Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines Administration Sets, Agilia Link, Agilia Duo
K210073 · Fresenius Kabi AG · FRN · Mar 1, 2022 · General Hospital
Device Facts
Record ID
K210073
Device Name
Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines Administration Sets, Agilia Link, Agilia Duo
Applicant
Fresenius Kabi AG
Product Code
FRN · General Hospital
Decision Date
Mar 1, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Agilia VP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids, medications, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, subcutaneous and intraosseous using dedicated administration sets. The Agilia VP Infusion System is also intended for neonatal care for the intermittent or continuous delivery of parenteral fluids for hydration and nutrition, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, and subcutaneous using dedicated administration sets. It is intended for use by trained healthcare professionals in healthcare facilities.
Device Story
Agilia VP Infusion System is a programmable, microcontroller-controlled volumetric linear peristaltic infusion pump. It delivers parenteral fluids, medications, and blood derivatives at programmed rates. Input involves manual programming by trained healthcare professionals in clinical settings. Device features include various infusion modes (continuous, intermittent, primary/secondary, KVO, bolus, loading dose, ramp-up/down, sequential) and mechanical sensors for air, occlusion, temperature, pressure, door status, and set installation. The system includes WiFi connectivity for server communication and is supported by accessories like the Agilia Link stacking rack and Agilia Duo power accessory. Output consists of controlled fluid delivery and system status/alarm notifications (e.g., battery, technical, infusion complete). The device benefits patients by providing precise, automated fluid and medication administration, reducing potential for manual dosing errors through integrated drug library software (Basic Profile). Healthcare providers use the pump interface to manage infusions, with the system providing clinical advisories and alarms to ensure safe operation.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including software verification and validation, human factors engineering (HFE/UE) studies, electrical safety (IEC 60601-1), EMC testing (IEC 60601-1-2), and cybersecurity assessment. Flow rate and bolus accuracy testing followed AAMI TIR101 2021.
Technological Characteristics
Volumetric linear peristaltic infusion pump; microcontroller-controlled. Dimensions: 5.3 x 7.5 x 6.7 inches; Weight: ~4.4 lbs. Power: 100-240 VAC. Connectivity: WiFi-enabled. Sensors: air, occlusion, temperature, pressure, door, set installation. Standards: IEC 60601-1 (3rd ed), IEC 60601-1-8, IEC 60601-1-2. Software: Basic Profile drug library within pump.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring intermittent or continuous delivery of parenteral fluids, medications, blood, and blood derivatives via intravenous, intra-arterial, subcutaneous, or intraosseous routes in healthcare facilities.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 1, 2022
Fresenis Kabi AG % Keith Dunn Director Regulatory Affairs Fresenius Kabi LLC, USA 3 Corporate Dr Suite 300 Lake Zurich, Illinois 60047
Re: K210073
Trade/Device Name: Agilia VP Infusion System, Agilia VP MC WIFI Infusion Pump, Volumat Lines Administration Sets, Agilia Link, Agilia Duo Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: FRN, MRZ, FPA Dated: January 8, 2021 Received: January 12, 2021
Dear Keith Dunn:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Courtney H. Lias, Ph.D. Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210073
Device Name Agilia VP Infusion System
#### Indications for Use (Describe)
The Agilia VP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids, medications, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, subcutaneous and intraosseous using dedicated administration sets.
The Agilia VP Infusion System is also intended for neonatal care for the intermittent or continuous delivery of parenteral fluids for hydration and nutrition, blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, and subcutaneous using dedicated administration sets.
It is intended for use by trained healthcare professionals in healthcare facilities.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of a blue abstract symbol on the left, resembling three vertical wavy lines. To the right of the symbol, the words "FRESENIUS" and "KABI" are stacked on top of each other, also in blue. The text is in a bold, sans-serif font.
# 510(k) SUMMARY
#### Date Prepared
February 28, 2022
#### Owner/Operator
Fresenius Kabi AG 61346 Bad Homburg Germany
Owner/Operator Number: 9027285
#### Contact Person
## Keith Dunn
Director, Regulatory Affairs Fresenius Kabi USA, LLC Three Corporate Drive, 2nd Floor Lake Zurich, IL 60047 USA
Phone: 224-817-2430 Fax: 847 550 2960 E-mail: keith.dunn(@fresenius-kabi.com
## Secondary Contact Person
#### Jason Ma
Sr. Manager, Regulatory Affairs Fresenius Kabi USA, LLC Three Corporate Drive, 2nd Floor Lake Zurich, IL 60047 USA
Phone: 224-817-4100 847 550 2960 Fax: E-mail: Jason.ma01(@fresenius-kabi.com
#### Device Trade Name
Agilia VP Infusion System
#### Common Name/Usual Name:
Infusion Pump and Accessories
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Fresenius Kabi logo. The logo consists of a blue graphic on the left and the words "FRESENIUS KABI" on the right. The graphic is made up of three blue, curved lines. The words "FRESENIUS KABI" are in blue, with "FRESENIUS" on top and "KABI" on the bottom.
## Classification Name
21 CR 880.5725 Infusion Pump
## Product Code and Classification Panel
The Agilia VP Infusion System (pump, set and accessories) has been classified as Class II under 21 CFR 880.5725, and reviewed by the General Hospital Devices Panel (80). For reference, the product code and classification of each device subsystem is identified in Table 1.
| Device Name | Regulation Number | Class | Description | Panel | Product<br>Code |
|-----------------------------------------------|-------------------|-------|---------------------------|-------|-----------------|
| Agilia VP MC WiFi Volumetric<br>Infusion Pump | 21 CFR 880.5725 | II | Infusion Pump | 80 | FRN |
| Agilia Link | 21 CFR 880.5725 | II | Infusion pump accessories | 80 | MRZ |
| Agilia Duo | 21 CFR 880.5725 | II | Infusion pump accessories | 80 | MRZ |
| Agilia USB Cable | 21 CFR 880.5725 | II | Infusion pump accessories | 80 | MRZ |
| Volumat Line administration<br>sets | 21 CFR 880.5725 | II | Infusion pump accessories | 80 | FPA |
## Table 1: Device Classification
## Legally Marketed Device Under Which Substantial Equivalence is Being Claimed
Agilia Infusion System (K121613)
## Device Description
The Agilia VP Infusion System includes the Agilia VP MC WiFi Volumetric Infusion Pump, which is a programmable electronic medical system dedicated to administering a pre-determined volume of an infusion product at a programmed rate, in combination with Volumat Line administration sets and optional accessories. The optional accessories are identified as follows:
- Agilia Link 4, 6 and 8 Stacking rack system intended to power and organize 4, 6 or ● 8 pumps at the patient bedside.
- Agilia Duo two-channel accessory designed to power two Agilia infusion pumps. ●
- Agilia USB Cable - intended to connect the Agilia VP infusion pump to a PC for serial communication.
## Statement of Intended Use/Indications For Use
The Agilia VP Infusion System is intended for adult and pediatric care for the intermittent or continuous delivery of parenteral fluids. medications, blood and blood derivatives through clinically accepted parenteral routes of administration. These routes
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Fresenius Kabi. The logo consists of a blue graphic on the left, which appears to be three stylized vertical lines. To the right of the graphic are the words "FRESENIUS KABI" in blue, with "FRESENIUS" on top of "KABI". There is a small subscript number 1 next to the letter S in Fresenius.
of administration include intravenous, intra-arterial, subcutaneous and intraosseous using dedicated administration sets.
The Agilia VP Infusion System is also intended for neonatal care for the intermittent or continuous delivery of parenteral fluids for hydration and nutrition, blood and blood derivatives through clinically accepted parenteral routes of administration. These routes of administration include intravenous, intra-arterial, and subcutaneous using dedicated administration sets.
It is intended for use by trained healthcare professionals in healthcare facilities.
#### Technological Comparison as Compared to the Predicate Device
A comparison between the predicate device and the subject device is provided in the Table 2.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Fresenius Kabi logo. The logo consists of a blue abstract symbol on the left and the words "FRESENIUS KABI" in blue on the right. The abstract symbol is made up of three vertical wavy lines.
| Characteristic | Predicate Device | Subject Device | Comment |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K121613 | K210073 | |
| Intended Use /<br>Indications for<br>Use | The Agilia Infusion System is<br>a transportable equipment<br>intended for use by trained<br>healthcare professionals in<br>healthcare facilities and<br>homecare environments on<br>adults, pediatrics and<br>neonate's human patients to<br>administer via a single<br>channel or mounted on a<br>multiple channels rack<br>accessory:<br>• Intermittent or continuous<br>delivery of parenteral fluids<br>(solutions, colloids,<br>parenteral nutrition) and<br>medications (including but<br>not limited to diluted drugs,<br>chemotherapy) through<br>clinically accepted IV routes<br>of administration.<br>• Transfusion of blood and<br>blood derivatives products. | The Agilia VP Infusion<br>System is intended for adult<br>and pediatric care for the<br>intermittent or continuous<br>delivery of parenteral fluids,<br>medications, blood and blood<br>derivatives through clinically<br>accepted parenteral routes of<br>administration. These routes<br>of administration include<br>intravenous, intra-arterial,<br>subcutaneous and<br>intraosseous using dedicated<br>administration sets.<br>The Agilia VP Infusion<br>System is also intended for<br>neonatal care for the<br>intermittent or continuous<br>delivery of parenteral fluids<br>for hydration and nutrition,<br>blood and blood derivatives<br>through clinically accepted<br>parenteral routes of<br>administration. These routes<br>of administration include<br>intravenous, intra-arterial,<br>and subcutaneous using<br>dedicated administration sets.<br>It is intended for use by<br>trained healthcare<br>professionals in healthcare<br>facilities. | Similar<br>The intended use<br>between the predicate<br>and proposed device<br>are the same in terms<br>of intended patient,<br>route of<br>administrations, and<br>treatment. Per FDA<br>guidance, the<br>statement was revised<br>to specifically name<br>the intended routes of<br>administration and to<br>differentiate the<br>intended patient<br>population . |
| Type of Pump | Volumetric, Linear Peristaltic<br>Pump | Volumetric, Linear Peristaltic<br>Pump | Same |
| Patient<br>Population | Adults, pediatrics, neonates | Adults, pediatrics, neonates | Same |
| Delivery<br>Modes | Continuous, Intermittent,<br>Primary, Secondary, KVO,<br>Bolus (Manual and<br>Programmable), Loading<br>Dose | Continuous, Intermittent,<br>Primary, Secondary, KVO,<br>Bolus (Manual and<br>Programmable), Loading<br>Dose, Ramp-up / Ramp<br>down, Sequential | Similar<br>Performance testing<br>demonstrates the<br>additional modes do<br>not introduce any new<br>issues of safety or<br>effectiveness. |
| Characteristic | Predicate Device | Subject Device | Comment |
| | K121613 | K210073 | |
| Infusion<br>Accuracy | $\pm$ 5% | $\pm$ 5% under most conditions | Similar<br>Detailed flow rate<br>accuracy disclosed in<br>the labeling |
| WIFI | No | Yes | Similar<br>Performance testing<br>demonstrates the<br>addition of WIFI<br>functionality does not<br>introduce any new<br>issues of safety or<br>effectiveness. |
| Pump Control | Microcontroller controlled | Microcontroller controlled | Same |
| Number of<br>Channels | 1 | 1 | Same |
| Mechanical<br>Sensors | air, occlusion, temperature,<br>pressure, door, set installation | air, occlusion, temperature,<br>pressure, door, set installation | Same |
| Barcoding | No | No | Same |
| Automated<br>Programming | No | No | Same |
| Manual<br>Programming | Yes | Yes | Same |
| Piggy back /<br>secondary | Yes | Yes | Same |
| Alarms | Air, occlusion, temperature,<br>system malfunction,<br>set installation, door,<br>low drug reservoir volume,<br>infusion complete, battery,<br>technical | Air, occlusion, temperature,<br>system malfunction,<br>set installation, door,<br>low drug reservoir volume,<br>infusion complete, battery,<br>technical | Same |
| Clinical<br>Advisories | Yes | Yes | Same |
| Infusion<br>Modes | Basic, continuous,<br>intermittent, primary /<br>secondary, KVO, Bolus<br>(direct and programmed),<br>Loading Dose | Basic, continuous,<br>intermittent, primary /<br>secondary, KVO, Bolus<br>(direct and programmed),<br>Loading Dose, Ramp-up /<br>Ramp Down, Sequential | Similar<br>Performance testing<br>demonstrates the<br>additional infusion<br>modes do not<br>introduce any new<br>issues of safety or<br>effectiveness. |
| Characteristic | Predicate Device | Subject Device | Comment |
| | K121613 | K210073 | |
| Features | Keypad lock | Keypad lock/automatic lock,<br>occlusion auto restart | Similar<br>Performance testing<br>demonstrates the<br>additional keypad<br>features do not<br>introduce any new<br>issues of safety or<br>effectiveness. |
| Dose Error<br>Reduction<br>Software /<br>Drug Library<br>Software<br>Compatibility | Basic Profile (within pump) | Basic Profile (within pump) | Same |
| Wireless<br>Server<br>Software | No | Yes | Similar<br>Performance testing<br>demonstrates the<br>addition of wireless<br>server software does<br>not introduce any new<br>issues of safety or<br>effectiveness. |
| Administration<br>Sets | Volumat Line administration<br>sets | Volumat Line administration<br>sets | Same |
| Racking /<br>Organization /<br>Power | Link+ Agilia | Agilia Link | Similar<br>Simpler, non-<br>communicating<br>accessory for proposed<br>device |
| | Agilia Duo | Agilia Duo | Same |
| Electrical<br>Safety | IEC 60601-1: 1998 +A1:1995<br>(2nd edition)<br>IEC 60601-1-8: 2003<br>+A1:2006<br>IEC 60601-1-2: 2007<br>ANSI/AAMI ID26:<br>2004/(R)2009 | IEC 60601-1: 2005<br>+A1:2012 (3rd edition)<br>IEC 60601-1-8: 2012<br>IEC 60601-1-2: 2014<br>ANSI/AAMI ID26:<br>2004/(R)2009 | Similar<br>Performance testing<br>demonstrates the<br>application of state-of-<br>the-art standards does<br>not introduce any new<br>issues of safety or<br>effectiveness. |
| Mechanical /<br>Power<br>Specifications | Pump Size: 5.3 x 7.5 x 6.7<br>inches<br>Pump Weight: App. 4.4 lbs.<br>Line Power: 100-240 VAC | Pump Size: 5.3 x 7.5 x 6.7<br>inches<br>Pump Weight: App. 4.4 lbs.<br>Line Power: 100-240 VAC | Same |
| Operating<br>Environment | Temperature: 41° F to 104° F<br>Relative Humidity: 20-90% | Temperature: 41° F to 104° F<br>Relative Humidity: 20-90% | Same |
| Characteristic | Predicate Device<br>K121613 | Subject Device<br>K210073 | Comment…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.