← Product Code [FMI](/productcode/FMI) · K210059

# Promisemed Insulin Pen Needle (K210059)

_Promisemed Hangzhou Meditech Co., Ltd. · FMI · Jul 21, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K210059

## Device Facts

- **Applicant:** Promisemed Hangzhou Meditech Co., Ltd.
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Jul 21, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Promisemed Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. It is sterile and for single use only. It is suitable for all age groups including neonate, infant, children and adult, and can be used by the patient at home or healthcare professionals at medical/health care centers.

## Device Story

Promisemed Insulin Pen Needle is a sterile, single-use, hypodermic needle designed for subcutaneous insulin delivery via pen injectors. The device consists of a needle tube, hub, container, shield, UV glue, and silicone lubricant. Operation is manual: the user attaches the needle to a pen injector and inserts it into the skin. The silicone lubricant facilitates needle insertion and rubber septum penetration. The device is intended for use by patients at home or by healthcare professionals in clinical settings. It is available in various gauge (29G-34G) and length (3.5mm-12mm) configurations. The device benefits patients by providing a sterile, disposable means for insulin administration.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified via ISO 9626, ISO 11608-2, and ISO 7864 testing. Biocompatibility testing (ISO 10993 series) confirmed no cytotoxicity, irritation, sensitization, or systemic toxicity. Hemolysis testing (ASTM F756-13) and particulate matter testing (USP <788>) met acceptance criteria. Package integrity and shelf-life (ASTM F1980-16) were validated.

## Technological Characteristics

Materials: X5CrNi18-10 stainless steel (tube), Polypropylene (hub/container), Polyethylene (shield). Lubricant: Silicone oil. Joint: UV glue. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Standards: ISO 7864, ISO 9626, ISO 11608-2, ISO 10993, ASTM F756-13, ASTM F88/F88-15, ASTM F1929-15, ASTM F1980-16. Dimensions: 29G-34G; 3.5mm-12mm lengths.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- Verifine Common Type Insulin Pen Needle ([K161950](/device/K161950.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the agency name in a sans-serif font.

July 21, 2021 Promisemed Hangzhou Meditech Co., Ltd. % Wei-Shan Hsu Regulatory Manager Vee Care (Asia) Limited 17th Chung Pont Commercial Building, 300 Hennessy Road Hong Kong, Hong Kong China

Re: K210059

Trade/Device Name: Promisemed Insulin Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Received: June 21, 2021

Dear Wei-Shan Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210059

Device Name Promisemed Insulin Pen Needle

### Indications for Use (Describe)

Promisemed Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. It is sterile and for single use only. It is suitable for all age groups including neonate, infant, children and adult, and can be used by the patient at home or healthcare professionals at medical/health care centers.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### 1 Date Prepared

July 21, 2021

### Submitter's Information 2

### Name of Sponsor:

Promisemed Hangzhou Meditech Co., Ltd.

# Address:

No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou

City, 311121 Zhejiang, China

# Contact Name:

Zearou Yang, Regulatory Affairs Manager

Telephone No.:

+86 571 88772985

### Fax No.:

+86 571 88772985

Email Address:

zearou.yang@promisemed.ca

#### ന Trade Name, Common Name, Classification

Trade/Product Name: Promisemed Insulin Pen Needle Common Name: Insulin Pen Needle Classification name: Needle, Hypodermic, Single Lumen Regulation Number: 21 CFR 880.5570 Device Class: Class II Product Code: FMI

- 4 Identification of Predicate Device(s) K161950: Verifine Common Type Insulin Pen Needle

{4}------------------------------------------------

### ഹ Description of the Device

Promisemed Insulin Pen Needle is manufactured by Promisemed Hangzhou Meditech Co., Ltd, which is designed for use with a pen injector for the subcutaneous injection of insulin. The user proceeds with inserting the needle into the skin manually.

The Promisemed Insulin Pen Needle consists of needle container, needle shield, needle tube, needle hub, UV glue and silicone oil. UV glue is used to glue needle tube and needle hub and the silicone oil is used to needle tube lubrication. Promisemed Insulin Pen Needle is the modification of the Verifine Common Type Insulin Pen Needle cleared in K161950 in extension of the range of gauge and needle length as well as change in material of needle shield. Promisemed Insulin Pen Needle is sterility Assurance Level (SAL) of 10 %, non-pyrogenic and single-use devices. It is supplied with several models. Different models are distinguished by needle gauge and length.

- 6 Indications for Use
Promisemed Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. It is sterile and for single use only. It is suitable for all age groups including neonate, infant, children and adult, and can be used by the patient at home or healthcare professionals at medical/health care centers.

- Similarities and Differences of the Proposed Device to the Predicate 7 Device
Promisemed Insulin Pen Needle is substantially equivalent in its technologies and functions to the Verifine Common Type Insulin Pen Needle which cleared under premarket notification number K161950.

Three design changes were made to the subject device, introducion of new gauge specification (34G), new needle length specification(3.5mm and 10mm) and Needle Shield material change. The following are comparisons between subject device and the predicate device.

A detailed comparison to the predicate is provided in Table 1.

|            | Subject Device       |         |     | Predicate Device (K161950)     |        |      |          |
|------------|----------------------|---------|-----|--------------------------------|--------|------|----------|
| Trade Name | Promisemed<br>Needle | Insulin | Pen | Verifine<br>Insulin Pen Needle | Common | Type | Comments |

{5}------------------------------------------------

| Manufacturer                            | Promisemed<br>Meditech Co., Ltd<br>Hangzhou                                                                                                                                                                                                                                                                                                                                      | Promisemed<br>Meditech Co., Ltd<br>Hangzhou                                                                                                                                                                                                           |                                                                                                                                                                                                                            |                  | Needle shield: Polyethylene<br>(PE)                                                                                                                                                                                                                                                                                                                                      | Needle shield:<br>Polypropylene (PP)                                                                                                                                                                                                                                                                                                                                         | Different<br>Both PE and PP belong<br>are commonly used<br>thermoplastic material.<br>This difference does not<br>affect the substantially<br>equivalence on the safety<br>and effectiveness. |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class                            | Class II                                                                                                                                                                                                                                                                                                                                                                         | Class II                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                       |                  | Joint medium: UV glue                                                                                                                                                                                                                                                                                                                                                    | Joint medium: UV glue                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                          |
| Product Code                            | FMI                                                                                                                                                                                                                                                                                                                                                                              | FMI                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                       |                  | Lubricant: Silicon oil                                                                                                                                                                                                                                                                                                                                                   | Lubricant: Silicon oil                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                          |
| Regulation<br>number                    | 880.5570                                                                                                                                                                                                                                                                                                                                                                         | 880.5570                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                       | Performance      | Seal: Dialyzer paper<br>Complied with ISO 7864, ISO<br>9626, ISO 11608-2                                                                                                                                                                                                                                                                                                 | Seal: Dialyzer paper<br>Complied with ISO 7864,<br>ISO 9626, ISO 11608-2                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                          |
| Regulation<br>Name                      | Needle, Hypodermic, Single<br>Lumen                                                                                                                                                                                                                                                                                                                                              | Needle, Hypodermic, Single<br>Lumen                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                       | Sterilization    | EO Sterilization                                                                                                                                                                                                                                                                                                                                                         | EO Sterilization                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                          |
| Intended Use/<br>Indications for<br>Use | Promisemed Insulin Pen<br>Needle is intended for use<br>with pen injector device for<br>subcutaneous injection of<br>insulin.<br>It is sterile and for single use<br>only. It is suitable for all age<br>groups including neonate,<br>infant, children and adult,<br>and can be used by the<br>patient at home or<br>healthcare professionals at<br>medical/health care centers. | The Common Type Insulin<br>Pen Needle is intended for<br>use with pen injector<br>device for subcutaneous<br>injection of insulin.                                                                                                                    | Different<br>The indications for Use is<br>similar except that the<br>subject device<br>specifically identifies<br>intended population,<br>sterile and single use,<br>and for home or<br>healthcare/health care<br>centers |                  | SAL:10-6                                                                                                                                                                                                                                                                                                                                                                 | SAL:10-6                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                          |
| Operating<br>Principle                  | The user proceeds with<br>inserting the needle into the<br>skin manually. The patient<br>end and the cartridge end of<br>the tube are lubricated using<br>a silicone based lubricant for<br>ease of injection and rubber<br>septum penetration.                                                                                                                                  | The user proceeds with<br>inserting the needle into<br>the skin manually. The<br>patient end and the<br>cartridge end of the tube<br>are lubricated using a<br>silicone based lubricant for<br>ease of injection and<br>rubber septum<br>penetration. | Same                                                                                                                                                                                                                       | Shelf Life       | 5 years                                                                                                                                                                                                                                                                                                                                                                  | 5 years                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                          |
| Gauge                                   | 29G, 30G, 31G, 32G,33G,<br>34G                                                                                                                                                                                                                                                                                                                                                   | 29G, 30G, 31G, 32G,33G                                                                                                                                                                                                                                | Different<br>New gauge specification<br>(34G) is introduced. This<br>difference does not<br>affect the substantially<br>equivalence on the safety<br>and effectiveness                                                     | Single use       | Yes                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                          |
| Needle Length                           | 3.5mm±0.4mm,<br>(4mm, 5mm, 6mm, 8mm,<br>10mm, 12mm) ±1.2mm                                                                                                                                                                                                                                                                                                                       | (4mm,5mm,6mm,8mm,<br>12mm) ±1.2mm                                                                                                                                                                                                                     | Different<br>Needle length of 3.5 mm<br>and 10mm were newly<br>introduced specifications.<br>This difference does not<br>affect the substantially<br>equivalence on the safety<br>and effectiveness.                       | Biocompatibility | Complied with ISO10993<br>series standards, and the<br>following tests are<br>performed<br>- Cytotoxicity: No<br>cytotoxicity<br>- Skin Irritation: No evidence<br>of skin irritation<br>- Skin Sensitization: No<br>evidence of sensitization<br>-Acute Systemic Toxicity:No<br>systemic toxicity<br>-Hemolysis: No evidence of<br>hemolysis<br>-Pyrogen: Non-pyrogenic | Complied with ISO10993<br>series standards, and the<br>following tests are<br>performed<br>- Cytotoxicity: No<br>cytotoxicity<br>- Skin Irritation: No<br>evidence of skin irritation<br>- Skin Sensitization: No<br>evidence of sensitization<br>-Acute Systemic<br>Toxicity:No systemic<br>toxicity<br>-Hemolysis: No evidence of<br>hemolysis<br>-Pyrogen: Non- pyrogenic | Same                                                                                                                                                                                          |
| Material                                | Needle Tube: X5CrNi18-10                                                                                                                                                                                                                                                                                                                                                         | Needle Tube: X5CrNi18-10                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                       |                  |                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                               |
|                                         | Needle Hub: Polypropylene<br>(PP)                                                                                                                                                                                                                                                                                                                                                | Needle Hub: Polypropylene<br>(PP)                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                       |                  |                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                               |
|                                         | Needle container:<br>Polypropylene (PP)                                                                                                                                                                                                                                                                                                                                          | Needle container:<br>Polypropylene (PP)                                                                                                                                                                                                               | Same                                                                                                                                                                                                                       |                  |                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                               |

{6}------------------------------------------------

### Discussion:

The Promisemed Insulin Pen Needle has the same intended use and technological characteristics as the predicate device. The basic technological and operating principles are the same for both devices. Both the subject and predicate devices are disposable, sterile, single patient use devices.

The difference in gauge size and material of needle shield does not affect the substantially equivalence on the safety and effectiveness. The Needle length of 3.5mm with tightened tolerance limits (±0.4mm), is within the same needle length range as the predicate 4mm (±1.20 mm), both meet the requirement of ISO 11608-2 to a 95%/95% C/R and substantially equivalent. The subject device has raised no different questions of safety and effectiveness.

{7}------------------------------------------------

#### 8 Performance Testing Summary

The bench testing performed verifies that the performance of the subject devices is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:

- · ISO 9626: Stainless steel needle tubing for the manufacture of medical devices -Requirements and test methods
- · ISO11608-2: Needle-based injection systems for medical use Requirements and test methods — Part 2: Needles
- ISO 7864: Sterile hypodermic needles for single use Requirements and test methods
- · Biocompatibility
	- a. ISO 10993-1:2009 Biological Evaluation of Medical Devices -- Part 1: Evaluation and testing within a risk management process
	- b. ISO 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5: Tests for in Vitro Cytotoxicity
	- c. ISO 10993-10:2010 Biological Evaluation of Medical Devices -- Part 10: Tests for Irritation and Skin Sensitization
	- d. ISO 10993-11:2006, Biological evaluation of medical devices Part 11: Tests for systemic toxicity-Acute systemic toxicity and pyrogen test
	- f. ASTM F756-13 Standard Practice for Assessment of Hemolytic Properties of Materials.
	- g. Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
	- h. Package integrity testing, after environmental conditioning and simulated transportation in accordance with ISTA 3A, was conducted on the final, packaged, and sterile devices. All packaging deemed acceptable for protection of product and sterility maintenance.
	- i. Sterile Barrier Packaging Testing performed on the proposed device: Seal strength ASTM F88/F88-15, Dye penetration ASTM F1929-15
	- i. Shelf life of 5 years is validated using the FDA recognized standard ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

{8}------------------------------------------------

#### Conclusion 9

The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The Promisemed Insulin Pen Needle is substancially equivalent to the Verifine Common Type Insulin Pen Needle with respect to the Indications for Use, target populations, treatment method, and technological characteristics.

---

**Source:** [https://fda-staging.innolitics.com/device/K210059](https://fda-staging.innolitics.com/device/K210059)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
