K210217 · Jiangsu Caina Medical Co.,Ltd · FPA · May 17, 2021 · General Hospital
Device Facts
Record ID
K210217
Device Name
Needleless Connector
Applicant
Jiangsu Caina Medical Co.,Ltd
Product Code
FPA · General Hospital
Decision Date
May 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Needleless Connector is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. It can be used for direct injection, intermittent infusion, continuous infusion or aspiration.
Device Story
Needleless connector provides sterile access to IV lines/catheters; activated by male luer. Used for direct injection, intermittent infusion, continuous infusion, or aspiration. Supports power injection of contrast media (max 325psi, 10ml/s). Used in clinical settings by healthcare providers. Device consists of shell, diaphragm valve, seat, and protecting cap. Benefits include needle-free access, reduced risk of needle-stick injuries, and maintenance of closed system integrity during IV therapy.
Clinical Evidence
No clinical data. Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, systemic toxicity, pyrogen, hemolysis), microbial ingress, seal integrity (ASTM F1886, F88, F1929), and mechanical performance (ISO 8536-12, ISO 80369-7, ISO 80369-20, burst pressure).
Technological Characteristics
Materials: Polycarbonate (shell/seat), Silicone rubber (septum), Polydimethylsiloxane (lubricant). Standards: ISO 10993 (biocompatibility), ISO 8536-12 (check valves), ISO 80369-7/20 (connectors), USP <85> (endotoxins), USP <151> (pyrogen), ASTM F756 (hemolysis). Single-use, sterile, 5ml flush volume, 7-day/100-activation limit. Power injection rated to 325psi/10ml/s.
Indications for Use
Indicated for needleless access to IV lines or catheters during IV therapy for direct injection, intermittent infusion, continuous infusion, or aspiration in patients requiring IV access.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 17, 2021
Jiangsu Caina Medical Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, Jiangsu 200120 China
Re: K210217
Trade/Device Name: Needleless Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: April 7, 2021 Received: April 15, 2021
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210217
Device Name Needleless Connector
Indications for Use (Describe)
The Needleless Connector is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. It can be used for direction, intermittent infusion, continuous infusion or aspiration.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exhibit #3 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K210217
- l. Date of Preparation: 4/7/2021
- 2. Sponsor Identification
Jiangsu Caina Medical Co., Ltd. No.23, Huanxi Rd, Zhutang Town, Jiangyin city, Jiangsu, 214415, China.
Establishment Registration Number: 3005670221
Contact Person: Jianwei Pan Position: Management Representative Tel: +86-510-8686 6666- 8027 Fax: +86-510-8686 6666- 8009 Email: Jianwei.Pan@cainamed.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
Mid-Link Consulting Co., Ltd. P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850 Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
### 4. Identification of Predicate Device
510(k) Number: K132413 Product Name: MZlO00 Product Code: FPA Regulation Number: 21 CFR 880.5440
ર્ Identification of Proposed Device
> Trade Name: Needleless Connector Common Name: IV Administration Set
Regulatory Information Classification Name: set, administration, intravascular Classification: II; Product Code: FPA Regulation Number: 21 CFR 880.5440 Review Panel: General Hospital
### Indication for Use:
The Needleless Connector is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. It can be used for direct injection, intermittent infusion, continuous infusion or aspiration.
### Device Description:
The proposed device, Needleless Connector, is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. It can be used for direct injection, intermittent infusion, continuous infusion or aspiration.
The proposed device is a needleless connector which can be activated by a male luer, and the flush volume for the device is 5ml. The device can be used for less than 7 days and 100 activations.
The Needleless Connector also suitable for power injection of contrast media into the central venous system only through IV line that is also indicated for power injection at a maximum pressure of 325psi and a flow rate of 10ml/s.
{5}------------------------------------------------
### 6. Summary of Technological Characteristics
| ITEM | Proposed Device | Predicate Device<br>K132413 | Remark |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Code | FPA | FPA | Same |
| Regulation No. | 21 CFR 880.5440 | 21 CFR 880.5440 | Same |
| Class | II | II | Same |
| Indication for Use | The Needleless Connector is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. It can be used for direct injection, intermittent infusion, continuous infusion or aspiration. | The MZIOOO Is a sterile single patient use connector for needleless access to the IV line and/or IV catheter during IV therapy. The MZ 1000 can be used for direct Injection, Intermittent infusion, continuous infusion or aspiration. | Same |
| Configuration | Shell | Shell | Same |
| | Diaphragm valve | Diaphragm valve | |
| | Seat | Seat | |
| | Protecting cap | Protecting cap | |
| Single Use | Yes | Yes | Same |
| Sterile | Yes | Yes | Same |
| Label/Labeling | Complied with 21 CFR part 801 | Complied with 21 CFR part 801 | Same |
Table 1 General Comparison
## Table 2 Performance Comparison
| ITEM | Proposed device | Predicate Device<br>K132413 | Remark |
|--------------------------------------------------------|-------------------------------|-------------------------------|-------------|
| Performance<br>Standard | ISO 8536-12<br>ISO 80369-7 | Unknown | Different 1 |
| Low Flush<br>Volume | 5ml | 5ml | Same |
| Utility time | Less than 7 days | 7 days | Different 2 |
| Activation times | 100 times | 200 times | |
| Maximum pressure and<br>flowrate of power<br>injection | 325psi&10ml/s | 325psi&10ml/s | Same |
| Disinfectant | 70% isopropanol antiseptic | 70% isopropanol antiseptic | Same |
| Interstitial or dead | No interstitial or dead space | No interstitial or dead space | Same |
{6}------------------------------------------------
| space? | internal to the connector | internal to the connector | |
|-----------|---------------------------|---------------------------|------|
| Hemolytic | Non-hemolytic | Non-hemolytic | Same |
Different 1- Performance Standard
The performance standard for the predicate device is unknown, the proposed device was tested according to the ISO 8536-12 and ISO 80369-7. The test results demonstrate that the diaphragm valve of the proposed device meets the requirements of check valve specified in ISO 8536-12. The test results demonstrate that the luer connector of the proposed device meets the requirements ofISO 80369-7.
The microbial ingress test was conducted on the proposed device to evaluate whether the needleless connector could prevent the ingress of microorganism by simulating clinical operation.
In addition, same as the predicate device, the proposed device also can be used for power injection of contrast media into the central venous system only through IV line. The test results demonstrate that the proposed device can provide power injection at a maximum pressure of 325psi and a flow rate of lOml/s.
Based on above analysis, although the performance standard for the predicate device is unknown, it does not raise new safety and effectiveness on the proposed device.
### Different 2- Service time and Activation times
The utility time and activation times of proposed device is different from the predicate device. The proposed device has undergone microbial ingress testing, and the test results demonstrate the proposed device can be used for no more than 7 days and 100 activations. Therefore, the difference will not affect the safety and effectiveness of the proposed device.
In addition, in order to ensure the end user to know the service time and activation times of the proposed device, the service time and activation times of proposed device are described in User Manual.
| Item | Proposed device | Predicate Device<br>K132413 | Remark |
|--------------------------|------------------------------------------------------------|-----------------------------|-------------|
| Patient-contact Material | | | |
| Shell | Polycarbonate (PC) (S-3001R)<br>Blue color; 8% | | |
| Septum valve | Silicone rubber (50% of 2000-70-A and<br>50% of 2000-70-B) | Unknown | Different 3 |
| Seat | Polycarbonate (PC) (Rx 1805)<br>Blue color; 8% | | |
| Lubricant | Polydimethylsiloxane | | |
| | | Table 3 Safety Comparison |
|--|--|---------------------------|
{7}------------------------------------------------
| Biocompatibility | | | |
|----------------------------|----------------------------|----------------------------|-------------|
| Cytotoxicity | No Cytotoxicity | No biocompatibility hazard | Different 4 |
| Intracutaneous | No Intracutaneous | | |
| Sensitization | No Sensitization | | |
| Systemic Toxicity | No Systemic Toxicity | | |
| Pyrogen | No Pyrogen | | |
| Hemolysis | No Hemolysis | | |
| Subacute systemic toxicity | No Toxicity | | |
| Sterile | Yes | Yes | Same |
| Disinfectant | 70% isopropanol antiseptic | 70% isopropanol antiseptic | Same |
| Single use | Single use | Single use | Same |
| Endotoxin Limit | 20 EU per device | 20 EU per device | Same |
### Different 3- Patient-contact Material
The patient-contact material of the predicate device is unknown, but the material of the proposed device was tested for biocompatibility and the test results show that there are no negative impacts from the materials that are used in the proposed device. Therefore, the difference will not affect the safety and effectiveness of the proposed device. Therefore, the difference does not affect equivalence of proposed device and predicate device.
### Different 4 Biocompatibility
The biocompatibility test items of the predicate device is unknown. However, the test items for the proposed device are sufficient and the test result show that there are no negative impacts from the materials that are used in the proposed device. Therefore, the difference will not affect the safety and effectiveness of the proposed device. Therefore, the difference does not affect equivalence of proposed device and predicate device.
{8}------------------------------------------------
### 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ► ISO 10993-7:2008 Biological Evaluation of Medical Device- Part 7: Ethylene Oxide Sterilization Residuals
- ♪ USP <85> Bacterial Endotoxin Limit
- ♪ ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ♪ ASTM F88/F88M-15, Standard Test Method For Seal Strength Of Flexible Barrier Materials. (Sterility)
- ▲ ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
- ▶ ISO 10993-5:2009 Biological evaluation of medical device- Part 5: Tests for in vitro cytotoxicity
- ♪ ISO 10993-10:2010 Biological evaluation of medical device- Part 10. Tests for irritaion and skin sensitization
- ♪ ISO 10993-11 :2017 Biological Evaluation of Medical Device- Part 11: Tests for Systemic Toxicity
- ♪ USP 42 NF 37 <151> Pyrogen Test
- ▶ ASTM F756-2017 Standard Practice for Assessment of Hemolytic Properties of Materials.
- ♪ ISO 8536-12:2007+A1:2013 Infusion equipment for medical use- Part 12: Check valves
- ▶ ISO 80369-7:2016 Small-bore connectors for liquids and gases in healthcare applications-Part 7: Connectors for intravascular or hypodermic applications
- ► ISO 80369-20:2015 Small-bore connectors for liquids and gases in healthcare applications Part 20: Common test methods
- ► PSI testing power inject or/burst pressure
- 8. Clinical Test Conclusion
No clinical study is included in this submission.
- 9. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the proposed subject device is as safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.