VueSonic One Contact Lens Cleaning System, VueSonic Advance 2 Contact Lens Cleaning System
Applicant
Vuesonic, LLC
Product Code
LYL · Ophthalmic
Decision Date
Apr 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
The VueSonic One Contact Lens Cleaning System is a cleaning system for Soft and Hybrid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology. The VueSonic Advance 2 Contact Lens Cleaning System is a cleaning system for Rigid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.
Device Story
Device cleans contact lenses using 3-D sonic vibration technology to mimic digital hand rubbing. Input: contact lenses placed in a holder with cleaning solution. Operation: 3-D multi-frequency modulated stereo power source and intelligent control circuit drive uniform three-dimensional relative movement between soft medical silicone damping heads and cleaning cushions. Output: cleaned contact lenses. Used in home environment by patients. Healthcare providers do not operate the device; output is visual cleanliness of lenses. Benefits: automated physical cleaning of lenses to remove debris/deposits.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-5, -10, -11), electrical safety/EMC (IEC 60601-1, -1-2, -1-11), usability (IEC 62366-1), protein/lipid removal, total extractive residues (21 CFR 177.2600), and lens compatibility (ISO 18369-2).
Technological Characteristics
Materials: medical grade silicone (lens holder, cushions, damping heads). Power: 1.5-3V (AAA batteries). Mechanism: 3-D multi-frequency modulated stereo power source with intelligent control circuit. Dimensions: limited to lenses ≤ 15mm diameter. Cleaning time: 2 minutes. Power density: 0.37-0.72 watt/ml.
Indications for Use
Indicated for cleaning soft, hybrid, and rigid contact lenses (diameter ≤ 15mm) by mimicking digital rubbing. Compatible with multi-purpose cleaning solutions; contraindicated for use with 3% Hydrogen Peroxide solutions (e.g., Clear Care).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 1, 2021
VueSonic, LLC % Marc Sanchez Regulatory Attorney Contract In-House Counsel and Consultants, LLC 1717 Pennsylvania Ave. NW, Suite #1025 Washington, D.C. 20006
Re: K210300
Trade/Device Names: VueSonic One Contact Lens Cleaning System, VueSonic Advance 2 Contact Lens Cleaning System Regulation Number: 21 CFR 886.5928 Regulation Name: Soft (Hydrophilic) Contact Lens Care Products Regulatory Class: Class II Product Code: LYL, MRC Dated: February 2, 2021 Received: February 3, 2021
Dear Marc Sanchez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
ence determination does not mean that FDA
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Angelo Green, PhD Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known)
K210300
#### Device Name
VueSonic One Contact Lens Cleaning System VueSonic Advance 2 Contact Lens Cleaning System
Indications for Use (Describe)
The VueSonic One Contact Lens Cleaning System is a cleaning system for Soft and Hybrid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.
The VueSonic Advance 2 Contact Lens Cleaning System is a cleaning system for Rigid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------|----------------------------------------------|
| <div></div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following information is provided as required by 21 CFR
807.92 for the VueSonic One Contact Lens Cleaning System and VueSonic Advance 2 Contact Lens Cleaning System 510(k) premarket notification.
| Sponsor: | VueSonic LLC<br>1234 Washington Ave, Suite 205<br>Miami Beach, FL 33139<br>Establishment Registration: to be determined<br>Ph: 855-770-0883 |
|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | VueSonic LLC<br>1234 Washington Ave, Suite 205<br>Miami Beach, FL 33139<br>Establishment Registration: to be determined<br>Ph: 855-770-0883 |
| Contact: | Marc C. Sanchez, Esq.<br>Contract In-House Counsel and Consultants, LLC<br>1717 Pennsylvania Ave. NW, Suite #1025<br>Washington, D.C. 20006<br>Ph: 202-765-4491<br>Email: msanchez@fdaatty.com |
Date of 510(k) Summary Preparation: March 18, 2021
Proprietary Names: VueSonic One Contact Lens Cleaning System, VueSonic Advance 2 Contact Lens Cleaning System
Common Name: Accessories, Solution, Ultrasonic Cleaners For Lenses
Regulation Number: 21 CFR 886.5928/21 CFR 886.5918
Regulatory Class: Class II
Product Code: LYL/MRC
Predicate Device(s): New Comfort Contact Lens Accessory (K974724) and AUTOLENS® System (K021699)
Device Description: The VueSonic One Contact Lens Cleaning System and VueSonic Advance 2 Contact Lens Cleaning Systems are cleaning systems for contact lenses. Both devices consist of a device body, a contact lens holder, which houses the cleaning cushions and the damping heads. The contact lens holder, cleaning cushions and damping heads are made of soft medical silicone, which will not damage the lens. The systems are designed to mimic human hand rubbing. This is done by the 3D
{4}------------------------------------------------
multi-frequency.
The proposed device consists of two versions:
- . For Rigid Lenses, VueSonic Advance 2 Contact Lens Cleaning System; and
- . For Soft and Hybrid Lenses, VueSonic One Contact Lens Cleaning System.
### VueSonic One:
The VueSonic One Contact Lens Cleaning System is a cleaning system for Soft and Hybrid contact lenses. It consists of a device body, a contact lens holder, which houses the cleaning cushions and the damping heads. The contact lens holder, cleaning cushions and damping heads are made of soft medical silicone, which will not damage the lens.
The system is de- signed to mimic human hand rubbing. This is done by the 3D multi-frequency modulated stereo power source and corresponding intelligent control circuit, which make a uniform three-dimensional relative movement between the damping heads and cleaning cushion.
The proposed device is compatible with Soft and Hybrid contact lens from all the brands with a diameter equal or less than 15mm. This limitation is due to the size of the device and in particular the damping head.
The proposed device is compatible with all "multi-purpose" cleaning solution such as Bausch & Lomb Multi-action Solution and Opti-Free. The only incompatible solution is Clear Care and Clear Care Plus because Clear Care requires its own standing case and contains 3% Hydrogen Peroxide, which may cause serious damage to eyes when used improperly.
### VueSonic Advance 2
The VueSonic Advance 2 Contact Lens Cleaning System is a cleaning system for Rigid contact lenses. It consists of a device body, a contact lens holder, which houses the cleaning cushions and the damping heads. The contact lens holder, cleaning cushions and damping heads are made of soft medical silicone, which will not damage the lens. The system is designed to mimic human hand rubbing. This is done by the 3D multi-frequency modulated stereo power source and corresponding intelligent control circuit, which make a uniform three-dimensional relative movement between the damping heads and cleaning cushion.
The proposed device is compatible with Rigid contact lens from all the brands with a diameter equal or less than 15mm. This limitation is due to the size of the device and in particular the damping head.
The proposed device is compatible with all "multi-purpose" cleaning solution such as Bausch & Lomb Multi-action Solution and Opti-Free. The only incompatible solution is Clear Care and Clear Care Plus because Clear Care requires its own standing case and contains 3% Hydrogen Peroxide, which may cause serious damage to eyes when used improperly.
{5}------------------------------------------------
## Indications for Use:
The VueSonic One Contact Lens Cleaning System is a cleaning system for Soft and Hybrid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.
The VueSonic Advance 2 Contact Lens Cleaning System is a cleaning system for Rigid contact lenses by mimicking digital rubbing with 3-D sonic Vibration Technology.
# Summary of Non-Clinical Test Reports
The following tests were performed on the subject devices and the test results show that the devices are substantially equivalent to the predicate devices on the market.
| Table 1<br>Summary of Non- Clinical Test Reports | | |
|--------------------------------------------------------|-------------------------------------------------------------------------------|---------------------|
| Biocompatibility | Ocular Irritation | ISO 10993-10, 12 |
| | In Vitro Cytotoxicity Test | ISO 10993-5 |
| | Sensitization | ISO 10993-10 |
| | Systemic Toxicity | ISO 10993-11, 12 |
| Electrical Safety and Electromagnetic<br>Compatibility | General<br>Requirements<br>electromagnetic<br>compatibility | IEC 60601-1-2 |
| | General Requirements electrical<br>safety | IEC 60601-1 |
| | General Requirements<br>home healthcare<br>environment | IEC 60601-1-11 |
| Usability | Human Factor Engineering and<br>Usability<br>Engineering<br>Evaluation Report | IEC62366-1:2015 |
| Protein and Lipid Removal Test | Protein and Lipid Removal Test | Citation |
| Total Extractive Residues | Total Extractive Residues | FDA 21 CFR 177.2600 |
| Software | Software Validation | FDA Guidance |
| Lens Compatibility Test | Lens Compatibility Test | ISO 18369-2:2017 |
{6}------------------------------------------------
Traditional 510(k) Submission
# Summary of Substantial Equivalence
The VueSonic One Contact Lens Cleaning System, VueSonic Advance 2 Contact Lens Cleaning System, and the predicates use a multipurpose contact lens solution and a physical mode of action to clean contact lenses. The proposed device uses a sophisticated multi-frequency modulated stereo power system, while K974724 uses ultrasonic waves and K021699 uses a tumbling motion. The different motions are all aimed at mimicking hand rubbing to physically clean contact lenses. There may be minor differences in labeling but this does not raise any new questions of safety or effectiveness.
| Device<br>Name/Model | Vue Sonic One<br>Contact Lens<br>Cleaning System,<br>Vue Sonic<br>Advance 2<br>Contact Lens<br>Cleaning System | New Comfort Contact<br>Lens Accessory | AUTOLENS® System |
|----------------------|----------------------------------------------------------------------------------------------------------------|---------------------------------------|------------------|
| 510(k) Number | K210300 | K974724 | K021699 |
### Table 2 Technological Characteristics
{7}------------------------------------------------
| Indications for Use | | | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The VueSonic<br>One Contact Lens<br>Cleaning System is a<br>cleaning system for Soft<br>and Hybrid contact<br>lenses by mimicking<br>digital rubbing with 3-D<br>sonic Vibration<br>Technology. The<br>VueSonic<br>Advance 2<br>Contact Lens Cleaning<br>System is a cleaning<br>system for Rigid contact<br>lenses by mimicking<br>digital rubbing with 3-D<br>sonic Vibration<br>Technology. | The New Comfort Contact<br>Lens Care Accessory is<br>intended for use in<br>conjunction with contact lens<br>solutions as an accessory in<br>the cleaning of contact lenses.<br>The New Comfort Contact<br>Lens Care Accessory is<br>indicated as aid for cleaning<br>as an accessory for soft<br>hydrophilic lenses and gas<br>permeable lenses when used<br>with the appropriate<br>Allergan® Soft Mate®<br>solutions such as the | The AUTOLENS® Automatic<br>Contact Lens Cleaning<br>Accessory is indicated for the<br>cleaning of soft (hydrophilic)<br>rigid gas permeable (RGP)<br>and hard<br>(PMMA) contact lenses<br>when used with<br>AUTOLENS® Multipurpose<br>Solution. This automatic<br>contact lens cleaning<br>accessory cleans contact<br>lenses without digital |
| | | | |
| | Consept® Cleaning and<br>Disinfection System which is<br>comprised of the<br>Allergan® Soft Mate®<br>Consept®-1 Cleaning and<br>Disinfecting solution and<br>Allergan® Soft Mate®<br>Consept®-2 Neutralizing and<br>Rinsing Solution or Spray.<br>The New Comfort Contact<br>Lens Care Accessory is<br>indicated for use for gas<br>permeable solutions such as<br>Allergan® Gas Permeable<br>Daily Cleaner and<br>Allergan® ComfortCare GP<br>Wetting and Soaking<br>Solution. The New<br>Comfort Contact<br>Lens Care Accessory may be<br>used for a receptacle for<br>chemical disinfection. | rubbing. The AUTOLENS®<br>Automatic Contact Lens<br>Cleaning Accessory may be<br>used as a receptacle for<br>chemical disinfection with<br>AUTOLENS® Multipurpose<br>Solution.<br>AUTOLENS®<br>Multipurpose Solution is<br>indicated for use in cleaning,<br>rinsing, chemical (not heat)<br>disinfecting and protein<br>removal, storing soft<br>(hydrophilic), rigid gas<br>permeable fluoro-silicone<br>acrylate and silicone acrylate)<br>and PMMA<br>contact lenses as<br>recommended<br>by your eye care practitioner. | |
| Mode of<br>Action | 3D multi-frequency<br>modulated stereo<br>power source and<br>corresponding<br>intelligent control<br>circuit, which make a<br>uniform three<br>dimensional relative<br>movement between<br>the damping heads and | Bubbles created from piezo<br>electric crystal (ultrasonic) | Tumble motion |
| | | | |
| Power Source | 1.5~3V<br>battery (one or two<br>"AAA"<br>batteries) | 110/120v (standard power<br>using four- foot UL cord) | 1.5v (three<br>"AA"<br>batteries) |
| Materials | Medical grade<br>silicone | ABS | Polypropylene resin plastic |
| Cleaning Time | 2 minutes | Unknown | Unknown |
| Watts of<br>Power per<br>Milliliter of<br>Fluid | 0.37 to 0.72 watt/ ml | Unknown | Unknown |
{8}------------------------------------------------
{9}------------------------------------------------
# Conclusion:
Therefore, taking into consideration Table 1 for Substantial Equivalence, an analysis of safety, indications, intended uses, performance, and technological properties, the VueSonic Contact Lens Cleaning System raises no new issues of safety or effectiveness and has been found to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.