1 ml Sterile Hypodermic Syringe For Single Use. With/Without Needle. Luer/Luer-Lock
K210200 · Set Medikal Sanayi VE Ticaret Anonim Sirketi · FMF · May 14, 2021 · General Hospital
Device Facts
Record ID
K210200
Device Name
1 ml Sterile Hypodermic Syringe For Single Use. With/Without Needle. Luer/Luer-Lock
Applicant
Set Medikal Sanayi VE Ticaret Anonim Sirketi
Product Code
FMF · General Hospital
Decision Date
May 14, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterile Hypodermic Syringe for Single Use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
Device Story
Sterile, single-use, 3-piece hypodermic syringe (barrel, plunger, piston) with or without needle. Used by medical professionals and trained caregivers in clinical settings for fluid injection or withdrawal. Manual operation: user pulls/pushes plunger within barrel to intake/expel fluid via luer or luer-lock connector. Device features transparent barrel with 0.1 ml graduation scale. Available in 1 ml size with various needle gauges (21G, 23G, 25G). Benefits include precise fluid delivery/aspiration and reduced risk of cross-contamination due to single-use design and EtO sterilization.
Clinical Evidence
Bench testing only. No clinical data. Device performance verified through conformance with ISO 7886-1 (syringes), ISO 7864 (needles), ISO 9626 (tubing), and ISO 10993 series (biocompatibility: cytotoxicity, irritation, sensitization, systemic toxicity, hemocompatibility).
Technological Characteristics
3-piece piston syringe. Materials: Polypropylene (barrel, plunger, needle hub), Isoprene rubber (piston), Polyethylene (plunger), SUS304 stainless steel (needle). Luer/Luer-lock 6% conical tip. EtO sterilized (SAL 10^-6). Complies with ISO 7886-1, ISO 7864, ISO 9626, and ISO 10993. Manual operation; no software or electronic components.
Indications for Use
Indicated for medical professionals and trained caregivers for the injection of fluids into or withdrawal of fluids from the body.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
SET MEDIKAL SANAYI VE TICARET ANONIM SIRKETI (K201284)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
Set Medikal Sanayi Ve Ticaret Anonim Sirketi % Mehmet Ormeci Consultant Medcer Uluslararasi Medikal Belgelendirme Anonim Sirketi Taspinar Mahallesi 2800. Caddesi A-2 Apt. No:6 B/49 Ankara. 06830 Turkey
Re: K210200
Trade/Device Name: 1 ml Sterile Hypodermic Syringe For Single Use. With/Without Needle. Luer/Luer-Lock Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, FMF, FMI Dated: April 7, 2021 Received: April 16, 2021
Dear Mehmet Ormeci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K210200
#### Device Name
Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock
Indications for Use (Describe)
The Sterile Hypodermic Syringe for Single Use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Ser Medikal. The logo consists of a blue square with a white "S" shape inside of it. To the right of the square is the text "ser medikal" in a sans-serif font.
### Special 510(k) Summary
| 510(k) Submitter Name | SET MEDIKAL SANAYI VE TICARET ANONIM SIRKETI |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter Address | Osmangazi Mahallesi Maresal Fevzi Cakmak Caddesi No:18 Esenyurt<br>İstanbul/TURKEY |
| 510(k) Submitter Telephone No | +90-212-622-0400 |
| Primary Correspondent Contact Information | Mehmet Fatih Ormeci (Consultant)<br>+90-532-397-3043<br>mformeci@gmail.com<br>Fax Number: +90 (212) 689 4 828 |
| Summary Preparation Date | 05/12/2021 |
| Trade or<br>Proprietary Name | Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock |
|------------------------------|--------------------------------------------------------------------------------|
| Common Name | Hypodermic Syringe |
| Classification Name | Piston Syringe |
| Regulation Number | 21 CFR 880.5860 |
| Product Code | FMF, FMI |
| | Subject Device Special<br>510k No | Predicate Device 510k<br>No | Predicate Device<br>Manufacturer |
|-----------------------------------------------------------------------------------------|-----------------------------------|-----------------------------|----------------------------------------------------|
| Sterile Hypodermic<br>Syringe for Single use,<br>with/without needle;<br>luer/luer-lock | K210200/S002 | K201284 | SET MEDIKAL SANAYI<br>VE TICARET ANONIM<br>SIRKETI |
### Indications for Use:
The Sterile Hypodermic Syringe for Single use with/without needle is intended to be used for medical purposes to inject fluid into or withdraw fluid from body.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SER Medikal. The logo consists of a blue square with a white "S" inside of it on the left side of the image. To the right of the square is the text "SER MEDIKAL" in a gray, sans-serif font. The text is slightly stylized, with the letters having a modern, geometric appearance.
## Special 510k Summary Piston Syringe: K210200/S002
| Syringe Size | Gauge | Length | Wall Type | Approval Number |
|--------------|-------|--------|-----------|-------------------------------|
| 1mL | 21 | 5/8" | TW | K210200 Current<br>Submission |
| 1mL | 23 | 1" | TW | |
| 1mL | 25 | 1" | RW | |
| 2mL | 20 | 1 1/2" | TW | |
| 2mL | 21 | 5/8" | TW | |
| 2mL | 22 | 1 1/4" | TW | |
| 2mL | 23 | 1 1/4" | TW | K201284 |
| 2mL | 25 | 1" | TW | |
| 2mL | 27 | 2" | RW | |
| 2mL | 27 | 1 1/2" | RW | |
| 2,5mL | 21 | 1 1/2" | TW | |
| 2,5mL | 21 | 5/8" | TW | |
| 2,5mL | 22 | 1 1/2" | TW | |
| 2,5mL | 22 | 1 1/4" | TW | |
| 2,5mL | 23 | 1 1/4" | TW | K201284 |
| 2,5mL | 25 | 1" | TW | |
| 2,5mL | 25 | 1" | RW | |
| 2,5mL | 27 | 1 1/2" | RW | |
| 2,5mL | 27 | 2" | RW | |
| 3mL | 21 | 1 1/2" | TW | |
| 3mL | 21 | 5/8 | TW | |
| 3mL | 22 | 1 1/4" | TW | |
| 3mL | 22 | 1 1/2" | TW | K201284 |
| 3mL | 23 | 1 1/4" | TW | |
| 3mL | 23 | 1" | TW | |
| 3mL | 25 | 1" | RW | |
| 5mL | 18 | 1 1/2" | TW | |
| 5mL | 20 | 1 1/2" | TW | |
| 5mL | 21 | 1 1/2" | TW | |
| 5mL | 21 | 5/8" | TW | |
| 5mL | 22 | 1 1/4" | TW | |
| 5mL | 22 | 1 1/2" | TW | |
| 5mL | 23 | 1 1/4" | TW | K201284 |
| 5mL | 23 | 1" | TW | |
| 5mL | 25 | 1" | RW | |
| Syringe Size | Gauge | Length | Wall Type | Approval Number |
| 10 ml | 18 | 1 1/2" | TW | K201284 |
| | 20 | 1 1/2" | TW | |
| | 21 | 5/8" | TW | |
| | 21 | 1 1/2" | TW | |
| | 22 | 1 1/4" | TW | |
| | 22 | 1 1/2" | TW | |
| | 23 | 1" | TW | |
| | 23 | 1 1/4" | TW | |
| 20 ml | 18 | 1 1/2" | TW | K201284 |
| | 20 | 1 1/2" | TW | |
| | 21 | 5/8" | TW | |
| | 21 | 1 1/2" | TW | |
| | 22 | 1 1/4" | TW | |
| | 22 | 1 1/2" | TW | |
| | 23 | 1" | TW | |
| | 23 | 1 1/4" | TW | |
| 50 ml | 15 | 1" | TW | K201284 |
| | 16 | 1" | TW | |
| | 18 | 1 1/2" | TW | |
| | 19 | 1 1/2" | TW | |
| | 20 | 1 1/2" | TW | |
| | 21 | 5/8" | TW | |
| | 21 | 1 1/2" | TW | |
| | 23 | 1 1/4" | TW | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SER Medikal. The logo consists of a blue diamond shape with a white "S" inside of it. To the right of the diamond is the text "SER medikal" in a gray, sans-serif font.
## Special 510k Summary Piston Syringe: K210200/S002
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for "Ser Medikal". The logo consists of a blue square with a stylized "S" inside it, followed by the text "Ser Medikal" in a gray, sans-serif font. The "S" in the square is also stylized and appears to be made up of multiple lines.
| Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock | | | |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| ITEM | Legally Marketed Device /<br>Predicate Device<br>K201284 | Modified Device<br>(Adding New Size, 1ml) | Comparison |
| Product name | Sterile Hypodermic Syringe<br>for Single use, with/without<br>needle; luer/luer-lock | Sterile Hypodermic Syringe<br>for Single use, with/without<br>needle; luer/luer-lock | Same |
| Product code | FMF | FMF | Same |
| Regulation No. | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | II | II | Same |
| Indications for use | The Sterile Hypodermic<br>Syringe for Single Use<br>with/without needle is<br>intended to be used for<br>medical purposes to inject<br>fluid into or withdraw fluid<br>from body. | The Sterile Hypodermic<br>Syringe for Single Use<br>with/without needle is<br>intended to be used for<br>medical purposes to inject<br>fluid into or withdraw fluid<br>from body. | Same |
| Configuration | Piston<br>Plunger<br>Barrel | Piston<br>Plunger<br>Barrel | Same |
| Sterility condition | EO Sterilized | EO Sterilized | Same |
| Environment of use | Prescription use | Prescription use | Same |
| Intended users | Medical professionals and<br>trained care givers | Medical professionals and<br>trained care givers | Same |
| Single use | Yes | Yes | Same |
| Operation mode | The plunger of syringe can be<br>pulled and pushed along inside<br>the barrel, allowing the syringe<br>to take in and expel the fluids<br>through the connector to the<br>patient. | The plunger of syringe can<br>be pulled and pushed along<br>inside the barrel, allowing<br>the syringe to take in and<br>expel the fluids through the<br>connector to the patient. | Same |
| Label/labeling | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | Same |
| Product<br>performance | Complied with<br>ISO 7886-1<br>ISO 7864<br>ISO 9626 | Complied with<br>ISO 7886-1<br>ISO 7864<br>ISO 9626 | Same |
| Volume | 2ml, 2,5ml, 3ml, 5ml, 6ml, 10ml,<br>12ml, 20ml, 24ml, 50ml | 1ml | Difference 1 |
| Sizes | 14G,15G,16G,18G,19G,20G,<br>21G,22G,23G,24G,25G,26G,27G,<br>28G,29G,30G | 21G, 23G, 25G | Same |
| Needle Lengths | 4-50 mm | 4-50 mm | Same |
| Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer-lock | | | |
| ITEM | Legally Marketed Device /<br>Predicate Device<br>K201284 | Modified Device<br>(Adding New Size, 1ml) | Comparison |
| Patient contact component and material | | | |
| Piston | Isoprene Rubber | Isoprene Rubber | Same |
| Barrel | PP | PP | Same |
| Plunger | PP/PE | PP/PE | Same |
| Needle | PP, SS304 | PP, SUS304 | Same |
| Sterilization | | | |
| SAL | 10-6 | 10-6 | Same |
| Endotoxin Limit | 20 EU per device | 20 EU per device | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for SER Medikal. The logo consists of a blue square tilted at a 45-degree angle with a white stylized letter "S" inside. To the right of the square is the text "SER MEDIKAL" in a modern, sans-serif font, with the letters in a light gray color.
#### Device Description
1 ml Sterile Hypodermic Syringe is 3-piece design. These 3 pieces are piston, plunger and barrel. The piston is made from isoprene rubber. Barrel is made from polypropylene. Plunger is made from polypropylene and polyethylene combination. Needle is made from stainless steel 304 grade and hub of needle is made from polypropylene. There are with or without needle configurations within this Special 510k submission. There are luer and luer-lock configurations within this special 510k submission. This special 510k submission only covers the 1ml syringe but several needle dimensions including 21Gx5/8", 23Gx1", and 25Gx1". The barrel marking scale interval is 0.1 ml and legible for all configurations covered by this Special 510k submission. Luer and Luer-lock have a 6% conical design. The barrel is transparent. The length (without needle), outside diameter are 99.30/100.2mm, 6.9/6.75 and 4.70/4.70, respectively for luer and luer-lock configurations. Intended users are medical professionals and trained caregivers. It is prescription use only.
It is EtO sterile and a single use device. Packaging combination is maintained from a transparent sterilization film and Tyvek paper appropriate to EtO sterilization.
The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient.
The products covered in this special 510k are in compliance with ISO 10993-1 and ISO 7886-1.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains a logo for "ser medikal". The logo consists of a blue square with a stylized "S" inside it on the left side. To the right of the square is the text "ser medikal" in a gray, sans-serif font.
# Special 510k Summary
Piston Syringe: K210200/S002
| Product Name | | Sterile Hypodermic Syringe for Single use, with/without needle; luer/luer lock | | |
|----------------------------------------------------------------------------------------------------|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|--|
| Instructions for use | | The Sterile Hypodermic Syringe for Single Use with/without needle is intended to be used for medical purposes to inject fluid into or | | |
| | | withdraw fluid from body. | | |
| Volume | | 1ml | | |
| Barrel Marking Specs (Scale interval) | | 0.1ml | | |
| Tip type | Luer | ✓ | | |
| | Luer lock | ✓ | | |
| Configuration | Barrel, plunger | ✓ | | |
| | Piston, plunger, barrel | ✓ | | |
| Barrel<br>Transparency | Transparent | ✓ | | |
| | UV Resistant | - | | |
| | | Opaque | - | |
| Needle (Needle configurations are mentioned in the<br>hypodermic needle product description table) | | With or without needle | | |
| Nozzle Type | | Luer | Luer Lock | |
| Length (without needle) | | 99.30 | 100.2 | |
| Diameter, Outside | | 6.90 | 6.75 | |
| Diameter, Inner | | 4.70 | 4.70 | |
| Plunger Material | | PE | | |
| Barrel Material | | PP | | |
| Piston Material | | Isoprene rubber | | |
| Nozzle Type | | Luer, Luer lock, 6% conical, 1,2mm | | |
| Lubricant Composition | | Silicone | | |
| Sterility Condition | | EO sterilized | | |
| Single Use | | Yes | | |
| Environment of use | | Prescription Use | | |
| Intended Users | | Medical professionals and trained care givers | | |
| Operation Mode | | The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient. | | |
| Graduation Legibility | | Legible | | |
| Endotoxin Limit | | 20 EU per device | | |
| Biocompatibility | | Complied with ISO 10993-4/5/10/11 (Cytotoxicity, Irritation, Sensitization Acute Systemic Toxicity, Hemolytical effect) | | |
| SAL | | 10-6 | | |
| Product Performance | | Complied with ISO 7886-1 | | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for "ser medikal". The logo consists of a blue diamond shape with a stylized "S" inside it on the left. To the right of the diamond is the text "ser medikal" in a gray, sans-serif font.
| Product Name | Sterile Hypodermic needle for Single use | | |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------|
| Sizes | 21G | 23G | 25G |
| Needle Lengths | $58$ " | 1" | 1" |
| (*): with 15 micron filter | | | |
| Needle Bond Strength | Min.<br>44N | Min.<br>34N | Min.<br>22N |
| Needle Cover Dimensions | $8×3$ | $8×3$ | $8×31$ |
| Needle Cover Strength | Min. 2N - max. 20N | | |
| Needle Cover Color | Transparent | | |
| Needle Tip Configuration | Back bevel needle | | |
| Configuration | Protective cap, Needle, Adhesives, Needle hub | | |
| Sterility Condition | EO sterilized | | |
| Single Use | Yes | | |
| Environment of use | Prescription Use | | |
| Intended Users | Medical professionals and trained care givers | | |
| Operation Mode | The plunger of syringe can be pulled and pushed along inside the barrel, allowing the syringe to take in and expel the fluids through the connector to the patient. | | |
| Graduation Legibility | Legible | | |
| Protective Cap Material | PP | | |
| Adhesive Material | Epoxy resin | | |
| Needle Hub Material | PP | | |
| Needle Material | PP (hub and cover), SUS304 Stainless Steel | | |
| Product Performance | Complied with ISO 7864, ISO 9626 | | |
| Endotoxin Limit | 20 EU per device | | |
| Biocompatibility | Complied with ISO 10993-4/5/10/11 (Cytotoxicity, Irritation, Sensitization Acute Systemic Toxicity, Hemolytical effect) | | |
| SAL | 10-6 | | |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for SER Medikal. The logo consists of a blue diamond shape with a white "S" inside of it. To the right of the diamond is the company name, "SER Medikal", in a gray, sans-serif font.
### Modification Discussion
In this Special 510(k) submission, a new size: 1ml syringe luer/luer lock with or without needle is covered. The intended use, raw material, manufacturing site, sterilization method, sterilization site, packaging material is same with cleared device K201284.
### Performance Testing
The sterile, hypodermic syringe described in this summary were tested and demonstrated to be in
conformance with the following FDA recognized standards:
- . ISO 9626 Second edition: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 7886-1 Second edition 2017-05 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use
- ISO 7864:2016 Fourth Edition: Sterile hypodermic needles for single use — Requirements and test methods
- . ISO 10993-4:2017 Biological Evaluation of Medical Devices — Part 4: Selection of Tests for Interactions with Blood
- ISO 10993-5:2009 Biological Evaluation of Medical Devices — Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10:2010 Biological Evaluation of Medical Devices — Part 10: Tests for Irritation and Skin Sensitization
- . ISO 10993-11:2017 Biological Evaluation of Medical Devices — Part 11: Tests for Systemic Toxicity
The sterile, hypodermic needle described in this summary were tested and demonstrated to be in
conformance with the following FDA recognized standards:
- . ISO 9626 Second edition: Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 7864:2016 Fourth Edition: Sterile hypodermic needles for single use — Requirements and test methods
- . ISO 10993-4:2017 Biological Evaluation of Medical Devices — Part 4: Selection of Tests for Interactions with Blood
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for "ser medikal". The logo consists of a blue, stylized letter "S" enclosed in a diamond shape, followed by the text "ser medikal" in a gray, sans-serif font. The logo is simple and modern, with a focus on the company's name.
- ISO 10993-5:2009 Biological Evaluation of Medical Devices — Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10:2010 Biological Evaluation of Medical Devices — Part 10: Tests for Irritation and Skin Sensitization
- . ISO 10993-11:2017 Biological Evaluation of Medical Devices — Part 11: Tests for Systemic Toxicity
#### Conclusion
The differences between the cleared (predicate) and the modified device do not raise any new or different questions of safety or effectiveness since modified devices are tested in accordance with ISO 7886-1 standard. The subject devices are substantially equivalent to the predicate devices with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.