>in the arch to be restored. |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha, followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font, and a small German flag is to the right of the word "IMPLANTS".
| | Antirotational Esthetic Abutment for implants<br>with internal hexagon connection<br>Manufacturer: AlphaDent | Antirotational Esthetic Abutment for implants<br>with conical connection.<br>Manufacturer: AlphaDent |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| 510K Number | K180968 | Current submission |
| Connection | Internal hexagon connection | Conical hexagon |
| Material | Grade 5 (Ti6AL4V-ELI) | Grade 5 (Ti6AL4V-ELI) |
| Size | AEA - Esthetic Antirotational<br>Diameter: 4.7 mm<br>Heights: 1, 2, 3 mm<br>Platform: 3.75 mm | AEAK - Antirotational Esthetic Abutment Konus<br>Diameter: 4.7 mm<br>Heights: 1, 2, 3 mm<br>Platform: 2.80 mm |
| Principle of Operation | Esthetic Screw Abutment is designed for the screw<br>retained rehabilitation process on single or<br>multiple units. | Esthetic Screw Abutment is designed for the screw<br>retained rehabilitation process on multiple units. |
| Similarities | The subject and primary predicate have same indications for use, functions, materials, surface treatment, general shape (design) and dimensions. | |
| | Esthetic Abutment for implants with internal<br>hexagon connection<br>Manufacturer: AlphaDent | Esthetic Abutment for implants with conical<br>connection.<br>Manufacturer: AlphaDent |
| 510K Number | K180968 | Current submission |
| Connection | Internal hexagon connection | Conical hexagon |
| Material | Grade 5 (Ti6AL4V-ELI) | Grade 5 (Ti6AL4V-ELI) |
| Size | TAE - Titanium Esthetic Abutments<br>Diameter: 4.3 mm<br>Heights: 1, 2, 3 mm<br>Platform: 3.75 mm | AEAK + - Esthetic Abutment Konus<br>Diameter: 4.3 mm<br>Heights: 1, 2, 3 mm<br>Platform: 2.80 mm |
| Principle of Operation | Esthetic Screw Abutment is designed for the screw<br>retained rehabilitation process on single or<br>multiple units. | Esthetic Screw Abutment is designed for the screw<br>retained rehabilitation process on multiple units. |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font, and to the right of the word "IMPLANTS" is a small image of the German flag.
| | One-Piece angular multi-unit for implants with | One-Piece angular multi-unit for implants with | | |
|------------------------|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--|--|
| Device Name: | internal hexagon connection | conical connection. | | |
| | Manufacturer: AlphaDent | Manufacturer: AlphaDent | | |
| | | | | |
| 510K Number | K180968 | Current submission | | |
| Material | Grade 5 (Ti6AL4V-ELI) | Grade 5 (Ti6AL4V-ELI) | | |
| Size | MUB - Multi-Unit Base | MUBK3 – One-Piece angular multi-unit | | |
| | Diameter: 4.3 mm | Diameter: 4.3 mm | | |
| | Heights: 1, 2, 3 mm | Heights: 1, 2, 3 mm | | |
| | Platform: 3.75 mm | Platform: 2.80 mm | | |
| | Angle: 17 /30 deg | Angle: 17/30 deg. | | |
| | Multi-unit Angled Base is a pre-manufactured | Multi-unit Angled Base is a pre-manufactured | | |
| Principle of Operation | prosthetic component directly connected to the | prosthetic component directly connected to the | | |
| | Endosseous dental implant and is intended for use | Endosseous dental implant and is intended for use | | |
| | as an aid in prosthetic rehabilitation. | as an aid in prosthetic rehabilitation. | | |
| Similarities | | The subject and primary predicate have same indications for use, functions, materials, surface treatment, | | |
| | general shape (design) and dimensions. | | | |
| | Locators for implants with internal hexagon | Locators for implants with conical connection. | | |
| | | | | |
| Device Name: | connection | Manufacturer: AlphaDent | | |
| | Manufacturer: A.B.Dental | | | |
| 510K Number | | Current submission | | |
| | K181381 | | | |
| | | | | |
| Material | | Grade 5 (Ti6AL4V-ELI) | | |
| | Grade 5 (Ti6AL4V-ELI) | | | |
| Size | P25 - A.B. LOC Abutments | LCK3 — Locator | | |
| | Diameter: 3.75 mm | Diameter: 3.75 mm | | |
| | Head diameter: 3.85 mm | Head diameter: 3.85 mm | | |
| | Heights: 0, 1, 2, 3, 4, 5, 6, 7 mm | Heights: 2, 3, 4, 5, 6 mm | | |
| | Platform: 3.75 mm | Platform: 2.80 mm | | |
| | Used for implant retained mucosa-supported | Used for implant retained mucosa-supported | | |
| | restorations, such as overdentures where the | restorations, such as overdentures where the | | |
| Principle of Operation | patient is fully edentulous in the arch to be | patient is fully edentulous in the arch to be | | |
## Similarities:
The subject device and primary predicate(K181138) have the same conical Hexagon Implant/abutment connection.
The subject device and reference predicate (K180968) have same indications for use, materials, surface treatment, general shape (design) and dimensions, function (implant/abutment range) and packaging (sterile barrier).
Differences: There is a slight difference in design and dimensions between primary predicate Neobiotech (K181138) and the subject device AlphaDent Konus.
The difference is in the implant/abutment connection
There is a minor difference in the type of internal thread M1.80-6H for the Neobiotech predicate device and 1-72 UNF 2B for Alpha Dent subject device. Both implants feature have identical thread outer diameter of 1.80 mm and therefore the thread type variation will not affect structural performance of the implant, furthermore both implants' threads offer same pitch design and therefore deemed equivalent.
# 6. NON-CLINICAL TEST
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font, and next to it is a small graphic that appears to be the German flag.
The Implants are packaged in clean room ISO CLASS 7 using materials known in the industry to be appropriate for medical device packaging and is provided with a minimum sterility assurance level of 10° validated in compliance with ANSI/AAMI/ISO 11137-1 Sterilization of healthcare products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
The components are manufactured from medical grade Titanium alloy (Ti 6Al 4V ELI) per ASTM F 136. Biocompatibility evaluation was conducted in accordance with ISO 10993-1, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing and, FDA guidance Use of International Standard ISO 10993-
1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
The biocompatibility evaluation established that all biological endpoints were met and that the device is biocompatible for its intended use.
SEM and Surface analysis (EDS) after Anodize process demonstrated the morphology and cleanliness of the final product.
Sterilization validation tests were conducted in compliance with ANSI/AAMI/ISO 11137-1 and EN ISO 11137-2 in order to ensure safety and effectiveness related to Alpha Dent Implants Dental Implants System. Test results have demonstrated that the SAL of 106 was achieved and all testing requirements were met.
Pyrogenicity information provided is based on FDA Guidance on "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submission for Devices Labeled as Sterile." The method used to determine that the device meets pyrogen limit specifications is LAL Endotoxin Analysis with testing limit of 20 EU/device, based on USP <161>.
Accelerated aging per ASTM-F-1980 have been applied on the final packaging followed by validating package integrity in order to substantiate 5 years shelf life.
Real Time shelf life study is ongoing.
Static and dynamic compression performance test was conducted per ISO 14801- Dentistry-Implants-Dynamic fatigue test for Endosseous Dental implants.
The worst case scenario was chosen based on the FDA guideline for Endosseous, root-form implants. The implant used in the testing was the smallest in diameter and the longest in length available with this system of implants. The 25° abutment was chosen as worst case combination prosthetic component.
Alpha Dent implants that are subject to this submission have performed as good as the predicate-Neobiotech. The measured fatigue limits fall similar, compared to the Neobiotech, and the run-out bending moment in these tests was within similar values than predicate devices Neobiotech (metal dental implants with a diameter of 3.5 mm).
Therefore, we are confident that Alpha Dent Implants can rely on these results to define Substantial equivalence to predicate device.
Abutments are supplied Non sterile and are to be sterilized by end user by moist heat sterilization. Moist heat Sterilization for Abutments was validated in compliance with ANSI/AAMI/ISO 17665-1 and the test results demonstrated that the sterilization process for abutments as stated by Alpha Dent Implants, was able to reduce 6 magnitudes of the biological indicator Geobacillus stearothermophilus based on the "full cycle" approach.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Alpha Dent Implants. The logo features the Greek letter alpha followed by the words "ALPHA DENT" in a bold, sans-serif font. Below the words "ALPHA DENT" is the word "IMPLANTS" in a smaller font. To the right of the word "IMPLANTS" is a small image of the German flag.
The results of the testing indicate that the Alpha Dent Implants System is substantial equivalent to the predicate devices sighted in this submission.
# CLINICAL TEST
No clinical studies were performed.
# CONCLUSION
Based on the analysis of the comparison between the predicate device and the subject device and the performance evaluation results contained in this premarket notification, it was concluded that the proposed device is substantially equivalent to the legally marketed predicate devices (primary and reference devices). Testing data shows that the subject device does not present questions of safety or effectiveness that were not addressed by Alpha Dent Implants. Substantial equivalence to the predicate devices were established based on the existing performance tests provided in this 510(k) Submission. The Active Bio (IAB), Active Konus (IAK), Classic Konus (ICK) implants and the predicate Implant IS-III Active that was previously cleared under 510(k) number K181138 are both intended for the same use, and are operated using equivalent clinical procedures. The reference to the Classic(IC) and Active(IS)implants, cleared under K180968, was made in order to support the indications for use, materials, surface treatment, general shape (design) and dimensions, function(implant/abutment range),packaging (sterile barrier)and sterilization.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.