The XO Cross is a support catheter intended to guide and support a guidewire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
Device Story
XO Cross Support Catheter is a single-lumen, laser-cut metal alloy tube catheter; used in peripheral vasculature access; supports guidewire placement; facilitates wire exchanges; provides conduit for saline or diagnostic contrast agents. Operated by physicians in clinical settings under fluoroscopy. Radiopaque shaft and tip allow tracking; hydrophilic coating on distal end reduces surface friction. Device available in various lengths (90-175cm) and diameters (2Fr-4Fr). Output is physical support and fluid delivery; aids navigation through vasculature; benefits patient by enabling successful guidewire placement and diagnostic imaging.
Clinical Evidence
Bench testing only. Verification and validation included biocompatibility testing (ISO 10993-1) for cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemolysis, and complement activation. Design verification included radiopacity, tensile testing, lubricity, durability, particulate testing, and simulated use testing per ISO 10555-1.
Technological Characteristics
Single-lumen laser-cut metal alloy shaft; tapered polymer tip with radiopaque filler; hydrophilic coating on distal portion. Sizes: 2Fr to 4Fr; lengths: 90cm to 175cm. Compatible with 0.014", 0.018", and 0.035" guidewires. Sterilization: Ethylene Oxide (EO). Complies with ISO 10555-1 and ISO 10993-1.
Indications for Use
Indicated for patients requiring peripheral vascular access, specifically for guidewire support, wire exchange, and delivery of saline or diagnostic contrast agents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
October 21, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Transit Scientific, LLC Spencer Walker Director of Regulatory Affairs University of Utah 10 North 1900 East, EHSL Rm. 22B Salt Lake City, Utah 84112
### Re: K210322
Trade/Device Name: XO Cross Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 24, 2021 Received: September 22, 2021
#### Dear Spencer Walker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/ training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210322
Device Name XO Cross Support Catheter
Indications for Use (Describe)
The XO Cross is a support catheter intended to guidewire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Transit Scientific, LLC |
| Contact Person: | Spencer Walker, MSc – Director of Regulatory Affairs<br>University of Utah/ Center for Medical Innovation<br>10 North 1900 East, Rm. 22B<br>Salt Lake City, UT 84112<br>(801) 581-5080<br>Spencer.walker@hsc.utah.edu |
| Date Prepared: | March 19, 2021 |
| Trade Name: | XO Cross Support Catheter |
| Classification Name: | Percutaneous Catheter<br>21 CFR §870.1250, Product Code DQY |
| Device Class: | Class II |
| Predicate Device: | 510(k) No.: K193420<br>Model: XO Cross Support Catheter<br>Manufacture: Transit Scientific, LLC<br>Classification: DQY |
#### Device Description:
The XO Cross Support Catheter is identical to the predicate device in that it is a single-lumen support catheter designed to support a guidewire during access of the peripheral vasculature, allowing for exchange of guidewires during vascular access procedures, and providing a conduit for the delivery of saline solutions or diagnostic contrast agents. The shaft and tip are both radiopaque to track the location of the XO Cross Support Catheter within the vasculature, while under fluoroscopy. The distal end of the subject catheter is also coated with a hydrophilic coating to reduce surface friction. It is available in various lengths (90cm, 135cm, 15cm) and ranges in diameter from 2 Fr (0.7 mm) to 4 Fr (1.35 mm) with a tapered polymer tip at its distal end.
{4}------------------------------------------------
| Product<br>Family | Catalog No. | Description |
|---------------------------------|-------------|-----------------------------------------------------------------------------------------------|
| XO Cross<br>Support<br>Catheter | C14-090 | XO Cross 014 – 0.014-inch diameter catheter shaft in<br>90cm, 135cm, 150cm and 175cm lengths. |
| | C14-135 | |
| | C14-150 | |
| | C14-175 | |
| | C18-090 | XO Cross 018 – 0.018-inch diameter shaft, in 90cm,<br>135cm, 150cm and 175cm lengths. |
| | C18-135 | |
| | C18-150 | |
| | C18-175 | |
| | C35-090 | XO Cross 035 - 0.035-inch diameter shaft, in 90cm,<br>135cm, 150cm and 175cm lengths. |
| | C35-135 | |
| | C35-150 | |
| | C35-175 | |
### Indications for Use:
The XO Cross is a support catheter intended to guide and support a guidewire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
## Comparative Analysis:
Modifications in design and materials of the previously 510(k) cleared XO Cross Support Catheter (K193420) did not result in new product codes. The following modifications were made to the subject catheter:
- . Polymer Tip Extrusion radiopaque filler % increased radiopacity for improved visibility
- Addition of Hydrophilic Coating on distal portion of catheter to reduce surface friction
- Micro-cut refinements for better manufacturability ●
It has been demonstrated that the modified XO Cross catheter is comparable to the predicate device in intended use, fundamental scientific technology, design, principles of operation and functional performance evaluations and is substantial equivalent as summarized in Table 2. Furthermore, the modified XO Cross catheter has been fully assessed within the Transit Scientific Risk Management and Design Controls systems. The differences raise no additional or different questions of safety or effectiveness from that already identified for the predicate device.
{5}------------------------------------------------
| Table 2: Substantial Equivalence Comparison Chart | | | | |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--|--|
| | Predicate - K193420<br>(XO Cross Support Catheter) | Subject Device - K210322<br>XO Cross Support Catheter | | |
| Ind. for Use | The XO Cross is a support catheter intended to<br>guide and support a guidewire during access of<br>the peripheral vasculature, allow for wire<br>exchanges and provide a conduit for the<br>delivery of saline solutions or diagnostic<br>contrast agents. | Same | | |
| Classification<br>Name | Cardiovascular Device - Percutaneous<br>Catheter<br>21 CFR §870.1250<br>Product Code: DQY<br>Class II | Same | | |
| Single Use | Yes | Same | | |
| Fundamental<br>Scientific<br>Technology | Provide support for guidewire placement and<br>fluid administration | Same | | |
| Prescription<br>(Rx Only) | Yes | Same | | |
| Anatomical<br>Access | Peripheral Vasculature | Same | | |
| Design | Single Lumen metal alloy tube laser cut<br>shaft w/ tapered tip and polymer outer<br>layer. | Single Lumen metal alloy tube<br>laser cut shaft w/ tapered tip and<br>polymer outer layer, with<br>hydrophilic coating. | | |
| Effective Length<br>(cm) | 90, 135, 150, 175 | Same | | |
| Catheter Sizes | 2Fr - 4Fr | Same | | |
| Distal Tip/<br>Crossing Profile | 1.6Fr, 1.9Fr, 3.2Fr | Same | | |
| Shaft Outer<br>Diameter | 0.029", 0.037", 0.051" | Same | | |
| Visibility | Radiopaque catheter body, polymer tip with<br>radiopaque filler | Radiopaque catheter body,<br>polymer tip with increased<br>radiopaque filler | | |
| Guidewire<br>Compatibility | 0.014", 0.018", 0.035" | Same | | |
| Sheath /<br>Introducer<br>Compatibility | 2.9Fr and 4Fr | Same | | |
| Max Infusion<br>Pressure | Tested for manual infusion, not for use with<br>power injectors. | Same | | |
| Sterility | Sterile - EO | Same | | |
| Biocompatibility | ISO 10993 | Same | | |
{6}------------------------------------------------
# Functional/Safety Testing:
Verification & Validation activities were performed on the subject XO Cross Support Catheter to demonstrate substantial equivalence to the predicate device:
- Biocompatibility - Biocompatibility testing was done due to the added hydrophilic coating to the catheter shaft. All tests were done per the same protocol and acceptance criteria as the predicate device, and as prescribed in ISO 10993-1 and in accordance with the 2020 FDA guidance document "Use of International Standard ISO 10993-1" for an external communicating device with limited exposure (≤ 24hrs) to circulating blood. The following tests were performed and passed:
- o Cytotoxicity
- Sensitization O
- Irritation/ Intracutaneous Toxicity O
- O Acute Systemic Toxicity
- Material Mediated Pyrogenicity O
- Hemolysis Assay O
- O Complement Activation
- LAL Pyrogenicity O
o
- Design Verification -The XO Cross models all meet or exceeded both Transit Scientifics in-. house requirements, and requirements listed in ISO 10555-1.
- Polymer Tip Extrusion radiopaque filler increase
- l Radiopacity Acceptability
- = Tip ID Testing
- Tip Tensile Testing l
- Hydrophilic Coating o
- Biocompatibility Testing .
- Catheter OD Testing l
- Coating Lubricity Testing
- l Coating Durability Testing
- Distal Catheter Tensile
- Coating Particulate Testing
- Simulated Use ■
- l Coating Integrity
- I Coaxial Testing
- Packaging The proposed changes to the XO Cross Support Catheter did not affect the ● packaging or its configuration.
#### Conclusion:
The modifications to the subject XO Cross Support catheter is substantially equivalent with respect to safety and effectiveness to the legally marketed predicate device for its intended use. The minor differences between the subject XO Cross Support catheter and the predicated device has no effect on safety or effectiveness, as established through various performance test.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.