K210166 · Medispec, Ltd. · ISA · Jan 27, 2023 · Physical Medicine
Device Facts
Record ID
K210166
Device Name
Omnispec ED1000
Applicant
Medispec, Ltd.
Product Code
ISA · Physical Medicine
Decision Date
Jan 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5660
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Omnispec ED1000 is intended for relief of minor muscle aches and pains, temporary increase in local blood circulation and activation of connective tissue.
Device Story
Omnispec ED1000 is a portable, self-contained shock wave therapy unit for therapeutic massage. Device uses electro-hydraulic method: electrode within water-filled stainless-steel reflector chamber (Shock Wave Applicator) ignites electrical discharge; water evaporation creates shock wave; wave reflects off chamber through flexible membrane into patient skin. Device provides focused pressure pulses; second focus (F2) occurs within applicator membrane. Used in clinical settings by healthcare providers. Output is focused pressure pulses transmitted to treatment site. Benefits include relief of minor muscle aches/pains, increased local blood circulation, and connective tissue activation. Provider operates device via touch screen panel and hand-held applicator.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation confirmed design specifications. Biocompatibility per ISO 10993-1; electrical safety per ANSI AAMI ES60601-1; EMC per IEC 60601-1-2 and 60601-2-36; acoustic parameters per IEC 61846.
Technological Characteristics
Electro-hydraulic shock wave generator; spark gap electrode; stainless-steel reflector chamber; silicone membrane; ultrasound gel coupling. Dimensions: 369x364x359mm. Connectivity: standalone. Software: moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, 60601-2-36, ISO 60601-1-6, ISO 62304, IEC 61846, ISO 10993-1.
Indications for Use
Indicated for relief of minor muscle aches and pains, temporary increase in local blood circulation, and activation of connective tissue in patients requiring therapeutic massage on upper extremities (shoulder to wrist) or lower extremities (hip to foot).
Regulatory Classification
Identification
A therapeutic massager is an electrically powered device intended for medical purposes, such as to relieve minor muscle aches and pains.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 27, 2023
Medispec Ltd. Avner Spector CEO 203 Perry Parkway, Suite 6 Gaithersburg, Maryland 20877
#### Re: K210166
Trade/Device Name: Omnispec ED1000 Regulation Number: 21 CFR 890.5660 Regulation Name: Therapeutic Massager Regulatory Class: Class I Product Code: ISA Dated: November 16, 2022 Received: December 21, 2022
#### Dear Avner Spector:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210166
Device Name Omnispec ED1000
Indications for Use (Describe)
The Omnispec ED1000 is intended for relief of minor muscle aches and pains, temporary increase in local blood circulation and activation of connective tissue.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | Commercial Use (If not for RFF, skip to next section) |
|--|-------------------------------------------------------|
| | In Situ Treatment with RFF |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Medispec logo. The logo features a stylized letter "M" with a blue circle around the top left portion of the letter. The rest of the logo is the word "MEDISPEC" in gray, block letters. The logo is simple and modern, and the colors are clean and professional.
# 510(k) Summary K210166
#### 1. SPONSOR
| Name: | Medispec, Ltd. |
|-----------------|------------------------------------------------|
| Address: | 203 Perry Pkwy Ste 6<br>Gaithersburg, MD 20877 |
| Phone: | 1-240-552-7310 |
| Fax: | 301-972-6098 |
| Contact Person: | Mr. Avner Spector |
| Email: | spector@medispec.com |
| Date Prepared: | July 12th, 2022 |
#### 2. DEVICE INFORMATION
| Proprietary Name: | Omnispec™ ED1000 |
|-------------------------|---------------------------------|
| Common Name: | Therapeutic Massager |
| Classification<br>Name: | Massager, Therapeutic, Electric |
| Regulation<br>Number: | 21 CFR 890.5660 |
| Product Code: | ISA |
| Device Class: | I |
#### 3. PREDICATE DEVICES
Equivalence is claimed to the OrthoGold 100™ of Tissue Regeneration Technologies, LLC, submissions number K210451, and D-ACTOR® 200 Vibration Massage System (K173692).
#### 4. INTENDED USE
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Medispec. The logo consists of a blue circle with a gray "M" inside of it. The text "MEDISPEC" is written in gray, with each letter connected to the next.
Omnispec™ ED1000TM is intended for the relief of minor muscle aches and pains, temporary increase in local blood circulation and activation of connective tissue.
#### 5. DEVICE DESCRIPTION
Omnispec™ ED1000 is a portable, self-contained Shock Wave Therapy unit. It employs an electro-hydraulic method of creating shock waves. With this technique, an electrode, located within a water-containing stainless-steel reflector chamber called Shock Wave Applicator (SWA), ignites an electrical discharge, evaporating a small portion of the water and creating a shock wave reflecting outward off the reflector through a flexible membrane. The result is a shock (pressure) wave that passes through a conducting medium (patient). The shock wave is transmitted through the skin surface of the patient to the treatment site. The device provides focused pressure pulses where the second focus (F2) occurs within the membrane of the applicator. Treatment should only be applied to the upper extremities, from the shoulder to the wrist, and to the lower extremities from the hip to the foot.
#### 6. PERFORMANCE TESTING
Verification and validation testing were performed and demonstrated that Omnispec™ ED1000 met the design specifications and it was safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed successfully.
Omnispec™ ED1000's software was validated and demonstrated to be of a moderate level of concern . Hazard analysis / risk management was performed and demonstrated that all risks were mitigated to an acceptable level.
The skin contacting component of the Shock Wave Applicator was tested for biocompatibility and found to conform to elements of ISO 10993-1.
Omnispec™ ED1000 was tested and demonstrated to conform to the general safety requirements of ANSI AAMI ES60601-1, as well as the electromagnetic compatibility requirements of IEC 60601-1-2 and 60601-2-36.
Ultrasonic energy parameters were evaluated using protocols from IEC 61846. Those tests demonstrated that the energy output of Omnispec™ ED1000 was within the safe limits and was demonstrated to be equivalent to the predicate device.
In addition, pressure stability, acoustic noise, cleaning and disinfection, drop and vibration, as well as Shock Wave Applicator shelf life verification and validation tests was also conducted. The performance testing has
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medispec logo. The logo features two overlapping circles, one in blue and one in gray. The word "MEDISPEC" is written in gray, with the "M" partially obscured by the circles. The logo is simple and modern, and the colors are clean and professional.
demonstrated that the Omnispec™ ED1000 is substantially equivalent to the predicate devices, and that it is safe and effective for its intended use. The table below compares Omnispec™ ED1000's characteristics to the predicate devices.
| Description | OrthoGold<br>100TM (Main<br>Predicate<br>Device) | D-ACTOR® 200<br>(Predicate<br>Device) | Omnispec ™<br>ED1000<br>(Subject<br>Device) | Comparison<br>Notes |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K210451 | K173692 | K210166 | NA |
| Product Code | ISA | ISA | ISA | Same |
| Regulation | 21 CFR<br>890.5660 | 21 CFR 890.5660 | 21 CFR<br>890.5660 | Same |
| Intended Use | Relief of minor<br>muscle aches<br>and pains,<br>temporary<br>increase in local<br>blood circulation<br>and activation of<br>connective<br>tissue | Relief of minor<br>muscle aches and<br>pains, temporary<br>increase in local<br>blood circulation<br>and activation of<br>connective tissue | Relief of minor<br>muscle aches<br>and pains,<br>temporary<br>increase in local<br>blood circulation<br>and activation<br>of connective<br>tissue | Same |
| Prescription/OTC<br>use | Prescription | Prescription | Prescription | Same |
| Mode of Action | Unfocused<br>pressure pulses | Radial (unfocused)<br>pressure waves, or<br>extracorporeal<br>pulse activation<br>respectively | Pressure pulses | Doesn't<br>affect safety<br>and<br>effectiveness |
| Mechanism of<br>Action | Extracorporeally<br>induced<br>unfocused<br>pressure pulses | Pneumatically<br>generated<br>vibrations +<br>unfocused<br>pressure pulses | Extracorporeally<br>induced<br>pressure pulses | Similar |
| Shockwave<br>Generation | Electrohydraulic,<br>spark gap<br>electrode | Pneumatic /<br>ballistic | Electrohydraulic,<br>spark gap<br>electrode | Same /<br>Similar,<br>respectively |
| Coupling<br>Solution | FDA-cleared<br>ultrasound gel | FDA-cleared<br>ultrasound gel | FDA-cleared<br>ultrasound gel | Same |
| User Interface | Touch screen<br>panel, touch<br>wheel, foot | Specific description<br>not available | Touch screen<br>panel, hand-<br>held applicator | Equivalent to<br>OrthoGold<br>100TM |
| Description | OrthoGold<br>100TM (Main<br>Predicate<br>Device) | D-ACTOR® 200<br>(Predicate<br>Device) | OmnispecTM<br>ED1000<br>(Subject<br>Device) | Comparison<br>Notes |
| | switch, hand-held applicator | | | |
| Main Unit<br>Dimensions<br>[mm] | 218 x 400 x 459<br>(H x W x D) | Specific value not<br>available | 369 x 364 x 359<br>(H x W x D) | Similar to<br>OrthoGold<br>100TM, both<br>suitable for<br>desktop use |
| Shockwave<br>Applicator<br>Dimensions<br>[mm] | 230 x Ø 70 | 4: 6mmOD,<br>15mmOD,<br>20mmOD,<br>35mmOD | 130 x Ø 90 | Similar, all<br>are intended<br>to be hand-held |
| Pulses<br>Frequency | Selectable, 1 - 8<br>Hz in steps of<br>0.5 Hz | Selectable, 1-21<br>Hz | Selectable, 2-<br>2.5 Hz, 120 or<br>160<br>pulses/minute | Similar,<br>OmnispecTM<br>ED1000's<br>frequencies<br>are within<br>the<br>Predicates'<br>range |
| Operating Mode | Continuous | Continuous | Continuous | Same as<br>OrthoGold<br>100TM |
| Peak compress-<br>ional acoustic<br>pressure pc<br>[Mpa] | 9.27 at energy<br>level 16 | 18.5 | 13.8 | Similar<br>OmnispecTM<br>ED1000's<br>pressure is<br>within the<br>range of D-<br>ACTOR®<br>200 |
| Peak<br>rarefactional<br>acoustic<br>pressure [Mpa] | -1.52 at energy<br>level 16 | 6.8MPa at 5 bar | 2.0 | Similar<br>OmnispecTM<br>ED1000's<br>pressure is<br>within the<br>range of D-<br>ACTOR®<br>200 |
| Positive peak<br>pressure<br>amplitude [Mpa] | 0.43 - 9.27 at<br>energy level 1 -<br>16 | 5 bar/18.5Mpa<br>3bar/13.4MPa | 13.8 | Similar<br>OmnispecTM<br>ED1000's<br>pressure is |
| Description | OrthoGoldTM 100TM (Main<br>Predicate<br>Device) | D-ACTOR® 200<br>(Predicate<br>Device) | OmnispecTM ED1000<br>(Subject<br>Device) | Comparison<br>Notes |
| | | | | within the<br>range of D-<br>ACTOR®<br>200 |
| Distance<br>between focus<br>and target<br>location | n/a | n/a | 65mm | Doesn't<br>affect safety<br>and<br>effectiveness |
| Maximum<br>penetration<br>depth (-6dB) | 25.4mm at<br>energy level 16 | 32 mm | 110mm | Doesn't<br>affect safety<br>and<br>effectiveness |
| Negative peak<br>pressure<br>amplitude (MPa) | -0.17 to -1.52 at<br>energy level 1 -<br>16 | 5bar/6.8MPa<br>3bar/5.0MPa | 2.0 | Similar<br>OmnispecTM<br>ED1000's<br>pressure is<br>within the<br>range of the<br>predicates |
| Energy flow<br>density | 0.00017-<br>0.04403mJ/mm2<br>at energy level<br>1-16 | 5bar/0.284mJ/mm2<br>3bar/0.176mJ/mm2 | 0.14 mJ/mm2 | Similar<br>OmnispecTM<br>ED1000's<br>energy flow<br>density is<br>within the<br>range of the<br>predicates |
| Derived focal<br>acoustic pulse<br>energy (mJ) | 0.022 - 2.278 at<br>energy level 1 -<br>16 | 5bar/6.5mJ<br>3bar/2.4mJ | 1.035 at 14.5<br>kV | Similar,<br>OmnispecTM<br>ED1000's<br>energy is<br>within<br>OrthoGold<br>100TM's<br>range |
| Rise time (ns) | 1.89 - 0.28 at<br>energy level 1 -<br>16 | Ultrasonic pulse:<br>2.5μs<br>Sonic pulse: 25μs<br>- 2.5ms | 39 | Similar<br>OmnispecTM<br>ED1000's<br>rise time is<br>within the<br>range of D- |
| Description | OrthoGold<br>100TM (Main<br>Predicate<br>Device) | D-ACTOR® 200<br>(Predicate<br>Device) | OmnispecTM<br>ED1000<br>(Subject<br>Device) | Comparison<br>Notes |
| | | | | ACTOR®<br>200 |
| Compress-ional<br>pulse duration<br>(μs) | 1.23 - 0.77 at<br>energy level 1 -<br>16 | Ultrasonic pulse:<br>5.0μs<br>Sonic pulse: 50μs<br>- 5.0ms | 0.136 | Similar |
| Applicator Wet<br>Shelf Life | Specific Value<br>Not Available | Specific Value Not<br>Available | 6 months | NA |
| Applicator Dry<br>Shelf Life | Specific Value<br>Not Available | Specific Value Not<br>Available | 18 months | NA |
| Applicator Use<br>Life | 100,000 pulses<br>at energy level 1<br>- 10; 70,000<br>pulses at energy<br>level 11 - 16 | Specific Value Not<br>Available | 110,000 pulses | Similar,<br>OmnispecTM<br>ED1000 has<br>only one<br>energy level |
| Patient<br>Contacting<br>Materials | Membrane:<br>Silicone<br>Coupling Media:<br>Ultrasound Gel | Coupling Media:<br>Ultrasound Gel | Membrane:<br>Silicone<br>Coupling Media:<br>Ultrasound Gel | Equivalent |
| Compliance with<br>Consensus<br>Standards | IEC 60601-1,<br>IEC 60601-1-2,<br>60601-2-36, ISO<br>60601-1-6, ISO<br>62304, Draft IEC<br>63045, ISO<br>10993-1 | IEC 60601-1, IEC<br>60601-1-2, ISO<br>62304, IEC 62304,<br>IEC 61000-3-2,<br>IEC 61000-3-3,<br>ISO 10993-1 | IEC 60601-1,<br>IEC 60601-1-2,<br>60601-2-36,<br>ISO 60601-1-6,<br>ISO 62304, IEC<br>61846, ISO<br>10993-1 | Equivalent,<br>Draft IEC<br>63045 is<br>equivalent to<br>IEC 61846<br>but applies<br>to unfocused<br>pressure<br>pulses |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Medispec. The logo consists of a stylized letter "M" with a blue circle around the top left portion of the letter. The rest of the word "MEDISPEC" is written in gray, sans-serif font.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Medispec. The logo consists of two overlapping circles, one in blue and one in gray, with the letter "M" in the center. The text "MEDISPEC" is written in gray, with the "M" connected to the circles. The logo is simple and modern, and the colors are calming and professional.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Medispec logo. The logo features a stylized letter "M" with a blue arc above it. The word "MEDISPEC" is written in gray, with the "M" being connected to the arc. The font is sans-serif and appears to be a modern design.
### 7. CLINICAL TESTING
Not applicable
#### 8. SUMMARY
Omnispec™ ED1000 has the same intended use and similar design features as compared with the predicate systems. The devices share similar core technologies and principle of operation. The performance testing has demonstrated that the performance characteristics of Omnispec™ ED1000 are equivalent to those of other legally marketed therapeutic massagers, and
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Medispec logo. The logo features a stylized symbol resembling interconnected circles in blue and gray. To the right of the symbol, the word "MEDISPEC" is written in gray, with each letter clearly visible. The overall design is clean and professional, suggesting a company in the medical or technology field.
therefore it supports a determination of Substantial Equivalence for the proposed intended use.
Any differences between Omnispec™ ED1000 and the predicate device do not affect the safety or effectiveness, or raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.