K210279 · Enbio Group AG · FLE · Oct 29, 2021 · General Hospital
Device Facts
Record ID
K210279
Device Name
Enbio S
Applicant
Enbio Group AG
Product Code
FLE · General Hospital
Decision Date
Oct 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.
Device Story
Enbio S is a table-top, pre-vacuum steam sterilizer for medical/dental instruments. User places load in aluminum chamber; selects program via TFT touch screen. Aluminum steam generator produces superheated steam; steam penetrates instruments at 121°C or 134°C for specified exposure times. Device pumps out steam; performs drying cycle; displays process completion to user. Used by healthcare providers in clinical settings. Output confirms sterilization status, ensuring instrument safety for patient use.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ANSI AAMI ST55:2016 standards, including vacuum tests, Bowie-Dick tests, empty/full/half-cycle thermal/pressure studies, and biological indicator testing (10^-6 SAL). Electrical safety and EMC verified per IEC 61010-1, 61010-2, IEC 60601-1-2, and IEC 61326-1.
Technological Characteristics
Table-top pre-vacuum steam sterilizer. Aluminum chamber with integrated heaters. 2.7L chamber volume. External water tank. TFT touch screen interface. USB port for data. Power: 110-120V, 60Hz, 15A. Complies with ANSI AAMI ST55:2016, IEC 61010-1/2, IEC 60601-1-2, IEC 61326-1.
Indications for Use
Indicated for sterilization of dental and medical instruments validated for steam sterilization. Suitable for solid, small porous, simple recessed, narrow-clearance items, dental handpieces, textiles, and plastics (121°C cycle). Not for liquid loads, biomedical waste, or incompatible materials.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 29, 2021
Enbio Group AG Lukasz Rogowski Corporate Quality Manager Eichengasse 3 Oensingen, CH-4702 Switzerland
Re: K210279
Trade/Device Name: Enbio S Regulation Number: 21 CFR 880.6880 Regulation Name: Steam sterilizer Regulatory Class: Class II Product Code: FLE Dated: September 24, 2021 Received: September 27, 2021
Dear Lukasz Rogowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, Ph.D. Assistant Director THT4B1:Sterility Devices Team DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
## K210279
Device Name Enbio S
### Indications for Use (Describe)
The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, biomedical waste or materials not compatible with sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.
Please refer to the table below for program name, load description, sterilization temperature, exposure time, drying time and maximum load.
| Program<br>Name | Load Description | Sterilization<br>Temperature | Sterilization<br>Time | Drying<br>Time | Maximum<br>Load |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----------------------|----------------|--------------------|
| 134°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>and textiles; wrapped and<br>unwrapped | 134°C (273°F) | 4 minutes | 3 minutes | 0.5 Kg/<br>1.1 lbs |
| 121°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>textiles, and plastics;<br>wrapped and unwrapped | 121°C (250°F) | 30 minutes | 5 minutes | 0.5 Kg/<br>1.1 lbs |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K210279
#### Sponsor/ Applicant 1.
Enbio Group AG Eichengasse 3 CH-4702 Oensingen, Switzerland
Mr. Lukasz Rogowski Corporate Quality Manager Email: lukasz.rogowski@enbio.com Phone: +48 605 058 629
Summary Preparation Date: October 29, 2021
#### 2. Device
| Trade Name | Enbio S |
|---------------------|------------------|
| Classification | Class 2 |
| Classification Name | Steam Sterilizer |
| Product Code | FLE |
| Regulation Number | 21 CFR 880.6880 |
| Review Panel | General Hospital |
#### 3. Predicate Device
Statclave G4 Chamber Autoclave (K190062)
#### 4. Device Description
The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It has a hermetically sealed, heated chamber made from aluminum, with two heaters sink inside chamber walls. Inside this chamber, the sterilized load is placed on a special perforated tray. After closing the chamber, the user selects the appropriate sterilization program through the TFT touch screen.
{4}------------------------------------------------
The actual sterilization phase starts after the pre-vacuum phase. The aluminum steam generator produces superheated steam and applies it inside the chamber. That steam penetrates the sterilized instruments. The set temperature is maintained inside the chamber depending on the selected sterilization cycle (121C° or 134°C), during a specified time (30 minutes or 4 minutes). After that time all the steam accumulated inside the chamber is pumped out and the drying cycle begins. When sterilization is finished, device displays to the user that process is completed, and that the load is sterile.
#### 5. Indications for Use
The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave.
Please refer to the table below for program name, load description, sterilization temperature, exposure time, drying time and maximum load.
| Program<br>Name | Load Description | Sterilization<br>Temperature | Sterilization<br>Time | Drying<br>Time | Maximum<br>Load |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----------------------|----------------|--------------------|
| 134°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>and textiles; wrapped and<br>unwrapped | 134°C (273°F) | 4 minutes | 3 minutes | 0.5 Kg/<br>1.1 lbs |
| 121°C | solid objects, small porous<br>objects, simple objects<br>recessed, narrow-clearance<br>items, dental handpieces,<br>textiles, and plastics;<br>wrapped and unwrapped | 121°C (250°F) | 30 minutes | 5 minutes | 0.5 Kg/<br>1.1 lbs |
{5}------------------------------------------------
#### Technological Characteristics Comparision Table 6.
Provided below is a technological comparison of the subject device with the predicate device.
| | Subject Device | Predicate Device | Comparison |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Trade Name | Enbio S | STATCLAVE G4 Chamber<br>Autoclave (K190062) | |
| Submitter | Enbio Group AG | SciCan Ltd. | - |
| Product Code | FLE | FLE | Same |
| Regulation Number | 21 CFR 880.6880 | 21 CFR 880.6880 | Same |
| Device Class | Class 2 | Class 2 | Same |
| Prescription / Over-The-Counter Use | Over-The-Counter | Over-The-Counter | Same |
| Intended Use | The Enbio S is an air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The Enbio S has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and/or damage to the autoclave. | The STATCLAVE G4 is a dynamic-air-removal (pre-vacuum) table-top steam sterilizer intended for use by a health care provider to sterilize medical products by means of pressurized steam. It is suitable for the sterilization of dental and medical instruments that are validated to be sterilized by steam. The STATCLAVE G4 has not been designed to sterilize liquid loads, bio-medical waste or materials not compatible with steam sterilization. The processing of such loads may result in incomplete sterilization and / or damage to the autoclave. | Same |
| Water tank | External | Internal reservoir | Different |
| Sterilization Chamber Volume | 2.7 L | 26 L | Different |
| Sterilization Chamber Dimensions | 292 x 192 x 45 mm<br>(L x W x H) | Diameter: 280 mm<br>Depth: 381 mm | Different |
| Device Dimensions (L x W x H) | 561 x 252 x 162 mm | 635 x 450 x 495 mm | Different |
| Weight | 15 kg (approximately) | 62.7kg (approximately) | Different |
| Power Rating | 110-120 V, 60Hz, 15A | 120V, 60Hz, 12 A | Similar |
| Wireless<br>Transmission<br>Capability | No | Yes | Different |
| USB Port | Yes | Yes | Same |
| Sterility and Shelf-<br>life | Not provided sterile.<br>No shelf-life claimed | Not provided sterile.<br>No shelf-life claimed | Same |
{6}------------------------------------------------
#### Non-clinical Bench (Performance) testing 7.
Provided below is the non-clinical testing to demonstrated the subject device meets the specification and accpetance criteria found in the Standards and test method listed below.
| Test Method | Purpose | Acceptance criteria | Result |
|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| ANSI AAMI ST55:2016<br>Vacuum Test | Verify air removal<br>performance | Average leak rate of 1<br>millimeter of mercury<br>(mmHg) (0.13 kPa) (0.019<br>psia) per min or less | Pass |
| Bowie Dick Test | Verify air removal<br>performance | The Bowie-Dick test<br>indicator sheet shall show<br>a uniform color change;<br>i.e., the color in the center<br>should be the same as<br>that at the outer edges. | Pass |
| ANSI AAMI ST55:2016<br>Empty Chamber Study<br>• 134°C (273°F) in 4 minutes | Verify pressure and<br>temperature – to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions<br>within the chamber | +3°C/-0°C<br>±0.3 bar | Pass |
| ANSI AAMI ST55:2016<br>Empty Chamber Study<br>• 121°C (250°F) in 30 minutes | Verify pressure and<br>temperature - to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions<br>within the chamber | | Pass |
| ANSI AAMI ST55:2016<br>Full Cycle Study<br>• 134°C (273°F) in 4<br>minutes | Verify pressure and<br>temperature - to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions | +3°C/-0°C<br>±0.3 bar | Pass |
| ANSI AAMI ST55:2016<br>Full Cycle Study<br>• 121°C (250°F) in 30<br>minutes | Verify pressure and<br>temperature - to ensure<br>that the sterilizer is<br>capable of providing<br>steady-state thermal<br>and pressure conditions<br>during the cycle | | Pass |
| ANSI AAMI ST55:2016<br>Half Cycle Study<br>• 134°C (273°F) in 4<br>minutes | Verify pressure and<br>temperature – to ensure<br>that the sterilizer is<br>capable of providing | +3°C/-0°C<br>±0.3 bar | Pass |
| ANSI AAMI ST55:2016<br>Half Cycle Study<br>• 121°C (250°F) in 30 minutes | steady-state thermal<br>and pressure conditions<br>during the cycle | | Pass |
| ANSI AAMI ST55:2016<br>Full Cycle Biological<br>Indicators<br>• 134°C (273°F) in 4 minutes | To verify biological<br>sterilization performance | The tested cycle has a<br>10-6 SAL | Pass |
| ANSI AAMI ST55:2016<br>Full Cycle Biological<br>Indicators<br>• 121°C (250°F) in 30 minutes | | | Pass |
| ANSI AAMI ST55:2016<br>Half Cycle Biological<br>Indicators<br>• 134°C (273°F) in 2 minutes | To ensure the efficacy of<br>the equipment and the<br>lethality of the<br>recommended<br>processing parameters<br>by biological challenge | The tested cycle has a<br>10-6 SAL | Pass |
| ANSI AAMI ST55:2016<br>Half Cycle Biological<br>Indicators<br>• 121°C (250°F) in 15 minutes | | | Pass |
| ANSI AAMI ST55:2016<br>Half Cycle Study /<br>Textile PCD<br>• 134°C (273°F) in 2 minutes | Verify half-cycle<br>sterilization of BI inside<br>textile PCD | The tested cycle has a<br>10-6 SAL | Pass |
| ANSI AAMI ST55:2016<br>Half Cycle Study /<br>Textile PCD<br>• 121°C (250°F) in 15 minutes | | | Pass |
| IEC 61010-1, 61010-2 | Verify electrical safety | Meets specifications of<br>standard | Pass |
| IEC 60601-1-2,<br>IEC 61326-1,<br>47 CFR 15 Subpart B, | Verify electromagnetic<br>compatibility | Meets specifications of<br>standards | Pass |
{7}------------------------------------------------
#### 8. Clinical Testing
The submission does not contain any data from clinical testing.
#### 9. Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the Enbio S is as safe, as effective, and performs as well as or better than the legally marketed predicate device, STATCLAVE G4 Chamber Autoclave (K190062).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.