Powder Free Black Nitrile Examination Gloves, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Resistant
K210330 · Wrp Asia Pacific Sdn. Bhd. · LZA · Mar 4, 2022 · General Hospital
Device Facts
Record ID
K210330
Device Name
Powder Free Black Nitrile Examination Gloves, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Resistant
Applicant
Wrp Asia Pacific Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
Mar 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable nitrile examination glove; worn on hand/finger by healthcare providers to prevent cross-contamination. Manufactured from synthetic rubber; surface treated for smooth donning without powder. Ambidextrous design. Tested for permeation resistance against various chemotherapy drugs and fentanyl citrate per ASTM D6978-05. Used in clinical settings; provides a physical barrier between patient and examiner. Output is the barrier protection itself; healthcare providers use the glove to maintain hygiene and safety during patient examinations and handling of hazardous substances. Benefits include reduced risk of contamination and protection against specific chemical exposures.
Clinical Evidence
No clinical data. Bench testing only, including physical properties (ASTM D412), dimensions (ASTM D3767), watertightness (ASTM D5151), residual powder (ASTM D6124), and chemotherapy/fentanyl permeation (ASTM D6978).
Indicated for use as a disposable medical glove worn by examiners to prevent cross-contamination between patient and examiner. Tested for use with specific chemotherapy drugs and fentanyl citrate.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K183287)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 4, 2022
WRP Asia Pacific SDN BHD % Saravanan Ramasamy HEAD of QARA Wrp Usa Inc 3700 Massillon Road. Suite 340 Uniontown, Ohio 44685
Re: K210330
Trade/Device Name: Dermagrip Powder Free Black Nitrile Examination Gloves, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Resistant Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO Dated: February 8, 2022 Received: February 14, 2022
Dear Saravanan Ramasamy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210330
### Device Name
DERMAGRIP POWDER FREE BLACK NITRILE PATIENT EXAMINATION GLOVE, NON STERILE, TESTED FOR USE WITH CHEMOTHERAPY DRUGS AND FENTANYL RESISTANCE
### Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs and Fentany] Citrate per ASTM D6978-05 (Re-approved 2013) Standard Practice for assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Chemotherapy Drug | Breakthrough Detection Time in Minutes |
|----------------------------|----------------------------------------|
| *Carmustine (BCNU) | 21.2 |
| Cisplatin | > 240 |
| Cyclophosphamide (Cytoxan) | > 240 |
| Dacarbazine | > 240 |
| Doxorubicin Hydrochloride | > 240 |
| Etoposide | > 240 |
| Fluorouracil | > 240 |
| Ifosfamide | > 240 |
| Methotraxate | > 240 |
| Mitomycin C | > 240 |
| Mitoxantrone | > 240 |
| Paclitaxel | > 240 |
| *Thiotepa | 34.6 |
| Vincristine Sulfate | > 240 |
* WARNING: Please note the following drugs have extremely low permeation times: Carmustine (BCNU): 21.1 minutes and Thiotepa: 34.6. Do not use with Carmustine and Thiotepa.
| Fentanyl Resistance | Breakthrough Detection Time in Minutes |
|----------------------------|----------------------------------------|
| Fentanyl Citrate Injection | No breakthrough up to 240 minutes |
Type of Use *(Select one or both, as applicable)*
| <div style="display:inline-block; vertical-align:middle;">☐</div> Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block; vertical-align:middle;">❌</div> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1.0 Submitter:
| Name: | Saravanan Ramasamy |
|------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | WRP Asia Pacific Sdn Bhd<br>Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi,<br>43900 Sepang, Selangor Darul Ehsan, MALAYSIA |
| Phone No.: | +60 3 8706 1486 |
| Fax No.: | +60 3 8706 1485 |
Date of Summary Prepared: 3 March 2022
#### 2.0 Identification of the subject device:
: Dermagrip Powder Free Black Nitrile Patient Examination Trade Name Glove, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Resistance Common Name : Patient Examination Gloves Classification Name : Patient Examination Gloves Device Classification : 1 Regulation Number : 21 CFR 880.6250 Product Code
### 3.0 Predicate Device:
## K183287
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.
Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate.
Company: Kossan International Sdn. Bhd.
#### 4.0 Description of The Device:
Powder Free Black Nitrile Patient Examination Gloves, Non-Sterile, Tested for Fentanyl Resistance meet all requirements of ASTM D6319, ASTM D6978, and FDA 21 CFR 880.6250.
The powder free nitrile examination glove is manufactured from nitrile, a synthetic rubber. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves with ease without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e. can be worn on right hand or left hand.
#### 5.0 Indication for use:
Dermagrip Powder Free Black Nitrile Patient Examination Glove, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Resistance is a disposable device made of synthetic rubber latex intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. The device has also been tested for use with fentanyl citrate.
These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs:
{4}------------------------------------------------
### Chemotherapy Drug Permeation
The following chemicals have been tested for the glove.
| Chemotherapy Drug | Concentration | Breakthrough Detection<br>Time in Minutes |
|----------------------------|---------------|-------------------------------------------|
| *Carmustine (BCNU) | 3.3mg/ml | 21.2 |
| Cisplatin | 1.0mg/ml | > 240 |
| Cyclophosphamide (Cytoxan) | 20.0mg/ml | > 240 |
| Dacarbazine | 10.0mg/ml | > 240 |
| Doxorubicin Hydrochloride | 2.0mg/ml | > 240 |
| Etoposide (Toposar) | 20.0mg/ml | > 240 |
| Fluorouracil | 50.0mg/ml | > 240 |
| Ifosfamide | 50.0mg/ml | > 240 |
| Mitoxantrone | 2.0mg/ml | > 240 |
| Paclitaxel (Taxol) | 6.0mg/ml | > 240 |
| *Thiotepa | 10.0mg/ml | 34.6 |
| Vincristine Sulfate | 1.0mg/ml | > 240 |
| Methotrexate | 25.0mg/ml | > 240 |
| Mitomycin C | 0.5mg/ml | > 240 |
*WARNING: Please note the following drugs have extremely low permeation times: Carmustine (BCNU): 21.1 minutes and Thiotepa: 34.6. Do not use with Carmustine and Thiotepa.
| Fentanyl Resistant | Concentration | Breakthrough Detection<br>Time in Minutes |
|----------------------------|---------------|-------------------------------------------|
| Fentanyl Citrate Injection | 100mcg/2mL | No breakthrough up to 240<br>minutes |
#### 6.0 Summary of the Technological Characteristics of the Device:
Dermagrip Powder Free Black Nitrile Patient Examination Glove, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Resistance are summarized with the following technological characteristics compared to ASTM D6319 or equivalent standards as shown in Table 1.
{5}------------------------------------------------
| Table 1 | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--|
| | | DEVICE PERFORMANCE | | | |
| CHARACTERISTICS | STANDARDS | PREDICATE | SUBJECT DEVICE | COMPARISON<br>ANALYSIS | |
| | | BLUE | DERMAGRIP POWDER FREE BLACK<br>NITRILE PATIENT EXAMINATION<br>GLOVE, NON-STERILE, TESTED FOR<br>USE WITH CHEMOTHERAPY DRUGS<br>AND FENTANYL RESISTANCE | | |
| 510(k) Number | - | K183287 | K210330 | | |
| Manufacturer(s) | - | Kossan International Sdn. Bhd. | WRP Asia Pacific | Same | |
| Material | ASTM D6319 | Nitrile | Nitrile | Same | |
| Color | - | Blue | Black | Same | |
| Texture | - | Fingertips textured | Hand textured (Fully textured<br>surface from fingertips to end of<br>palm ) | Different | |
| Physical Properties | ASTM D6319 | Meets | Meets | | |
| Before Aging<br>Tensile Strength: | | 28Mpa min | 14MPa min | Different but | |
| Ultimate Elongation: | | 500% min | 500% min | within the ASTM<br>standard | |
| After Aging<br>Tensile Strength:<br>Ultimate Elongation: | | 30Mpa min<br>480% min | 14MPa min<br>400% min | Different but<br>within the ASTM<br>standard | |
| Thickness :<br>- Finger<br>- Palm<br>- Cuff | ASTM D6319 | 0.10mm<br>0.06mm<br>0.04mm | 0.05mm min<br>0.05mm min<br>0.05mm min | Different but<br>within the ASTM<br>standard | |
| Powder Free | ASTM D6124 | Less than 2mg per glove | Less than 2mg per glove | Same | |
| CHARACTERISTICS STANDARDS | DEVICE PERFORMANCE | | | COMPARISON ANALYSIS | |
| | PREDICATE | SUBJECT DEVICE | | | |
| | BLUE | DERMAGRIP POWDER FREE BLACK<br>NITRILE PATIENT EXAMINATION<br>GLOVE, NON-STERILE, TESTED FOR<br>USE WITH CHEMOTHERAPY DRUGS<br>AND FENTANYL RESISTANCE | | | |
| Primary Skin Irritation<br>- ISO 10993-10:2010<br>(E) & Consumer<br>Product Safety<br>Commission Title 16.<br>Chapter II, Part 1500 | Non Irritating | Passes (Not a primary skin<br>irritant)<br>There was no erythema or<br>oedema noted on test site after<br>(24±2), (48±2) and (72±2)<br>hours. The primary Irritation<br>Index (PII) was "0" | | Similar | |
| Biocompatibility | Dermal Sensitization-<br>ISO 10993-10: 2010<br>(E) & Consumer<br>Product Safety<br>Commission, Title 16,<br>Chapter II, Part<br>1500.3 (c) (4) | Non Sensitization | Passes (Not a contact<br>sensitizer).<br>There was no positive allergic<br>reaction observed during the<br>challenge phase (at 0, 24<br>hours and 48 hours) in animals<br>treated with the test material<br>and negative control. | Similar | |
| | Cytotoxicity - MEM<br>Elution, ISO 10993-5:<br>2009 (E) | Non Cytotoxic | Acute systemic toxicity testing<br>performed to demonstrate<br>safety. | Different | |
{6}------------------------------------------------
{7}------------------------------------------------
| CHARACTERISTICS | STANDARDS | DEVICE PERFORMANCE | | COMPARISON<br>ANALYSIS |
|---------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| | | PREDICATE<br>BLUE | SUBJECT DEVICE | |
| | | | DERMAGRIP POWDER FREE BLACK<br>NITRILE PATIENT EXAMINATION<br>GLOVE, NON-STERILE, TESTED FOR<br>USE WITH CHEMOTHERAPY DRUGS<br>AND FENTANYL RESISTANCE | |
| Biocompatibility | Acute Systemic<br>Toxicity, ISO 10993-<br>11:2017 (E) | Not Available | It is concluded that the product<br>did not induce any systemic<br>toxicity. | Different |
| Watertight (1000ml) | ASTM D5151:2019 | Inspection Level 1, AQL 1.5 | Inspection Level 1, AQL 1.5 | Same |
| Intended use | - | A patient examination glove is a<br>disposable device intended for<br>medical purposes that is worn<br>on the examiner's hand to<br>prevent contamination between<br>patient and examiner. | A patient examination glove is<br>a disposable device intended<br>for medical purposes that is<br>worn on the examiner's hand<br>or finger to prevent<br>contamination between patient<br>and examiner. | Same |
| Size | Medical Glove<br>Guidance Manual -<br>Labeling | X Small<br>Small<br>Medium<br>Large<br>Extra Large | X Small<br>Small<br>Medium<br>Large<br>Extra Large | Same |
| Single use | Medical Glove<br>Guidance Manual -<br>Labeling | Single use | Single use | Same |
{8}------------------------------------------------
| | | DEVICE PERFORMANCE | | |
|----------------------------------------|---------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| CHARACTERISTICS | STANDARD | PREDICATE | SUBJECT DEVICE | COMPARISON ANALYSIS |
| | S | BLUE | DERMAGRIP POWDER FREE BLACK<br>NITRILE PATIENT EXAMINATION<br>GLOVE, NON-STERILE, TESTED FOR<br>USE WITH CHEMOTHERAPY DRUGS<br>AND FENTANYL RESISTANCE | |
| Chemotherapy Drugs | | Minimum Breakthrough | Minimum Breakthrough | |
| Permeation Test | | Detection Time in Minutes | Detection Time in Minutes | |
| Carmustine (BCNU) (3.3mg/ml) | | 10.1 | 21.1 | Different |
| Cisplatin (1.0mg/ml) | | >240 | >240 | Same |
| Cyclophosphamide (Cytoxan) (20.0mg/ml) | | >240 | >240 | Same |
| Cytarabine (100.0mg/ml) | | >240 | Not Available | Different |
| Dacarbazine (10.0mg/ml) | | >240 | >240 | Same |
| Doxorubicin Hydrochloride (2.0mg/ml) | | >240 | >240 | Same |
| Etoposide (Toposar) (20.0mg/ml) | | >240 | >240 | Same |
| Fluorouracil (50.0mg/ml) | ASTM<br>D6978 | >240 | >240 | Same |
| Ifosfamide (50.0mg/ml) | | >240 | >240 | Same |
| Methotrexate (25.0mg/ml) | | >240 | >240 | Same |
| Mitoxantrone (2.0mg/ml) | | >240 | >240 | Same |
| Mitomycin C (0.5mg/ml) | | >240 | >240 | Same |
| Paclitaxel (Taxol) (6.0mg/ml) | | >240 | >240 | Same |
| *Thiotepa (10.0mg/ml) | | 30.2 | 34.6 | Different |
| Vincristine Sulfate (1.0mg/ml) | | >240 | >240 | Same |
| Warning Statement | | Please note that Carmustine (BCNU)<br>has extremely low permeation times<br>of 10.1 minutes | WARNING: Please note the following<br>drugs have extremely low permeation<br>times: Carmustine (BCNU): 21.1 minutes<br>and Thiotepa: 34.6. Do not use with<br>Carmustine and Thiotepa. | Different |
| Fentanyl Resistant (100mcg/2ml) | ASTM D6978 | No breakthrough up to 240 minutes | No breakthrough up to 240 minutes | Same |
{9}------------------------------------------------
There are no significant differences between the two products and are identical in terms of intended use, materials design, manufacturing methods except for color, physical properties and thickness.
### 7.0 Summary of Non-Clinical Testing
The performance test data of the non-clinical test for this powder free nitrile examination glove is summarized as per below.
{10}------------------------------------------------
| Test<br>Method | Standard | Purpose of Testing | Acceptance Criteria | | Results | | |
|------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|------------------------|-----------------|-----------------|------|------|
| Physical<br>Properties | ASTM D412<br>(Standard Test<br>Method for<br>Vulcanized Rubber<br>and Thermoplastic<br>Elastomers-Tension) | To evaluate the<br>tensile (tension)<br>properties of glove. | Tensile<br>strength | Min 14.0<br>MPa | Min 14.0<br>MPa | Pass | Pass |
| | | | Ultimate<br>elongation | Min<br>500% | Min<br>400% | Pass | Pass |
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | | Results |
|----------------|-------------------------------------|-------------------------------------|--------------------------|---------------------|---------|
| Dimension | ASTM D3767<br>Standard Practice for | To measure the<br>length, width and | Length<br>(all sizes) | Min 240 mm | Pass |
| | Rubber- | thickness<br>of | Width | XS: ≤ 80 mm | Pass |
| | Measurement of | glove | | S: 80 ± 10 | Pass |
| | Dimensions | | | M: 95 ± 10 | Pass |
| | | | | L: 110 ± 10 | Pass |
| | | | | XL: ≥ 110 | Pass |
| | | | Thickness<br>(all sizes) | Finger: Min 0.05 mm | Pass |
| | | | | Palm: Min 0.05 mm | Pass |
| | | | | Cuff: Min 0.05 mm | Pass |
{11}------------------------------------------------
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results |
|----------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|---------|
| Watertight | ASTM D5151<br>(Standard Test<br>Method for<br>Detection of Holes<br>in Medical Gloves) | To detect holes<br>that leak water<br>and thereby<br>compromise the<br>usefulness of the<br>glove. | Sample size: 500 pcs<br>Inspection level: G1<br>AQL: 1.5, Acceptance No. 10,<br>Found 2 | Pass |
| Test<br>Method | Standard | Purpose of<br>Testing | Acceptance Criteria | Results |
|--------------------|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------|---------|
| Residual<br>Powder | ASTM D6124<br>(Standard Test<br>Method for Residual<br>Powder on Medical<br>Gloves) | To determine the<br>amount of<br>residual powder<br>and non-powder<br>solids found on<br>gloves. | Less than 2 mg per glove | Pass |
{12}------------------------------------------------
### 8.0 Summary of Clinical Testing:
No clinical study is included in this submission.
### 9.0 Conclusion
The conclusion drawn from the non-clinical tests demonstrate that the subject device (Dermagrip Powder Free Black Nitrile Patient Examination Glove, Non-Sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Resistance) is as safe, as effective, and performs as well as or better than the legally marketed predicate devices K183287.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.