K210479 · Minxray, Inc. · IZL · Mar 19, 2021 · Radiology
Device Facts
Record ID
K210479
Device Name
IMPACT and X-Ranger
Applicant
Minxray, Inc.
Product Code
IZL · Radiology
Decision Date
Mar 19, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1720
Device Class
Class 2
Indications for Use
This is a portable X-ray system with following limitations of use: The device may be used for handheld diagnostic imaging of body extremities. The device may be used for stand mounted diagnostic imaging of head, abdomen, or extremities. The device may be used for stand mounted imaging of the chest when used without a grid. - Not to be used on bariatric patients, unless imaging body extremities - Not for mammography use - This device is not intended to replace a stationary radiographic system, which may be required for full optimization of image quality and radiation exposure for different exam types.
Device Story
Portable battery-operated digital X-ray system; consists of generator (K182207) and wireless digital X-ray acquisition components (K191451). Inputs: X-ray radiation through patient anatomy. Operation: High-frequency generator produces X-rays; wireless 14x17 inch digital panel captures images; data transmitted to laptop (Dell Precision 3550 for IMPACT; Dell Latitude 7424 for X-RANGER) via 300Mbps Mini Wireless N USB Adapter. Used in clinical settings; operated by healthcare professionals. Output: Digital radiographic images displayed on laptop for diagnostic review. Benefits: Provides portable diagnostic imaging capability for extremities and specific body regions; facilitates point-of-care diagnostics where stationary systems are unavailable.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including software validation, risk analysis, and compliance with IEC 60601 series standards (safety, EMC, radiation protection, usability). System integration and usability testing confirmed successful assembly and image acquisition.
Technological Characteristics
High-frequency generator; 1.35 kW; 40-90 kVp; 10-20 mA. Lithium-ion battery (57.6V DC). Wireless 14x17 inch digital receptor (154 μm pixel pitch). Connectivity: Wireless USB adapter to laptop. Software: Unmodified from K191451. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, IEC 62304, IEC 62366.
Indications for Use
Indicated for diagnostic X-ray imaging of body extremities (handheld or stand-mounted) and head, abdomen, or chest (stand-mounted only). Contraindicated for bariatric patients (except extremities) and mammography.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
March 19, 2021
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MinXray, Inc. % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K210479
Trade/Device Name: IMPACT and X-Ranger Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL, MQB Dated: March 3, 2021 Received: March 9, 2021
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210479
Device Name IMPACT and X-Ranger
#### Indications for Use (Describe)
This is a portable X-ray system with following limitations of use: The device may be used for handheld diagnostic imaging of body extremities.
The device may be used for stand mounted diagnostic imaging of head, abdomen, or extremities. The device may be used for stand mounted imaging of the chest when used without a grid.
- Not to be used on bariatric patients, unless imaging body extremities
- Not for mammography use
- This device is not intended to replace a stationary radiographic system, which may be required for full optimization of image quality and radiation exposure for different exam types.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary: 510(k) Number K210479 MinXray, Inc. 3611 Commercial Avenue Northbrook, Illinois 60062, USA Toll Free 1-800-221-2245 (USA & Canada) Tel. 1-847-564-0323 Fax 1-847-564-9040 Date Prepared: March 15, 2021 Contact: Keith Kretchmer, President
- 1. Identification of the Device: Trade/Device Names: IMPACT and X-Ranger Regulation Number: 21CFR892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: II Product Code: IZL, MQB Common/Usual Name: Mobile Diagnostic X-Ray System
- 2. Equivalent legally marketed device: K182207 Trade/Device Name: TR90BH Manufacturer: MinXray, Inc. Regulation Number: 21CFR892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: II Product Code: IZL Common/Usual Name: Mobile Diagnostic X-Ray System
- 3. Reference device: The IMPACT and X-Ranger use the same digital receptor panel and software, unmodified, as found in: Trade/Device Name: CMDR 2CW Submission: K191451 Manufacturer: MinXray, Inc. Regulation Number: 21CFR892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB, and LLZ. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 4. Indications for Use (intended use): This is a portable X-ray system with following limitations of use: The device may be used for handheld diagnostic imaging of body extremities. The device may be used for stand mounted diagnostic imaging of head, abdomen, or extremities. The device may be used for stand mounted imaging of the chest when used without a grid. - Not to be used on bariatric patients, unless imaging body extremities
- Not for mammography use
- This device is not intended to replace a stationary radiographic system, which may be reguired for
{4}------------------------------------------------
full optimization of image quality and radiation exposure for different exam types.
- 5. Description of the Device: This is a complete portable battery operated digital x-ray system with the indications for use stated above. It consists of the battery operator described in K182207 paired with the wireless battery operated digital x-ray acquisition components described in K191451. This is therefore a new combination of all previously cleared components. The digital x-ray receptor system uses a wireless panel and a Dell laptop computer fitted with a 300Mbps Mini Wireless N USB Adapter. The generator can be hand held (within the limitations of the indications), tripod mounted, or mobile stand mounted. Two models are proposed: IMPACT and X-RANGER. The two models differ in their targeted markets and in the type of computer supplied. The IMPACT uses a DELL PRECISION 3550 Laptop whereas the X-RANGER uses a DELL LATITUDE 7424 RUGGED LAPTOP. The IMPACT is aimed at the commercial market whereas the X-RANGER is aimed at the military market. The software is unmodified from that employed in our reference device (K191451) and it remains at a moderate level of concern. The image receptor (cleared under K191451) is wireless and measures 14x17 inches.
- 6. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation.
| | TR90BH K182207 | IMPACT AND X-RANGER | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Intended<br>Use: | The TR90BH is a portable X-ray system<br>with following limitations of use: The<br>device may be used for handheld<br>diagnostic imaging of body extremities.<br>The device may be used for stand<br>mounted diagnostic imaging of head,<br>abdomen, or extremities. The device<br>may be used for stand mounted imaging<br>of the chest when used without a grid. -<br>Not to be used on bariatric patients,<br>unless imaging body extremities – Not<br>for mammography use – The TR90BH is<br>not intended to replace a stationary<br>radiographic system, which may be<br>required for full optimization of image<br>quality and radiation exposure for<br>different exam types. | The same, only the model number is changed.<br>This is a portable X-ray system with following<br>limitations of use: The device may be used for<br>handheld diagnostic imaging of body<br>extremities. The device may be used for stand<br>mounted diagnostic imaging of head,<br>abdomen, or extremities. The device may be<br>used for stand mounted imaging of the chest<br>when used without a grid.<br>- Not to be used on bariatric patients, unless<br>imaging body extremities<br>- Not for mammography use<br>- This device is not intended to replace a<br>stationary radiographic system, which may be<br>required for full optimization of image quality<br>and radiation exposure for different exam<br>types. | | |
| Generator Characteristics | | | | |
| Weight: | 7.5kgs | SAME | | |
| Size / | 219 x 350 x 190 mm | SAME | | |
| Energy<br>Source: | Lithium-ion Rechargeable Battery<br>(57.6DC), 300 exposures per charge. | SAME | | |
## 7. Substantial Equivalence Chart: Below.
{5}------------------------------------------------
| | TR90BH K182207 | IMPACT AND X-RANGER |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Use<br>Interface: | Soft touch push buttons | SAME |
| Exposure<br>times: | 0.01 sec - 1.0 sec : 0.01 sec Step<br>High Power Mode<br>0.01 sec - 0.3 sec : 0.01 sec Step | SAME |
| mA: | 20 mA @ 40 kVDC - 60 kVDC (2 kVP<br>steps)<br>15 mA @ 62 kVDC - 80 kVDC (2 kVP<br>steps)<br>10 mA @ 82 kVDC - 90 kVDC (2 kVP<br>steps)<br>High Power Mode<br>15 mA @ 82 kVDC - 90 kVDC (2 kVP<br>steps) | SAME |
| Memory<br>settings<br>(technique) | 5 memories via pushbutton | SAME |
| HF Generator | High Frequency | SAME |
| kW | 1.35 kW | SAME |
| kVp: | 40 - 90kVp | SAME |
| X-ray Tube | D-0814 | SAME |
| FDA<br>Performance<br>Standard | Complies | SAME |
| Collimator | Mikasa BLD34L | SAME |
| Digital Image Capture | | |
| Panel | Not supplied | Same as used in our reference device<br>K191451, 14" x 17" Wireless<br>Pixel Pitch 154 μm<br>2304 × 2816 pixels |
| Panel<br>Performance | Not supplied | Panel MTF ~63%..........(@ 1lp/mm)<br>Panel DQE~62%..........(@ 0lp/mm) |
| Software | Not supplied | Same as used in our reference device<br>K191451 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows two different setups of the TR90BH K182207 IMPACT AND X-RANGER. The left side of the image shows the TR90BH K182207 IMPACT AND X-RANGER by itself. The right side of the image shows the TR90BH K182207 IMPACT AND X-RANGER with the optional X200S Tripod, a laptop, and a carrying bag.
- 8. Summary of non-clinical testing: The following testing still applies to our predicate generator since it has not been modified: Software validation and risk analysis was performed. Laboratory testing was performed according to the following standards:
IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) Medical Electrical Equipment, Part 1: General Requirements for Safety
IEC 60601-1-2:2007 Medical Electrical Equipment-Part 1-2: General Requirements for Safety – 2. Collateral Standard-Electromagnetic compatibility - Requirements and tests
IEC 60601-1-3:2008 (Second Edition) for use with IEC 60601-1: 2005 (Third Edition) Medical electrical equipment Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray Equipment IEC 60601-1-6:2010 (Third Edition) for use in conjunction with IEC 60601-1: 2005 (Third Edition)Medical electrical equipment Part 1-6: General requirements for safety - Collateral Standard: Usability IEC 60601-2-28:2010 (Second Edition) for use in conjunction with IEC 60601-1: 2005 (Third edition) Medical electrical equipment Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
IEC 60601-2-54 (First Edition): 2009 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
IEC 62304:2006 (First Edition) Medical device software: Software life-cycle processes
IEC 62366: 2007 (First Edition) + A1: 2014 Medical devices – Application of usability engineering to medical devices.
The test results showed compliance with the above standards. We also confirmed overall operation by taking and reviewing test images using the imaging system and software from our reference device.
Previously performed: Battery performance testing: A 30 Party Testing Lab was engaged to do safety performance testing on the rechargeable lithium-ion battery pack, done according to the United Nations "Recommendations on the Transport of Dangerous Goods Manual of Tests and Criteria." The following testing was successfully performed: Altitude simulation, Thermal test, Vibration, Shock, External short circuit, Impact, Overcharge, and Forced discharge. In addition, battery life testing was performed to establish how many x-ray exposures can be made on one battery charge. The AC/DC adapter used with the battery charger is medical grade UL Listed (E356265).
{7}------------------------------------------------
During the original design process, we took the following FDA guidance into account: Guidance for Industry and FDA Staff Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use; Document issued on December 24, 2008. This involved employing extra shielding and precautionary information in the labeling.
For the modified system representing the combination of our predicate generator and our reference imaging system, we performed integration and usability testing to confirm the ability of the user to assemble, operate, and get diagnostic images using our reference digital imaging components. Assembly and system test procedures have been written and approved.
- 9. Summary of clinical testing: Not applicable. Clinical testing was not deemed to show substantial equivalence. We relied on non-clinical testing and compliance with standards.
- 10. Conclusion: After analyzing bench tests, it is the conclusion of MinXray Inc. that the IMPACT AND X-RANGER Portable X-Ray Equipment is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.