The WorkMate Claris™ System is indicated for the use during clinical electrophysiology procedures. The WorkMate Claris System is a fully computerized system for capturing and measuring physiological data in the clinical electrophysiology (EP) laboratory. It provides digital signal acquisition and display of those electrical signals on high resolution monitors.
Device Story
WorkMate Claris System is a computerized diagnostic tool for clinical electrophysiology (EP) labs. It acquires intracardiac signals from catheters and surface ECG signals via cables. An integrated amplifier conditions these signals before computer processing. During procedures, an automated software waveform detector identifies cardiac activation on preselected leads. The system computes temporal interval measurements (intra-channel and inter-channel) on a beat-by-beat basis, displaying them in real-time. Users can review continuous digitized signals, retrieve past study passages, and interface with external imaging or ablation equipment. The system aids clinicians in monitoring and analyzing cardiac electrical activity to support diagnostic decision-making during EP procedures.
Clinical Evidence
Bench testing only. Software verification testing and design validation studies were performed to ensure the system meets requirements, remains compatible with EP lab equipment, and that the installation process for software/firmware updates is effective. Risk management review confirmed no new hazards were introduced.
Technological Characteristics
Programmable diagnostic computer (Class II, DQK). System components include display workstations (HP z440, Onyx-1723) running Windows 10. Connectivity includes integration with EP catheters, surface ECG leads, and ancillary modules (Scribe, Unity). Software includes McAfee Application Control for security. Sterilization is not applicable as the system is a diagnostic computer.
Indications for Use
Indicated for use during clinical electrophysiology procedures in the EP laboratory for capturing and measuring physiological data.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 11, 2021
Abbott Ed Sandberg Senior Regulatory Affairs Specialist One St. Jude Medical Drive St. Paul, Minnesota 55117
Re: K210392
Trade/Device Name: WorkMate Claris System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: February 9, 2021 Received: February 10, 2021
Dear Ed Sandberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh Deoras Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210392
Device Name WorkMate Claris™ System
#### Indications for Use (Describe)
The WorkMate Claris™ System is indicated for the use during clinical electrophysiology procedures.
The WorkMate Claris System is a fully computerized system for capturing and measuring physiological data in the clinical electrophysiology (EP) laboratory. It provides digital signal acquisition and display of those electrical signals on high resolution monitors.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K210392, WorkMate Claris™ System
Image /page/3/Picture/1 description: The image shows the Abbott logo. The logo consists of a blue, stylized letter "A" on the left and the word "Abbott" in black, bold font on the right. The blue "A" is made up of three horizontal lines that are connected by a vertical line on the left side.
| 510(k) Information | |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K210392 |
| 510(k) Type | Special 510(k) |
| Date Prepared | February 9th, 2021 |
| Submitter Information | |
| Manufacturer<br>Name/Address | Abbott<br>One St. Jude Medical Drive<br>St. Paul, MN 55117 |
| Contact Person | Ed Sandberg<br>Regulatory Affairs Specialist<br>Phone: 651-245-8398<br>ed.sandberg@abbott.com |
| Device Information | |
| Trade Name | Workmate Claris™ System v1.2 |
| Common Name | Programmable Diagnostic Computer |
| Class | II |
| Classification Name | 870.1425, computer, diagnostic, programmable |
| Product Code | DQK |
| Predicate Device | Workmate Claris™ System v1.1 (K151911) |
| Reference Applications | N/A |
| Device Description | The WorkMate Claris™ System is a fully computerized system for capturing and measuring physiological data in the clinical electrophysiology (EP) laboratory. It provides digital signal acquisition and display of those electrical signals on high resolution monitors.<br><br>The WorkMate Claris™ System is connected to electrophysiology catheters that are guided into various locations within the heart, and to surface electrocardiogram (ECG) cables. Intracardiac and ECG signals are then acquired from electrodes on the indwelling catheters and ECG leads connected to the amplifier, which amplifies and conditions the signals before they are received by the WorkMate Claris System computer for display, measurement and storage. The cardiac stimulator integrated with the parent recording system, sends electrical impulses to indwelling catheters through the Amplifier.<br><br>During the procedure, cardiac signals are acquired and an automated software waveform detector (trigger) performs online recognition of cardiac activation on preselected leads. Temporal interval measurements are computed on a beat-by-beat basis on multiple channels and dynamically posted on the Real Time display. Intervals are calculated between waveforms from the same source on a specific channel (intra- channel measurements) and from multi-source signals across two or more channels (inter-channel measurements).<br><br>Signals are also presented on a review monitor for measurement and analysis. Continuous capture of the digitized signals can be invoked, and the user can also retrieve and display earlier passages of the current study without interruption of the real-time display. The system can also acquire, display and record data from other interfaced devices in use during the procedure, such as imaging devices and |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Abbott logo. The logo consists of a blue symbol on the left and the word "Abbott" in black on the right. The blue symbol is a stylized letter "a" made up of three horizontal lines and a curved line. The word "Abbott" is in a bold, sans-serif font.
| ablation generators. | | | |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Indications for Use | Workmate Claris™ System v1.2<br>WorkMate Claris™ System is indicated for the use during clinical electrophysiology procedures.<br>WorkMate Claris™ System is a fully computerized system for capturing and measuring physiological data in the clinical electrophysiology (EP) laboratory. It provides digital signal acquisition and display of those electrical signals on high resolution monitors. | | |
| Submission History | No prior submissions have been made to FDA for the device that is the subject of this submission. | | |
| Predicate Comparison | | | |
| Comparison | The WorkMate Claris™ System v.1.2, subsystems, and optional ancillary modules, WorkMate™ Scribe™ Module and WorkMate™ Unity™ Review Module, use the same fundamental scientific technology, intended use, and indications as the predicate device. The changes are limited to the modification of system software to support an operating system update to Windows® 10, to manage end of life components, security updates/enhancements, and address software bug fixes.<br>There are no new or increased risks that result from the modifications and the changes do not raise any new questions of safety and effectiveness in regards to the subject device.<br>The modifications for the subject device are outlined below. | | |
| | Characteristic | Predicate Device (K151911) | Subject Device (K210392) |
| | | H700123, H700124; WorkMate Claris™ System and H700120, H700121<br>WorkMate™ Scribe™ Module | |
| | Trade /<br>Proprietary<br>Name | WorkMate Claris™ System | Same |
| | Device Class | II | Same |
| | Classification<br>Name | Computer, Diagnostic,<br>Programmable | Same |
| | Product Code | DQK | Same |
| | Indications for<br>use | The WorkMate Claris™ System is indicated for use during clinical electrophysiology procedures.<br><br>The WorkMate Claris™ System is a fully computerized system for capturing and measuring physiological data in the clinical electrophysiology (EP) laboratory. It provides digital signal acquisition and display of those electrical signals on high resolution monitors. | Same |
| Characteristic | Predicate Device<br>(K151911) | Subject Device<br>(K210392) | |
| H700977; WorkMate Claris Display Workstation | | | |
| Software<br>Version | v1.1.1 | v1.2 | |
| Workstation<br>Hardware | HP z620 | HP z440 | |
| H701060, WorkMate™ Scribe™ Display Workstation | | | |
| Software<br>Version | v1.1.1 | v1.2 | |
| Workstation<br>Hardware | HP 8200 | Onyx-1723 | |
| H800007; WorkMate Claris™ System Software Upgrade Kit v1.2 (z440) | | | |
| Operating<br>System | Windows 7 | Windows 10 | |
| Software<br>Version | v1.1.1 | v1.2 | |
| Acrobat Reader | v.11.0.06 | v.20.006.20042 | |
| Bomgar Client<br>(SJMConnect) | v. 14.3.1 | v19.1.7 | |
| McAfee<br>Application<br>Control | v.8.0.0.817 | v.8.2.1.407 | |
| Microsoft Word | v. 15.0.4420.1017 | v. 16.0.10359.20023 | |
| Microsoft<br>Power Point | v. 15.0.4420.1017 | v. v. 16.0.10359.20023 | |
| H800008; WorkMate™ Scribe™ Module Software Version 1.2 | | | |
| Operating<br>System | Windows 7 | Windows 10 | |
| Software<br>Version | v1.1.1 | v1.2 | |
| Acrobat<br>Reader | v.11.0.06 | v.20.006.20042 | |
| Bomgar Client<br>(SJMConnect) | v. 14.3.1 | v19.1.7 | |
| McAfee<br>Application<br>Control | v.8.0.0.817 | v.8.2.1.407 | |
| Microsoft Word | v. 15.0.4420.1017 | v. 16.0.10359.20023…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.