K210129 · Luvo Medical Technologies, Inc. · PBX · Jul 20, 2021 · General, Plastic Surgery
Device Facts
Record ID
K210129
Device Name
RF Thermal System
Applicant
Luvo Medical Technologies, Inc.
Product Code
PBX · General, Plastic Surgery
Decision Date
Jul 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation
Device Story
The RF Thermal System is a prescription-use, microprocessor-controlled console with an attached handpiece that delivers monopolar radiofrequency (RF) energy to human tissue. The device is operated by a clinician to elevate tissue temperature for therapeutic purposes, specifically for pain relief, muscle spasms, and local circulation improvement. The system includes a temperature sensor in the handpiece to monitor and maintain tissue temperature between 40°C and 44°C. The clinician selects treatment parameters via the console interface. By delivering controlled RF energy, the device induces localized heating, which provides therapeutic benefits to the patient. The system is designed for clinical use.
Clinical Evidence
Bench testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and high-frequency surgical equipment safety (IEC 60601-2-2). Biocompatibility was verified per ISO 10993. Mechanical performance testing confirmed output accuracy within ±20%. Human thermal testing (n=9) on the forearm, thigh, and abdomen demonstrated the device's capability to raise tissue temperature to 40°C and maintain it between 40°C–44°C for 10 minutes, confirming therapeutic efficacy.
Technological Characteristics
Monopolar RF generator; 4 MHz frequency; microprocessor-controlled console; handpiece with integrated temperature sensor. Materials comply with ISO 10993-1, 10993-5, 10993-10, and 10993-12. Packaging validated per ASTM D 4169:2016. Electrical safety per AAMI/IEC 60601-1:2005 + AMD 1:2012.
Indications for Use
Indicated for patients requiring elevated tissue temperature for temporary relief of pain, muscle spasms, and increased local circulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.
Jul1y 20, 2021
LUVO Medical Technologies, Inc. Mr. Gregory Berzak Director of Regulatory Affairs and Quality Compliance 125 Fleming Drive Cambridge, Ontario N1T2B8 Canada
Re: K210129
Trade/Device Name: RF Thermal System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: PBX Dated: June 10, 2021 Received: June 22, 2021
Dear Mr. Berzak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210129
Device Name
Thermal RF System
Indications for Use (Describe)
The Thermal RF System is indicated for:
Heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> <span> ☑ </span> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> <span> ☐ </span> </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510K Summary RF Thermal System K210129
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Applicant: | LUVO Medical Technologies, Inc. |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | LUVO Medical Technologies, Inc.<br>125 Fleming Dr<br>Cambridge, Ontario, Canada N1T 2B8 |
| Contact Person: | Mr. Gregory Berzak<br>Director of Regulatory Affairs and Quality Compliance |
| Contact Information: | 519-620-3900- phone<br>gregoryb@clarionmedical.com |
| Preparation Date: | July 16, 2021 |
| Device Trade Name: | RF Thermal System |
| Common Name: | Radio Frequency generator |
| Regulation Name: | 21 CFR 878.4400, Electrical cutting and coagulation device and<br>accessories; Massager, Vacuum, Radio Frequency Induced Heat |
| Product Code: | PBX - Massager, Vacuum, Radio Frequency Induced Heat |
| Legally Marketed<br>Predicate: | K200241 - Tempsure |
| Regulatory Class: | Class II Prescription Use |
| Description of Thermal RF: | The Thermal RF System is a Class II Medical Device that utilizes<br>monopolar RF energy for the purpose of elevating tissue<br>temperature for selected medical conditions. The system is<br>comprised of a micro-processor-controlled and user-friendly<br>console that houses the power supply, the electronics and the<br>user interface. It then has a handpiece that is attached to the<br>console, and through the user interface can be selected for<br>administering the treatment. |
| Indications for Use | Heating for the purpose of elevating tissue temperature for<br>selected medical conditions such as temporary relief of pain,<br>muscle spasms, and increase in local circulation. |
{4}------------------------------------------------
### 510K Summary RF Thermal System K210129
Comparison to Predicate Device:
| | Thermal RF | Predicate: TempSure | |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | K210129 | K200241 | |
| Indication for Use | Heating for the purpose of<br>elevating tissue<br>temperature for selected<br>medical conditions such as<br>temporary relief of pain,<br>muscle spasms, and<br>increase in local circulation | Heating for the purpose of<br>elevating tissue<br>temperature for selected<br>medical conditions such as<br>temporary relief of pain,<br>muscle spasms, and<br>increase in local circulation | Same |
| Technical<br>Specification | | | |
| Frequency (MHz) | 4 | 4 | Same |
| Operation Type | Monopolar | Monopolar | Same |
| Output Power | Up to 60W | 120-300W | Similar |
| TRF-E Tip | | | |
| Treatment Area | 0.38 cm² | 18mm | Similar |
| TRF-R Tip | | | |
| Treatment Area | 2.54 cm² | 25mm | Similar |
| TRF-S tip | | | |
Performance Data:
The following performance data was provided in support of the substantial equivalence determination:
IEC 60601-1 Test for Medical Electrical equipment was performed for General Requirements for basic safety and essential performance.
IEC 60601-1-2 Test for Medical Equipment for General Requirements for basic safety and essential performance: electromagnetic compatibility
IEC 60601-2-2 Test for Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
ISO 10993-1 Biological Evaluation of Medical Devices - Part 1:Evaluation and Testing Within a Risk Management Process
{5}------------------------------------------------
### 510K Summary RF Thermal System K210129
Thermal Testing was conducted on nine human subjects. The testing was conducted on three different parts of the body, the forearm, the thigh and the abdomen. Results showed that the Thermal RF system is capable of raising the temperature to 40°C and then maintain the temperate between 40°C – 44°C for 10 minutes, which meets the requirements for a therapeutic treatment.
Performance Test Summary:
The RF Thermal System completed performance testing to ensure safety and efficacy of the device.
Package Validation testing was performed successfully according to ASTM D 4169 : 2016. Assessment to national (National standard AAMI/IEC 60601-1:2005 + AMD 1:2012) and international regulations for Electrical devices was also performed successfully to Electrical Safety (IEC 60601), Electromagnetic compatibility (IEC 60601-1-2) and for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories (IEC 60601-2-2).
The RF Thermal handpiece TIP comes into contact with the patient' skin was subject to Biocompatibility testing using the following standards to determine compliance: ISO 10993-1, ISO 10993-12, ISO 10993-5 & ISO 10993-10. The Raw material compliance to the Biocompatibility standards demonstrate compliance and that it is safe to contact skin as intended.
Additionally, mechanical performance was done by measuring the output accuracy of the TRF TIP, all results demonstrate accuracy within ±20%. Thermal therapeutic testing was also conducted for the RF Thermal System to show substantial equivalence and safety by showing the correlation between output energy and the corresponding temperature for the intended use. The quantitative data report ensures safety and effectiveness. The RF Thermal handpiece also has a temperature sensor which was tested successfully to demonstrate the device meets design specifications and performance requirements.
#### Conclusion:
The RF Thermal System has the same technology, principle of operation, indications for Use, and similar technical specifications as the predicate device. Performance test results also demonstrated the subject device can perform the same intended use as safely and effectively as the predicate device. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.