← Product Code [DRF](/productcode/DRF) · K210313

# EasyFinder Fixed Curve Diagnostic Catheter, Forlnk Cable for Mapping Catheter (K210313)

_Shanghai Microport EP Medtech Co., Ltd. · DRF · Dec 30, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210313

## Device Facts

- **Applicant:** Shanghai Microport EP Medtech Co., Ltd.
- **Product Code:** [DRF](/productcode/DRF.md)
- **Decision Date:** Dec 30, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

EasyFinder™ Fixed Curve Diagnostic Catheter is used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies.

## Device Story

EasyFinder™ Fixed Curve Diagnostic Catheter is a sterile, single-use device featuring a high-torque shaft and fixed distal curve with an array of platinum electrodes. It is used by physicians in professional healthcare environments to perform electrophysiological mapping of cardiac structures. The catheter captures cardiac electrical signals (electrograms) and delivers stimulation pulses; these are transmitted via the FORLNK™ Cable for Mapping Catheter to external, separately cleared diagnostic electrophysiology equipment. The high-torque shaft allows manual rotation of the curved tip for precise positioning within the heart. The output, displayed on the external equipment, assists clinicians in diagnosing cardiac arrhythmias and guiding clinical decision-making during electrophysiology studies.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1), mechanical requirements (ISO 10555-1), electrical safety and electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-27), radiopacity validation, and sterilization validation (ISO 11135, AAMI TIR 28).

## Technological Characteristics

Sterile, single-use diagnostic catheter with platinum electrodes and high-torque shaft. Fixed curve design available in 28 models/6 curve types. Connectivity via FORLNK™ accessory cable to external EP equipment. Complies with ISO 10555-1, ISO 10993-1, IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-27. Sterilized via Ethylene Oxide (EO).

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- IBI-1100(TM) ELECTROPHYSIOLOGY CATHETER ([K946333](/device/K946333.md))

## Reference Devices

- Cordis Webster Fixed Curve Catheter ([K992965](/device/K992965.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.

December 30, 2022

Shanghai MicroPort EP MedTech Co., Ltd. Min Zhu RA Engineer Building 23&28, Lane 588, Tianxiong Rd. Shanghai, Shanghai 201318 People's Republic of China

Re: K210313

Trade/Device Name: EasyFinder™ Fixed Curve Diagnostic Catheter; FORLNK™ Cable for Mapping Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode recording catheter or electrode recording probe Regulatory Class: Class II Product Code: DRF Dated: January 25, 2021 Received: February 3, 2021

Dear Xia Tian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210313

#### Device Name

EasyFinder™ Fixed Curve Diagnostic Catheter FORLNK™ Cable for Mapping Catheter

Indications for Use (Describe)

EasyFinder™ Fixed Curve Diagnostic Catheter is used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the EverPace logo. The logo consists of the letters 'e' and 'o' connected together, followed by the word 'EverPace' in blue. The 'e' is blue, and the 'o' is red. There is a trademark symbol in the upper right corner of the logo.

## 510(k) Summarv

#### Submitter Information

- Shanqhai MicroPort EP MedTech Co., Ltd. A. Company Name: B. Company Address: Building 23&28, Lane 588, Tianxiong Rd., 201318 Shanghai, PEOPLE'S REPUBLIC OF CHINA C. Company Phone: +86 21 38954600*53701 D. Company Facsimile: +86 21 20903925 E. Email: zhumin@everpace.com F. Contact Person: Zhu, Min December 30, 2022 G. Summary Prepared on:
## Identification of Proposed Device

| A. Device Trade Name:   | EasyFinder™ Fixed Curve Diagnostic Catheter<br>FORLNK™ Cable for Mapping Catheter |
|-------------------------|-----------------------------------------------------------------------------------|
| B. Device Common Name:  | Electrode recording catheter                                                      |
| C. Classification Name: | Catheter, Electrode Recording, Or Probe,<br>Electrode Recording                   |
| D. Regulation Number    | 21 CFR 870.1220                                                                   |
| E. Product Code:        | DRF                                                                               |
| F. Device Class:        | Class II                                                                          |
| G. Review Panel         | Cardiovascular                                                                    |

#### Device Description

The EasyFinder™ Fixed Curve Diagnostic Catheter is a sterile, single-use diagnostic catheter with multiple electrodes and a fixed distal curve. It has been designed to facilitate electrophysiological mapping of cardiac structures in 28 models and 6 kinds of curves.

The catheter has a high-torque shaft with a fixed curve section containing an array of platinum electrodes. The high-torque shaft allows the plane of the curved tip to be rotated to facilitate accurate positioning of the catheter tip at the desired site.

The catheter is connected to separately cleared diagnostic electrophysiology equipment via an accessory, the FORLNK™ Cable for Mapping Catheter.

# Indications for Use Statement

EasyFinder™ Fixed Curve Diagnostic Catheter is used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies.

#### Identification of Predicate Device

| A. Product Name: | IBI-1100(TM) ELECTROPHYSIOLOGY<br>CATHETER |
|------------------|--------------------------------------------|
|------------------|--------------------------------------------|

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the EverPace logo. The logo consists of a stylized "E" and "O" in blue and red, respectively, followed by the word "EverPace" in blue. A small "TM" symbol is located to the upper right of the word "EverPace".

- B. Manufacturer: IRVINE BIOMEDICAL, INC.
- C. 510(k) Number: K946333
- 21 CFR 870.1220 D. Regulation Number
- E. Product Code: DRF
- F. Device Class: Class II

# Identification of Reference Device

- A. Product Name: Cordis Webster Fixed Curve Catheter
- B. Manufacturer: Cordis Webster, Inc.
- K992965 C. 510(k) Number
- D. Requlation Number 21 CFR 870.1220
- E. Product Code: DRF
- F. Device Class: Class II

# Non-Clinical Performance Testing

Non-clinical performance testing was completed for the EasyFinder™ Fixed Curve Diagnostic Catheter and accessory FORLNK™ Cable for Mapping Catheter to support their substantial equivalence to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- (1) Biocompatibility Verification: The biological safety of the catheter was verified as per the requirements of ISO 10993-1:2018 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process and FDA's modified ISO quidelines in accordance with the FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".
- (2) Mechanical Verification: Mechanical testing was performed to verify compliance of the catheter with ISO 10555-1:2013 Intravascular catheters - Sterile and single-use catheters - Part 1: General requirements.
- (3) Electrical Verification: Electrical testing was performed to verify compliance of the catheter and accessory cable with applicable IEC 60601-1: 2005 Medical electrical equipment – Part1: General requirements for basic safety and essential performance, IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances – Requirements and tests and IEC 60601-2-27:2011 Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for EverPace. The logo consists of a stylized "E" and "O" connected together, with the "E" in blue and the "O" in red. To the right of the symbol is the word "EverPace" in blue font. A small trademark symbol is located in the upper right corner of the word "EverPace".

- (4) Bench Validation: Validation testing of the catheter was performed to validate the design of the device with regards to radiopacity. Validation testing of the accessory cable was performed to evaluate the design and function of the cable.
- (5) Cleaning and Sterilization Validation: Validation testing was performed to demonstrate that the catheter and accessory cable could be cleaned and sterilized in accordance with ISO 11135:2014 Sterilization of health care products-Ethylene Oxide: Requirements for development, validation and routine control of a sterilization process for medical device, AAMI TIR 28:2009 Product Adoption and Process Equivalency for Ethylene Oxide sterilization and the FDA guidance document.
- (6) Shelf Life Validation: Validation testing was performed to demonstrate the shelf life of the catheter and accessory cable are three years.

## Clinical Tests Conclusion

No clinical study is included in this submission.

| Description               | Proposed Device                                                                                                                                                   | Predicate Device<br>(K946333)                                                                                                                                    | Remark  |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Product Code              | DRF                                                                                                                                                               | DRF                                                                                                                                                              | SE      |
| Regulation No.            | 21 CFR 870.1220                                                                                                                                                   | 21 CFR 870.1220                                                                                                                                                  | SE      |
| Class                     | Class II                                                                                                                                                          | Class II                                                                                                                                                         | SE      |
| Indications for<br>Use    | EasyFinder™ Fixed Curve<br>Diagnostic Catheter is used<br>for electrogram recording and<br>cardiac stimulation during<br>diagnostic electrophysiology<br>studies. | IBI-1100(TM)<br>ELECTROPHYSIOLOGY<br>CATHETER is used for<br>electrogram recording and<br>cardiac stimulation during<br>diagnostic electrophysiology<br>studies. | SE      |
| Configuration             | Cable connector                                                                                                                                                   | Cable connector                                                                                                                                                  | SE      |
|                           | Tip electrode                                                                                                                                                     | Tip electrode                                                                                                                                                    | SE      |
|                           | Ring electrode                                                                                                                                                    | Ring electrode                                                                                                                                                   | SE      |
|                           | High-torque shaft                                                                                                                                                 | High-torque shaft                                                                                                                                                | SE      |
| Dimensional<br>Comparison | Catheter effective length                                                                                                                                         | Catheter effective length                                                                                                                                        | Similar |
|                           | Length of color code                                                                                                                                              | Length of color code                                                                                                                                             | SE      |
|                           | Electrode length                                                                                                                                                  | Electrode length                                                                                                                                                 | Similar |
|                           | Electrode diameter                                                                                                                                                | Electrode diameter                                                                                                                                               | SE      |
|                           | Electrode spacing                                                                                                                                                 | Electrode spacing                                                                                                                                                | SE      |

## Comparison to Predicate Device

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for EverPace. The logo consists of a stylized "E" and "O" connected together, with the "E" in blue and the "O" in red. To the right of the symbol is the word "EverPace" in blue, with a trademark symbol in the upper right corner.

|                                       | Outside diameter of catheter shaft     | Outside diameter of catheter shaft     | SE      |
|---------------------------------------|----------------------------------------|----------------------------------------|---------|
|                                       | Curve shape                            | Curve shape                            | Similar |
| Functional<br>Performance             | Comply with ISO 10555-1                | Comply with ISO 10555-1                | SE      |
| Biological<br>Characteristics         | Comply with ISO 10993-1                | Comply with ISO 10993-1                | Similar |
| Intended<br>Healthcare<br>Environment | Professional healthcare<br>environment | Professional healthcare<br>environment | SE      |
| Sterilization<br>Method               | EO sterilized                          | EO sterilized                          | SE      |
| Single Use                            | Yes                                    | Yes                                    | SE      |
| Shelf Life                            | 3 years                                | 3 years                                | SE      |

# Substantially Equivalent (SE) Conclusion

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

The proposed and predicate devices share the same indications for use and fundamental scientific technology, including principles of operation and mechanism of action. Design and technological differences between the proposed and predicate devices do not raise any new concerns of safety and effectiveness. The results of verification and validation testing demonstrate that the EasyFinder™ Fixed Curve Diagnostic Catheter and the accessory cable are as safe, as effective, and perform in a manner that is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K210313](https://fda-staging.innolitics.com/device/K210313)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
