Home Use IPL Beauty Device, Model(s): MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6, MELSYA-M7.
K210311 · Shenzhen Mismon Technology Co., Ltd. · OHT · Jul 23, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210311
Device Name
Home Use IPL Beauty Device, Model(s): MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6, MELSYA-M7.
Applicant
Shenzhen Mismon Technology Co., Ltd.
Product Code
OHT · General, Plastic Surgery
Decision Date
Jul 23, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Home Use IPL Beauty Device is an over-the-counter device intended for removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs.
Device Story
The Home Use IPL Beauty Device is a personal, light-based hair reduction device for over-the-counter use. It utilizes a Xenon flashlamp to deliver intense pulsed light (IPL) to the skin. The device consists of a main body with a non-removable treatment window and an external power adapter. The user operates the device by manually triggering pulses via a finger switch. The light energy targets hair follicles to achieve hair reduction. It is intended for home use by the consumer. The device is available in multiple models (MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6, MELSYA-M7) differing in appearance and treatment window size (3.0 cm² or 3.6 cm²).
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2), photobiological safety (IEC 62471), and performance testing per IEC 60601-2-57. Software verification and validation were performed for a moderate level of concern.
Technological Characteristics
Xenon arc flashlamp source; energy density 2.0-5.1 J/cm²; pulse duration 9-12 ms; spot sizes 3.0 cm² or 3.6 cm²; finger-switch activation; external power adapter. Standards: ISO 10993-1/5/10, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471.
Indications for Use
Indicated for over-the-counter removal of unwanted hair in small areas (underarms, facial hair below chin line) and large areas (legs).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Home Use IPL Beauty Device/MELSYA-MA, MS-206B (K180016)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 23, 2021
Shenzhen Mismon Technology Co., Ltd. Rain Yip Registered Engineer 4th Floor, Building B, Zone A, Longquan Technology Park, Tongfuyu Phase2, Tongsheng Community Shenzhen, Guangdong 518000 China
Re: K210311
Trade/Device Name: Home Use IPL Beauty Device, Model(s): MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6, MELSYA-M7. Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: April 3, 2021 Received: June 21, 2021
Dear Rain Yip:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210311
#### Device Name
Home Use IPL Beauty Device Model(s): MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6, MELSY A-M7
#### Indications for Use (Describe)
The Home Use IPL Beauty Device is an over-the-counter device intended for removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### "510(k) Summary" as required by 21 CFR Part 807.92.
### Date: 2021-01-25
#### I. Submitter
Shenzhen Mismon Technology Co., Ltd. 4th Floor, Building B, Zone A, Longquan Technology Park, Tongfuyu Phase 2, Tongsheng Community, Dalang Street, Longhua District, Shenzhen, China 518000 Tel.: +86 755 23732187
Vicky Yang General Manager Tel: +86 158 1871 5190 Email: vicky@mismon.com
#### II. Device
Name of Device: Home Use IPL Beauty Device Model(s): MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6, MELSYA-M7 Common or Usual Name: Light Based Over-The-Counter Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810
#### III. Predicate Device
The predicate devices are listed as below:
- 1) Primary: K180016, Home Use IPL Beauty Device/MELSYA-MA, MS-206B, manufactured by Shenzhen Mismon Technology Co., Ltd., and cleared March 20, 2018
- 2) K192432, IPL Home Use Hair Removal Device/D-1128, manufactured by Shenzhen Bosidin Technology Co., Ltd., and cleared November 11, 2019
#### IV. Device Description
The Home Use IPL Beauty Device is a personal, light-based, hair reduction device intended to be sold over-the-counter directly to the end user. The device provides hair reduction using IPL technology. Of which, the Device includes MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6 and MELSYA-M7 six models. All have adopted the same structure design, consisting of IPL beauty device main body and power adapter two parts, and one non-removable lamp head (light-emitting treatment window) located in the main body which is the source of
{4}------------------------------------------------
optical radiation, namely a Xenon flashlamp. Meanwhile, the device is power from power adapter via an external power.
The difference of all models is mainly the exception of appearance and spot size (the size of treatment window) which do not affect the intended use. The spot size of model MS-208B, MS212B and MELSYA-M6 is designed to 3.6cm², model MS-216B, MS-218B and MELSYA-M7 is designed to 3.0 cm2, as well as the energy density of the all models is in the range of 2.0~5.1J/cm².
# V. Indications for Use
The Home Use IPL Beauty Device is an over-the-counter device intended for removal of unwanted hair such as but not limited to small areas such as underarm and facial hair below the chin line and large areas such as legs.
# VI. Comparison of Technological Characteristics With the Predicate Device
The Home Use IPL Beauty Device has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate devices do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the Home Use IPL Beauty Device may be found substantially equivalent to its predicate device.
Home Use IPL Beauty Device is compared with the following Predicate Devices in terms of intended use, design, material, specifications, and performance.
- 1) (Primary) K180016, "Home Use IPL Beauty Device/MELSYA-MA, MS-206B", manufactured by "Shenzhen Mismon Technology Co., Ltd." in Guangdong, China
- 2) K192432, "IPL Home Use Hair Removal Device/D-1128", manufactured by "Shenzhen Bosidin Technology Co.,Ltd" in Guangdong, China
| Comparison<br>Elements | Subject Device | Predicate Device 1<br>(Primary) | Predicate Device 2 |
|------------------------|-----------------------------------------------|-----------------------------------------------------|--------------------------------------------------|
| K Number | Pending | K180016 | K192432 |
| Common name | Light Based Over-The-<br>Counter Hair Removal | Light Based Over-The-<br>Counter Hair Removal | Light Based Over-<br>The-Counter Hair<br>Removal |
| Trade name | Home Use IPL Beauty<br>Device | Home Use IPL Beauty<br>Device/MELSYA-MA,<br>MS-206B | IPL Home Use Hair<br>Removal Device/D-<br>1128 |
| Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp |
| Pulse duration | 9-12 milliseconds | 9-11 milliseconds | 7.5-14 milliseconds |
{5}------------------------------------------------
| Comparison<br>Elements | Subject Device | Predicate Device 1<br>(Primary) | Predicate Device 2 |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Energy density | 2.6~5.1J/cm2 | 5.5J/cm2(Max.) | 2.0~4.0J/cm2 |
| Spot size (size of<br>treatment<br>window) | MS-<br>208B/MS212B/MELSYA-<br>M6: 3.6cm2<br>MS-216B/MS-<br>218B/MELSYA-M7:<br>3.0cm2 | MELSYA-M5: 3.9cm2<br>MS-206B: 3.0cm2 | Regulator window:<br>4.5cm2 ? 2.0cm2<br>Filter window: 2.5cm2 |
| Delivery device | Direct illumination to<br>tissue | Direct illumination to<br>tissue | Direct illumination to<br>tissue |
| Pulsing control | Finger switch | Finger switch | Finger switch |
| Indication for<br>use/Intended use | The Home Use IPL<br>Beauty Device is an over-<br>the-counter device<br>intended for removal of<br>unwanted hair such as but<br>not limited to small areas<br>such as underarm and<br>facial hair below the chin<br>line and large areas such<br>as legs. | Home Use IPL Beauty<br>Device is an over-the-<br>counter device<br>intended for removal of<br>unwanted hair such as<br>but not limited to small<br>areas such as underarm<br>and facial hair below<br>the chin line and large<br>areas such as legs. | IPL Home Use Hair<br>Removal Device is an<br>over-the-counter<br>device intended for<br>removal of unwanted<br>body and/or facial<br>hair. |
| Location for use | OTC | OTC | OTC |
## VII.Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the subject device was conducted in accordance with the "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process", as recognized by FDA. The battery of testing was performed to, and passed, including:
- > ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices –Par t 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10:2010, Biological Evaluation of Medical Devices –Par t 10: Tests for Irritation and Skin Sensitization
{6}------------------------------------------------
#### 2) Electrical Safety and Eye Safety
Electrical safety and Eye safety testing was performed to, and passed, the following standards:
- > IEC 60601-1-2:2014 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility
- > IEC 60601-1:2005, MOD Medical electrical equipment -Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-11:2015 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- > IEC 60601-2-57:2011 Medical electrical equipment -Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
#### 3) Eye Safety
- > IEC 62471:2006 Photobiological safety of lamps and lamp systems
#### 4) Software Verification and Validation
Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
#### Summary
Based on the above performance as documented in this application, the subject device Home Use IPL Beauty Device was found to have a safety and effectiveness profile that is similar to the predicate device.
#### VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device Home Use IPL Beauty Device is to be concluded substantial equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.