NFC-600 provides non-mydriatic color retina and external images of the eye as an aid to clinicians in the evaluation and diagnosis of eye disease. NFC-600 is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions. NFC-600 is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health.
Device Story
NFC-600 is a portable, non-contact fundus camera for capturing, storing, and displaying 12MP color retinal and external eye images. Device uses NIR LED for alignment and white LED flash for image capture. Features include auto 3D tracking, auto exposure, auto-focusing, and internal 10-point fixation. Unlike the predicate (NFC-700), which is an all-in-one unit, the NFC-600 removes the embedded PC module, connecting via USB 2.0 to an external user-supplied PC or laptop for operation. Operated by clinicians in clinical settings; proprietary software on the external PC manages image capture, storage, retrieval, and export. Output provides high-resolution images to aid clinicians in evaluating and documenting ocular health, facilitating diagnosis of eye disease.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 15004-1, ANSI Z80.36, and ISO 10940 (optical design) was verified. Patient-contact materials were evaluated per ISO 10993. Software was validated per IEC 62304.
Technological Characteristics
Non-contact fundus camera; 12MP CMOS sensor; NIR LED (alignment) and white LED (flash) illumination. Dimensions: 330x260x330mm; weight ≤12kg. Connectivity: USB 2.0 to external PC. Materials: ABS (chinrest), TPE (forehead rest). Standards: IEC 60601-1, IEC 60601-1-2, ISO 15004-1, ANSI Z80.36, ISO 10940, ISO 10993, IEC 62304.
Indications for Use
Indicated for in-vivo viewing of the posterior and external area of the eye in patients requiring non-mydriatic retinal and external eye imaging for evaluation, diagnosis, and documentation of ocular health.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
NFC-700 Non-Mydriatic Auto Fundus Camera (K182199)
Submission Summary (Full Text)
{0}------------------------------------------------
April 13, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized human figure, while the FDA acronym and full name are written in blue, with the acronym in a blue square.
Crystalvue Medical Corporation Oliver Lin Director of Quality Assurance No.116, Ln.956, Zhongshan Rd., Taoyuan Dist., Taovuan. 33072 Taiwan
Re: K210197
Trade/Device Name: NFC-600 Automated Portable Retinal Camera Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: March 18, 2021 Received: March 23, 2021
### Dear Oliver Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) f or combination products; and, if applicable, the electronic product radiation control provisions (Sections 53 1 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elvin Y. Ng -S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210197
Device Name NFC-600 Automated Portable Retinal Camera
Indications for Use (Describe)
NFC-600 is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions.
NFC-600 is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health.
Type of Use (Select one or both, as applicable)
| <span>☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------|------------------------------------------------------------|
|-------------------------------------------------------------|------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Crystalvue. The word "Crystalvue" is written in a sans-serif font. The first part of the word, "Crystal", is written in black, while the second part, "vue", is written in blue. To the left of the word "Crystalvue" is a graphic that looks like a crystal.
# 510(k) Summary
Crystalvue Medical Corporation NFC-600 Automated Portable Retinal Camera
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.
### Submitter
| Company name: | Crystalvue Medical Corporation |
|------------------|-----------------------------------------------------------------------------|
| Company Address: | No.116, Ln.956, Zhongshan Rd.,<br>Taoyuan Dist., Taoyuan City 33072, Taiwan |
| Contact Person: | Oliver Lin<br>Director of Quality Assurance |
| Phone: | +886 3 360 7711 Ext.2051 |
| Fax: | +886 3 360 7722 |
| E-mail: | Oliver.Lin@crystalvue.com.tw |
| Date prepared: | February 19, 2021 |
## Device Information
| Classification: | Class II |
|----------------------|-------------------------------------------|
| Trade Name: | NFC-600 Automated Portable Retinal Camera |
| Common Name: | Fundus Camera |
| Classification Name: | Ophthalmic Camera, AC power |
| Product Code: | HKI |
| Regulation Number: | 21 CFR § 886.1120 |
## Predicate Devices
Trade Name: NFC-700 Non-Mydriatic Auto Fundus Camera (Crystalvue Medical Corp.) Classification Name: HKI, Ophthalmic Camera, AC power 510(k) Number: K182199
#### Intended Use
NFC-600 provides non-mydriatic color retina and external images of the eye as an aid to clinicians in the evaluation and diagnosis of eye disease.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Crystalvue" in a stylized font. The "C" in Crystal is a blue circle with a white design inside, resembling a crystal. The rest of the word "Crystal" is in black, while "vue" is in a gradient from black to blue.
## Indication for Use
NFC-600 is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions.
NFC-600 is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health.
## Device Description
NFC-600 is a non-contact fundus camera for capturing, storing and displaying the color fundus images with 12 million pixels. It was designed a non-contact, images provide high resolution digital imaging device, auto 3D tracking, auto exposure, and auto capture functions, a portable device with auto-focusing technique and easy operation.
Different from the all-in-one design (built-in PC module) of previous cleared product. NFC-600 removed the embedded computer module and designed to use the USB 2.0 port which is located behind the base to connect to the user's PC or laptop, making the device smaller and lighter to achieve a portable design. The dimensions of NFC-600 is about 330mm (L)x 260mm (W)x 330mm (H) and the whole weight not more than 12 kg.
Same as the predicate device, NFC-600 uses NIR LED as illumination during alignment to the retina of patients' eyes. All operations are performed through the proprietary software, which will be installed in the user's PC or laptop to capture, store, view, retrieve, and export ophthalmic images.
#### Purpose of the Special 510(k)
The purpose for this 510(k) notification is to modify the current legally marketed device of Crystalvue Medical Corporation: NFC-700 Non-mydriatic Auto Fundus Camera (K182199). The modified device is referred to as NFC-600 Automated Portable Retinal Camera.
#### Summary of Modification
The difference between subject device and predicate device is that NFC-600 removed the embedded computer module from the device and changes it to an external PC or laptop, thereby making the appearance of NFC-600 smaller and lighter. In order to remove the embedded computer module from the device, the control circuit and PCB outline of control board need to do design modification to be used in NFC-600.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Crystalvue" in a stylized font. The "C" in Crystal is a circle with a pattern inside, resembling a crystal or lens. The rest of the word "Crystal" is in black, while "vue" is in a gradient of blue, transitioning from a darker to a lighter shade.
## Performance Data
As describe in Summary of Modification paragraph, in order to evaluate whether the changed control circuit and PCB control board of NFC-600 affect the safety and effectiveness, non-clinical tests including electrical safety and electromagnetic compatibility were conducted. It has been tested and found to comply with the consensus standards IEC 60601-1:2005 (MOD) and IEC 60601-1-2:2014.
The optical design of NFC-600 is exactly the same as the predicate device. It had been tested and complies with the following FDA- recognized consensus standards: ISO 15004-1:2006, ANSI Z80.36:2016 and ISO 10940:2009.
The materials of contact parts with patients in the NFC-600 are the same as predicate device and had been evaluated according to ISO10993 series; the software used in the NFC-600 is identical to the predicate device, and it was validated according to the standard IEC 62304:2015.
Above tests all meet the requirement and without any deviation, the modified NFC-600 as intended and produced the expected results, demonstrating that the safety and effectiveness of the modified device is as same as the predicate.
#### Substantial Equivalence
The subject device NFC-600 Automated Portable Retinal Camera has the following similarities to the predicate NFC-700 Auto Non-Mydriatic Fundus Camera cleared under 510(k) K182199:
- Have the same intended use / indications for use •
- . Uses the same operating principle
- . With the same basic technological characteristics and performance
- . With the same materials
The modification was evaluated by Crystalvue to determine if they could affect the safety or effectiveness of the subject device and it was determined that it does not raise any new questions of safety and effectiveness. Verification and validation was performed to ensure that the device performs as intended.
In summary, we believe the NFC-600 is substantially equivalent to predicate device NFC-700. A table comparing the technological characteristics with the predicate device is provided below.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Crystalvue" in a stylized font. The "C" is a light blue circle with a white center, resembling a crystal ball. The rest of the word is written in a combination of black and blue, with "crystal" in black and "vue" in blue.
| Product<br>Item | Subject Device NFC-600 | Predicate Device (K182199) | Comparison |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Intended Use | NFC-600 provides non-mydriatic color retina and external images of the eye as an aid to clinicians in the evaluation and diagnosis of eye disease. | NFC-700 provides non-mydriatic color retina and external images of the eye as an aid to clinicians in the evaluation and diagnosis of eye disease. | No changes |
| Indication for Use | NFC-600 is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions. NFC-600 is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health. | NFC-700 is a non-contact, high resolution digital imaging device which is suitable for photographing, displaying and storing images of the retina and external areas of the eye to be evaluated under non-mydriatic conditions. NFC-700 is indicated for in-vivo viewing of the posterior and external area of the eye and the images are intended for use as an aid to clinicians in the evaluation, diagnosis and documentation of ocular health. | No changes |
| Operation Principle | The optical design of fundus camera is based on the principle of monocular indirect ophthalmoscopy.<br><br>1. Fundus observation:<br>A build in light ray from the infrared light LED source to illuminate the fundus. Alignment of the device is performed by build in eye tracking indicator and working distance indicator to adjust system to best XYZ position automatically.<br><br>2. Image capture:<br>System use split-image technique to do image focus adjustment automatically to capture the best quality of image. White light from LEDs Flash module irradiates the fundus. The light reflected from eye portions forms an image, and the image is captured by built-in color CMOS camera module for fundus image capture. | | No changes. |
| Appearance | | | |
| -Shape | External PC or laptop | All in one (embedded PC) | NFC-700 is an All-In-One device; NFC-600 adopts external PC or laptop, making NFC-600 smaller and lighter. |
| -Dimensions (mm) | W260 x D330 x H330 | W282 x D485 x H492 | |
| -Weight | 12 Kg | 17 Kg | |
| Technological Characteristics | | | |
| -Image (resolution) | 12 MP | | No changes. |
| -Eye Fixation | Internal 10 points | | No changes. |
| -Light source | Observation: NIR LED / Flash: White LED | | No changes. |
| -Alignment | Fully automatic 3D tracking | | No changes. |
| Materials | | | |
| -Chinrest | ABS | | No changes. |
| -Forehead rest | TPE | | No changes. |
| Performance | | | |
| -Field of View | 45° | | No changes. |
| -Minimum Pupil Size | 4.0 mm | | No changes. |
| -Working Distance | 25 mm | | No changes. |
| -Focus Adjustment<br>Range | 1.Without compensation lens: -15 to +10 D<br>2. With compensation lens: -30D to -10D or +5D to +30D | | No changes. |
## Comparison of Technological Characteristics with the Predicate Device
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Crystalvue" in a stylized font. The "C" in "Crystal" is replaced with a circular graphic that resembles a lens or a crystal ball. The word "Crystal" is in black, while "vue" is in a gradient of blue, transitioning from a darker shade to a lighter one.
## Conclusion
Based on the performance data, identical intended use and technological characteristics and the similarities in functional design, the subject device NFC-600 is substantially equivalent to the NFC-700 (K182199) and the modification does not raise any new questions regarding safety and effectiveness. Performance and safety data demonstrate that the modified device is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.