The Pulse Oximeter is a non-invasive device intended for spot checking and/or continuous data collection of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in home environments and clinical institutions except acute clinical environment.
Device Story
Portable pulse oximeter; measures SpO2 and pulse rate via red (660 nm) and infrared (940 nm) light absorption through perfused finger tissue; utilizes MCU and measurement circuits to process optical signals; powered by internal lithium rechargeable battery; intended for adult patients in home or clinical settings (excluding acute care); operated by patient or clinician; data transmitted via Bluetooth; output displayed numerically; assists clinicians in monitoring oxygenation status; benefits patients through non-invasive, continuous or spot-check monitoring.
Clinical Evidence
Controlled desaturation study conducted on 12 healthy adult subjects (light-to-dark-skinned) under induced hypoxia. Subject device SpO2 accuracy compared to CO-oximeter (control). Results: Arms of 1.57% for 70-100% SpO2 range, meeting ISO 80601-2-61 requirements (≤2%). No adverse effects reported.
Technological Characteristics
Materials: PC (enclosure), Silicone (finger sleeve). Sensing: Red (660 nm) and IR (940 nm) LEDs. Power: Internal lithium rechargeable battery. Connectivity: Bluetooth. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61, ISO 10993-1/5/10. Software: Moderate level of concern.
Indications for Use
Indicated for adult patients for non-invasive spot checking and/or continuous monitoring of functional oxygen saturation (SpO2) and pulse rate (PR) in home and clinical settings, excluding acute clinical environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
November 8, 2022
Shenzhen Viatom Technology Co., Ltd. Weifeng Tao Regulatory Manager 4E, Building 3, Tingwei Industrial Park, No.6 Liufang Road Block 67, Xin'an Street, Baoan District, 5 Shenzhen, Guangdong 518101 China
Re: K203812
Trade/Device Name: Oxyfit Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: November 7, 2022 Received: November 7, 2022
Dear Weifeng Tao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# James J. Lee -S
James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203812
Device Name Pulse Oximeter
Indications for Use (Describe)
The Pulse Oximeter is a non-invasive device intended for spot checking and/or continuous data collection of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in home environments and clinical institutions except acute clinical environment.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <span style="font-family: Symbol;">×</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span>□</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K203812
This summary of 510(k) information is submitted as required by requirements of SMDA and21 CFR §807.92.
#### 1. Submission sponsor
| Submission Date | Dec.18.2020 |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Name | Shenzhen Viatom Technology Co., Ltd. |
| Address | 4E, Building 3, Tingwei Industrial Park, No.6 Liufang Road, Block 67, Xin'an Street, Baoan District, Shenzhen, 518101, Guangdong, China |
| Contact person | Zhou Saixin |
| Title | General Manager |
| E-mail | zhousaixin@viatomtech.com |
| Tel | +86-0755-86638929 |
#### 2. Submission correspondent
| Name | Shenzhen Viatom Technology Co., Ltd. |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Address | 4E, Building 3, Tingwei Industrial Park, No.6 Liufang Road, Block 67, Xin'an Street, Baoan District, Shenzhen, 518101, Guangdong, China |
| Contact person | Mr.Weifeng Tao |
| E-mail | taoweifeng@viatomtech.com |
#### 3. Subject Device Information
| Trade/Device Name | Oxyfit Pulse Oximeter |
|-------------------|-----------------------------|
| Model | PO6,PO6A |
| Common name | Pulse Oximeter |
| Regulatory Class | Class II |
| Classification | 21CFR 870.2700/Oximeter/DQA |
| Submission type | Traditional 510(K) |
{4}------------------------------------------------
#### 4. Predicate Device Information 4.1Primary Predicate Device
510(k) Number: K200414 Trade/ Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Manufacturer: Shenzhen Aeon Technology Co., Ltd.
#### 4.2Reference Predicate Device
510(k) Number: K150869 Trade/ Device Name: Checkme Pro Health Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: II Product Code: MWI Manufacturer: Shenzhen Viatom Technology Co., Ltd.
### 5. Device Descriptions
The Pulse Oximeter is a non-invasive device intended for spot checking and/or continuous data collection of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in home environments and clinical institutions except acute clinical environment.
The Pulse Oximeter has two models.Model PO6 and Model PO6A are almost the same (include principle, structure, material, production process, performance) except appearance, details see below:
| Model | Appearance color |
|-------|------------------|
| PO6 | Blue |
| PO6A | Gray |
{5}------------------------------------------------
The device consists of main unit, SpO2 sensor and charging cable. The main unitis mainly composed of MCU, power management circuit,SpO2 measurement circuit,display control circuit, etc.
The device is powered by internal battery. The device is not for life supporting or life-sustaining, not for implant. The device or sensor is not sterile, the sensor does not needsterilization, and the sensor is reusable but does not need re-sterilization since it is notsterile. The device is for prescription. The device does not contain drug or biological products.
#### 6. Intended Use/ Indications for Use
The Pulse Oximeter is a non-invasive device intended for spot checking and/or continuous data collection of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in home environments and clinical institutions except acute clinical environment.
| Character<br>istics | Subject Device | Primary Predicate<br>Device | Reference Predicate<br>Device | Remark |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Name of<br>the device | Oxyfit Pulse<br>Oximeter<br>PO6,PO6A | Pulse Oximeter<br>A310B | Checkme Pro Health<br>Monitor | --- |
| Manufactu<br>rer | Shenzhen Viatom<br>Technology Co.,<br>Ltd | Shenzhen Aeon<br>Technology Co.,<br>Ltd. | Shenzhen Viatom<br>Technology Co.,<br>Ltd | --- |
| 510(K)<br>Number | N/A | K200414 | K150869 | --- |
| Product<br>code | 21 CFR 870.2700,<br>DQA | 21 CFR 870.2700,<br>DQA | 21 CFR 870.2300,<br>MWI<br>Secondary product<br>codes: 21 CFR<br>870.2700, DQA<br>21 CFR 870.2340, DPS<br>21 CFR 880.2910, FLL<br>21 CFR 870.2300, DRT | Substantial<br>equivalent |
| Classificat<br>ion | II | II | II | Substantial<br>equivalent |
| Indication<br>for Use | The Pulse Oximeter<br>is a non-invasive<br>device intended for<br>spot checking<br>and/or continuous<br>data collection of<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2)<br>and pulse rate (PR).<br>This portable device<br>is indicated for use<br>in adult patients in<br>home environments<br>and clinical<br>institutions except<br>acute clinical<br>environment. | The Pulse Oximeter<br>is a non-invasive<br>device intended for<br>spot checking of<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2)<br>and pulse rate (PR).<br>This portable device<br>is indicated for use in<br>adult patients in<br>clinical institution<br>and home<br>environments. | The Checkme Pro<br>Health Monitor is<br>intended to be used<br>for measuring,<br>displaying, reviewing<br>and storing of ECG<br>(adults only), oxygen<br>saturation and pulse<br>rate (adults only for<br>continuous data<br>collection and<br>recording, adults and<br>pediatrics for spot-<br>checking) and<br>temperature in the<br>home or in healthcare<br>facilities. This device<br>is not intended to<br>substitute for a<br>hospital diagnostic<br>ECG device and not<br>to be used on patients<br>with implanted<br>cardiac devices, such<br>as pacemakers and/or<br>implanted cardio-<br>defibrillators (ICDs). | Different<br>1) |
| Rx or<br>OTC | Rx | Rx | Rx | Substantial<br>equivalent |
| Patient<br>type | Adults | Adults | Adult and pediatric | Different 2) |
| Intended<br>applicatio<br>n site | Finger | Finger | Finger | Substantial<br>equivalent |
| Principle | The device displays<br>numerical values<br>for functional<br>oxygen<br>saturation of arterial<br>hemoglobin (SpO2)<br>and pulse rate by<br>measuring the<br>absorption of red<br>and infrared (IR)<br>light passing | The device displays<br>numerical values for<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2)<br>and pulse rate by<br>measuring the<br>absorption of red<br>and infrared (IR)<br>light passing<br>through perfused | The device displays<br>numerical values for<br>functional oxygen<br>saturation of arterial<br>hemoglobin (SpO2)<br>and pulse rate by<br>measuring the<br>absorption of red<br>and infrared (IR) light<br>passing<br>through perfused | Substantial<br>equivalent |
| | | | | |
| | through perfused<br>tissue. Changes<br>in the absorption<br>caused by the<br>pulsation of blood<br>in the vascular<br>bed are used to<br>determine oxygen<br>saturation and pulse<br>rate. | tissue. Changes<br>in the absorption<br>caused by the<br>pulsation of blood in<br>the vascular<br>bed are used to<br>determine oxygen<br>saturation and pulse<br>rate. | tissue. Changes<br>in the absorption<br>caused by the<br>pulsation of blood in<br>the vascular<br>bed are used to<br>determine oxygen<br>saturation and pulse<br>rate. | |
| LED<br>wavelength | Red=660 nm;<br>Infrared=940 nm; | Red=660 nm;<br>Infrared=905 nm; | Red=660 nm;<br>Infrared=940 nm; | Different 3 |
| SpO2<br>display<br>range | 0%-100% | 0%-100% | 0%-100% | Substantial<br>equivalent |
| SpO2<br>measurem<br>ent<br>accuracy | 70%-100%: ±2%<br>(Arms:1.57)<br>70%-80%: ±3%<br>80%-90%: ±2%<br>90%-100%: ±2%<br>0%-69%: not defined | 70-100%, ±3%;<br>0-69%, unspecified; | 70%-100%: ±2%<br>(Arms:1.88)<br>70%-80%: ±3%<br>80%-90%: ±2%<br>90%-100%: ±2%<br>0%-69%: not<br>defined | Substantial<br>equivalent |
| SpO2<br>resolution | 1% | 1% | 1% | Substantial<br>equivalent |
| Pulse rate<br>measurem<br>ent range | 30 bpm~250 bpm | 30 bpm~250 bpm | 30 bpm~250 bpm | Substantial<br>equivalent |
| Pulse rate<br>accuracy | ±2bpm or ±2%<br>(whichever is greater) | ±2bpm | ±2bpm or ±2%<br>(whichever is greater) | Substantial<br>equivalent |
| Pulse rate<br>resolution | 1bpm | ±2bpm | 1bpm | Substantial<br>equivalent |
| Power<br>supply | Lithium<br>rechargeable<br>battery | 2 AAA<br>alkaline<br>batteries | Lithium<br>rechargeable<br>battery | Different 4 |
| Wireless | Bluetooth | Bluetooth | Bluetooth | Substantial<br>equivalent |
| Contacting<br>duration | Prolonged (24h to 30<br>day) | Less than 24h | Less than 24h | Substantial<br>equivalent |
| Contacting<br>type | Skin surface-<br>contacting | Skin surface-contacting | Skin surface-contacting | Substantial<br>equivalent |
| Biocompati<br>bility of<br>patient<br>contact<br>parts | Comply with ISO<br>10993-1 | Comply with ISO<br>10993-1 | Comply with ISO<br>10993-1 | Substantial<br>equivalent |
| Standard | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-61<br>ISO 10993-5<br>ISO 10993-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-61<br>ISO 10993-5<br>ISO 10993-10 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-61<br>ISO 10993-5<br>ISO 10993-10 | Substantial<br>equivalent |
#### 7. Substantial equivalence Comparisons to the 510(k) Cleared Devices(Predicate Devices)
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Shenzhen Viatom Technology Co., Ltd. Product: Pulse Oximeter
VOL 008: 001 510(k) Summary Version: A/1
{9}------------------------------------------------
#### Justification of difference:
Different 1):
Compared with Primary Predicate Device:
The subject device is used for spot-check and/or continuous data collection whereas the primary predicate device is used in adults for spot-checking.
The difference is the function "and/or continuous data collection".But the reference predicate device is also used in adults for spot-check and continuous data collection. And the subject device is based on the reference predicate device. So the difference does not affect the safety and effectiveness of the subject device.And the subject device together meets the requirements of the IEC 60601-1,IEC 60601-1-11,ISO 80601-2-61 and IEC 60601-1-2.
Compared with Reference Predicate Device:
The subject device is used for spot-check and/or continuous data collection in adults, whereas the reference predicate device is used in adults only for continuous data collection and recording, as well as adults and pediatrics for spot-checking.The subject device is based on the reference predicate device, So they are substantially equivalent, just different expression.
Both the subject device and the reference predicate device have similar function including measuring, displaying, storing and transmitting of pulse oxygen saturation(SpO2) and Pulse Rate.The reference predicate device is a Cardiac Monitor and has more function (such as ECG and Temperature).The function of the subject device is obviously within the range of the Reference Predicate Device. So the differences will not raise new questions of safety and effectiveness. The subject devices passed the test of bench test and ISO 80601-2-61.
Different 2):
Compared with Primary Predicate Device:
Both the subject device and the primary predicate device are same.
Compared with Reference Predicate Device:
The subject device is used for adult patients whereas the reference predicate device is used in a wider range of patient types like adults and pediatric patients. The Patient type of the subject device is obviously within the range of the Reference Predicate Device.The differences does not affect the safety and effectiveness of the subject device.
{10}------------------------------------------------
#### Different 3):
Compared with Primary Predicate Device:
Infrared LED Wavelength of the subject device is 940nm, and primary predicate's wavelength is 905nm. These two wavelengths are both adopted widely in measurement of SpO2. What's more, the reference predicate's wavelength is 940nm. And the subject device together meets the requirements of the IEC 60601-1, IEC 60601-1-11, ISO 80601-2-61 and IEC 60601-1-2.The differences does not affect the safety and effectiveness of the subject device.
Compared with Reference Predicate Device:
Both the subject device and the reference predicate device are same.
Different 4):
Compared with Primary Predicate Device:
The power supply of the subject device is Lithium rechargeable battery, and the power supply of the primary predicate device is 2 AAA alkaline batteries. And the subject device together meets the requirements of the IEC 60601-1.IEC 60601-1-11,ISO 80601-2-61 and IEC 60601-1-2.What's more,the power supply of the reference predicate device is Lithium rechargeable battery.So the differences does not affect the safety and effectiveness of the subject device.
Compared with Reference Predicate Device:
Both the subject device and the reference predicate device are same.
{11}------------------------------------------------
#### 8. Brief discussions of the non-clinical tests
The subject device conforms to the following guidances and standards:
- IEC 60601-1: Medical Electrical Equipment Part 1: General Requirements for ● Basic Safety and essential performance
- IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- ISO 80601-2-61: Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.
- IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- I SO 10993-1 Fourth edition 2009-10-15 Biological evaluation of nedical devices - Part 1: Evaluation and testing within a risk nanagement process [Including: Technical Corrigendum 1 (2010)]
- I SO 10993-5 Third edition 2009-06-01 Biological evaluation of nedical devices - Part 5: Tests for in vitro cytotoxicity
- I SO 10993-10 Third Edition 2010-08-01 Biological evaluation of nedical devices -Part 10: Tests for irritation and skin sensitization
- I SO 14971 Risk Minagement
- 47 CFR FCC PART 15. Subpart C Unintentional Radiators/ Mscellaneous ● Wrel ess Commit at i ons Ser vi ce
- FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
- FDA Guidance for Pulse Oximeters - Premarket Notification Submissions [510(k)s]
- FDA Guidance for Content of Premarket Submissions for Management of Cybersecurity in Medical Devices.
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It is concluded that the subject Pulse Oximeter device is in compliance with the requirements of the aforementioned tests.
{12}------------------------------------------------
#### 9. Brief discussions of clinical tests
The pulse oximeter(PO6、PO6A) was subjected to clinical testing. The functional oxygen saturation (SpO2) measurement has been validated in accordance with ISO 80601-2-61.
Clinical testing (controlled desaturation study) was conducted, during induced hypoxia studies ona total of 12 healthy adult ( healthy, non-smoking, light-to-dark-skinned ) in an independent research laboratory. The measured arterial hemoglobin saturation value (SpO2) of the subject device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter (control device). The accuracy of the subject device is in comparison with the control device over the SpO2 range of 70~100%.
The SpO2 accuracy results include:
• 1.57% which meets the accuracy requirement of less than or equal to 2% under no motion condition. The accuracy specification is reported as accuracy root mean square (Arms).
The result met the criteria specified in the ISO 80601-2-61; In addition, there were no reported adverse effects during these investigations.
#### 10. Brief discussions of Biocompatibility tests
The patient-contacting components and materials are listed together with the corresponding test items in the following table. The contact duration is limited to be within 24 hours to 30 days, and the type of contact belongs to surface medical device-intact skin with prolonged duration
| Patient-contacting<br>Components | Contact Materials | Leverage<br>biocompatibility<br>test items |
|----------------------------------|-------------------|----------------------------------------------------|
| Enclosure | PC | Cytotoxicity |
| Finger sleeve | Silicone | Skin Sensitization<br>Test<br>Skin Irritation Test |
The PC, Silicone of the test article is identical to the medical device in its final finished form in formulation, processing, sterilization, and geometry, and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents).
Based on International Standard ISO 10993 and FDA Biological evaluation guidance,the material are tested for vitro cytotoxicity, irritation and Skin Sensitization.
{13}------------------------------------------------
The biocompatibility test results demonstrated that there's no cytotoxic potential, no evidence of significant irritation nor evidence of sensitization. Thus, the device meets the requirement of Biocompatibility.
#### 11. Brief discussions of Bluetooth tests
The device Bluetooth will be enabled automatically after it's turned on.
To establish a Bluetooth connection:
- Keep the device Bluetooth enabled.
- Make sure the phone Bluetooth is enabled.
And the testing included the Wireless Coexistence tests via Bluetooth, results of which demonstrate the Bluetooth performance of the subject device.
#### 12. Software information
The software level of concern for the subject device is MODERATE. According to FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff", the software validation documentation summarizes the required information for a MODERATE Level of Concern device. And a testing included the Cybersecurity tests, results of which demonstrate the Cybersecurity of the subject device.
## 13. Other information (such as required by FDA guidance):
No other information.
#### 14. Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, Shenzhen Viatom Technology Co., Ltd. concludes that the subject device has the same intended use as the predicate device, and the technological differences do not raise different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.