K210239 · Acuitive Technologies, Inc. · MAI · Feb 24, 2021 · Orthopedic
Device Facts
Record ID
K210239
Device Name
CITRESPLINE and CITRELOCK ACL Implants
Applicant
Acuitive Technologies, Inc.
Product Code
MAI · Orthopedic
Decision Date
Feb 24, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Citrespline™ and Citrelock™ ACL Implants are intended for soft tissue reattachment, i.e. fixation of ligament and tendon graft tissue in surgeries of the shoulder, elbow, foot/ankle, knee, and hand/wrist. More specifically: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tendon Reattachment, Acromion-Clavicular Separation Repair, Deltoid Repair, Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle. Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction. Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty). Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist.
Device Story
Citrespline and Citrelock ACL Implants are bioresorbable soft tissue fixation devices; used for attaching ligament/tendon grafts to bone. Devices feature cannulated designs: 'twist-ribbed' (Citrelock) or 'straight spline' (Citrespline). Implants are placed into surgically prepared bone sites using reusable instruments; provide press-fit, interference fixation. Material is CITREGEN, a biocomposite of 60% unsintered hydroxyapatite and 40% citrate-based polyester; undergoes surface erosion via hydrolysis. Used by orthopedic surgeons in clinical settings (OR). Output is mechanical fixation of soft tissue; facilitates healing by securing graft during tissue integration. Benefits include secure graft attachment and gradual resorption of implant material.
Clinical Evidence
Bench testing only. Performance validated via pull-out testing at time-zero and after 26 weeks of soaking in 37°C PBS. Fatigue testing performed after 5,000 cycles of simulated physiological loading. Sterility, packaging, and shelf-life (1 year) validated. Bacterial endotoxin testing confirmed compliance with limits (<20 EUs/procedure).
Technological Characteristics
Materials: CITREGEN (60% unsintered hydroxyapatite, 40% citrate-based polyester). Design: Cannulated, press-fit interference fixation; diameters 7-12mm, lengths 23-33mm. Principle: Mechanical fixation via spline or twist-ribbed geometry; bioresorbable via hydrolysis. Provided sterile.
Indications for Use
Indicated for soft tissue reattachment and fixation of ligament and tendon graft tissue in orthopedic surgeries of the shoulder, elbow, foot/ankle, knee, and hand/wrist in patients requiring such repairs.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
February 24, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Acuitive Technologies, Inc. % Robert Poggie President BioVera, Inc 65 Promenade Saint Louis Notre-Dame-de-L'lle-Perrot, QC J7V7P2 Canada
#### Re: K210239
Trade/Device Name: Citrespline™ and Citrelock™ ACL Implants Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: MAI Dated: January 28, 2021 Received: January 29, 2021
## Dear Robert Poggie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K210239
Device Name
Citrespline™ and Citrelock™ ACL Implants
#### Indications for Use (Describe)
The Citrespline™ and Citrelock™ ACL Implants are intended for soft tissue reattachment, i.e. fixation of ligament and tendon graft tissue in surgeries of the shoulder, elbow, foot/ankle, knee, and hand/wrist. More specifically:
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tendon Reattachment, Acromion-Clavicular Separation Repair, Deltoid Repair, Capsulolabral Reconstruction.
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle.
Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction. Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty). Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## CITRESPLINE™ and Citrelock™ACL Implants
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information summarizes the safety and effectiveness of Acuitive Technologies' CITRESPLINE™ and CITRELOCK™ ACL Implants.
# A. SUBMITTERS INFORMATION
| Submitter Name: | BioVera, Inc. |
|----------------------|-----------------------------------------------------------------------------|
| Submitter Address: | 65 Promenade Saint Louis, Notre-Dame-de-L'lle-Perrot, QC<br>J7V 7P2, CANADA |
| Contact Person: | Robert A Poggie, PhD |
| Phone & Fax Numbers: | 514-901-0796 |
| Date of Submission: | January 28, 2021 |
# B. DEVICE IDENTIFICATION & MANUFACTURER
| Manufacturer Name: | Acuitive Technologies, Inc. |
|-----------------------|-------------------------------------------------------------------------------------------------------|
| Manufacturer Address: | 50 Commerce Drive, Allendale, NJ 07401, USA |
| Registration Number: | 10079115 |
| Contact Name: | Matthew Poggie |
| Title: | Sr. VP R&D Operations |
| Device Trade Name: | CITRESPLINE™ and CITRELOCK™ ACL Implants |
| Device Common Name: | Fastener, Fixation, Non-Degradable, Soft Tissue |
| Classification Name: | Single/multiple component metallic bone fixation appliances<br>and accessories (21 C.F.R. § 888.3030) |
| Classification Code: | MAI |
| Classification Panel: | Orthopedic |
| Regulation Number: | 21 C.F.R. § 888.3030 |
## C. PRIMARY PREDICATE DEVICE
| K200725 | CITREGEN™ Tendon Interference Screw (TIS) |
|---------|-------------------------------------------|
| | CITRELOCK™ Tendon Fixation Device |
{4}------------------------------------------------
## D. Indications for Use
The CITRESPLINE™ and CITRELOCK™ ACL Implants are intended for soft tissue reattachment, i.e. fixation of ligament and tendon graft tissue in surgeries of the shoulder, elbow, foot/ankle, knee, and hand/wrist. More specifically:
Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tendon Reattachment, Acromion-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle.
Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist.
## E. Device Description
The CITRESPLINE™ and CITRELOCK™ ACL Implants are soft tissue fixation devices ranging in diameter from 7 to 12 mm in 1 mm increments, and lengths of 23, 28, and 33 mm. There are two designs for the system: 1) a "twist-ribbed" cannulated, self-locking device (trade names CITRELOCK ACL and CITRELOCK) and 2) a "straight spline" style device (trade name CITRESPLINE). Both designs include a cannula through the central axis to allow for alignment into the surgically prepared site. Like the predicate devices, surgical placement of the implants in bone is facilitated with a reusable set of instruments. The CITRESPLINE and CITRELOCK ACL Implants are intended to attach soft tissue to bone for procedures in the foot/ankle, knee, shoulder, elbow, and hand/wrist.
The CITRESPLINE and CITRELOCK ACL Implants are made from CITREGEN biomaterial, which is a homogeneous biocomposite comprised of 60 wt. -% unsintered hydroxyapatite (HA) and 40 wt.-% polyester. CTREGEN's polymer component is a citrate-based network of completely amorphous polymer chains crosslinked together to form an elastomeric material. As water penetrates the subject device, surface erosion of the polymer phase occurs through hydrolysis of the ester bonds located between the monomers and at the crosslink sites.
## F. Comparison of Technological Characteristics
The CTRESPLINE and CITRELOCK ACL Implants have the same intended and indications for use, are manufactured from the same Citregen biomaterial using the same manufacturing processes, possess similar technological characteristics, and have the same principles of
{5}------------------------------------------------
operation as the primary predicate device (K200725). The subject and predicate СтяЕGENbased devices have comparable technological characteristics that include:
- Composed of the same CITREGEN resorbable biocomposite material, ●
- Same clinical indications for use and intended use, ●
- Press-fit, interference fixation designs,
- Same general surgical technique and reusable instrument set, and ●
- Provided sterile to the end user. ●
The two differences in technological characteristics of the subject and predicate devices are: (1) the subject devices include sizes that are larger in diameter and longer in length, and (2) the CTRESPLINE Implant is characterized by a spline design, versus the screw and twist-ribbed designs of the primary predicate device (K200725). Verification and validation data (e.g. 'pull-out' testing) demonstrated that the CirRESPLINE and CITRELOCK ACL Implants are substantially equivalent to the CITREGEN TIS and CITRELOCK predicate devices.
Further to the above comparison of device characteristics, the subject and predicate devices are comprised of CITREGEN biomaterial, which is a homogeneous biocomposite comprised of unsintered hydroxyapatite (HA) and polyester that is bioresorbed over time. The design and size options of the subject CITRESPLINE and CITRELOCK ACL Implants do not present a new worst case relative to simulated soft tissue fixation in bone, which is attributed to the larger diameter and longer length of engagement in bone associated with the subject devices.
## G. Performance Data
Verification and validation (V&V) activities were guided by the FDA quidance document "Premarket Notification (510(k)) Submissions for Bone Anchors", with pull-out testing performed for worst case subject devices at time-zero and after soaking in 37C PBS through 26 weeks, and pull out testing after 5,000 cycles of simulated physiological loading, The results of these tests showed the worst case subject devices (e.g. smallest diameter, shortest length) to possess similar or significantly higher pull out strengths relative to the primary predicate devices (K200725), which is attributed to the larger diameter and longer engagement length of the subject devices.
V&V activities included adoption of previously performed validations for sterility of implants, packaging, and shipping and handling. Packaging and shelf life tests using real and accelerated time aging were performed with passing results. Bacterial endotoxin testing (BET) showed CITREGEN test articles to meet the set endotoxin limits (<20 EUs / surgical procedure). The labeled shelf-life of the product is one year based on the stability data analyzed to date.
# H. Conclusion
Based on the indications for use, technological characteristics, and the V&V activities summarized herein, Acuitive Technologies, Inc. has determined that the subject CTRESPLINE and CITRELOCK ACL Implants are substantially equivalent to the legally marketed CiTREGEN TIS and CITRELOCK devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.