← Product Code [HKY](/productcode/HKY) · K203850

# CATS®-D Tonometer Prism (K203850)

_Cats Tonometer, LLC · HKY · Jun 14, 2021 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K203850

## Device Facts

- **Applicant:** Cats Tonometer, LLC
- **Product Code:** [HKY](/productcode/HKY.md)
- **Decision Date:** Jun 14, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1930
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The CATS®-D Tonometer Prism is intended to be used with Goldmann type tonometers for the measurement of intraocular pressure of the human eye.

## Device Story

Optical image prism for Goldmann applanation tonometers; facilitates intraocular pressure (IOP) measurement. Device contacts cornea; transforms applanation force into IOP reading (1.0g = 10mmHg). Used in clinical settings by eye care professionals. Disposable design features detachable tip; snaps onto prism housing. Replaces adhesive-based reusable predicate. Benefits include reduced cross-contamination risk via single-use, sterile delivery.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization, irritation). Device confirmed non-cytotoxic, non-sensitizing, and non-irritating. Design verification and sterilization validation completed.

## Technological Characteristics

Material: PMMA. Dimensions: 28.75mm length, 6.5mm contact diameter. Optical: 30-degree bi-prism angle, 1.4906 refractive index. Connectivity: None (mechanical optical accessory). Sterilization: Ethylene Oxide. Form factor: Disposable prism with detachable tip.

## Regulatory Identification

A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.

## Predicate Devices

- CATS® Reusable Tonometer prism ([K173904](/device/K173904.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

CATS Tonometer LLC Paul Kramsky Consultant 1517 N. Wilmot Suite 143 Tucson, Arizona 85712

Re: K203850

Trade/Device Name: CATS®-D Tonometer Prism Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and accessories Regulatory Class: Class II Product Code: HKY Dated: May 4, 2021 Received: May 6, 2021

Dear Paul Kramsky:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203850

Device Name

CATS®-D Tonometer Prism

Indications for Use (Describe)

The CATS®-D Tonometer Prism is intended to be used with Goldmann type tonometers for the measurement of intraocular pressure of the human eye.

| Type of Use (Select one or both, as applicable)                                                                                                   |                                                                                                                                                  |
|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;">     Prescription Use (Part 21 CFR 801 Subpart D)   </span> </span> | <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;">     Over-The-Counter Use (21 CFR 801 Subpart C)   </span> </span> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

### 510(k) Summary for the CATS®-D Tonometer Prism

#### 5.1 Submission Sponsor:

CATS Tonometer, LLC 1517 N. Wilmot, Suite 143 Tucson, Arizona 85712 Telephone: 520- 975-4550

#### 5.2 Contact:

CATS Tonometer, LLC 1517 N. Wilmot, Suite 143 Tucson, Arizona 85712 Telephone: 520- 975-4550 Contact: Nannon Roosa, President Email: nannon(@catsiop.com

#### 5.3 Date Prepared:

June 8, 2021

#### 5.4 Device Name:

| Trade/Proprietary Name: | CATS®-D Tonometer Prism   |
|-------------------------|---------------------------|
| Classification Name:    | Tonometer and Accessories |
| Product Code:           | HKY                       |
| Regulation Number:      | 21 CFR 886.1930           |

#### ર્સ્ટ Predicate Devices:

The CATS®-D Tonometer Prism is substantially equivalent in terms of intended use and technological characteristics to the CATS® Reusable Tonometer prism (K173904).

#### ર્ડ.6 Device Description:

The CATS®-D Tonometer Prism is used as an optical image prism for Goldmann applanation style tonometers. It is made of PMMA, the corneal contact diameter is 6.5 mm and the total length of the prism is 30 mm.

#### 5.7 Indications for Use:

The CATS® Tonometer Prism is intended to be used with Goldmann type tonometers for the measurement of intraocular pressure of the human eye.

#### 5.8 Performance Data:

Design verification and sterilization validation and shelf-life testing were performed. Cytotoxicity (MEM Elution), Sensitization (Kligman Maximization Test), and Irritation (Intracutaneous irritation test) were performed per ISO 10993-5 and ISO 10993-10. The device was reported to be non-cytotoxic, non-sensitizing and non-irritating.

{4}------------------------------------------------

#### 5.9 Substantial Equivalence:

The CATS®-D Tonometer Prism is substantially equivalent to the 510(k) cleared CATS® reusable prism with the exception being that the disposable prism is supplied with a sterilized detachable, which snaps on to the prism housing, whereas the 510(k) cleared reusable prism was connected to the body with adhesive. This change resulted in a very slight decrease in overall length of the prism, which does not affect the safety and effectiveness of the device. In addition, the packaging was changed to a cardboard box consisting of a PETG tray with Tyvek lid, and sterilization by Ethylene Oxide. The claim of substantial equivalence to the previously cleared CATS® Reusable Tonometer Prism (K173904) is supported by the following comparison of characteristics in Table 1 and non-clinical performance data. Therefore, the CATS®-D Tonometer Prism is substantially equivalent to the predicate device.

| Characteristic              | CATS®-D Tonometer<br>Prism                    | CATS® Reusable<br>Tonometer Prism<br>(Predicate) |
|-----------------------------|-----------------------------------------------|--------------------------------------------------|
| Overall length              | 28.75 mm                                      | 30.0 mm                                          |
| Base diameter               | 10.9 mm                                       | 10.9 mm                                          |
| Contact surface diameter    | 6.5 mm                                        | 7.0 mm                                           |
| Bi-Prism diameter           | 6.0 mm                                        | 6.0 mm                                           |
| Bi-Prism length             | 10.67 mm                                      | 10.67 mm                                         |
| Applanation Diameter        | 3.29 ± 0.02                                   | 3.29 ± 0.02                                      |
| Bi-prism angle              | 30 degrees                                    | 30 degrees                                       |
| Index of refraction         | 1.4906                                        | 1.4906                                           |
| Prism construction material | PMMA                                          | PMMA                                             |
| Prism component adhesive    | Class 9                                       | Class 9                                          |
| Prism contact surface       | Concave-convex<br>contoured                   | Concave-convex<br>contoured                      |
| Prism surface finish        | Optical polish                                | Optical polish                                   |
| Force to IOP conversion     | 1.0g = 10mm Hg                                | 1.0g = 10mm Hg                                   |
| Prism tip                   | Detachable                                    | Connected to body                                |
| Packaging                   | 4 PETG trays w/Tyvek<br>lids in cardboard box | Cardboard box                                    |
| Sterilization               | Ethylene Oxide                                | Non-sterile                                      |

## Table 1 - Comparison of CATS® Disposable and CATS® Reusable (Predicate) Tonometer Prisms

---

**Source:** [https://fda-staging.innolitics.com/device/K203850](https://fda-staging.innolitics.com/device/K203850)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
