← Product Code [KRA](/productcode/KRA) · K210114

# Rebar Micro Catheter (K210114)

_Micro Therapeuatics, Inc. D/B/A Ev3 Neurovascular · KRA · Feb 16, 2021 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210114

## Device Facts

- **Applicant:** Micro Therapeuatics, Inc. D/B/A Ev3 Neurovascular
- **Product Code:** [KRA](/productcode/KRA.md)
- **Decision Date:** Feb 16, 2021
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Rebar™ Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy.

## Device Story

Endhole, single-lumen microcatheter; introduced over steerable guidewire into vasculature. Proximal end features standard luer adapter; semi-rigid proximal shaft transitions to flexible distal shaft for navigation. Radiopaque markers at distal end allow fluoroscopic visualization. Outer surface coated for lubricity. Used in clinical settings by physicians; inserted via guide catheter with hemostatic side arm adapter. User flushes lumen with heparinized saline; advances guidewire/catheter assembly to target site. Once positioned, guidewire is removed; syringe attached to hub for infusion of contrast media or delivery of interventional devices. Facilitates targeted vascular access; enables diagnostic imaging or therapeutic intervention.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by non-clinical bench testing, including visual inspection, hub integrity (ISO 80369-7), liquid leak/air aspiration (ISO 10555-1), dynamic/static burst pressure, and tensile strength testing.

## Technological Characteristics

Single-lumen catheter; Pebax/stainless steel helical coil proximal shaft; PTFE inner lining; Trogamid hub; Dymax UV adhesive; Platinum-Iridium marker bands. Dimensions: 130-153 cm length; 2.4F-2.8F OD. Sterilization: Ethylene Oxide (EO). Connectivity: None (manual).

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- Rebar™ Micro Catheter ([K093750](/device/K093750.md))

## Reference Devices

- Marathon™ Flow Directed Micro Catheter ([K202318](/device/K202318.md))
- Riptide™ Aspiration System (React 71 Catheter) ([K182101](/device/K182101.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 16, 2021

Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Catherine Chiou Associate Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618

Re: K210114

Trade/Device Name: Rebar Micro Catheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA, QJP Dated: January 15, 2021 Received: January 19, 2021

Dear Catherine Chiou:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210114

Device Name Rebar™ Micro Catheter

Indications for Use (Describe)

The Rebar™ Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy.

| Type of Use (Select one or both, as applicable)                                           |
|-------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> For activities that DO NOT MAKE claims or charges</label> |
| <label><input type="checkbox"/> For activities that MAKE claims or charges</label>        |

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary [21 CFR 807.92]

## K210114

| 510(k) Owner:   | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular<br>9775 Toledo Way<br>Irvine, CA 92618<br>Establishment Registration: 2029214 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Catherine Chiou<br>Associate Regulatory Affairs Specialist<br>Telephone: (562)286-3057<br>Email: catherine.chiou@medtronic.com |

| Date Summary<br>Prepared: | February 15, 2021                         |
|---------------------------|-------------------------------------------|
| Trade Name of Device:     | Rebar™ Micro Catheter                     |
| Common Name of<br>Device: | Vascular Microcatheter                    |
| Review Panel:             | Cardiovascular; Neurovascular             |
| Product Code:             | KRA; QJP                                  |
| Regulation Number:        | 21 CFR 870.1210; 21 CFR 870.1250          |
| Regulation Name:          | Continuous Flush Catheter                 |
| Device Classification     | Class II                                  |
| Predicate Device:         | Rebar™ Micro Catheter<br>510(K)s: K093750 |

#### Predicate Device

Rebar™ Micro Catheter (K093750)

#### Reference Device

Marathon™ Flow Directed Micro Catheter (K202318)

Riptide™ Aspiration System (React 71 Catheter) (K182101)

## Device Description

The Rebar" Micro Catheter is an endhole, single-lumen catheter designed to be introduced over a steerable guidewire into the vasculature. The proximal end of the catheter incorporates a standard luer adapter to facilitate the attachment of accessories. The catheter has a semi-rigid proximal shaft which transitions into the flexible distal shaft to facilitate the advancement of the catheter in the anatomy. Single or dual radiopaque markers at the distal end facilitate fluoroscopic visualization. The outer surface of the catheter is coated to increase lubricity.

## Indications for Use Statement

The Rebar™ Micro Catheter is intended for the delivery of interventional devices or contrast media into the vasculature of the peripheral and neuro anatomy.

{4}------------------------------------------------

# Technical Comparison with Predicate Device:

A comparison of the Rebar™ Micro Catheter to the predicate device is provided in the table below:

| Attribute                             | Predicate: Rebar™ Micro Catheter<br>(K093750)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Subject: Rebar™<br>Micro Catheter                                                                                                                                                |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use<br>(IFU) Statement | The Rebar Micro Catheter is intended for the controlled<br>selective infusion of physician-specified therapeutic<br>agents or contrast media into the vasculature of the<br>peripheral and neuro anatomy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | The Rebar Micro<br>Catheter is intended<br>for the delivery of<br>interventional devices<br>or contrast media into<br>the vasculature of the<br>peripheral and neuro<br>anatomy. |
| Principles of<br>Operations           | The Rebar™ Micro Catheter is inserted within a guide<br>catheter that has already been advanced to the desired<br>location. The user connects a hemostatic side arm<br>adapter to the guide catheter to prevent backflow of<br>blood during insertion of the micro catheter. The user<br>also connects a one-way stopcock to the hemostatic<br>side arm adapter to continuously flush the guide<br>catheter with saline. The outer surface of the catheter<br>must be hydrated, and the catheter lumen must be<br>flushed with heparinized saline by attaching a saline<br>filled syringed to the catheter hub.<br>The user selects an appropriate steerable guidewire,<br>inserts it into the hub of the micro catheter and<br>advances the guidewire in the lumen. The stopcock is<br>closed, and the hemostatic valve loosened. The<br>guidewire and micro catheter are introduced as a unit<br>through the hemostatic port of the hemostatic sidearm<br>adapter into the lumen of the guiding catheter. The<br>guidewire/catheter assembly is advanced to the distal<br>tip of the guiding catheter or until the desired site has<br>been reached. The valve around the micro catheter is<br>tightened enough to prevent backflow but allowing<br>some movement through the valve by the micro<br>catheter. The stopcock is then opened. When ready to<br>infuse, the user withdraws the guidewire completely<br>from the micro catheter. A syringe with the desired<br>infusate is connected to the hub of the micro catheter.<br>The user infuses the site as required and then removes<br>and discards the micro catheter. | Same                                                                                                                                                                             |
| Proximal Tubing                       | Pebax with stainless steel helical coil                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                             |
| Distal Tubing                         | Adhesive, UV Cure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                             |
| Adhesive                              | Adhesive, Loctite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Dymax UV Adhesive                                                                                                                                                                |
| Attribute                             | Predicate: Rebar™ Micro Catheter<br>(K093750)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Subject: Rebar™<br>Micro Catheter                                                                                                                                                |
| Strain Relief                         | Inner: Elvax<br>Outer: Dynaflex                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                             |
| Inner lining                          | PTFE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                             |
| Outer Coating                         | Pebax                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                             |
| Hub Resin                             | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Trogamid                                                                                                                                                                         |
| Marker Bands                          | Platinum-Iridium Alloy                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                             |
| Usable Length                         | Rebar™ 18:<br>105-5081-130: 130 cm<br>105-5081-153: 153 cm<br>105-5083-153: 153 cm<br>Rebar™ 27:<br>105-5082-130: 130 cm<br>105-5082-145: 145 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                             |
| Proximal Section OD<br>(hub)          | Rebar™ 18:<br>2.7F – 0.035" Maximum<br>Rebar™ 27:<br>2.8F – 0.043" Maximum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                             |
| Distal Section OD (at<br>tip)         | Rebar™ 18:<br>2.4F – 0.037" Maximum<br>Rebar™ 27:<br>2.8F – 0.043" Maximum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                             |
| Distal Section ID (at<br>tip)         | Rebar™ 18:<br>0.020" Minimum<br>Rebar™ 27:<br>0.026" Minimum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                             |
| Carton                                | Natural, PTFE Tubing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                             |
| Pouch                                 | High-Density Polyethylene (HDPE)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                             |
| Shelf Life                            | 2 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                             |
| Sterilization                         | Ethylene Oxide (EO)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                             |
| Use                                   | Single Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                             |

{5}------------------------------------------------

# Performance Data- Bench

The following non-clinical bench testing was performed to evaluate the performance of the Rebar™ Micro Catheter. The passing results of the testing demonstrated that the changes do not raise new questions of safety and effectiveness, supporting the substantial equivalence to the predicate device.

| Bench Testing                                | Test Method Summary                                                                                                                 | Results Summary                                                                                               |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Safety Visual Inspection                     | Safety Visual Inspection: External surface of<br>catheter should be free from process and<br>surface defects and extraneous matter. | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for visual inspection.                            |
| Bench Testing                                | Test Method Summary                                                                                                                 | Results Summary                                                                                               |
| Catheter Hub Integrity A                     | Catheter Hub Integrity A: Catheter hub<br>transition section shall be approximately 6°<br>angle and length ≥ 0.5"                   | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Catheter Hub Integrity A.                     |
| Catheter Hub Integrity B                     | Catheter Hub Integrity B: Catheter shall be<br>made of a clear thermal plastic material<br>without internal ledges and voids        | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Catheter Hub Integrity B.                     |
| Catheter Hub Integrity C                     | Catheter Hub Integrity C: Per ISO 80369-<br>7:2018, the hub must meet ISO gauging<br>requirement of 6% luer taper                   | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Catheter Hub Integrity C.                     |
| Catheter Hub Integrity D<br>(Liquid Leak)    | Catheter Hub Integrity D (Liquid Leak): The<br>Rebar™ Micro Catheter was evaluated per<br>ISO 10555-1: 2013 Annex C                 | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Catheter Hub Integrity D<br>(Liquid Leak).    |
| Catheter Hub Integrity E (Air<br>Aspiration) | Catheter Hub Integrity E (Air Aspiration): The<br>Rebar™ Micro Catheter was evaluated per<br>ISO 10555-1: 2013 Annex D              | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Catheter Hub Integrity E<br>(Air Aspiration). |
| Catheter Pressurization<br>(Dynamic Burst)   | Catheter Pressurization Dynamic Burst:<br>Catheter shall withstand a<br>dynamic pressure of ≥ 700 psi without<br>rupturing/leaking  | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Catheter Pressurization<br>(Dynamic Burst).   |
| Catheter Pressurization<br>(Static Burst)    | Catheter Pressurization Static Burst:<br>Catheter shall withstand a static pressure of ≥<br>400 psi without rupturing/leaking       | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Catheter Pressurization<br>(Static Burst).    |
| ISO 80369-7:2016<br>Dimensional Testing      | ISO 80369-7:2018 Dimensional Testing: The<br>Rebar™ Micro Catheter was evaluated per<br>ISO 80369-7:2016                            | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for ISO 80369-7:2016<br>Dimensional Testing.      |
| ISO 80369-7:2016<br>Performance Testing      | ISO 80369-7:2018 Performance Testing: The<br>Rebar™ Micro Catheter was evaluated per<br>ISO 80369-7:2016                            | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for ISO 80369-7:2016<br>Performance Testing.      |
| Bench Testing                                | Test Method Summary                                                                                                                 | Results Summary                                                                                               |
| Tensile Strength                             | Tensile Strength: The Rebar™ Micro Catheter<br>was evaluated per ISO 10555-1: 2013                                                  | The Rebar™ Micro Catheter<br>met the acceptance criteria<br>for Tensile Strength.                             |

{6}------------------------------------------------

{7}------------------------------------------------

## Biocompatibility

Biocompatibility test results for subject Rebar Micro Catheter will be leveraged and adopted for the design change to the hub. The new hub material is Trogamid, which is found within other cleared Medtronic Neurovascular product lines, and is considered indirect patient contact. Biocompatibility data can be leveraged per ISO 10993-1 from the React™ 71 Catheter (K182101), as the modified hub has the same nature and duration of contact, has similar geometry, and is identical in formulation, processing, and sterilization. Slight dimensional changes that were made to the inside of the hub, outside of the hub, and luer thread are minor and do not impact biocompatibility data. Additionally, the new UV adhesive included in the manufacturing process for the hub is nonpatient contacting, thereby not impacting biocompatibility results of subject Rebar™ Micro Catheter. Both materials were found to be non-hemolytic and non-cytotoxic per Marathon™ Flow Directed Micro Catheter (K202318).

#### Performance Data- Animal:

Animal testing was not included in the subject submission. The differences in technological characteristics do not raise new questions of safety and effectiveness as demonstrated through non-clinical bench testing using wellestablished scientific methods.

## Performance Data - Clinical:

Clinical testing was not included in the subject submission. The differences in technological characteristics do not raise new questions of safety and effectiveness as demonstrated through non-clinical bench testing using wellestablished scientific methods.

#### Conclusion

The differences in technological characteristics do not raise new questions of safety and effectiveness as demonstrated through non-clinical bench testing using well-established scientific methods. The information provided in this submission supports the proposal of substantial equivalence for the Rebar™ Micro Catheter.

---

**Source:** [https://fda-staging.innolitics.com/device/K210114](https://fda-staging.innolitics.com/device/K210114)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).