s-Clean SQ-SL Implant System Regular is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single- or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one- and two-stage surgical procedures. This system is intended for delayed loading.
Device Story
Thread-type endosseous dental implant system; fixture placed in alveolar bone to replace missing tooth function; provides connection for prosthetic devices via internal hex. Used in one- or two-stage surgical procedures; intended for delayed loading. Operated by dental professionals in clinical settings. Device consists of fixture and cover screw; fixture surface treated with SLA (sandblasted, large-grit, acid-etched). Output is a stable base for dental restorations; aids in restoring masticatory function and aesthetics for edentulous patients.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing leveraged from predicate devices, including sterilization validation (ISO 11137), shelf-life testing (ASTM F1980), biocompatibility (ISO 10993 series), and bacterial endotoxin testing (ANSI/AAMI ST72, USP <161>, USP <85>). Surface modification equivalence confirmed via surface roughness, composition analysis, and SEM imaging.
Indicated for partially or fully edentulous mandibles and maxillae in patients requiring single- or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Dentis Co., Ltd. % April Lee Consultant Withus Group Inc. 106 Superior Irvine, California 92620
Re: K210132
Trade/Device Name: s-Clean SQ-SL Implant System Regular Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: January 8, 2021 Received: January 19, 2021
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or post-marketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K210132
Device Name s-Clean SQ-SL Implant System Regular
#### Indications for Use (Describe)
s-Clean SQ-SL Implant System Regular is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single- or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one- and two-stage surgical procedures. This system is intended for delayed loading.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> X Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "DENTIS" in bold, black letters. To the left of the word is an orange graphic that resembles a flame or a stylized letter "d". The graphic is a gradient of orange, with the darkest shade at the bottom and the lightest at the top.
- Plant 1: 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Plant 2: 6, Yuram-ro, Dong-gu, Daegu, Korea Tel: 82.53.582.2804 / Fax: 82.53.583.2806
#### 510(k) Summary K210132
●
#### Submitter
Dentis Co., Ltd. Gyu Ri Kim 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 Korea Email: kgr1026@dentis.co.kr Tel. +82-53-589-3541 Fax. +82-53-289-7922
#### Official Correspondent
Withus Group Inc. April Lee 106 Superior Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
#### Device Information
- Trade Name: s-Clean SQ-SL Implant System Regular ●
- . Common Name: Dental Implant System
- Classification Name: Implant, Endosseous, Root-Form
- Product Code: DZE
- Panel: Dental
- Regulation Number: 872.3640
- Device Class: Class II ●
- Date Prepared: 01/08/2021
#### Predicate Devices:
The subject device is substantially equivalent to the following predicate devices:
Primary Predicate
- . K192688, s-Clean SQ-SL Implant System Regular by Dentis Co., Ltd.
#### Indications for Use:
s-Clean SQ-SL Implant System Regular is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single- or multiple-unit restorations including: cemented retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. This system is dedicated for one- and two-stage surgical procedures. This system is intended for delayed loading.
#### Device Description:
This submission is to add new implants to the previously cleared device, s-Clean SQ-SL Implant System Regular (K192688). The newly added implant is "s-Clean SQ-SL Fixture with Ø5.2mm diameter".
The s-Clean SQ-SL Implant System Regular is composed of a fixture and cover screw. The s-Clean SQ-SL Fixture is a thread-type implant made of Commercial Pure Titanium Grade 4 according to ASTM F67
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the word "DENTIS" in bold, sans-serif font. To the left of the word is an orange, abstract graphic. The graphic is curved and has a gradient effect, with the color transitioning from a lighter orange to a darker orange.
Plant 1: 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea Plant 2: 6, Yuram-ro, Dong-gu, Daegu, Korea . Tel: 82.53.582.2804 / Fax: 82.53.583.2806
which is placed in the alveolar bone to replace the function of the missing tooth. This device has a connection between the upper prosthesis and the internal Hex.
.
The Fixture's surface is treated with SLA (Sandblasted with Large-grit and Acid-etching). It is the only part to be implanted into bone, to provide connection of prosthetic devices or other components of a dental implant set within the human body (mandibular or maxillary bone). The dimensions are as follows:
| Device Name | Diameter (mm) | Length (mm) |
|--------------------------|---------------|---------------------------|
| s-Clean SQ-SL<br>Fixture | $ 5.2 $ | 7.0, 7.5, 9.5, 11.5, 13.5 |
The fixture is packaged with a cover screw cleared under K192688. The fixture is provided sterile.
#### Materials:
The Fixture is fabricated from CP Titanium Grade 4 (Conforming to ASTM Standard F67).
#### Summaries of Technology Characteristics:
| | Subject Device | Primary Predicate Device |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | K210132 | K192688 |
| Manufacturer | Dentis Co., Ltd | Dentis Co., Ltd |
| Trade Name | s-Clean SQ-SL Implant Regular System | s-Clean SQ-SL Implant Regular System |
| Description | Image: s-Clean SQ-SL Implant Regular System | Image: s-Clean SQ-SL Implant Regular System |
| Indications for Use | s-Clean SQ-SL Implant System Regular is<br>indicated for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of single- or multiple-unit<br>restorations including: cemented retained,<br>screw retained, or overdenture restorations,<br>and terminal or intermediate abutment<br>support for fixed bridgework. This system<br>is dedicated for one- and two-stage<br>surgical procedures. This system is<br>intended for delayed loading. | s-Clean SQ-SL Implant System Regular is<br>indicated for use in partially or fully<br>edentulous mandibles and maxillae, in<br>support of single- or multiple-unit<br>restorations including: cemented retained,<br>screw retained, or overdenture restorations,<br>and terminal or intermediate abutment<br>support for fixed bridgework. This system is<br>dedicated for one- and two-stage surgical<br>procedures. This system is intended for<br>delayed loading. |
| Diameter (mm) | ⌀5.2 | ⌀4.1, 4.35, 4.8, 5.8, 6.8, 7.8 |
| Length (mm) | 7, 7.5, 9.5, 11.5, 13.5 | 7, 7.5, 9.5, 11.5, 13.5 |
| Surface Treatment | SLA | SLA |
| Material | CP Titanium Gr. 4 (ASTM F67) | CP Titanium Gr. 4 (ASTM F67) |
| Sterilization | Gamma irradiation | Gamma irradiation |
| Shelf Life | 8 years | 8 years |
- 1) s-Clean SQ-SL Fixture
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DENTIS. The logo features an orange abstract shape to the left of the word "DENTIS", which is written in a bold, sans-serif font. The abstract shape appears to be a stylized letter "D" or a flame-like design. The overall design is simple and modern.
| Plant 1: | 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea |
|----------|---------------------------------------------|
| Plant 2: | 6, Yuram-ro, Dong-gu, Daegu, Korea |
| Tel: | 82.53.582.2804 / Fax: 82.53.583.2806 |
| Comparison | The subject device and the primary predicate device have the same characteristics such as<br>Indications for Use, design, length, surface treatment, material, abutment connection, and<br>sterilization method. The difference between the two devices is the diameter. Since the<br>diameter of the subject device is in range of the diameters of the primary predicate, it does<br>not affect product performance. Therefore, the subject device is substantially equivalent. |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
●
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the word "DENTIS" in bold, black font. To the left of the word is an orange graphic that resembles a stylized flame or a curved shape with a circle inside. The overall impression is a logo or brand name.
Plant 1: 99, Seongseoseo-ro, Dalseo-gu, Daegu, Korea ● Plant 2: 6, Yuram-ro, Dong-gu, Daegu, Korea Tel: 82.53.582.2804 / Fax: 82.53.583.2806
#### Non-Clinical Test Data
No new non-clinical testing was performed for the subject device since the subject device compared to the predicate device are substantially equivalent in indications, fundamental technology, material, and design. Any testing provided in the predicate device may be leveraged for the subject device of using the same materials, manufacturing methods, and sterilization procedures. Although the diameters are slightly different, it does not impact the ability to determine substantial equivalence of the subject device because the diameter of the subject device is within the range of diameters of the primary predicate.
The tests below were performed for the predicate devices and leveraged for the subject device:
- Sterilization Validation Test for subject fixtures according to ISO 11137-1, 2, 3 referenced in K192688
- Shelf Life Test for subject fixtures according to ASTM F1980 referenced in K153639 ●
- Biocompatibility testing according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-● 5:2009, ISO 10993-6:2007, ISO 10993-10:2010, and ISO 10993-11:2006 referenced in K153639
- . Bacterial Endotoxin Test Report for subject fixtures according to ANSI/AAMI ST72:2011, USP <161>, and USP <85> referenced in K192688
The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device.
The surface modification information with SLA (Sandblasted with Large-grit and Acid-etching) was provided. To compare surface modification between the subject and predicate devices, K153639, surface roughness, surface composition analysis, and SEM imaging were provided and it demonstrated substantial equivalence.
The fatigue testing per ISO 14801 was not conducted as the subject device is not compatible with any angled abutments.
Non-clinical tests followed the recommendations in the "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
#### Conclusion
The s-Clean SQ-SL Implant System Regular constitutes a substantially equivalent medical device, meeting all the declared requirements of its intended use. This system has the same intended use and fundamental scientific technology as its predicate devices. Therefore, the s-Clean SO-SL Implant System Regular is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.